Validation of a Vital Signs Monitoring Wristband
NCT05566886 · Status: COMPLETED · Type: OBSERVATIONAL · Enrollment: 100
Last updated 2025-05-04
Summary
Today, continuous monitoring of vital signs remains a challenge since it generally requires the patient to be connected to multiple wired sensors, which restricts patient mobility in the intra-mural setting and complicates home monitoring in the extra-mural setting. Wearable devices on the wrist, although emerging, are often not clinically validated or limited to the monitoring of one or two vital signs.
This study aims to validate the Corsano CardioWatch 287-2 for the continuous monitoring of heart rate at ≤ 4 bpm root mean squared error (RMSE); interbeat intervals at ≤ 50 ms RMSE; breathing rate at ≤ 2 brpm RMSE; and peripheral oxygen saturation at ≤ 3 percentage point RMSE. Also, this study aims to validate the Corsano CardioWatch 287-2 for the measurement of non-invasive blood pressure according to ISO 81060-2:2018.
Conditions
Interventions
- DEVICE
-
Continuous measurement of photoplethysmography at the wrist
Patients receive a wristband that uses multi-color LEDs to measure pulsations at the wrist.
Sponsors & Collaborators
-
Corsano Health B.V.
lead INDUSTRY
Eligibility
- Sex
- ALL
- Healthy Volunteers
- No
Timeline & Regulatory
- Start
- 2022-10-19
- Primary Completion
- 2023-02-07
- Completion
- 2023-02-07
- FDA Device
- Yes
Countries
- Netherlands
Study Locations
More Related Trials
-
Clinical Application Study of Wearable Multi-channel Electrocardiogram Acquisition Equipment in the Diagnosis of Myocardial Ischemia in Patients with Suspected CHD
NCT06868940 ·Status: NOT_YET_RECRUITING
-
Remote Blood Pressure Monitoring With the Wearable SENBIOSYS Photoplethysmographic Device
NCT04379986 ·Status: COMPLETED ·Phase: NA
-
CardioSenseSystem Compared Study Regarding Efficacy and Safety in the Monitoring of ECG
NCT03610529 ·Status: UNKNOWN ·Phase: NA
-
Clinical Data Validation of Novel Wireless Cortrium C3 Monitor for Continuous Electrocardiogram Measurements
NCT02714907 ·Status: COMPLETED ·Phase: NA
-
Contactless Vital Signs Measurement Study (CVMS)
NCT05886686 ·Status: UNKNOWN
-
Necklace-Shaped Sensor for Non-Invasive Remote Monitoring of Vitals
NCT02337790 ·Status: COMPLETED
-
Validation Study in Patients of a Non-invasive Blood Pressure Monitor for Perioperative Use
NCT03795831 ·Status: UNKNOWN
-
Relevance of Recording an ECG Trace With a Connected Watch
NCT06272396 ·Status: RECRUITING ·Phase: NA
-
Atrial Fibrillation Detection Using Garmin Wearable Technology
NCT03566836 ·Status: UNKNOWN
-
Measuring Accuracy of Wristband for Registration of ECG RR-Intervals for One Hand, SenceBand in Comparison to Holter
NCT03747861 ·Status: COMPLETED ·Phase: NA
-
Heartbeat Sensitivity Calibration Using the ECG Accessory
NCT01268007 ·Status: COMPLETED
-
Suitability of Unobtrusive Bed Monitoring and Wrist-worn Heart Rate Monitor for Patient Monitoring
NCT03572751 ·Status: COMPLETED
-
Automated Diagnostic Performance of Smartwatch ECG for Arrhythmia Detection Using the PulseAI Neural Network.
NCT05668377 ·Status: UNKNOWN
-
Advanced Wireless, Wearable Sensors for Continuous Hemodynamic Monitoring
NCT04362553 ·Status: COMPLETED
-
Development and Validation of a Telemonitoring System for High-risk Cardiovascular Patients
NCT05485532 ·Status: ACTIVE_NOT_RECRUITING
-
Validation of a Sensor for Non-Invasive Measurements
NCT03463148 ·Status: COMPLETED
-
Comparison of the Non-invasive Biobeat Device With an Invasive Arterial Line
NCT03603860 ·Status: COMPLETED
-
Effects of Remote Patient Monitoring on Heart Failure Management
NCT00778986 ·Status: COMPLETED ·Phase: NA
-
Optimizing Radiation Dose and Utilizing Wearable Devices to Reduce Arrhythmia Risk in Patients Undergoing Thoracic Radiotherapy: A Prospective Cohort Study
NCT06897189 ·Status: NOT_YET_RECRUITING
-
Heart at Home- a Self-care Study
NCT01759368 ·Status: COMPLETED ·Phase: NA
-
Feasibility of a New Ambulatory Multi-vital Signs Monitor
NCT05562011 ·Status: TERMINATED ·Phase: NA
-
Validity and Reliability of Optical Heart Rate Sensors
NCT05525000 ·Status: COMPLETED ·Phase: NA
-
Clinical Outcomes After Implementation of COntinuous Vital Sign Monitoring On the General Ward.
NCT04189653 ·Status: COMPLETED
-
Contactless Measurement of Heart Rate, Heart Rate Variability, Breathing Rate and Blood Pressure Using Remote Photoplethysmography
NCT06829615 ·Status: COMPLETED ·Phase: NA
-
A Study on the Effectiveness of the Application of an Artificial Intelligence Algorithm for Calibrating PPG With ECG to Improve the Accuracy of Atrial Fibrillation Burden Estimation
NCT06552468 ·Status: COMPLETED