COR-INSIGHT: Optimizing Cardiovascular and Cardiopulmonary Outcomes with AI-Driven Multiplexed Indications Using COR ECG Wearable
NCT06763549 · Status: ENROLLING_BY_INVITATION · Type: OBSERVATIONAL · Enrollment: 15000
Last updated 2025-01-08
Summary
The COR-INSIGHT trial aims to evaluate the effectiveness of Peerbridge COR advanced ambulatory ECG wearables (COR 1.0 and COR 2.0) in accurately and non-invasively detecting cardiovascular and cardiopulmonary conditions using AI-based software (CardioMIND and CardioQSync). The study devices offer non-invasive, multiplexed, AI-enabled direct-from-ECG detection as a novel alternative to traditional diagnostic methods, including imaging, hemodynamic monitoring systems, catheter-based devices, and biochemical assays. Continuous COR ECG data collected in hospital, outpatient clinic, or home settings will be analyzed to evaluate the predictive accuracy, sensitivity, specificity, and performance of these devices in differentiating between screen-positive and screen-negative subjects.
The panel of screened indications encompasses a broad spectrum of clinically relevant cardiovascular, cardiopulmonary, and sleep-related diagnostic parameters, which are critical for advanced patient assessment and management. In the cardiovascular domain, the protocol emphasizes the detection and classification of heart failure, assessment of ejection fraction severity, and identification of myocardial infarction, including pathological Q-waves and STEMI. It further addresses diagnostic markers for arrhythmogenic conditions such as QT interval prolongation, T-wave alternans, and ventricular tachycardia, as well as insights into ischemia, atrial enlargement, ventricular activation time, and heart rate turbulence. Additional parameters, such as heart rate variability, pacing efficacy, electrolyte imbalances, and structural abnormalities, including left ventricular hypertrophy, contribute to comprehensive cardiovascular risk stratification.
In the non-invasive cardiopulmonary context, the protocol incorporates metrics like respiratory sinus arrhythmia, cardiac output, stroke volume, and stroke volume variability, providing critical insights into hemodynamic and autonomic function. The inclusion of direct-from-ECG metrics for sleep-related disorders, such as the apnea-hypopnea index, respiratory disturbance index, and oxygen saturation variability, underscores the protocol's utility in addressing the intersection of cardiopulmonary and sleep medicine. This multifaceted approach establishes a robust framework for precision diagnostics and holistic patient management.
The COR 1.0 and COR 2.0 wearables provide multi-lead ECG recordings, with COR 2.0 offering extended capabilities for cardiopulmonary metrics and longer battery life (up to 14 days). COR 2.0 supports tri-modal operations:
(i) Extended Holter Mode: Outputs Leads II and III, mirroring the functionality of COR 1.0 for broader ECG monitoring applications.
(ii) Cardiopulmonary Mode: Adds real-time recording of Lead I, V2, respiratory impedance, and triaxial accelerometer outputs, providing advanced cardiopulmonary insights.
(iii) Real-Time Streaming Mode: Streams data directly to mobile devices or computers via Bluetooth Low Energy (BLE), enabling real-time waveform rendering and analysis.
The COR 2.0 units are experimental and not yet FDA-cleared.
Primary endpoints include sensitivity (true positive rate) \> 80%, specificity (true negative rate) \> 90%, and statistical agreement with reference devices for cardiovascular, cardiopulmonary, and sleep metrics. Secondary endpoints focus on predictive values (PPV and NPV) and overall diagnostic performance. The study employs eight distinct sub-protocols (A through H) to address a variety of cardiovascular, cardiopulmonary, and sleep-related diagnostic goals. These sub-protocols are tailored to specific clinical endpoints, varying in duration (30 minutes to 14 days) and type of data collection. Up to 15,000 participants will be enrolled across multiple sub-protocols. Screening ensures eligibility, and subjects must provide informed consent before participation. Dropouts and non-compliant subjects will be excluded from final analyses.
