Capillary Blood Volume and Pain Perception Obtained in a Process of Puncturing With Different Safety Lancets

NCT03752229 · Status: COMPLETED · Phase: NA · Type: INTERVENTIONAL · Enrollment: 300

Last updated 2018-11-26

No results posted yet for this study

Summary

The study was a randomized, single blind, uni-center pivotal study in male and female healthy volunteers, to determine capillary blood volume and pain perception obtained in a process of puncturing with different safety lancets.

A group of three hundred (300) Caucasian male and female Subjects was randomized (aged \> 18 to \< 65 years old) who were in good physical and mental condition. The Subjects had 4 (four) fingers (two fingers from each hand) lanced each with a different type of safety lancet. Overall, 4 (four) different safety lancets were investigated per each Subject.

Overall, 48 (forty eight) different safety lancets (4 different safety lancets per one Subject) were investigated, the capillary blood volume and pain were measured.

Conditions

  • Healthy

Interventions

DEVICE

Acti-Lance Lite

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICE

Acti-Lance Universal

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICE

Acti-Lance Special

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICE

Medlance Plus Super Lite

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICE

Medlance Plus Lite

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICE

Medlance Plus Universal

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICE

Medlance Plus Extra

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICE

Medlance Plus Special

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICE

Prolance Micro Flow

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICE

Prolance Low Flow

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICE

Prolance Normal Flow

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICE

Prolance High Flow

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICE

Prolance Max Flow

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICE

Prolance Pediatric

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICE

Haemolance Micro Flow

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICE

Haemolance Low Flow

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICE

Haemolnace Normal Flow

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICE

Haemolance High Flow

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICE

Haemolance Max Flow

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICE

Haemolance Pediatric

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICE

MediSafe Solo 29 Gauge

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICE

MediSafe Solo 23 Gauge

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICE

ergoLance Micro Flow

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICE

ergoLance Normal Flow

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICE

ergoLance High Flow

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICE

Medlance (light blue)

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICE

Medlance (blue)

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICE

Medlance (dark blue)

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICE

Medlance (yellow)

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICE

Medlance (orange)

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICE

Medlance (red)

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICE

Unistik 3 Extra

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICE

Unistik 3 Comfort

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICE

Unistik Touch

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICE

ReadyLance (blue)

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICE

ReadyLance (orange)

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICE

AssureLance

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICE

AssureLance MicroFlow

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICE

Surgilance

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICE

Surgilance Lite (grey)

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICE

Surgilance Lite (purple)

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICE

SteriLance Lite2

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICE

SensiLance (pink)

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICE

SteriLance Press

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICE

SensiLance (orange)

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICE

mylife

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICE

VeriFine

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICE

Microdot

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Sponsors & Collaborators

  • HTL-Strefa S.A.

    lead INDUSTRY

Study Design

Allocation
RANDOMIZED
Purpose
OTHER
Masking
SINGLE
Model
PARALLEL

Eligibility

Min Age
18 Years
Max Age
65 Years
Sex
ALL
Healthy Volunteers
Yes

Timeline & Regulatory

Start
2017-12-08
Primary Completion
2017-12-17
Completion
2018-01-30

Countries

  • Poland

Study Locations