Accuracy of ScanWatch Pulse Oximeter With Profound Hypoxia

NCT04380389 · Status: COMPLETED · Phase: NA · Type: INTERVENTIONAL · Enrollment: 14

Last updated 2026-05-04

Study results available
· View outcomes & findings →

Summary

Performance testing of pulse oximeters. The protocol involves brief stable arterial oxygen desaturation in healthy volunteers and sampling arterial blood when a stable level of hypoxia has been attained. The blood sample is analyzed for oxygen saturation with a gold standard bench CO-oximeter. This protocol is aligned with the latest ISO and FDA guidance documents for pulse oximeter testing.

Conditions

  • Oximetry

Interventions

DIAGNOSTIC_TEST

hypoxia

Brief stable arterial oxygen desaturation in healthy volunteers and sampling arterial blood when a stable level of hypoxia has been attained.

Sponsors & Collaborators

  • Withings

    lead INDUSTRY

Study Design

Allocation
NA
Purpose
DIAGNOSTIC
Masking
NONE
Model
SINGLE_GROUP

Eligibility

Min Age
18 Years
Max Age
49 Years
Sex
ALL
Healthy Volunteers
Yes

Timeline & Regulatory

Start
2020-03-11
Primary Completion
2020-03-12
Completion
2020-03-12
FDA Device
Yes

Countries

  • United States

Study Locations