User Performance Evaluation of Contour Plus, Accu-Chek Active, Accu-Chek Performa and OneTouch Select Simple Blood Glucose Monitoring Systems Following ISO 15197:2013

NCT02461394 · Status: COMPLETED · Phase: NA · Type: INTERVENTIONAL · Enrollment: 150

Last updated 2016-01-26

No results posted yet for this study

Summary

The user performance evaluation shows whether people with diabetes are able to obtain accurate measurement results with a blood glucose monitoring system. In this study, user performance evaluation will be performed for Contour Plus (Bayer Healthcare Diabetes Care), Accu-Chek Active (Roche Diagnostics), Accu-Chek Performa (Roche Diagnostics) and OneTouch Select Simple (LifeScan) following ISO 15197:2013, clause 8.

The study will be performed in 2 parts. Each part will comprise testing of 2 BGMS.

For each BGMS, measurement procedures for user performance evaluation will be performed with 1 test meter and 1 reagent system lot by the study subjects.

The same meter and an additional test meter will be used for double measurements performed by study personnel (with the same reagent system lot used by subjects).

Conditions

Interventions

DEVICE

Blood glucose monitoring systems for self-testing

Sponsors & Collaborators

  • Bayer

    collaborator INDUSTRY
  • Institut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm

    lead OTHER

Eligibility

Min Age
18 Years
Sex
ALL
Healthy Volunteers
Yes

Timeline & Regulatory

Start
2015-05-31
Primary Completion
2015-12-31

Countries

  • Germany

Study Locations

More Related Trials