Respiratory Rate Validation Study

NCT05856422 · Status: COMPLETED · Type: INTERVENTIONAL · Enrollment: 24

Last updated 2024-11-26

Study results available
· View outcomes & findings →

Summary

The purpose of this study is to conduct a Respiratory Rate accuracy validation on pediatrics population comparing the Gabi SmartCare Gabi Band to the Reference, an FDA cleared End Tidal Carbon Dioxide monitor (GE Datex- Ohmeda) by manually scoring the collected waveform for data analysis.

Conditions

  • Measure Accuracy of Respiratory Rate

Interventions

DEVICE

Respiratory rate measure based on PPG

Testing was conducted under normal office environment conditions. Each subject was connected to the reference EtCO2 monitor (GE Datex-Ohmeda) and the Device Under Test, the Gabi SmartCare Gabi Band. A waveform collected from an EtCO2 measurement was used to determine performance of respiratory rate metrics. Each subject was given a nasal cannula that allowed for measurement of the EtCO2 respiratory rate. The reference respiratory rate was measured simultaneously with the Device Under Test. Data was marked for stable quiet periods that were useable for analysis.

Sponsors & Collaborators

  • Gabi SmartCare

    lead INDUSTRY

Study Design

Allocation
NA
Purpose
OTHER
Masking
NONE
Model
SINGLE_GROUP

Eligibility

Min Age
1 Minute
Max Age
12 Years
Sex
ALL
Healthy Volunteers
Yes

Timeline & Regulatory

Start
2022-12-05
Primary Completion
2022-12-29
Completion
2022-12-29

Countries

  • United States