Validation of Brain Oxygenation Monitor
NCT00815490 · Status: COMPLETED · Phase: NA · Type: INTERVENTIONAL · Enrollment: 18
Last updated 2013-04-16
Summary
A cerebral oximeter is a device that uses light to measure the amount of oxygen within the brain. It is similar to the device that measures the level of oxygen in the tip of the finger, known as a pulse oximeter. The cerebral oximeter consists of a sensor placed on the forehead that both emits and detects the amount of light absorbed. This study will determine how accurate the device is by comparing the displayed value on the monitor with blood samples taken simultaneously from the arterial blood in the wrist and venous blood in the neck. In order to test the device over a suitable range, the level of oxygen within the blood will be reduced in a controlled manner by reduction of the inspired oxygen concentration. This is the equivalent of ascending to an altitude of 16,000 feet. The study will be conducted in healthy volunteers.
Conditions
- Healthy
Interventions
- DEVICE
-
Desaturation
Reduction in blood oxygen saturation by sequential reduction in inspired gas composition. Steps are of 6 minutes duration with reduction in pulse oximeter oxygen saturation from 100 to 70%.
Sponsors & Collaborators
-
Nonin Medical, Inc
collaborator INDUSTRY - lead OTHER
Principal Investigators
-
David B MacLeod, FRCA · Duke Health
Study Design
- Allocation
- NA
- Purpose
- DIAGNOSTIC
- Masking
- NONE
- Model
- SINGLE_GROUP
Eligibility
- Min Age
- 21 Years
- Max Age
- 35 Years
- Sex
- ALL
- Healthy Volunteers
- Yes
Timeline & Regulatory
- Start
- 2008-06-30
- Primary Completion
- 2008-11-30
- Completion
- 2012-04-30
Countries
- United States
Study Locations
More Related Trials
-
Validation of Brain Oxygenation Monitor on Pediatric Patients
NCT00849940 ·Status: TERMINATED ·Phase: NA
-
Assessing Cerebrovascular Reactivity Based on Cerebral Oximetry: a Pilot Study
NCT01483495 ·Status: COMPLETED
-
Desaturation Validation of INVSENSOR00028
NCT03704636 ·Status: COMPLETED ·Phase: NA
-
Use of Near Infrared Spectroscopy to Measure Intracranial ScvO2
NCT01746576 ·Status: COMPLETED ·Phase: NA
-
Accuracy of Detection of Methemoglobin With Pulse Oximetry
NCT03869840 ·Status: COMPLETED
-
SpO2 Accuracy Low Saturation Validation of the OxySoftN Sensor
NCT04811963 ·Status: COMPLETED
-
A Prospective Study of a New Device for Monitoring Cerebral Oxygenation on Healthy Volunteers
NCT00598819 ·Status: COMPLETED ·Phase: PHASE1
-
Accuracy of Noninvasive Pulse Oximeter Sensor (MightySat RX)
NCT03125135 ·Status: COMPLETED ·Phase: NA
-
Desaturation Validation of INVSENSOR00037
NCT04002973 ·Status: COMPLETED ·Phase: NA
-
Accuracy of Noninvasive Pulse Oximeter Sensor (Rainbow DCI)
NCT03124823 ·Status: COMPLETED ·Phase: NA
-
Magnetic Resonance Spectroscopy Studies of Acute Hypoxic and Hyperoxic Breathing in Healthy Volunteers
NCT03106376 ·Status: WITHDRAWN ·Phase: NA
-
Brain Oxygenation During Prehospital Anesthesia
NCT03948711 ·Status: COMPLETED
-
Desaturation Validation of INVSENSOR00036
NCT04002960 ·Status: COMPLETED ·Phase: NA
-
Accuracy of Noninvasive Pulse Oximeter Sensor (LNCS DBI)
NCT03124979 ·Status: COMPLETED ·Phase: NA
-
Validation of the O3 Regional Somatic Tissue Oxygenation Monitor
NCT04584788 ·Status: COMPLETED ·Phase: NA
-
Accuracy of Pulse Oximeters With Profound Hypoxia At Rest (NIHO17)
NCT05745662 ·Status: COMPLETED
-
Influence of Variations in Systemic Blood Flow and Blood Pressure on Cerebral Oxygenation Measured by Cerebral Oximetry
NCT01424800 ·Status: COMPLETED ·Phase: NA
-
Determination of SpO2 and PR Accuracy Specifications at Rest (71Ag_Vital-0031)
NCT05297500 ·Status: ACTIVE_NOT_RECRUITING
-
Accuracy of Pulse Oximeters With Profound Hypoxia at Rest
NCT04814823 ·Status: COMPLETED
-
Evaluation of MABOS for Improved Pulse Oximeter Accuracy in Patients of Color
NCT05995613 ·Status: COMPLETED
-
N600X Low Saturation Accuracy Validation
NCT05532670 ·Status: COMPLETED
-
Hypoxia at Multiple Wavelengths of Light
NCT01957891 ·Status: COMPLETED
-
Accuracy of SpO2 for Noninvasive Pulse Oximeter Sensor (RD DCI)
NCT03124797 ·Status: COMPLETED ·Phase: NA
-
Accuracy of SpO2 for Noninvasive Pulse Oximeter Sensor (LNCS ADTX)
NCT03125018 ·Status: COMPLETED ·Phase: NA
-
SpO2 Accuracy Validation of the OxySoft Sensor Via Reference CO-Oximetry Motion Study
NCT04559763 ·Status: COMPLETED