Conditions
- Cardiopulmonary Failure
- Myocardial Infarction (MI)
- Heart Decompensation
- Heart Failure
- HFrEF - Heart Failure with Reduced Ejection Fraction
- HFpEF - Heart Failure with Preserved Ejection Fraction
- Syncopation
- Syncope
- Ischemic Cardiovascular Disease
- STEMI
- STEMI (ST Elevation MI)
- Atrial Fibrillation (AF)
- Atrial Enlargement
- LVF
- Conduction Defect
- Conduction Abnormalities
- Heart Block
- Valvular Diseases
- Cardiac Output, Low
- Stroke Volume
- Stroke Volume Variation
- Hyperkalemia
- Hypercalcemia
- Hypocalcemia
- LV Dysfunction
- QT Prolongation
- Sudden Cardiac Death Due to Cardiac Arrhythmia
- Ventricular Arrhythmia
- Pacing
- Pacing Induced Dyssynchrony
- Silent Ischemia
- Pericarditis
- Sleep Related Breathing Disorder
- RSA
- Apnea, Obstructive
- Cardiac Output Measurement
- Respiratory Impedance
- CRT And/or ICD
- Infarction
- Cardiomyopathies, Primary
- Hypertrophy
Interventions
- DEVICE
-
SUBPROTOCOL A
A 30-minute COR ECG recording will be performed, including 15 minutes of continuous sitting data used for clinical indication analysis through CardioMIND software. Data collection can occur in-clinic, via PHS, or using a shipped device with telehealth support. Sleep, HRV, and AFIB risk analysis will not be included.
- DEVICE
-
SUBPROTOCOL B
Subjects will undergo 24 hours of COR ECG recording, which must include at least one 15-minute unsupervised sitting session. Data analysis will include sleep patterns, HRV, and AFIB risk. ECG acquisition may take place in-clinic, via Patient Home Setup (PHS), or with a shipped device.
- DEVICE
-
SUBPROTOCOL C
COR 1.0 ECG recordings will be conducted over 2 to 7 days, with at least one 15-minute unsupervised sitting session. At least 25% of randomly selected trial participants subjects will also wear a cleared wrist or finger device to track SpO2 during sleep. PI or Sub-Investigator will assign the number of wear days to the subject. The analysis will include sleep data, HRV, and AFIB risk assessment.
- DEVICE
-
SUBPROTOCOL D
COR 2.0 (CORMDX) recordings will be conducted over 2 to 14 days, with at least one 15-minute unsupervised sitting session. At least 25% of randomly selected trial participants will also wear a cleared wrist or finger device to track SpO2 during sleep nights. PI or Sub-Investigator will assign the number of wear days to the subject. The analysis will include sleep data, HRV, and AFIB risk assessment.
- DEVICE
-
SUBPROTOCOL E
Subjects will complete a 30-minute CORMDx (COR 2.0) ECG recording, with 15 minutes of continuous sitting data used for clinical indication analysis. Sleep, HRV, and AFIB risk analysis will not be performed.
- DEVICE
-
SUBPROTOCOL F
A 30-minute COR 2.0 ECG recording will be completed, along with a 12-lead resting ECG using an FDA-cleared reference device. Sleep, HRV, and AFIB analysis will not be performed.
- DEVICE
-
SUBPROTOCOL G
A 40-minute COR 2.0 recording will be conducted in two phases: 15 minutes of upright sitting and 15 minutes in a supine position. Hemodynamic monitoring, including blood pressure, cardiac output, and stroke volume variation, will be conducted using an FDA-cleared reference device. No sleep, HRV, or AFIB analysis will be performed.
- DEVICE
-
SUBPROTOCOL H
Subjects will undergo a 40-minute COR 2.0 recording, followed by a 40-minute COR 1.0 recording. Hemodynamic monitoring using an FDA-cleared reference device will be performed during both sessions. The transition between devices must occur within 60 minutes if the same adhesive is reused. Sleep, HRV, and AFIB analysis will not be performed.
