Accuracy of the O2 MedTech PMS-3000 Monitor
NCT01418079 · Status: COMPLETED · Phase: PHASE2 · Type: INTERVENTIONAL · Enrollment: 42
Last updated 2012-05-23
Summary
A cerebral oximeter is a device that uses light to measure the amount of oxygen within the brain. It is similar to the device that measures the level of oxygen in the tip of the finger, known as a pulse oximeter. The cerebral oximeter consists of a sensor placed on the forehead that both emits and detects the amount of light absorbed. This study will determine how accurate the device is by comparing the displayed value on the monitor with blood samples taken simultaneously from the arterial blood in the wrist and venous blood in the neck. In order to test the device over a suitable range, the level of oxygen within the blood will be reduced in a controlled manner by reduction of the inspired oxygen concentration. This is the equivalent of ascending to an altitude of 16,000 feet. The study will be conducted in healthy volunteers.
Conditions
- Healthy
Interventions
- DEVICE
-
O2 MedTech cerebral oximeter (Oxygen desaturation)
Reduction in blood oxygen saturation by sequential reduction in inspired gas composition. Steps are of 6 minutes duration with reduction in pulse oximeter oxygen saturation from 100 to 70%.
Sponsors & Collaborators
- collaborator OTHER
-
O2 MedTech, Inc.
lead INDUSTRY
Principal Investigators
-
David B MacLeod, FRCA · Duke University
Study Design
- Allocation
- NON_RANDOMIZED
- Purpose
- TREATMENT
- Masking
- NONE
- Model
- SINGLE_GROUP
Eligibility
- Min Age
- 18 Years
- Max Age
- 50 Years
- Sex
- ALL
- Healthy Volunteers
- Yes
Timeline & Regulatory
- Start
- 2011-08-31
- Primary Completion
- 2012-02-29
- Completion
- 2012-02-29
Countries
- United States
Study Locations
More Related Trials
-
Determination of SpO2 and PR Accuracy Specifications at Rest (71Ag_Vital-0031)
NCT05297500 ·Status: ACTIVE_NOT_RECRUITING
-
Accuracy of SpO2 for Noninvasive Pulse Oximeter Sensor (RD DCI)
NCT03124797 ·Status: COMPLETED ·Phase: NA
-
Pulse Oximeter Accuracy in Healthy Humans During Hypoxia
NCT06142019 ·Status: RECRUITING
-
Accuracy of Pulse Oximeters With Profound Hypoxia At Rest (NIHO17)
NCT05745662 ·Status: COMPLETED
-
Pulse Oximeter Hypoxia
NCT04562584 ·Status: COMPLETED ·Phase: NA
-
Accuracy of Noninvasive Pulse Oximeter Sensor (LNCS DBI)
NCT03124979 ·Status: COMPLETED ·Phase: NA
-
Correlation of Venous Oxygen Saturation Measurements by Non-invasive Mespere Venous Oximeter and Central Venous Catheter
NCT01908790 ·Status: COMPLETED
-
Accuracy of Pulse Oximeters in Profound Hypoxia
NCT05920278 ·Status: COMPLETED
-
SpO2 Accuracy Validation of the OxySoft Sensor Via Reference CO-Oximetry Motion Study
NCT04559763 ·Status: COMPLETED
-
Alternate Methodology of Pulse Oximeter Validation
NCT03628560 ·Status: SUSPENDED ·Phase: NA
-
Use of Near Infrared Spectroscopy to Measure Intracranial ScvO2
NCT01746576 ·Status: COMPLETED ·Phase: NA
-
Accuracy of Noninvasive Pulse Oximeter Sensor (Rainbow DCI)
NCT03124823 ·Status: COMPLETED ·Phase: NA
-
Accuracy of Oxygen Saturation (SpO2) Noninvasive Pulse Oximeter Sensor (RD Disposable) Under Motion Conditions
NCT03124784 ·Status: COMPLETED ·Phase: NA
-
Accuracy of SpO2 for Noninvasive Pulse Oximeter Sensor (LNCS ADTX)
NCT03125018 ·Status: COMPLETED ·Phase: NA
-
Validation of the O3 Regional Somatic Tissue Oxygenation Monitor
NCT04584788 ·Status: COMPLETED ·Phase: NA
-
Magnetic Resonance Spectroscopy Studies of Acute Hypoxic and Hyperoxic Breathing in Healthy Volunteers
NCT03106376 ·Status: WITHDRAWN ·Phase: NA
-
Oxitone Sp02 Hypoxia Test Versus Reference Pulse Oximetry
NCT02870179 ·Status: COMPLETED
-
SpO2 Accuracy Low Saturation Validation of the OxySoftN Sensor
NCT04811963 ·Status: COMPLETED
-
Accuracy of Pulse Oximeters With Profound Hypoxia
NCT02571686 ·Status: COMPLETED
-
Accuracy of Detection of Methemoglobin With Pulse Oximetry
NCT03869840 ·Status: COMPLETED
-
OM2 Motion Verification Study
NCT06415786 ·Status: COMPLETED
-
Accuracy of Methemoglobin Measurement for Rainbow DCI Sensors
NCT03124927 ·Status: COMPLETED ·Phase: NA
-
Accuracy of Pulse Oximeters With Profound Hypoxia at Rest
NCT04814823 ·Status: COMPLETED
-
N600X Low Saturation Accuracy Validation
NCT05532670 ·Status: COMPLETED
-
Accuracy of Pulse Oximeters With Profound Hypoxia At Rest (NIHO16)
NCT05745675 ·Status: COMPLETED