3D MR Versus 3D CT for Glenohumeral Arthritis

NCT03379545 · Status: COMPLETED · Phase: NA · Type: INTERVENTIONAL · Enrollment: 31

Last updated 2021-03-25

Study results available
· View outcomes & findings →

Summary

This study aims to assess the efficacy of three-dimensional magnetic resonance (MR) compared to three-dimensional computed tomography (CT) imaging. This proposed study is a prospective, unrandomized, single-blinded, self-controlled, and single-armed diagnostic radiological evaluation study.

Conditions

Interventions

DIAGNOSTIC_TEST

3-dimensional (3D) magnetic resonance (MR) imaging

MRI is performed using 3T scanners with a dedicated 16-channel shoulder array coils. The MRI sequences include 3-mm slice thickness and 0.5-mm gap width with a field of view of 14 or 15 cm. There were 6 diagnostic sequences with axial, coronal, and sagittal proton density weighting as well as coronal T2 with frequency selective fat suppression and sagittal T1 images.

DIAGNOSTIC_TEST

3D computed tomography (CT) imaging

The CT protocol consists of 3-mm axial images of the glenoid reconstructed into 1-mm sagittal and coronal 2D reconstructions using the following parameters: 120 kV, 280 mA, and pitch of 0.9. The CT data were also used to produce a 3D reconstruction of each glenoid.

Sponsors & Collaborators

Principal Investigators

  • Joseph D Zuckerman, MD · NYU Langone Health

Study Design

Allocation
NA
Purpose
DIAGNOSTIC
Masking
NONE
Model
SINGLE_GROUP

Eligibility

Min Age
18 Years
Sex
ALL
Healthy Volunteers
No

Timeline & Regulatory

Start
2017-01-11
Primary Completion
2018-12-21
Completion
2020-06-05
FDA Device
Yes

Countries

  • United States

Study Locations

More Related Trials

Entities

Read the full study record

This page highlights key information. For complete eligibility criteria, study locations, investigator contacts, and the full protocol, visit the original record on ClinicalTrials.gov.