A Prospective, Controlled, Single-Center Confirmatory Study to Evaluate the Accuracy and Safety of Core Temperature Measurement Using the XST600 Electronic Thermometer Compared to the 3M SpotOn Thermometer in Adult Patients
NCT07432269 · Status: NOT_YET_RECRUITING · Type: OBSERVATIONAL · Enrollment: 117
Last updated 2026-04-20
Summary
The body temperature measured by the skin is easily affected by the measurement environment such as temperature or humidity, and in some cases, the difference from the central body temperature is particularly large. Therefore, there has been an attempt to develop a new method of measuring skin body temperature to predict the central body temperature by measuring body temperature in the skin. The purpose of this study is to evaluate the effectiveness and safety of the core temperature measured using this medical device by comparing the accuracy of the body temperature measured through the XST600 electronic thermometer applying the HF technology developed by Chois Technology with the body temperature measured by the Gold Standard 3M Bair Hugger Temperature Monitoring System Model 370. - Number of target subjects: 117 in total - Basis for calculation: This clinical trial is an corroborating clinical trial to confirm the clinical accuracy and effectiveness of the test equipment (XST600), so it is set based on the requirements of ISO 80601-2-56 (the "Medical Device Standard" 50. Electronic thermometer). 1. Total number of target subjects: 117 (105 + 10%) \* At least 105 participants are required in accordance with ISO 80601-2-56:2017, and 117 participants are targeted by applying the normal research dropout rate of 10%. \* If the research procedure is completed in 105 participants without dropping out, the study is terminated. ① Normal body temperature subject: 82 (73 participants + 10%): Those with a core temperature of 36.1℃ (10) or more and 38.0℃ or less when measured by RCT. \* If the research procedure is completed in 73 participants without dropping out, the study is terminated. ② Number of subjects with heating: 35 (32 participants + 10%): Subjects with a core temperature of 38.0℃ or higher when measured by RCT \* According to ISO 80601-2-56, the maximum number of subjects with heating is 32, by applying the standard that should not exceed 30%, and the target is 35, by applying the normal research dropout rate of 10%. \* If the research procedure is completed without dropping out of only 32 subjects, the study is terminated.
Conditions
- All Patients Who Require Monitoring of Their Temperatures Over the Age of 19
Sponsors & Collaborators
-
Yonsei University
lead OTHER
Eligibility
- Min Age
- 19 Years
- Max Age
- 80 Years
- Sex
- ALL
- Healthy Volunteers
- No
Timeline & Regulatory
- Start
- 2026-05-20
- Primary Completion
- 2027-01-15
- Completion
- 2027-01-15
More Related Trials
-
Measurement of Core Temperatures During Therapeutic Hypothermia
NCT01292148 ·Status: COMPLETED
-
Feasibility of Wearable-Based Continuous Temperature Monitoring
NCT07397988 ·Status: NOT_YET_RECRUITING
-
Validation of the Measurement Accuracy of the Tcore Thermometer
NCT04489927 ·Status: TERMINATED
-
Intraoperative Body Core Temperature Monitoring: Oesophageal Probe vs Heated Controlled Servo Sensor
NCT03820232 ·Status: COMPLETED
-
Zero Heat Flux Thermometry System Comparison Trial
NCT01670760 ·Status: COMPLETED ·Phase: NA
-
Non-invasive Forehead Skin Temperature in Cardiac Surgery
NCT04160845 ·Status: UNKNOWN
-
Feasibility and Accuracy of Core Temperature Measurements Using the Esophageal Temperature Probe Inserted Through the Gastric Lumen of Supraglottic Airway Device in Pediatrics
NCT05705206 ·Status: COMPLETED
-
Accuracy and Precision Comparison of Temptraq
NCT03264833 ·Status: COMPLETED
-
Comparison of Upper and Lower Body Air Warming in Patients Undergoing Thoracic Surgery
NCT02993666 ·Status: COMPLETED ·Phase: NA
-
Correlation Between Core Temperature and Skin Temperature in Pediatrics
NCT04057612 ·Status: COMPLETED
-
Accuracy of Measurement of Surface Body Temperature Using the Health e Pod Device (Card Guard) Compared to Gold Standard Equipment
NCT00850083 ·Status: UNKNOWN
-
The Clinical Performance Validation of Electronic Thermometer and Infrared Thermometers
NCT06353490 ·Status: ACTIVE_NOT_RECRUITING ·Phase: NA
-
Does Environmental Factors Affect Accuracy in Pre-hospital Non-invasive Temperature Measurement?
NCT02274597 ·Status: COMPLETED
-
Can the SpotOn™ Zero-Heat-Flux-Thermometry Sensor Accurately Measure Core Temperature in Children?
NCT03157609 ·Status: COMPLETED ·Phase: NA
-
Core Body Temperature Measurement During Hot and Cold Environmental Exposure
NCT01793337 ·Status: COMPLETED ·Phase: NA
-
Correlation, Accuracy, Precision and Practicability of Zero Heat Flux Temperature Monitoring
NCT02031159 ·Status: COMPLETED
-
Temporal Artery Thermometer in Patient Transport: Reliability and Validity.
NCT03920943 ·Status: WITHDRAWN ·Phase: NA
-
A Comparison of Temperature Measurement During Living Donor Liver Transplantation
NCT04943263 ·Status: UNKNOWN ·Phase: NA
-
Evaluation of the Tyto Thermometer When Used in Clinical Care Setting
NCT03452020 ·Status: UNKNOWN ·Phase: NA
-
Study of a New Clinical Device for Reducing Body Core Temperature
NCT01996982 ·Status: TERMINATED ·Phase: NA
-
Comparison of a Novel Non-Invasive, Non-Touch Infrared Thermometer With Routine Thermometry in Routine Clinical Practice
NCT02805322 ·Status: COMPLETED ·Phase: PHASE4
-
Continuous Temperature Monitoring for tHe Early Recognition of Febrile Neutropenia in Haematological MALignancies
NCT05203809 ·Status: UNKNOWN
-
Performance Study of the Thermodiag Solution for Body Temperature Measurement
NCT06703931 ·Status: NOT_YET_RECRUITING ·Phase: NA
-
Principles and Test Methods of Non-contact Body Thermometry
NCT05247736 ·Status: UNKNOWN
-
Intraoperative Body Temperature Measurement Using Infrared Camera: Comparison With Esophageal, and Skin Temperatures
NCT04964232 ·Status: UNKNOWN ·Phase: NA