Capillary Blood Volume and Pain Perception Obtained in a Process of Puncturing With Different Safety Lancets
NCT03752229 · Status: COMPLETED · Phase: NA · Type: INTERVENTIONAL · Enrollment: 300
Last updated 2018-11-26
Summary
The study was a randomized, single blind, uni-center pivotal study in male and female healthy volunteers, to determine capillary blood volume and pain perception obtained in a process of puncturing with different safety lancets.
A group of three hundred (300) Caucasian male and female Subjects was randomized (aged \> 18 to \< 65 years old) who were in good physical and mental condition. The Subjects had 4 (four) fingers (two fingers from each hand) lanced each with a different type of safety lancet. Overall, 4 (four) different safety lancets were investigated per each Subject.
Overall, 48 (forty eight) different safety lancets (4 different safety lancets per one Subject) were investigated, the capillary blood volume and pain were measured.
Conditions
- Healthy
Interventions
- DEVICE
-
Acti-Lance Lite
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
- DEVICE
-
Acti-Lance Universal
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
- DEVICE
-
Acti-Lance Special
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
- DEVICE
-
Medlance Plus Super Lite
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
- DEVICE
-
Medlance Plus Lite
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
- DEVICE
-
Medlance Plus Universal
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
- DEVICE
-
Medlance Plus Extra
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
- DEVICE
-
Medlance Plus Special
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
- DEVICE
-
Prolance Micro Flow
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
- DEVICE
-
Prolance Low Flow
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
- DEVICE
-
Prolance Normal Flow
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
- DEVICE
-
Prolance High Flow
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
- DEVICE
-
Prolance Max Flow
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
- DEVICE
-
Prolance Pediatric
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
- DEVICE
-
Haemolance Micro Flow
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
- DEVICE
-
Haemolance Low Flow
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
- DEVICE
-
Haemolnace Normal Flow
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
- DEVICE
-
Haemolance High Flow
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
- DEVICE
-
Haemolance Max Flow
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
- DEVICE
-
Haemolance Pediatric
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
- DEVICE
-
MediSafe Solo 29 Gauge
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
- DEVICE
-
MediSafe Solo 23 Gauge
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
- DEVICE
-
ergoLance Micro Flow
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
- DEVICE
-
ergoLance Normal Flow
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
- DEVICE
-
ergoLance High Flow
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
- DEVICE
-
Medlance (light blue)
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
- DEVICE
-
Medlance (blue)
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
- DEVICE
-
Medlance (dark blue)
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
- DEVICE
-
Medlance (yellow)
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
- DEVICE
-
Medlance (orange)
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
- DEVICE
-
Medlance (red)
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
- DEVICE
-
Unistik 3 Extra
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
- DEVICE
-
Unistik 3 Comfort
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
- DEVICE
-
Unistik Touch
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
- DEVICE
-
ReadyLance (blue)
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
- DEVICE
-
ReadyLance (orange)
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
- DEVICE
-
AssureLance
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
- DEVICE
-
AssureLance MicroFlow
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
- DEVICE
-
Surgilance
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
- DEVICE
-
Surgilance Lite (grey)
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
- DEVICE
-
Surgilance Lite (purple)
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
- DEVICE
-
SteriLance Lite2
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
- DEVICE
-
SensiLance (pink)
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
- DEVICE
-
SteriLance Press
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
- DEVICE
-
SensiLance (orange)
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
- DEVICE
-
mylife
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
- DEVICE
-
VeriFine
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
- DEVICE
-
Microdot
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
Sponsors & Collaborators
-
HTL-Strefa S.A.
