Intracranial Hypertension and Optic Nerve Sheath Diameter

NCT01796015 · Status: COMPLETED · Phase: NA · Type: INTERVENTIONAL · Enrollment: 230

Last updated 2025-09-04

No results posted yet for this study

Summary

Intracranial hypertension (ICH) is a mortality risk factor in severe traumatic brain injury (TBI), in purulent meningitis, in hepatic encephalopathy and in Reye's syndrome. It is also a risk factor for severe neurologic sequelae in survivors. Intracranial pressure (ICP) monitoring is likely to guide therapeutics, and certain research on adults or on children, suggest that IH therapeutic approach, for instance for bacterial meningitis, would improve the prognosis. Two monitoring techniques are currently recommended. They are reference methods for ICP measure :

* monitoring with intraventricular catheter,
* intra-parenchymal monitoring using optical fiber catheter. Non invasive methods have been suggested, including ultrasound measurement of optic nerve sheath diameter (ONSD) which is the most interesting one. The ONSD measured ultrasonically is correlated with ICP level in adults with severe TBI. A diameter over 5,9 mm predicts ICH within the first 24 hours. In children, ONSD average values have been worked out, and an ONSD increase is found in children suffering from hydrocephalus with IH and in children with TBI. ICH precocious detection is fundamental in children sensitive to ICH because their cerebral development is not finished yet. Difficulties met for ICP monitoring implementation in infants and its invasive nature are often disliked by clinicians. A non-invasive exam is then essential to allow a better care of children with ICH in intensive care unit.

Conditions

  • Cerebral Stroke
  • Cerebral Venous Thrombosis

Interventions

OTHER

ONSD ultrasound

Description of the ONSD ultrasound: ONSD is measured 3 mm behind the globe and perpendicularly to the optic nerve axis (2 dimensions mode). For each optic nerve, 2 measures are performed: 1 sagittal and 1 transverse. The patient is positioned in 30° dorsal elevation. Gel is applied on the eyelid and a 7,5 MHz probe is used. At the same time and on the same side, a transcranial Doppler is performed using a 2 MHz probe placed on the temple. Are measured: systolic velocity, mean velocity, diastolic velocity, resistance index and pulsatility index. As ONSD measures cannot be done blinded (regarding ICP monitoring), a centralised evaluation (by 2 experts) of the recorded sonograms is planned.

Sponsors & Collaborators

  • Hospices Civils de Lyon

    lead OTHER

Principal Investigators

  • Etienne JAVOUHEY, MD · HOPITAL FEMME MERE ENFANT HOSPICES CIVILS DE LYON

Study Design

Allocation
NON_RANDOMIZED
Purpose
DIAGNOSTIC
Masking
NONE
Model
SINGLE_GROUP

Eligibility

Min Age
2 Months
Max Age
16 Years
Sex
ALL
Healthy Volunteers
No

Timeline & Regulatory

Start
2011-04-30
Primary Completion
2017-02-28
Completion
2017-02-28

Countries

  • France

Study Locations

More Related Trials

Read the full study record

This page highlights key information. For complete eligibility criteria, study locations, investigator contacts, and the full protocol, visit the original record on ClinicalTrials.gov.

View NCT01796015 on ClinicalTrials.gov