Data Collection for Validation of an NIBP Algorithm in Pregnant Women
NCT05789849 · Status: UNKNOWN · Phase: NA · Type: INTERVENTIONAL · Enrollment: 170
Last updated 2023-03-29
Summary
The purpose of the study is to collect per subject blood pressure data on the device's intended population as described in the International Organization for Standardization (ISO) Standard 81060-2:2018+A1:2020 Non-invasive sphygmomanometers.
Conditions
- Blood Pressure
Interventions
- DEVICE
-
Collecting blood pressure measurements with noninvasive blood pressure device
Noninvasive blood pressure measurements will collected based on the recommendations of the International Organization for Standardization (ISO) 81060-2:2018+A1:2020 Non-invasive sphygmomanometers -Clinical investigation of intermittent automated measurement type. This study has only 1 arm; investigational NIBP measurements will be collected using a NIBP Auscultatory Algorithm with alternating NIBP measurements collected using a reference sphygmomanometer.
Sponsors & Collaborators
-
GE Healthcare
lead INDUSTRY
Principal Investigators
-
Nandini Raghuraman, MD · Division of Maternal-Fetal Medicine Washington University in St Louis School of Medicine
-
Adithya Bhat, MD · Washington University School of Medicine
Study Design
- Allocation
- NA
- Purpose
- OTHER
- Masking
- NONE
- Model
- SINGLE_GROUP
Eligibility
- Min Age
- 18 Years
- Sex
- FEMALE
- Healthy Volunteers
- No
Timeline & Regulatory
- Start
- 2022-09-14
- Primary Completion
- 2023-09-30
- Completion
- 2023-09-30
- FDA Device
- Yes
Countries
- United States
Study Locations
More Related Trials
-
Non-invasive Validation of Non-invasive Central Blood Pressure Measurements Using Oscillometric Pulse Wave Analysis
NCT03049709 ·Status: COMPLETED
-
Larsen & Toubro (L&T) Noninvasive Blood Pressure Module Performance Validation
NCT01097603 ·Status: COMPLETED
-
Comparison of Measurement of Arterial Blood Pressure by L&T NIBP Module With Invasive Arterial Blood Pressure in Neonates
NCT01196806 ·Status: COMPLETED
-
Validation of a Noninvasive Automated Blood Pressure Device
NCT03566888 ·Status: COMPLETED
-
Somnotouch-NIBP Compared to Standard Ambulatory 24 Hours Blood Pressure Measurement
NCT03054688 ·Status: COMPLETED ·Phase: NA
-
NIBP (Non-invasive Blood Pressure) Algorithm Enhancements (Characterize the Robustness of Motion Artifact Suppression)
NCT05185037 ·Status: TERMINATED
-
Fresenius Noninvasive Blood Pressure Validation
NCT04573595 ·Status: COMPLETED ·Phase: NA
-
Non-Invasive Blood Pressure Monitor Accuracy in Neonatal Subjects
NCT02642198 ·Status: UNKNOWN
-
Guiding Hypertension Management Using Different Blood Pressure Monitoring Strategies
NCT03578848 ·Status: UNKNOWN ·Phase: NA
-
Virtual Cardiac Wellness Program Following Hypertensive Disorders of Pregnancy
NCT04998942 ·Status: UNKNOWN ·Phase: NA
-
Blood Pressure Assessment in Waiting Room During Pregnancy
NCT06248463 ·Status: NOT_YET_RECRUITING
-
Vital Signs Blood Pressure Trial
NCT06272318 ·Status: NOT_YET_RECRUITING
-
Non-invasive Blood Pressure (BP) Monitoring Compared to A-Line BP Monitoring
NCT00739700 ·Status: COMPLETED
-
Home Assessment of Blood Pressure in PregnancY
NCT06491199 ·Status: NOT_YET_RECRUITING
-
Somnotouch-NIBP Compared to Standard Ambulatory 24-hours Blood Pressure Measurement
NCT03975582 ·Status: COMPLETED
-
App-based Remote Blood Pressure Monitoring
NCT05595629 ·Status: COMPLETED ·Phase: NA
-
Mercury Versus Automated Blood Pressure Monitoring
NCT01753895 ·Status: COMPLETED
-
Accuracy Testing of Validated and Non-validated Home BP Devices Sold on the Online Market.
NCT06688409 ·Status: RECRUITING ·Phase: NA
-
CONNEQT Pulse Validation Study
NCT06002971 ·Status: TERMINATED
-
NIBP Cuff Comparison Study
NCT07024472 ·Status: COMPLETED
-
Wrist-based Blood Pressure Monitoring Study
NCT03661112 ·Status: COMPLETED
-
Clinical Study About Blood Pressure Measurement Efficacy on Multifunction KEITO
NCT01359748 ·Status: COMPLETED ·Phase: NA
-
Invasive Validation of Non-invasive Central Blood Pressure Measurements Using Oscillometric Pulse Wave Analysis
NCT03046264 ·Status: COMPLETED
-
Autonomic Regulation of Blood Pressure in Premature and Early Menopausal Women
NCT04439370 ·Status: RECRUITING
-
Development and Validation of a Short Tool to Assess the Awareness of Pregnancy-induced Hypertension
NCT05859282 ·Status: COMPLETED