SpO2 Comparison Between Commercial Smartwatch and Medical-grade Pulse Oximeter During Hypoxia
NCT05657288 · Status: UNKNOWN · Phase: NA · Type: INTERVENTIONAL · Enrollment: 40
Last updated 2023-03-17
Summary
The aim of the project is to experimentally compare SpO2 measured on a wrist by commercial Smartwach and on a finger measured by a medical-grade pulse oximeter in healthy humans during gradual desaturation in the range of 60-100% SpO2 and simultaneously during a step change from the hypoxic phase to the recovery phase (return to physiological values).
Conditions
- Hypoxia
- Pulse Oximetry
Interventions
- OTHER
-
Breathing mixtures of O2 and N2 and monitoring SpO2 on finger and left wrist
Volunteers sequentially inhale three hypoxic gas mixtures each for 2.5 minute. Every measurement will begin with a two-minute stabilization phase, during which the physiological values of the volunteer will be checked. During the whole experiment, non-invasive SpO2 measurements will be performed on the middle finger and on the wrist of left hand.
Sponsors & Collaborators
-
Czech Technical University in Prague
lead OTHER
Principal Investigators
-
Simon Walzel, MSc. · Czech Technical University
Study Design
- Allocation
- NA
- Purpose
- BASIC_SCIENCE
- Masking
- NONE
- Model
- SINGLE_GROUP
Eligibility
- Min Age
- 18 Years
- Max Age
- 60 Years
- Sex
- ALL
- Healthy Volunteers
- Yes
Timeline & Regulatory
- Start
- 2022-12-28
- Primary Completion
- 2023-05-20
- Completion
- 2023-12-20
Countries
- Czechia
Study Locations
More Related Trials
-
Pulse Oximeter Accuracy in Healthy Humans During Hypoxia
NCT06142019 ·Status: RECRUITING
-
Hand Perfusion and Pulse Oximetry Performance
NCT06925646 ·Status: NOT_YET_RECRUITING
-
Accuracy of CardioWatch 287-2 During Profound Hypoxia
NCT05542732 ·Status: COMPLETED ·Phase: NA
-
The Effect of Hypercapnia on Physiological Parameters During Short-term Breathing
NCT05441007 ·Status: COMPLETED ·Phase: NA
-
Accuracy of Pulse Oximeters During Profound Hypoxia
NCT05790616 ·Status: UNKNOWN ·Phase: NA
-
Finger Pulse Oximeters in Hosptial Setting
NCT06994663 ·Status: COMPLETED
-
Accuracy of Pulse Oximeters in Profound Hypoxia
NCT05920278 ·Status: COMPLETED
-
Accuracy of Pulse Oximeters With Profound Hypoxia
NCT05139693 ·Status: COMPLETED
-
Accuracy Validation of a Pulse Oximetry Monitor
NCT02506010 ·Status: WITHDRAWN
-
Comparison of Pulse Oximetry (SpO2) With Different Oximeters and Arterial Saturation (SaO2): Oxygap2 Study
NCT06102499 ·Status: RECRUITING
-
Hypoxia Blood Validation in Multiple Pulse Oximeters
NCT00881829 ·Status: UNKNOWN
-
Accuracy Validation of the Cadwell Pulse Oximetry System
NCT02842476 ·Status: COMPLETED
-
Accuracy of Pulse Oximeters With Profound Hypoxia
NCT06022523 ·Status: UNKNOWN ·Phase: NA
-
SpO2 Measurement Depending on Different Pulse Oximeter Averaging Time Settings
NCT04933773 ·Status: COMPLETED ·Phase: NA
-
SpO2 Accuracy of Noninvasive Disposable Pulse Oximeter Sensor With Motion Conditions
NCT05264116 ·Status: COMPLETED ·Phase: NA
-
SpO2 Accuracy of Noninvasive Disposable Pulse Oximeter Sensor
NCT05245526 ·Status: COMPLETED ·Phase: NA
-
Oxitone Sp02 Hypoxia Test Versus Reference Pulse Oximetry
NCT02870179 ·Status: COMPLETED
-
Pulse Oximeter Accuracy During Stable Hypoxia Plateaus
NCT06460246 ·Status: COMPLETED ·Phase: NA
-
Accuracy of Evie Ring Pulse Oximeter in Profound Hypoxia
NCT06101654 ·Status: COMPLETED
-
Accuracy of ScanWatch Pulse Oximeter With Profound Hypoxia
NCT04380389 ·Status: COMPLETED ·Phase: NA
-
Determination of SpO2 and PR Accuracy Specifications at Rest (71Ag_Vital-0031)
NCT05297500 ·Status: ACTIVE_NOT_RECRUITING
-
The Effect of Continuous Pulse Oximetry Monitoring and Early Intervention Using Wearable Device on the Incidence of Postoperative Hypoxemia
NCT04694599 ·Status: UNKNOWN ·Phase: NA
-
SpO2 Accuracy Validation of Pulse Oximetry Systems During Motion and Non-Motion Conditions
NCT02475668 ·Status: COMPLETED
-
Accuracy of Pulse Oximeters With Profound Hypoxia At Rest (NIHO17)
NCT05745662 ·Status: COMPLETED
-
Controlled Desaturation Study for TipTraQ SpO2 Performance Validation
NCT06634693 ·Status: COMPLETED