The Sunrise SpO2 and Pulse Rate Accuracy Validation Study
NCT05386836 · Status: COMPLETED · Phase: NA · Type: INTERVENTIONAL · Enrollment: 12
Last updated 2022-06-28
Summary
The purpose of this study is to evaluate the SpO2 accuracy performance of the Sunrise Sensor 2 placed at the tip of the chin during non-motion conditions over the range of 70-100% by comparison to SaO2 values determined from arterial blood sample specimen analyzed by a CO-oximeter. Testing is conducted under normal office environment conditions. It is expected that the Accuracy Root Mean Square (ARMS) performance of the oximetry system will meet the required specification of ARMS of 3.5% or less allowed for reflectance technology in non-motion conditions for the range of 70-100% SaO2.
A secondary goal is to evaluate the pulse rate performance simultaneously collected over the SpO2 range covered.
Conditions
- Validation Study
Interventions
- DEVICE
-
Pulse oximeter validation population
Evaluation of the SpO2 accuracy performance of the Sunrise Sensor 2 placed at the tip of the chin during non-motion conditions over the range of 70-100% by comparison to SaO2 values determined from arterial blood sample specimen analyzed by a CO-oximeter.
Sponsors & Collaborators
-
Sunrise
lead INDUSTRY
Study Design
- Allocation
- NA
- Purpose
- DIAGNOSTIC
- Masking
- NONE
- Model
- SINGLE_GROUP
Eligibility
- Min Age
- 18 Years
- Max Age
- 50 Years
- Sex
- ALL
- Healthy Volunteers
- Yes
Timeline & Regulatory
- Start
- 2022-05-24
- Primary Completion
- 2022-06-21
- Completion
- 2022-06-21
Countries
- Belgium
Study Locations
More Related Trials
-
Desaturation Validation of INVSENSOR00036
NCT04002960 ·Status: COMPLETED ·Phase: NA
-
SpO2 Accuracy Comparison of Smart Sock V. 2 SpO2 to Arterial Blood CO-Oximetry
NCT03630016 ·Status: COMPLETED ·Phase: NA
-
SpO2 Accuracy Validation of Pulse Oximetry Systems During Motion and Non-Motion Conditions
NCT02475668 ·Status: COMPLETED
-
Clinical Performance of Masimo INVSENSOR00040
NCT05332392 ·Status: COMPLETED ·Phase: NA
-
SpO2 Validation of Noninvasive Red Diamond Disposable Pulse Oximeter Sensor
NCT03124602 ·Status: COMPLETED ·Phase: NA
-
Accuracy Validation of Belun Oxygen Saturation Pulse Oximeter FDA Submission Study
NCT03439306 ·Status: COMPLETED
-
Hypoxia Blood Validation in Multiple Pulse Oximeters
NCT00881829 ·Status: UNKNOWN
-
Belun SpO2 Accuracy Comparison to Arterial Blood CO-Oximetry
NCT04873947 ·Status: COMPLETED
-
SpO2 Hypoxia Accuracy Validation Study
NCT03348189 ·Status: COMPLETED
-
Controlled Desaturation Study for Perin Health Patch Validation
NCT06680700 ·Status: RECRUITING
-
SpO2 Accuracy Validation of the OxySoft Sensor Via Reference CO-Oximetry Motion Study
NCT04559763 ·Status: COMPLETED
-
Accuracy of Continuous SpO2 Measurement in Adults
NCT01183338 ·Status: WITHDRAWN
-
Desaturation Validation of INVSENSOR00037
NCT04002973 ·Status: COMPLETED ·Phase: NA
-
Induced Hypoxia Study for Validation of SpO2 Accuracy
NCT00788983 ·Status: COMPLETED
-
Accuracy of SpO2 for Noninvasive Pulse Oximeter Sensor (LNCS ADTX)
NCT03125018 ·Status: COMPLETED ·Phase: NA
-
SpO2 Accuracy of Noninvasive Disposable Pulse Oximeter Sensor
NCT05245526 ·Status: COMPLETED ·Phase: NA
-
An Invasive Controlled Desaturation Study to Determine the Accuracy of the CART-I Pulse Oximeter
NCT05693961 ·Status: COMPLETED ·Phase: NA
-
Accuracy of Pulse Oximeter With Profound Hypoxia
NCT05693168 ·Status: COMPLETED ·Phase: NA
-
SpO2 Accuracy of Noninvasive Disposable Pulse Oximeter Sensor With Motion Conditions
NCT05264116 ·Status: COMPLETED ·Phase: NA
-
Accuracy of Noninvasive Pulse Oximeter Sensor (MightySat RX)
NCT03125135 ·Status: COMPLETED ·Phase: NA
-
Desaturation Validation of INVSENSOR00009
NCT03396835 ·Status: COMPLETED ·Phase: NA
-
SpO2 Accuracy of Noninvasive Pulse Oximeter Sensor
NCT05141500 ·Status: COMPLETED ·Phase: NA
-
Performance of Masimo MightySat Pulse Oximeter in Differing Levels of Skin Pigmentation
NCT05083351 ·Status: COMPLETED ·Phase: NA
-
RDS MultiSense® SpO2 Calibration
NCT05466942 ·Status: COMPLETED ·Phase: NA
-
Comparison of Normal Readings and Repeatability Across Skin Colors
NCT01326819 ·Status: COMPLETED