Trial Outcomes & Findings for Clinical Performance of Masimo INVSENSOR00048 (NCT NCT06289556)

NCT ID: NCT06289556

Last Updated: 2026-08-25

Results Overview

The outcome measure is determination of the Masimo INVSENSOR00048 temperature performance by comparing the Clinical Bias and Limits of Agreement (LOA) between the corresponding adjustment mode temperature of the INVSENSOR00048 and reference thermometer in degrees celsius. Clincal Bias is defined as mean difference between output temperatires of a clinical thermometer and a reference clinical thermometer for the intended reference body site with specified limitis of agreement when measured from selected group of subjects . Limits of agreement is defined as the magnitude of a potential disagreement between outputs of two clinical thermometers equal to double the standard deviation of output temperate differences when used on the same human subject .

Recruitment status

TERMINATED

Target enrollment

180 participants

Primary outcome timeframe

Approximately 90 minutes

Results posted on

2026-08-25

Participant Flow

Participant milestones

Participant milestones
Measure
INVSENSOR00048
All subjects are enrolled into this arm and will have temperature measurements obtained. Masimo INVSENSOR00048: Masimo INVSENSOR00048 will be used to collect temperature measurements.
Overall Study
STARTED
180
Overall Study
COMPLETED
109
Overall Study
NOT COMPLETED
71

Reasons for withdrawal

Reasons for withdrawal
Measure
INVSENSOR00048
All subjects are enrolled into this arm and will have temperature measurements obtained. Masimo INVSENSOR00048: Masimo INVSENSOR00048 will be used to collect temperature measurements.
Overall Study
Withdrawal by Subject
3
Overall Study
Data Transfer Failure
12
Overall Study
Met Data Exclusion Criteria
56

Baseline Characteristics

Clinical Performance of Masimo INVSENSOR00048

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
INVSENSOR00048
n=109 Participants
All subjects are enrolled into this arm and will have temperature measurements obtained. Masimo INVSENSOR00048: Masimo INVSENSOR00048 will be used to collect temperature measurements.
Age, Categorical
<=18 years
109 Participants
n=31 Participants
Age, Categorical
Between 18 and 65 years
0 Participants
n=31 Participants
Age, Categorical
>=65 years
0 Participants
n=31 Participants
Sex: Female, Male
Female
71 Participants
n=31 Participants
Sex: Female, Male
Male
38 Participants
n=31 Participants
Race (NIH/OMB)
American Indian or Alaska Native
1 Participants
n=31 Participants
Race (NIH/OMB)
Asian
10 Participants
n=31 Participants
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
2 Participants
n=31 Participants
Race (NIH/OMB)
Black or African American
4 Participants
n=31 Participants
Race (NIH/OMB)
White
66 Participants
n=31 Participants
Race (NIH/OMB)
More than one race
5 Participants
n=31 Participants
Race (NIH/OMB)
Unknown or Not Reported
21 Participants
n=31 Participants

PRIMARY outcome

Timeframe: Approximately 90 minutes

The outcome measure is determination of the Masimo INVSENSOR00048 temperature performance by comparing the Clinical Bias and Limits of Agreement (LOA) between the corresponding adjustment mode temperature of the INVSENSOR00048 and reference thermometer in degrees celsius. Clincal Bias is defined as mean difference between output temperatires of a clinical thermometer and a reference clinical thermometer for the intended reference body site with specified limitis of agreement when measured from selected group of subjects . Limits of agreement is defined as the magnitude of a potential disagreement between outputs of two clinical thermometers equal to double the standard deviation of output temperate differences when used on the same human subject .

Outcome measures

Outcome measures
Measure
INVSENSOR00048
n=109 Participants
All subjects are enrolled into this arm and will have temperature measurements obtained. Masimo INVSENSOR00048: Masimo INVSENSOR00048 will be used to collect temperature measurements.
Clinical Performance of Masimo INVSENSOR00048
Clinical Bias
0.19 Degrees Celsius
Clinical Performance of Masimo INVSENSOR00048
Limits of Agreement
0.76 Degrees Celsius

Adverse Events

INVSENSOR00048

Serious events: 0 serious events
Other events: 1 other events
Deaths: 0 deaths

Serious adverse events

Adverse event data not reported

Other adverse events

Other adverse events
Measure
INVSENSOR00048
n=180 participants at risk
All subjects are enrolled into this arm and will have temperature measurements obtained. Masimo INVSENSOR00048: Masimo INVSENSOR00048 will be used to collect temperature measurements.
Injury, poisoning and procedural complications
Itchiness and redness at device placement site
0.56%
1/180 • 1-5 hours

Additional Information

Chelsea Frank

Masimo Corporation

Phone: 949-297-7000

Results disclosure agreements

  • Principal investigator is a sponsor employee
  • Publication restrictions are in place