The Sunrise SpO2 and Pulse Rate Accuracy Validation Study

NCT05386836 · Status: COMPLETED · Phase: NA · Type: INTERVENTIONAL · Enrollment: 12

Last updated 2022-06-28

No results posted yet for this study

Summary

The purpose of this study is to evaluate the SpO2 accuracy performance of the Sunrise Sensor 2 placed at the tip of the chin during non-motion conditions over the range of 70-100% by comparison to SaO2 values determined from arterial blood sample specimen analyzed by a CO-oximeter. Testing is conducted under normal office environment conditions. It is expected that the Accuracy Root Mean Square (ARMS) performance of the oximetry system will meet the required specification of ARMS of 3.5% or less allowed for reflectance technology in non-motion conditions for the range of 70-100% SaO2.

A secondary goal is to evaluate the pulse rate performance simultaneously collected over the SpO2 range covered.

Conditions

  • Validation Study

Interventions

DEVICE

Pulse oximeter validation population

Evaluation of the SpO2 accuracy performance of the Sunrise Sensor 2 placed at the tip of the chin during non-motion conditions over the range of 70-100% by comparison to SaO2 values determined from arterial blood sample specimen analyzed by a CO-oximeter.

Sponsors & Collaborators

  • Sunrise

    lead INDUSTRY

Study Design

Allocation
NA
Purpose
DIAGNOSTIC
Masking
NONE
Model
SINGLE_GROUP

Eligibility

Min Age
18 Years
Max Age
50 Years
Sex
ALL
Healthy Volunteers
Yes

Timeline & Regulatory

Start
2022-05-24
Primary Completion
2022-06-21
Completion
2022-06-21

Countries

  • Belgium