25010 - ABL90 FLEX PLUS - Precision Capillary Adult - Clinical Study Protocol
NCT07307352 · Status: COMPLETED · Type: OBSERVATIONAL · Enrollment: 39
Last updated 2026-03-04
Summary
Conducting the ABL90 FLEX PLUS Clinical Precision Study for 15 Parameters in Adult Whole Blood is to validate performance claims for precision in heparinized whole blood for 15 parameters: pH, pCO2, pO2, cCa2+, cCl-, cK+, cNa+, cGlu, cLac, ctHb, sO2, FO2Hb, FMetHb, FCOHb, and FHHb, while being influenced by testing environment that is representative of intended use environment, site variability, and operator variability in a point-of-care (POC) setting.
The investigational device is the ABL90 FLEX PLUS incl. consumables, running SW3.5 MR2, manufactured by Radiometer Medical ApS The study is being conducted in Denmark and in total, a minimum of 30 subjects are to be enrolled to provide successful measurement values from 2 sites.
Conditions
- Diagnostic Tests
Interventions
- DIAGNOSTIC_TEST
-
ABL90 FLEX PLUS analyser running SW3.5 MR2.
The ABL90 FLEX PLUS analyzer is a portable, automated analyzer that measures pH, blood gases, electrolytes, metabolites, and oximetry in whole blood - utilizing 3 measuring modes (syringe, short probe, and capillary mode) to measure samples from a syringe, test tube or a capillary tube. The analyzer provides results for 17 parameters in 35 seconds using 65 μL heparinized whole blood.
Sponsors & Collaborators
-
Radiometer Medical ApS
lead INDUSTRY
Principal Investigators
-
Klaus T Kristiansen, MD · Hvidovre Hospital, ICU
-
Lars Peter k Andersen, MD · Sjællands Universitetshospital Køge, Centre for Anaesthesiology
Eligibility
- Min Age
- 18 Years
- Sex
- ALL
- Healthy Volunteers
- No
Timeline & Regulatory
- Start
- 2026-01-20
- Primary Completion
- 2026-01-30
- Completion
- 2026-01-30
- FDA Device
- Yes
Countries
- Denmark
Study Locations
More Related Trials
-
AQT90 FLEX NTproBNP2 Test Kit Matrix Comparison and Sample Stability Verification Study
NCT06905392 ·Status: COMPLETED
-
Evaluation of the New Acceleromyograph TOF 3D
NCT05005676 ·Status: UNKNOWN ·Phase: NA
-
Clinical Performance of Capillary Blood Samples Collected Using the Tasso+ Serum Gel Capillary Blood Collection System for Clinical Chemistry Testing of 14 Analytes on an Abbott Alinity Analyzer
NCT07260799 ·Status: RECRUITING
-
DxFLEX 10C Clinical Study
NCT05467449 ·Status: COMPLETED
-
Swab Sample Collection for the Detection of Bacterial Proteases
NCT01881815 ·Status: COMPLETED
-
Safety and Performance of Short- and Long-term Baseplates for Attaching the UNEEG Episight Recorder - Part I
NCT05111847 ·Status: COMPLETED ·Phase: NA
-
Swab Sample Collection for the Detection of Bacterial Proteases in Wounds Assay
NCT01750203 ·Status: COMPLETED
-
Use of Rotational Thromboelastometry (ROTEM) to Characterize Coagulation Abnormalities in Burn Patients
NCT02388776 ·Status: TERMINATED ·Phase: NA
-
Comparison of Acceleromyography (AMG) and Mechanomyography (MMG) for Establishing Potency of Neuromuscular Blocking Agents
NCT00660413 ·Status: COMPLETED ·Phase: PHASE4
-
Clinical Evaluation of i-STAT Analyzer Glucose, Hematocrit and Sodium for Capillary Samples
NCT02897154 ·Status: COMPLETED
-
The Equivalency of the Imp SFB7+ (L-Dex U400) With the Imp XCA Device
NCT00766935 ·Status: COMPLETED
-
Electromagnetic Tracking of Devices During Interventional Procedures
NCT01279148 ·Status: WITHDRAWN
-
PRP Yield and Platelet Distribution With EDTA
NCT07084272 ·Status: NOT_YET_RECRUITING
-
Apixaban Validation Study - Additional Study on Fresh Samples
NCT03073265 ·Status: COMPLETED
-
Clinical Evaluation of the iMOST-X1 Hematology Analyzer - Matrix Study
NCT04778553 ·Status: COMPLETED
-
HemoSphere Alta Study
NCT07148323 ·Status: RECRUITING
-
Evaluation of Potassium (K) and Ionized Calcium (iCa) Tests Using Capillary Specimens
NCT07287618 ·Status: RECRUITING ·Phase: NA
-
Clinical Evaluation of Precision for the Potassium and Ionized Calcium Tests in Capillary Specimens
NCT07287631 ·Status: RECRUITING ·Phase: NA
-
Autonomous Blood Drawing Optimization and Performance Testing
NCT05878483 ·Status: RECRUITING ·Phase: NA
-
Evaluation of Tasso+ Blood Self-Collection for Clinical Diagnostic Assessment of Various Biomarkers
NCT05852925 ·Status: RECRUITING ·Phase: NA
-
A Comparative Clinical Study to Assess the Accuracy of TensorTip MTX, a Non Invasive Device
NCT02766010 ·Status: TERMINATED ·Phase: NA
-
A Study to Compare Analyte Levels in Blood Collected Using an Investigational Collection Device With a Commercial Predicate
NCT01934842 ·Status: WITHDRAWN ·Phase: NA
-
Controlled Acute Hypoxia Study - Abbreviated Sensor Line
NCT02216344 ·Status: COMPLETED
-
Biomarkers for the Detection of Lymphatic Insufficiency
NCT02375165 ·Status: ENROLLING_BY_INVITATION
-
abioSCOPE IgE Assay to Aid in the Diagnosis of Allergies
NCT04442932 ·Status: WITHDRAWN