Trial Outcomes & Findings for VicorderCS NIBP Validation Study to Review The Ability to Capture Blood Pressure Measurements (NCT NCT06560034)

NCT ID: NCT06560034

Last Updated: 2026-06-29

Results Overview

A mean difference was measured between the VicorderCS noninvasive blood pressure measurement method and the auscultation method using a Korotkoff sounds device with human observers' readings.

Recruitment status

COMPLETED

Target enrollment

86 participants

Primary outcome timeframe

1 Day

Results posted on

2026-06-29

Participant Flow

Recruitment will be performed using email and a flier. Text of the recruitment emails will contain the following information: * Brief description of the study. * Importance of the proposed research * The study will be conducted entirely at the aforementioned location and last for approximately 40 minutes

As described further in the following section on recruitment methods, participants will be sent an email providing information on the study and providing steps to enroll. Members who respond to recruitment emails will be directed to a link where they can schedule their study time and sign a consent form. They will have the opportunity to contact study personnel with questions if necessary. During "sign up", participants will be taken through a questionnaire to collect relevant information.

Participant milestones

Participant milestones
Measure
Cross-Sectional
Blood pressure data were obtained from a total of 86 subjects
Overall Study
STARTED
86
Overall Study
COMPLETED
86
Overall Study
NOT COMPLETED
0

Reasons for withdrawal

Withdrawal data not reported

Baseline Characteristics

Race and Ethnicity were not collected from any participant.

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
Cross-Sectional
n=86 Participants
Blood Pressure
Age, Continuous
40.2 Years
STANDARD_DEVIATION 16.23 • n=86 Participants
Sex: Female, Male
Female
44 Participants
n=86 Participants
Sex: Female, Male
Male
42 Participants
n=86 Participants
Region of Enrollment
United States
86 Participants
n=86 Participants

PRIMARY outcome

Timeframe: 1 Day

A mean difference was measured between the VicorderCS noninvasive blood pressure measurement method and the auscultation method using a Korotkoff sounds device with human observers' readings.

Outcome measures

Outcome measures
Measure
Cross-Sectional
n=86 Participants
Blood pressure data were obtained from a total of 86 subjects
Blood Pressure
Systolic Blood Pressure
-4.63 mmHg
Standard Deviation 5.32
Blood Pressure
Diastolic Blood Pressure
2.27 mmHg
Standard Deviation 4.93

Adverse Events

Cross-Sectional Group

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Serious adverse events

Adverse event data not reported

Other adverse events

Adverse event data not reported

Additional Information

Dr. Nitin Vaswani

Manos Health

Phone: 7577263449

Results disclosure agreements

  • Principal investigator is a sponsor employee
  • Publication restrictions are in place