N600X Low Saturation Accuracy Validation
NCT05532670 · Status: COMPLETED · Type: OBSERVATIONAL · Enrollment: 15
Last updated 2022-11-23
Summary
To compare pulse oximetry saturation measurements, during normal to low saturation, to saturation measurements made by a multi-wavelength CO-oximeter, taken from arterial blood samples from healthy human subjects.
Conditions
Interventions
- DEVICE
-
Non-Invasive Controlled Hypoxia
Following arterial cannulation, hypoxia is induced by reducing the percent of inspired oxygen the subject breathes while simultaneous measurements are recorded from the test pulse oximeters and arterial blood samples are collected at targeted levels of saturation.
Sponsors & Collaborators
-
Medtronic - MITG
lead INDUSTRY
Principal Investigators
-
Arthur Cabrera, MD · Clinimark, LLC
Eligibility
- Min Age
- 18 Years
- Max Age
- 50 Years
- Sex
- ALL
- Healthy Volunteers
- Yes
Timeline & Regulatory
- Start
- 2022-09-29
- Primary Completion
- 2022-10-07
- Completion
- 2022-11-01
- FDA Device
- Yes
Countries
- United States
Study Locations
More Related Trials
-
Desaturation Validation of INVSENSOR00028
NCT03704636 ·Status: COMPLETED ·Phase: NA
-
Controlled Desaturation Study for TipTraQ SpO2 Performance Validation
NCT06634693 ·Status: COMPLETED
-
Induced Hypoxia Study for Validation of SpO2 Accuracy
NCT00788983 ·Status: COMPLETED
-
RDS MultiSense® SpO2 Calibration
NCT05466942 ·Status: COMPLETED ·Phase: NA
-
SpO2 Hypoxia Accuracy Validation Study
NCT03348189 ·Status: COMPLETED
-
SpO2 Accuracy Validation of Pulse Oximetry Systems During Motion and Non-Motion Conditions
NCT02475668 ·Status: COMPLETED
-
Desaturation Validation of INVSENSOR00037
NCT04002973 ·Status: COMPLETED ·Phase: NA
-
Accuracy of SpO2 for Noninvasive Pulse Oximeter Sensor (LNCS ADTX)
NCT03125018 ·Status: COMPLETED ·Phase: NA
-
SpO2 Validation of Noninvasive Red Diamond Disposable Pulse Oximeter Sensor
NCT03124602 ·Status: COMPLETED ·Phase: NA
-
SpO2 Accuracy Comparison of Medline ReNewal Sensors to Arterial Blood CO-Oximetry
NCT03565679 ·Status: COMPLETED ·Phase: NA
-
Pulse Oximeter Accuracy During Stable Hypoxia Plateaus
NCT06460246 ·Status: COMPLETED ·Phase: NA
-
Accuracy of Pulse Oximeter With Profound Hypoxia
NCT05693168 ·Status: COMPLETED ·Phase: NA
-
Accuracy of Noninvasive Pulse Oximeter Sensor (LNCS DBI)
NCT03124979 ·Status: COMPLETED ·Phase: NA
-
Accuracy of Pulse Oximeters With Profound Hypoxia
NCT05322382 ·Status: COMPLETED ·Phase: NA
-
Pulse Oximeter Accuracy in Healthy Humans During Hypoxia
NCT06142019 ·Status: RECRUITING
-
Controlled Desaturation Study for Perin Health Patch Validation
NCT06680700 ·Status: RECRUITING
-
The Sunrise SpO2 and Pulse Rate Accuracy Validation Study
NCT05386836 ·Status: COMPLETED ·Phase: NA
-
SpO2 Accuracy of Noninvasive Disposable Pulse Oximeter Sensor
NCT05245526 ·Status: COMPLETED ·Phase: NA
-
Accuracy Validation of the Cadwell Pulse Oximetry System
NCT02842476 ·Status: COMPLETED
-
Clinical Performance of Masimo INVSENSOR00040
NCT05332392 ·Status: COMPLETED ·Phase: NA
-
Accuracy of CardioWatch 287-2 During Profound Hypoxia
NCT05542732 ·Status: COMPLETED ·Phase: NA
-
Pulse CO-Oximetry Noninvasive Monitoring and Laboratory Measurements of Subject Blood Samples
NCT01638455 ·Status: COMPLETED ·Phase: NA
-
Validation of Next Generation Cerebral and Tissue Oximeter
NCT03128372 ·Status: COMPLETED ·Phase: NA
-
Accuracy of Oxygen Saturation (SpO2) Noninvasive Pulse Oximeter Sensor (RD Disposable) Under Motion Conditions
NCT03124784 ·Status: COMPLETED ·Phase: NA
-
Accuracy of Continuous SpO2 Measurement in Adults
NCT01183338 ·Status: WITHDRAWN