Sponsors & Collaborators
-
Peerbridge Health, Inc
lead INDUSTRY
Principal Investigators
-
Sandeep Gulati, PhD · Peerbridge Health, Inc
Eligibility
- Min Age
- 18 Years
- Sex
- ALL
- Healthy Volunteers
- Yes
Timeline & Regulatory
- Start
- 2024-11-16
- Primary Completion
- 2026-04-15
- Completion
- 2026-04-15
- FDA Device
- Yes
Countries
- United States
Study Locations
More Related Trials
-
Clinical Safety and Efficacy Testing of a Medical Device Kardi Ai, an ECG Telemetry Monitoring System
NCT07018648 ·Status: RECRUITING ·Phase: NA
-
Monitoring Patients With Acute Cardiac Disease Using a Wireless, Wearable, Non-invasive Monitor
NCT04635371 ·Status: COMPLETED
-
Optimizing Radiation Dose and Utilizing Wearable Devices to Reduce Arrhythmia Risk in Patients Undergoing Thoracic Radiotherapy: A Prospective Cohort Study
NCT06897189 ·Status: NOT_YET_RECRUITING
-
mobiCARE™ ECG Monitoring System
NCT06896734 ·Status: COMPLETED
-
Vis-Rx Post-Market Evaluation
NCT04533503 ·Status: COMPLETED
-
Study Comparing In-laboratory Polysomnography Electrocardiogram (PSG ECG) to Simultaneously Recorded In-laboratory ECG on the CPC M1 Device
NCT01234077 ·Status: COMPLETED ·Phase: NA
-
The Long COVID-19 Wearable Device Study
NCT05741112 ·Status: RECRUITING ·Phase: NA
-
Evaluation of an Ultraportable ECG Recording Device and Its Visualization Software for the Detection of Atrial Fibrillation
NCT03975725 ·Status: UNKNOWN ·Phase: NA
-
Link-HF: Multisensor Non-invasive Telemonitoring System for Prediction of Heart Failure Exacerbation
NCT03037710 ·Status: COMPLETED
-
Study on Cardiac Output Evaluation Based on Wearable Monitoring Data
NCT06938893 ·Status: NOT_YET_RECRUITING
-
Assessment of the BodyGuardian Remote Monitoring Platform in Elderly Healthy Subjects
NCT01808053 ·Status: COMPLETED
-
Monitoring Arrhythmia Patients in in Hospital Monitoring
NCT00919646 ·Status: COMPLETED
-
EVAluation of the Non Inferiority of Interpretability of Cardiokin Compared to Holter
NCT03932227 ·Status: COMPLETED ·Phase: NA
-
Multi-site Study of the Clinical Impact of an AI-assisted Approach to Referring Patients With Interstitial Lung Disease for Diagnostic Evaluation of Pulmonary Hypertension
NCT06911632 ·Status: RECRUITING ·Phase: NA
-
A Multi-center, National, Open-label, Prospective Study to Evaluate the Performance of the V-Lap™ System
NCT06147336 ·Status: RECRUITING ·Phase: NA
-
Performance Assessment of a Remote Patient Monitoring Device
NCT02570906 ·Status: COMPLETED
-
CardioSenseSystem Compared Study Regarding Efficacy and Safety in the Monitoring of ECG
NCT03610529 ·Status: UNKNOWN ·Phase: NA
-
WATCH (Wearable Artificial inTelligence for Cardiac Function and Health Monitoring)
NCT07058064 ·Status: RECRUITING ·Phase: NA
-
Clinical Performance of Withings Move ECG Software for Atrial Fibrillation Detection
NCT04493749 ·Status: ACTIVE_NOT_RECRUITING
-
Evaluation of ECG Transmission and AI Models Using Apple Watch ECGs and Symptoms Data Collected Using a Mayo iPhone App
NCT05324566 ·Status: COMPLETED
-
An Evaluation of Concordance of Smartwatch ECG and One Clinical ECG and Comparison of The Two ECGS in Terms of Predictive Risks
NCT05617391 ·Status: ACTIVE_NOT_RECRUITING
-
Proactive Risk Evaluation for Cardiac Implantable Electronic Device Strategy Using AI-ECG
NCT07217236 ·Status: RECRUITING ·Phase: NA
-
Assess Measurements of Wireless Cardiac Output Device
NCT02252757 ·Status: COMPLETED
-
Automated Diagnostic Performance of Smartwatch ECG for Arrhythmia Detection Using the PulseAI Neural Network.
NCT05668377 ·Status: UNKNOWN
-
Remote Monitoring of Multiple Indicators of Heart Failure
NCT04267744 ·Status: ACTIVE_NOT_RECRUITING ·Phase: NA