Evaluating Myocardial Injury for Defibrillation Threshold Testing Methods for ICD Implantation
NCT02111993 · Status: COMPLETED · Phase: NA · Type: INTERVENTIONAL · Enrollment: 60
Last updated 2019-05-22
Summary
The objective of this study is to evaluate myocardial injury, if any, as quantified by cardiac markers (Troponin-T) in defibrillation threshold (DFT) testing during implantation of implantable cardioverter defibrillators (ICDs) using the upper limit of vulnerability (ULV) method vs. standard defibrillation threshold method.
Conditions
- Cardiomyopathy
- Heart Failure
- Cardiac Arrhythmia
Interventions
- DEVICE
-
Standard Defibrillation Testing
Once the ICD is implanted, Standard Defibrillation Threshold (DFT) testing will begin. A low energy shock will be delivered to the heart at the peak of the T wave to induce ventricular fibrillation (VF). Once VF is induced, a rescue shock of 25J will be delivered to restore sinus rhythm. This process will be repeated once more after 5 minutes.
- DEVICE
-
Upper Limit of Vulnerability Testing
Once the ICD is implanted, Upper Limit of Vulnerability (ULV) testing will commence. An 18 J shock will be delivered will at different points from the t-wave. If VF is not induced with any of these shocks, then the ULV will be considered to be 18J. If VF is induced, then a 25 J rescue shock will be delivered.
Sponsors & Collaborators
- collaborator INDUSTRY
-
University of California, San Diego
lead OTHER
Principal Investigators
-
Ulrika Birgersdotter-Green, MD · University of California, San Diego
Study Design
- Allocation
- RANDOMIZED
- Purpose
- OTHER
- Masking
- NONE
- Model
- PARALLEL
Eligibility
- Min Age
- 18 Years
- Sex
- ALL
- Healthy Volunteers
- No
Timeline & Regulatory
- Start
- 2011-08-31
- Primary Completion
- 2015-04-30
- Completion
- 2015-06-30
Countries
- United States
Study Locations
More Related Trials
-
T-wave Alternans and Intrathoracic Impedance Measurements
NCT00669682 ·Status: COMPLETED
-
Comparison of a MAGNET Protocol Versus "off-on" Protocol for Patients With Implantable Cardioverter Defibrillators
NCT01599065 ·Status: COMPLETED
-
Clinical Investigation of the Medtronic EnTrust™ Implantable Cardioverter Defibrillator (ICD), Model D153ATG
NCT00273195 ·Status: COMPLETED ·Phase: NA
-
Trop-Shock DFT-testing Versus None
NCT01230086 ·Status: COMPLETED ·Phase: PHASE4
-
Reduction of Inappropriate Shocks of Implantable Cardioverter Defibrillators (ICDs) With Enhanced SVT Discriminators
NCT01963286 ·Status: COMPLETED ·Phase: NA
-
Optimised Decrement Evoked Potential (DeEP) Mapping to Guide Ventricular Tachycardia (VT) Ablation in Patients With Structural Heart Disease VT
NCT06937983 ·Status: RECRUITING ·Phase: NA
-
MAgnetic Resonance Imaging-guided implanTation of Cardioverter DEFibrillators
NCT06269692 ·Status: NOT_YET_RECRUITING ·Phase: NA
-
A PRospective, rAndomizEd Comparison of subcuTaneOous and tRansvenous ImplANtable Cardioverter Defibrillator Therapy
NCT01296022 ·Status: UNKNOWN ·Phase: NA
-
Effect of Right Ventricular Lead Position on Defibrillation Threshold
NCT02844127 ·Status: WITHDRAWN ·Phase: NA
-
Programming Implantable Cardioverter Defibrillators in Patients With Primary Prevention Indication
NCT00743522 ·Status: COMPLETED
-
Right Ventricular Defibrillation Lead Select Site Study
NCT00385749 ·Status: UNKNOWN ·Phase: NA
-
Further Acute Research of a Subcutaneous Implantable Defibrillator (S-ICD) System
NCT00399217 ·Status: COMPLETED
-
Defibrillation Testing During Implantable Cardioverter Defibrillator (ICD) Replacement
NCT04206371 ·Status: COMPLETED ·Phase: NA
-
Novel ICD Arrhythmia Detection Algorithm
NCT04981496 ·Status: ACTIVE_NOT_RECRUITING
-
Evaluation of Inducible Monomorphic Ventricular Tachycardia (MMVT) in Patients With St. Jude Medical (SJM) Implantable Cardioverter Defibrillator (ICD) Systems or Cardiac Resynchronization Therapy Defibrillation (CRT-D) Systems.
NCT02584595 ·Status: WITHDRAWN
-
Comparison of Right Ventricular Septal and Right Ventricular Apical Pacing in Patients Receiving a CRT-D Device
NCT00833352 ·Status: COMPLETED ·Phase: NA
-
Triggers of Ventricular Arrhythmias (TOVA) Study
NCT00005243 ·Status: COMPLETED
-
Clinical and Device Functional Assessment of Real World ICD Patients
NCT02341768 ·Status: COMPLETED
-
The Use of Subcutaneous Implantable Cardiac Defibrillators (ICDs) Following Extraction of Conventional ICDs
NCT01665131 ·Status: UNKNOWN ·Phase: NA
-
Decision Support for Adults Facing Implantable Cardioverter-Defibrillator Pulse Generator Replacement
NCT02668900 ·Status: COMPLETED ·Phase: NA
-
Deactivation of Implantable Cardioverter Defibrillator During Non-thoracic Surgical Procedures
NCT00806377 ·Status: COMPLETED
-
Study of the Implantable Cardioverter Defibrillator (ICD) Intracardiac Electrograms
NCT00916435 ·Status: COMPLETED
-
Prevention Of Sudden Cardiac Death After Myocardial Infarction by Defibrillator Implantation
NCT05665608 ·Status: RECRUITING ·Phase: NA
-
Antiarrhythmics Versus Implantable Defibrillators (AVID)
NCT00000531 ·Status: COMPLETED ·Phase: PHASE3
-
Reduction of Inappropriate Implantable Cardioverter Defibrillator (ICD) Therapies in Primary Prevention Patients
NCT01217528 ·Status: COMPLETED ·Phase: NA