RDS MultiSense® SpO2 Calibration

NCT05466942 · Status: COMPLETED · Phase: NA · Type: INTERVENTIONAL · Enrollment: 16

Last updated 2024-12-19

No results posted yet for this study

Summary

Data will be collected with the MultiSense® pulse oximetry system during non-motion conditions over the range of 70-100%. A Clinimark reference system, previously correlated to arterial blood CO-Oximetry will be the basis for comparison. A minimum of fifteen (15) up to fifty (50) healthy adult subjects, ranging in pigmentation from light to dark, with at least six (6) subjects with dark pigmentation (Fitzpatrick 5-6), will be enrolled in the study to meet the study design requirements defined by ISO 80601-2-61 and by the FDA's Guidance for Pulse Oximeters (March 4, 2013). SpO2 data will be evaluated during non-motion conditions. SpO2 data will be evaluated during stable step plateaus of induced hypoxic levels. The investigational device will be placed on the thorax of the subjects (patch on the upper back and external electrode on the right pectoral). Simultaneous data collection will be set up for the reference and the systems under test.

Conditions

Interventions

DEVICE

Pulse Oximeter, Respiratory Rate, Heart Rate

Measurement of physiological parameters

Sponsors & Collaborators

  • Clinimark, LLC

    collaborator OTHER
  • Rhythm Diagnostic Systems

    lead INDUSTRY

Study Design

Allocation
NA
Purpose
OTHER
Masking
NONE
Model
SINGLE_GROUP

Eligibility

Min Age
18 Years
Max Age
50 Years
Sex
ALL
Healthy Volunteers
Yes

Timeline & Regulatory

Start
2022-03-15
Primary Completion
2022-03-17
Completion
2022-10-20
FDA Device
Yes

Countries

  • United States