Trial Outcomes & Findings for Milky Way Sensor Motion Validation (NCT NCT07228416)
NCT ID: NCT07228416
Last Updated: 2026-08-17
Results Overview
The ivWatch Model 400 issues Check IV notifications to communicate the need for a clinician to check an IV site. A red notification suggests an increased likelihood that an IV infiltration may be occurring, with a greater likelihood relative to a yellow notification. This measure describes the average number of red notifications issued by the ivWatch device in a given day when monitoring normal, non-infiltrated tissues. The 95% confidence interval for the red notification rate was calculated using a negative binomial regression model.
COMPLETED
NA
20 participants
24 Hours
2026-08-17
Participant Flow
Participant milestones
| Measure |
Non-Infiltrated Tissue
The ivWatch Model 400 with Milky Way Sensor monitored a common peripheral IV site over a 24 hour observation period.
|
|---|---|
|
Overall Study
STARTED
|
20
|
|
Overall Study
COMPLETED
|
20
|
|
Overall Study
NOT COMPLETED
|
0
|
Reasons for withdrawal
Withdrawal data not reported
Baseline Characteristics
Race and Ethnicity were not collected from any participant.
Baseline characteristics by cohort
| Measure |
Non-Infiltrated Tissue
n=20 Participants
The ivWatch Model 400 with Milky Way Sensor monitored a common peripheral IV site over a 24 hour observation period.
|
|---|---|
|
Age, Continuous
|
48 years
n=20 Participants
|
|
Sex: Female, Male
Female
|
13 Participants
n=20 Participants
|
|
Sex: Female, Male
Male
|
7 Participants
n=20 Participants
|
PRIMARY outcome
Timeframe: 24 HoursPopulation: Check IV notifications are counted per day for each research subject
The ivWatch Model 400 issues Check IV notifications to communicate the need for a clinician to check an IV site. A red notification suggests an increased likelihood that an IV infiltration may be occurring, with a greater likelihood relative to a yellow notification. This measure describes the average number of red notifications issued by the ivWatch device in a given day when monitoring normal, non-infiltrated tissues. The 95% confidence interval for the red notification rate was calculated using a negative binomial regression model.
Outcome measures
| Measure |
Non-Infiltrated Tissue
n=20 Participants
The ivWatch Model 400 with Milky Way Sensor monitored a common peripheral IV site over a 24 hour observation period.
|
|---|---|
|
Red Check IV Notification Rate Per Day
|
0.208 Check IV notifications per day
Interval 0.078 to 0.553
|
Adverse Events
Non-Infiltrated Tissue
Serious adverse events
Adverse event data not reported
Other adverse events
Adverse event data not reported
Additional Information
Results disclosure agreements
- Principal investigator is a sponsor employee
- Publication restrictions are in place