Accuracy and Safety of Anytime 5Pro and Anytime 4Pro CGM Systems in Adults With Diabetes
NCT07645313 · Status: NOT_YET_RECRUITING · Phase: NA · Type: INTERVENTIONAL · Enrollment: 70
Last updated 2026-06-12
Summary
This post-market clinical follow-up (PMCF) investigation evaluates the accuracy and safety of two CE-marked continuous glucose monitoring (CGM) systems, the Anytime 5Pro and the Anytime 4Pro, in adults with type 1 or type 2 diabetes. Each participant simultaneously wears one Anytime 5Pro sensor and one Anytime 4Pro sensor, one on each upper arm, for the full labelled wear period of each device (up to 16 and 15 days, respectively). CGM readings are compared against venous blood glucose measured by a laboratory reference analyzer (YSI 2500) during four in-clinic sampling sessions that include controlled glucose manipulation. Participants are masked to real-time CGM readings throughout the study. The investigation is conducted to confirm device performance and safety under real-world conditions of use in a European population, in line with the manufacturer's post-market clinical follow-up obligations under the EU Medical Device Regulation.
Conditions
- Diabetes Mellitus, Type 1 | Diabetes Mellitus, Type 2
Interventions
- DEVICE
-
Reference venous blood sampling (YSI 2500)
During the four in-clinic reference sampling sessions, participants undergo controlled manipulation of blood glucose to obtain paired CGM-reference measurements spanning a wide glucose range, including hypoglycaemic and hyperglycaemic levels. Glucose is raised and/or lowered under continuous medical supervision using standard clinical means (e.g., carbohydrate intake and/or insulin administration) according to the protocol, with predefined safety limits and stopping rules. Each session lasts up to 9 hours, with venous blood sampled every 15 (+/-3) minutes via an indwelling catheter for reference glucose measurement on the YSI 2500 analyzer. The purpose is to ensure adequate representation of low, mid-range, and high glucose values for the accuracy endpoints (MARD, agreement rates, and Consensus Error Grid analysis), rather than to treat the participant's diabetes.
Sponsors & Collaborators
-
Jiangsu Yuwell POCTech Biotechnology Co., Ltd..
collaborator UNKNOWN -
Yuwell Group
lead INDUSTRY
Study Design
- Allocation
- NA
- Purpose
- DIAGNOSTIC
- Masking
- NONE
- Model
- SINGLE_GROUP
Eligibility
- Min Age
- 18 Years
- Sex
- ALL
- Healthy Volunteers
- No
Timeline & Regulatory
- Start
- 2026-07-20
- Primary Completion
- 2027-01-31
- Completion
- 2027-01-31
More Related Trials
-
Effectiveness and Safety Study of the DexCom™ G4 Continuous Glucose Monitoring System
NCT01111370 ·Status: COMPLETED ·Phase: NA
-
Continuous Glucose Monitoring in Hospitalized Patients With Diabetes Mellitus
NCT04653454 ·Status: COMPLETED ·Phase: NA
-
System Accuracy Evaluation of 2 CE-marked Blood Glucose Monitoring System
NCT02710617 ·Status: COMPLETED ·Phase: NA
-
CGM Use in COVID19 Patients (Infection With Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2))
NCT04756141 ·Status: COMPLETED ·Phase: NA
-
To Evaluate the Efficacy and Safety of Multi-parameter Non-Invasive Glucose Sensors for for Blood Glucose Measurement
NCT06017349 ·Status: COMPLETED ·Phase: NA
-
Standardized Procedure for the Assessment of New-to-market Continuous Glucose Monitoring Systems
NCT01751932 ·Status: COMPLETED ·Phase: NA
-
Continuous Monitoring and Control of Hypoglycemia
NCT03340831 ·Status: UNKNOWN
-
Safety and Effectiveness of A Novel Continuous Glucose and Ketone Monitoring System
NCT07139808 ·Status: RECRUITING ·Phase: NA
-
Evaluation of the Safety and Effectiveness of the Dexcom Glucose Monitoring (CGM) System
NCT04645641 ·Status: UNKNOWN ·Phase: NA
-
Continuous Glucose Monitors (CGMs) for All: Studying the Impact of Expanding CGM Access
NCT06658067 ·Status: RECRUITING ·Phase: NA
-
Effectiveness and Safety Study of the Dexcom G4™ Continuous Glucose Monitoring System
NCT01514292 ·Status: COMPLETED ·Phase: NA
-
Evaluate the Safety and Efficacy of Sinocare CGM System Regarding Real Time Glucose Level Monitoring
NCT05995756 ·Status: COMPLETED
-
Effectiveness and Safety of the Dexcom™ G6 Continuous Glucose Monitoring System
NCT02880267 ·Status: COMPLETED ·Phase: NA
-
CGM-Assisted Community Management for Type 2 Diabetes
NCT07097415 ·Status: COMPLETED ·Phase: NA
-
Digital Health Solutions for Patients With Type 2 Diabetes
NCT07642830 ·Status: ENROLLING_BY_INVITATION ·Phase: NA
-
User Performance of the CONTOUR NEXT and CONTOUR TV3 Blood Glucose Monitoring System (BGMS)
NCT03730480 ·Status: COMPLETED ·Phase: NA
-
Assessing Continuous Glucose Monitor Use in Underserved Primary Care Patients
NCT05766488 ·Status: ENROLLING_BY_INVITATION ·Phase: NA
-
Evaluation of Glycemic Control Using Adjunctive Continuous Glucose Monitoring With the Dexcom G4 System
NCT01627210 ·Status: COMPLETED
-
Effect of CGM on Glucose Control in Non-insulin-treated Patients With Type 2 Diabetes Mellitus
NCT06587438 ·Status: ACTIVE_NOT_RECRUITING ·Phase: NA
-
Reliability of the Freestyle Libre CGM in the Inpatient Setting During the COVID-19 Surge
NCT04417270 ·Status: COMPLETED ·Phase: NA
-
Accuracy Comparison of Two CGMs in Hospitalized Patients
NCT05081817 ·Status: COMPLETED
-
Evaluation of the Safety and Effectiveness of the Dexcom Continuous Glucose Monitoring (CGM) System
NCT04794478 ·Status: COMPLETED ·Phase: NA
-
Develop Novel Methods to Display, Report and Analyze CGM Data for Clinical Decision-Making in People With Diabetes
NCT00465881 ·Status: COMPLETED
-
Effectiveness and Safety Study of the Next-Generation DexCom™ SEVEN® Continuous Glucose Monitoring System
NCT00722241 ·Status: COMPLETED
-
Continuous Glucose Monitoring (CGM) in Subjects With Type 2 Diabetes
NCT01341067 ·Status: COMPLETED