lead INDUSTRY
Study Design
- Allocation
- RANDOMIZED
- Purpose
- OTHER
- Masking
- SINGLE
- Model
- PARALLEL
Eligibility
- Min Age
- 18 Years
- Max Age
- 65 Years
- Sex
- ALL
- Healthy Volunteers
- Yes
Timeline & Regulatory
- Start
- 2017-12-08
- Primary Completion
- 2017-12-17
- Completion
- 2018-01-30
Countries
- Poland
Study Locations
More Related Trials
-
The Effect of Mechanical Vibration and ShotBlocker on Pain Levels During Heel Lance in Healthy Term Neonates
NCT06380556 ·Status: COMPLETED ·Phase: NA
-
Effects of Vibration, Heel Warming and Shotblocker on Pain Levels During Heel Blood Collection
NCT06646848 ·Status: COMPLETED ·Phase: NA
-
Vibration Device Effect on Post-Injection Pain in the Upper Extremity
NCT05651139 ·Status: COMPLETED ·Phase: NA
-
Ultrasound Needle Transducer for Regional Anesthesia Validation Study
NCT04726930 ·Status: RECRUITING
-
Variable Approaches of Intradiscal O3-O2 Injection
NCT03038529 ·Status: COMPLETED ·Phase: NA
-
Effect of Patterned Versus Plain Fixation on Pain During IV Therapy in Children
NCT07316036 ·Status: ACTIVE_NOT_RECRUITING ·Phase: NA
-
The Effect of Shotblocker on Pain Violence and Injection Satisfaction
NCT05647239 ·Status: COMPLETED ·Phase: NA
-
IM İnjection-PAIN-Cold Spray and Stress Ball
NCT07233694 ·Status: NOT_YET_RECRUITING ·Phase: NA
-
ShotBlocker, Cold Needling, and Both Methods in Combination Pain and Satisfaction
NCT07258953 ·Status: COMPLETED ·Phase: NA
-
The Effect of Vibration on Pain During Subcutaneous Heparin Injection
NCT06469112 ·Status: COMPLETED ·Phase: NA
-
The Antalgic Effect of Pulse Frequency and Pad Size of TENS on Blunt Pressure Pain
NCT01999595 ·Status: COMPLETED ·Phase: NA
-
Effect of TENS Application After Thoracic Surgery
NCT05927168 ·Status: UNKNOWN ·Phase: NA
-
Transcutaneous vs Percutaneous Electrical Stimulation of the Radial Nerve
NCT07247721 ·Status: COMPLETED ·Phase: NA
-
Syringe Pressure Analysis: A Critical Examination of the Operator's Burden
NCT06476353 ·Status: COMPLETED ·Phase: NA
-
Effects of Electrotherapy on Pain, Anxiety, Motion and Disability in Adults With Neck or Upper Body Pain.
NCT05382039 ·Status: COMPLETED ·Phase: NA
-
Burst Crossover Trial
NCT05372822 ·Status: RECRUITING ·Phase: NA
-
Effects of Transcutaneous Electrical Nerve Stimulation for Postoperative Quality of Recovery After Shoulder Surgery
NCT05833113 ·Status: COMPLETED ·Phase: NA
-
Does Trans-Cutaneous Electrical Nerve Stimulation (TENS) Alleviate The Pain Experienced During Bone Marrow Sampling?
NCT02005354 ·Status: COMPLETED ·Phase: NA
-
Effectiveness of Mechano-Analgesia and Cold Application on Ecchymosis, Pain and Satisfaction at Subcutaneous Heparin Injection
NCT03645369 ·Status: COMPLETED ·Phase: NA
-
Effects of Shotblocker and Manual Pressure on Pain and Satisfaction
NCT06947525 ·Status: NOT_YET_RECRUITING ·Phase: NA
-
Effect of Cold Application to Different Side Neck Region on Venous Cannulation Pain
NCT06253299 ·Status: COMPLETED
-
Which Technique is Effective in Reducing the Pain of Peripheral Intravenous Catheterization in Pediatric Patients, Infra-Red Light or Buzzy®?
NCT05523518 ·Status: UNKNOWN ·Phase: NA
-
Normative Values Laser Nociceptive Evoked Potentials Depending on the Age
NCT02531048 ·Status: COMPLETED ·Phase: NA
-
Ultrasound-Guided Technique for Thoracic Epidural Insertion
NCT01449214 ·Status: COMPLETED ·Phase: NA
-
TENS Unit To Decrease Pain After Laminaria Insertion
NCT06529003 ·Status: COMPLETED ·Phase: NA