Comparison Evaluation of Blood Glucose Meter Systems in a Clinical Setting

NCT01614613 · Status: COMPLETED · Phase: NA · Type: INTERVENTIONAL · Enrollment: 146

Last updated 2016-02-29

Study results available
· View outcomes & findings →

Summary

The purpose of this study was to test the performance of an investigational Bayer Blood Glucose (BG) meter and five additional Blood Glucose Monitoring Systems (BGMSs). The goal was to obtain glucose samples from subjects with diabetes across the meter operating ranges. A capillary sample was collected from each subject at 3 different times during the visit to obtain natural capillary blood samples with a range of glucose concentrations. Some blood samples were glycolyzed to lower the glucose concentration levels and glucose solution was added to other samples to raise glucose concentration levels. All meter BG results were compared with plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer). All lancings and testing were performed by study staff.

Conditions

Interventions

DEVICE

Bayer G3 / Tatsu System

Study staff performed Blood Glucose (BG) tests with capillary fingerstick blood collected from subjects with diabetes. All meter BG results were compared with plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer).

DEVICE

Accu-Chek® Aviva Nano Meter/Accu-Chek® Aviva Test Strips

Study staff performed Blood Glucose (BG) tests with capillary fingerstick blood collected from subjects with diabetes. All meter BG results were compared with plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer).

DEVICE

Freestyle Lite® Meter and Test Strips with ZipwikTM tabs

Study staff performed Blood Glucose (BG) tests with capillary fingerstick blood collected from subjects with diabetes. All meter BG results were compared with plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer).

DEVICE

OneTouch® Ultra®2 / OneTouch® Ultra® Blue Test Strips

Study staff performed Blood Glucose (BG) tested with capillary fingerstick blood collected from subjects with diabetes. All meter BG results were compared with plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer).

DEVICE

One Touch® VerioTM Pro/One Touch® VerioTM Test Strips

Study staff performed Blood Glucose (BG) tests with capillary fingerstick blood collected from subjects with diabetes. All meter BG results were compared with plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer).

DEVICE

Truetrack® Meter/Truetrack® Test Strips

Study staff performed Blood Glucose (BG) tests with capillary fingerstick blood collected from subjects with diabetes. All meter BG results were compared with plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer).

Sponsors & Collaborators

  • Ascensia Diabetes Care

    lead INDUSTRY

Principal Investigators

  • Leslie J Klaff, MD · Rainier Clinical Research Center Inc

Study Design

Allocation
NA
Purpose
DIAGNOSTIC
Masking
NONE
Model
SINGLE_GROUP

Eligibility

Min Age
18 Years
Sex
ALL
Healthy Volunteers
No

Timeline & Regulatory

Start
2012-05-31
Primary Completion
2012-06-30
Completion
2012-06-30

Countries

  • United States

Study Locations

More Related Trials

Entities

Diseases

Read the full study record

This page highlights key information. For complete eligibility criteria, study locations, investigator contacts, and the full protocol, visit the original record on ClinicalTrials.gov.

View NCT01614613 on ClinicalTrials.gov