Desaturation Validation of INVSENSOR00038

NCT04017611 · Status: COMPLETED · Phase: NA · Type: INTERVENTIONAL · Enrollment: 13

Last updated 2020-10-20

Study results available
· View outcomes & findings →

Summary

This study is designed to evaluate the trending accuracy of a noninvasive regional oximetry measurement of somatic oxygen saturation. One investigational sensor (INVSENSOR00038) will be placed on the somatic site of the volunteer. The values obtained by the test sensor will be compared to the reference value obtained from an FDA-cleared sensor placed on the somatic site on the contralateral side. Data will be collected from healthy adult subjects while undergoing a desaturation procedure wherein the concentration of oxygen inhaled is slowly reduced until the subject's oxygen saturation is approximately 70%. The subject will then be returned to inhaling room air.

Conditions

  • Healthy

Interventions

DEVICE

INVSENSOR00038

Noninvasive regional oximeter

Sponsors & Collaborators

  • Masimo Corporation

    lead INDUSTRY

Study Design

Allocation
NA
Purpose
OTHER
Masking
NONE
Model
SINGLE_GROUP

Eligibility

Min Age
18 Years
Max Age
50 Years
Sex
ALL
Healthy Volunteers
Yes

Timeline & Regulatory

Start
2019-07-08
Primary Completion
2019-07-19
Completion
2019-07-19
FDA Device
Yes

Countries

  • United States

Study Locations

More Related Trials

Read the full study record

This page highlights key information. For complete eligibility criteria, study locations, investigator contacts, and the full protocol, visit the original record on ClinicalTrials.gov.