Accuracy of Pulse Oximeters During Profound Hypoxia
NCT05790616 · Status: UNKNOWN · Phase: NA · Type: INTERVENTIONAL · Enrollment: 66
Last updated 2023-03-30
Summary
This study is designed to calibrate and determine the accuracy for SpO2, pulse rate and respiratory rate of the newly in-house build Test Device wrist 1 (TDw1, or EVA) at Philips.
SpO2, pulse rate and respiratory rate during hypoxia will be calculated by using data of well-known reference devices, including:
* A reference SpO2 sensor of Nellcor placed at a fingertip, that reflects also continuously the pulse rate Will be used to compare with the test device.
* A reference respiratory rate device that calculates the respiratory rate based on detection of end-tidal CO2 peaks by capnography.
* Oxygen saturation in arterial blood samples (SaO2), determined by a co-oximeter will be used to calculate the accuracy of the test device.
During the study the following devices will be additionally used by the volunteers:
* AppleWatch 7
* TDw2, watch build by philips, using the PPG and software technology developed by Philips
* A smartphone that detects reflected PPG signals from the reflected screen at the handpalm, by the build in frontfacing camera (TDc) of the smartphone
Volunteers will undergo progressive hypoxia (9 min/% O2) from 21 to 10% O2 in an altitude room, resulting in a volunteer's SpO2 of 73%. During this deliberated hypoxia, the volunteers wear the test and reference devices.
This study consists of 4 sub-studies (NI = non-invasive; IN = invasive with an arterial line):
* NI (Fast-Sitting): volunteers are seated in the hypoxia room in which the ambient oxygen concentration decreases at a speed of 9 min/% O2. If the volunteer reached a SpO2 ≤73% for more than 1 minute, he/she leaves the hypoxia room. And will breath air with 21% oxygen. Volunteers wear TDw1 and TDw2 and the reference devices.
* NI(Fast-Lying): identical to NI(Fast-Sitting) but volunteers lay on a mattress. Volunteers wear TDw1 and AppleWatch 7 and the reference devices.
* NI (Slow-Sitting): identical to NI (Fast-Sitting), but after one of the volunteers reaches a SpO2 ≤73% for more than one minute, oxygen in the room increases at a speed of 9 min/% O2 until normal ambient air oxygen concentration of 21%. Volunteers wear TDw1 and AppleWatch 7 and the reference devices.
* IN(Fast-Sitting): identical to NI(Fast-Sitting) but the volunteer's oxygen saturation in blood samples withdrawn via an arterial line is measured in the laboratory.
The NI studies include 18 healthy participants in each sub-study. After the first volunteers have completed the study, small adaptations in the software of the study devices is still possible, e.g. to increase the quality of the PPG-signals. After the three sub-studies are completed, the algorithm for conversion of raw PPG signals to SpO2, pulse rate and respiratory rate will be defined and fixed for the test devices.
During the IN-study, which can only be started after completion of all NI studies, an arterial catheter will be inserted in the radial artery of the 12 participating volunteers, in order to take several blood samples to measure oxygen saturation in the blood (25 samples at well-defined moments during the study per volunteer). Using these results of arterial oxygen saturation, the accuracy of the test devices can be calculated.
Conditions
Interventions
- PROCEDURE
-
Arterial blood draw
For the IN(Fast-Sitting) arm arterial blood will be collected to derive "Gold-standard" SaO2 reference values.
