Clinical Evaluation of the Performance of Bionet Sphygmomanometer Modules Based on Protocol Presented by the International Organization for Standardization
NCT06609486 · Status: COMPLETED · Phase: NA · Type: INTERVENTIONAL · Enrollment: 86
Last updated 2025-03-03
Summary
According to INTERNATIONAL STANDARD (hereafter ISO) blood pressure (BP) monitor clinical evaluation protocol (ISO 81060-2:2018, Non-invasive sphygmomanometers- Part2: Clinical investigation of intermittent automated measurement type) for BN1 (BNiBP), an electronic sphygmomanometer module newly developed by Bionet Co., Ltd., the clinical group and evaluation criteria presented in, the effectiveness of blood pressure measurement performance was evaluated.
The auscultation method used as a reference (gold standard) of a non-invasive blood pressure (NIBP) measurement method was used. The first blood pressure measurement was performed once each (reference BP not used for evaluation, R0; test BP not used for evaluation, T0) with auscultation and the test medical device (BM5 equipped with BN1 (BNiBP)). The blood pressure used for evaluation was measured four times by auscultation (first reference BP to fourth reference BP, R1 to R4) and three times by the test medical device (first test BP to third test BP, T1 to T3) alternately. A minimum interval of 1 min was allowed between each blood pressure measurement.
Three data pairs (mean value of R1, R2 \& T1 / R2, mean value of R3 \& T2 / R3, mean value of R4 \& T3) were obtained from one study subject. This study aims to verify that the mean value of the differences and standard deviations of data pairs meet the criteria of ISO 81060-2:2018.
Conditions
- Blood Pressure
Interventions
- DEVICE
-
Blood pressure measurement group
In this study, the protocol applied to automatic electronic blood pressure system module BN1 for performance evaluation is ISO's proposed "ISO 81060-2:2018 Non-invasive sphygmomanometers- Part2: Clinical investment of internal automatic measurement type". The initial BP measurement is performed alternately, measuring the blood pressure by the measurer and the test device once each, and measuring blood pressure by the measurer four times by the observer and three times by the test device.
Sponsors & Collaborators
-
Bionet Co., Ltd
collaborator INDUSTRY -
Hoon-Ki Park
lead OTHER
Principal Investigators
-
Hoon-Ki Park, M.D., Ph.D. · Department of Family Medicine, Hanyang University Seoul Hospital
Study Design
- Allocation
- NA
- Purpose
- DIAGNOSTIC
- Masking
- NONE
- Model
- SINGLE_GROUP
Eligibility
- Min Age
- 3 Years
- Sex
- ALL
- Healthy Volunteers
- Yes
Timeline & Regulatory
- Start
- 2020-06-01
- Primary Completion
- 2021-07-21
- Completion
- 2021-07-23
Countries
- South Korea
Study Locations
More Related Trials
-
Clinical Study About Blood Pressure Measurement Efficacy on Multifunction KEITO
NCT01359748 ·Status: COMPLETED ·Phase: NA
-
Clinical Trial of YuWell YE900 Electronic Sphygmomanometer
NCT04379323 ·Status: COMPLETED ·Phase: NA
-
Data Collection for Validation of an NIBP Algorithm in Pregnant Women
NCT05789849 ·Status: UNKNOWN ·Phase: NA
-
Clinical Study of Blood Pressure Monitor in Conformance With the ANSI/AAMI SP10 Standard
NCT01295528 ·Status: COMPLETED
-
Clinical Trial of Blood Pressure Monitor in Conformance With the ANSI/AAMI ISO 81060-2 Standard
NCT03588559 ·Status: COMPLETED ·Phase: NA
-
Clinical Validation of The Combei BP 880 W
NCT05883605 ·Status: COMPLETED ·Phase: NA
-
Clinical Study of Blood Pressure Monitor in Conformance With the ANSI/AAMI SP10
NCT01502553 ·Status: COMPLETED ·Phase: NA
-
Clinical Trial of Wrist Ambulatory Blood Pressure Monitor in Conformance With the EN ISO 81060-2:2018+AMD1:2020
NCT06573216 ·Status: COMPLETED ·Phase: NA
-
Screening for Masked Hypertension With Smart Wearable Devices (SMART)
NCT05409898 ·Status: UNKNOWN
-
Feasibility and Effectiveness of a Ring-type Blood Pressure Measurement Device
NCT06084065 ·Status: RECRUITING ·Phase: NA
-
Calibration of a Wrist Cuff Blood Pressure Device, According to the AAMI/ESH/ISO Universal Standard
NCT04869826 ·Status: UNKNOWN
-
Comparison of Blood Pressure Measurements Between Transdermal Optical Imaging and Standard of Care
NCT04539860 ·Status: UNKNOWN
-
Clinical Trial on the Accuracy of Blood Pressure Measurement by YE990 Medical Automatic Electronic Sphygmomanometer
NCT06565702 ·Status: NOT_YET_RECRUITING
-
Valid-B.Well BP25 Study
NCT07340957 ·Status: NOT_YET_RECRUITING ·Phase: NA
-
Acute Effect of Photobiomodulation in Individuals with Hypertension
NCT06678698 ·Status: RECRUITING ·Phase: NA
-
Clinical Trial of Wrist Blood Pressure Monitor in Conformance With the EN ISO 81060-2:2019/A1:2020 Standard
NCT05482035 ·Status: COMPLETED ·Phase: NA
-
Blood Pressure Monitor Clinical Test (Cuff Range: 22 Cm-36cm)
NCT01584583 ·Status: COMPLETED
-
A Clinical Evaluation of Omron Digital Blood Pressure Monitor
NCT03130686 ·Status: COMPLETED
-
Validation of a Noninvasive Automated Blood Pressure Device
NCT03566888 ·Status: COMPLETED
-
Intelligent Hypertension Intervention Study
NCT05526300 ·Status: COMPLETED ·Phase: NA
-
Blood Pressure Monitor Clinical Test (Cuff Range: 13.5 Cm-22cm)
NCT01594307 ·Status: COMPLETED
-
Etiological Classification-guided Individual Intervention in Primary Hypertension
NCT06941935 ·Status: NOT_YET_RECRUITING ·Phase: NA
-
Evaluation of Continuous Non Invasive Blood Pressure With Adult Volunteers
NCT04876664 ·Status: UNKNOWN ·Phase: NA
-
Primary Technology Enhanced Care in Hypertension
NCT03698890 ·Status: UNKNOWN ·Phase: NA
-
Clinical Test for Armed-used Blood Pressure Monitor With Preformed Cuff
NCT01001455 ·Status: COMPLETED