A Clinical Trial to Verify the Accuracy of Seers Technology's Pulse Oximeter (mobiCARE +Pulse, MP100W) in Healthy Volunteers
NCT05895682 · Status: COMPLETED · Phase: NA · Type: INTERVENTIONAL · Enrollment: 12
Last updated 2023-06-08
Summary
It is important to evaluate the accuracy of the pulse oximeter, which is commonly used in clinical practice. The US FDA only allows the use of devices that meet the ISO 80601-2-61:2017 standard in the United States. According to the ISO 80601-2-61:2017 standard, the accuracy of the oxygen saturation (SpO2) of the pulse oximeter should have an error range of less than 4.0% compared to the actual arterial blood oxygen saturation (SaO2) in the range of 70-100%. To prove compliance with this requirement, it is necessary to conduct a clinical trial that induces hypoxemia in healthy adults and compares and evaluates the arterial blood oxygen saturation values of the pulse oximeter's SpO2 and the arterial blood oxygen saturation values of the carbon monoxide-oxygen meter (CO-oximeter).
Conditions
- Oxygen Saturation
Interventions
- DEVICE
-
mobiCARE +Pulse (MP100W)
Volunteers were monitored using end-tidal carbon dioxide partial pressure and fraction of inspired oxygen. Each volunteer was placed in a semi-Fowler's position and connected to a breathing circuit to administer the nitrogen-air-carbon dioxide mixtures. A nose clip was applied to prevent breathing of room air. For frequent blood sampling, an arterial cannula was placed in the radial artery of each volunteer. Pulse oximeter probes were simultaneously attached to each volunteer's fingers. Each volunteer was exposed to various levels of induced hypoxia from 70\~100% of SaO2. Each plateau of oxygen saturation was maintained for at least 30 s until stabilization, after which 1 ml of arterial blood was drawn into a heparinized syringe. The study period consisted of two rounds of hypoxia, and the volunteers were maintained on room air between each round. SaO2 measurements using a CO-oximeter were used as a reference for the SpO2 accuracy.
Sponsors & Collaborators
-
Seers Technology Co., Ltd.
lead INDUSTRY
Study Design
- Allocation
- NA
- Purpose
- OTHER
- Masking
- NONE
- Model
- SINGLE_GROUP
Eligibility
- Min Age
- 20 Years
- Max Age
- 50 Years
- Sex
- ALL
- Healthy Volunteers
- Yes
Timeline & Regulatory
- Start
- 2023-04-29
- Primary Completion
- 2023-05-14
- Completion
- 2023-05-15
Countries
- South Korea
Study Locations
More Related Trials
-
Postmarket Clinical Follow-up Study to Confirm the Performance of the Wearable Pulse-Oximeter BrOxy M (SOMBRERO)
NCT04810221 ·Status: COMPLETED ·Phase: NA
-
Clinical Trial Testing the BioSerenity Pulse Oximeter (EOXY) Efficacy, Performance and Safety in Healthy Volunteer
NCT03614416 ·Status: COMPLETED ·Phase: NA
-
Comparison of SpO2 Measurement Accuracy of Different Types of Smartwatches
NCT05789563 ·Status: COMPLETED ·Phase: NA
-
Pulse Oximetry- Performance During Severe Signal Interference
NCT01720355 ·Status: COMPLETED
-
RDS MultiSense® SpO2 Validation Study
NCT05359328 ·Status: COMPLETED ·Phase: NA
-
Philps FAST2022 picoSAT Pulse Oximetry Desaturation Study
NCT05683366 ·Status: COMPLETED
-
Vital USA SpO2 Accuracy Comparison to Arterial Blood CO-Oximetry
NCT04285411 ·Status: COMPLETED ·Phase: NA
-
Accuracy of Smartwatches in Measuring Oxygen Levels in Patients With Pulmonary Hypertension: A Pilot Study
NCT07311135 ·Status: RECRUITING
-
Feasibility of a New Ambulatory Multi-vital Signs Monitor
NCT05562011 ·Status: TERMINATED ·Phase: NA
-
Validation Study of Selfie Technology for Assessment of Vital Signs
NCT04688112 ·Status: UNKNOWN
-
Performance Comparison Between Masimo W1™ and Apple Watch Series 8
NCT05968079 ·Status: COMPLETED
-
Validity and Reliability of a Clinical Method for Assessment of Cardiorespiratory Fitness Using Seismocardiography
NCT05356871 ·Status: COMPLETED
-
Desaturation Validation of INVSENSOR00038
NCT04017611 ·Status: COMPLETED ·Phase: NA
-
Non-contact Blood Pressure and SpO2 Measurement Based on Image Sensor
NCT06204822 ·Status: RECRUITING
-
Clinical Performance of Masimo INVSENSOR00057 for Heart Rate Measurements
NCT05462886 ·Status: COMPLETED ·Phase: NA
-
EmbracePlus SpO2 Accuracy Validation
NCT04964609 ·Status: COMPLETED ·Phase: NA
-
SpO2 Validation Study - Philips FAST 2024
NCT06484452 ·Status: COMPLETED
-
Comparison of Tissue Oxygenation Measurement Using Multimodal Devices
NCT05784103 ·Status: COMPLETED ·Phase: NA
-
Pulse Oximetry_Performance Testing- PCBA-1 (Printed Circuit Board Assembly)
NCT01747460 ·Status: COMPLETED
-
Validity of Measuring Preoperative Fitness Using Seismofit
NCT05943730 ·Status: UNKNOWN ·Phase: NA
-
Diagnostic Study of Huami Smart Wearable Device
NCT04429906 ·Status: UNKNOWN
-
Estimating Peak Oxygen Uptake in People Living With Coronary Heart Disease
NCT05505344 ·Status: COMPLETED
-
Observational, Non-Interventional Study Supporting Validation of VO2Max Estimation Methods Using Results in Patients Receiving Standard of Care Cardiopulmonary Exercise Tests (CPET)
NCT05678530 ·Status: RECRUITING
-
Home Pulse Oximeter Use in Patients With COVID-19
NCT04373161 ·Status: COMPLETED ·Phase: NA
-
Contactless Vital Signs Measurement Study (CVMS)
NCT05886686 ·Status: UNKNOWN