- OTHER
-
Oxygen reduction
Volunteers undergo an ambient oxygen decrease from 21% to 10%
Sponsors & Collaborators
-
Philips Clinical & Medical Affairs Global
lead INDUSTRY
Principal Investigators
-
Raf De Jongh, MD · Hypoxia Centre of the Complementary Medical Centre (CMC) Genk
Study Design
- Allocation
- NON_RANDOMIZED
- Purpose
- SCREENING
- Masking
- NONE
- Model
- SINGLE_GROUP
Eligibility
- Min Age
- 18 Years
- Max Age
- 65 Years
- Sex
- ALL
- Healthy Volunteers
- Yes
Timeline & Regulatory
- Start
- 2023-01-28
- Primary Completion
- 2023-11-01
- Completion
- 2023-11-01
Countries
- Belgium
Study Locations
More Related Trials
-
Accuracy of Pulse Oximeters With Profound Hypoxia
NCT05322382 ·Status: COMPLETED ·Phase: NA
-
SpO2 Accuracy Validation of the OxySoft Sensor Via Reference CO-Oximetry Motion Study
NCT04559763 ·Status: COMPLETED
-
Controlled Acute Hypoxia Study Comparing Pulse Oximetry to Arterial Blood Samples During Motion
NCT02247765 ·Status: COMPLETED ·Phase: PHASE3
-
Hypoxia Blood Validation in Multiple Pulse Oximeters
NCT00881829 ·Status: UNKNOWN
-
Pulse Oximeter Accuracy in Healthy Humans During Hypoxia
NCT06142019 ·Status: RECRUITING
-
Accuracy Validation of a Pulse Oximetry Monitor
NCT02506010 ·Status: WITHDRAWN
-
SpO2 Comparison Between Commercial Smartwatch and Medical-grade Pulse Oximeter During Hypoxia
NCT05657288 ·Status: UNKNOWN ·Phase: NA
-
Functionality of the Pulse Oximeter and Accelerometer Portions of a Novel Wearable Device
NCT05197790 ·Status: COMPLETED ·Phase: NA
-
Project Mountain - Comparing SpO2 and SaO2 for Accuracy
NCT06372106 ·Status: RECRUITING
-
Determination of SpO2 and PR Accuracy Specifications at Rest (71Ag_Vital-0031)
NCT05297500 ·Status: ACTIVE_NOT_RECRUITING
-
Pulse Oximeter Accuracy During Stable Hypoxia Plateaus
NCT06460246 ·Status: COMPLETED ·Phase: NA
-
Accuracy of Pulse Oximeter With Profound Hypoxia
NCT05693168 ·Status: COMPLETED ·Phase: NA
-
N600X Low Saturation Accuracy Validation
NCT05532670 ·Status: COMPLETED
-
SpO2 Accuracy Low Saturation Validation of the OxySoftN Sensor
NCT04811963 ·Status: COMPLETED
-
Accuracy of Pulse Oximeters With Profound Hypoxia
NCT06022523 ·Status: UNKNOWN ·Phase: NA
-
Hand Perfusion and Pulse Oximetry Performance
NCT06925646 ·Status: NOT_YET_RECRUITING
-
Alternate Methodology of Pulse Oximeter Validation
NCT03628560 ·Status: SUSPENDED ·Phase: NA
-
Accuracy of Pulse Oximeters With Profound Hypoxia At Rest (NIHO18)
NCT05807464 ·Status: COMPLETED
-
Accuracy of SpO2 for Noninvasive Pulse Oximeter Sensor (LNCS ADTX)
NCT03125018 ·Status: COMPLETED ·Phase: NA
-
SpO2 Hypoxia Accuracy Validation Study
NCT03348189 ·Status: COMPLETED
-
Accuracy Validation of the Cadwell Pulse Oximetry System
NCT02842476 ·Status: COMPLETED
-
Finger Pulse Oximeters in Hosptial Setting
NCT06994663 ·Status: COMPLETED
-
Accuracy of Pulse Oximeters With Profound Hypoxia At Rest (NIHO16)
NCT05745675 ·Status: COMPLETED
-
Vital Sign Monitoring With Continuous Pulse Oximetry and Wireless Clinician Notification After Surgery
NCT02907255 ·Status: COMPLETED ·Phase: NA
-
Fingertip Pulse Oximeter Clinical Test
NCT01605955 ·Status: COMPLETED