Version Testing of EnzySystem Version A for Hemophilia A
NCT06273865 · Status: SUSPENDED · Type: OBSERVATIONAL · Enrollment: 40
Last updated 2025-08-15
Summary
Background of the study:
To measure blood clotting, blood is taken from a vein. This blood is processed in the laboratory and then tested. A new device has been developed that requires only a very small volume of blood (5-10 drops of blood) to perform the laboratory tests. The long-term goal is that this device can be used by a doctor or at home to quickly measure blood clotting. In this research we want to compare this new system with the standard methods - measurement in the laboratory - and evaluate whether the correct value is determined.
Objective of the study:
The primary objective of this study is to demonstrate that the EnzySystem HemA version A can record thrombin generation TG and quantify FVIII activity levels within a time frame of 60 min in fresh whole blood samples of healthy volunteers and patients with hemophilia A in the Enzyre laboratory (for healthy volunteers) and the Radboudumc (for patients with hemophilia A).
Study design:
This is a cross-sectional observational study. All participants are asked to fill a questionnaire prior to blood collection. The blood of healthy volunteers will be collected in an office of Enzyre BV, the blood of patients will be collected in the Radboudumc. Blood collection, by venepuncture, will be conducted by a Radboudumc research nurse or physician of the research team in both locations. In total, four blood tubes with citrate as anticoagulant will be drawn (a total of around 11 mL).
Study population:
The study population consists of 20 healthy volunteers: evenly distributed between male and female; ages spread over the range from 20 to 70 years old; recruited by Enzyre via advertisement. 20 Patients: 5 severe hemophilia A; 5 moderate hemophilia A; 10 mild hemophilia A; recruited from the Hemophilia Treatment Center (HTC) Nijmegen-Eindhoven-Maastricht (NEM) (location Radboudumc).
Primary study parameters/outcome of the study:
Demonstrate that the EnzySystem HemA version A can record TG and quantitative FVIII activity levels within a time frame of 60 min in fresh blood samples of healthy volunteers and patients with hemophilia A.
Secondary study parameters/outcome of the study (if applicable):
Secondary study parameters are composed whether the measured values comply with the desired assay specificity and accuracy.
Outcomes are analysed for equivalence compared to one-stage FVIII assay, FVIII chromogenic assay, thrombin generation via the Nijmegen Hemostasis Assay, and possibly via the Technoclone assay.
Is it possible to measure FVIII activity with the EnzySystem HemA version A in fresh blood samples of healthy volunteers and patients with hemophilia A, compared to the gold standard with;
* Precision in the normal range (60-140%): min. 30%
* Precision in the low range (3-10%): min. 50%
* Limit of Detection range min. 100 % FVIII activity
* Limit of Detection low range min. 3 % FVIII activity Is it possible to measure TG with the EnzySystem HemA version A in fresh blood samples of healthy volunteers and patients with hemophilia A, compared to the gold standard with;
* Precision in the normal range (60-140%) of control samples: min. 30%
* Precision in patient with hemophilia A: min. 50%
* Limit of Detection, high range \> 400 nM thrombin activity
* Limit of Detection measured with Plasma, low range \< 50 nM thrombin
Other study parameters All samples will also be tested for other hemostasis specific parameters as these parameters may affect a proper measurement of both FVIII activity and Thrombin Generation. The following parameters will be measured in plasma obtained from the whole blood vacutainers. Moreover, left over samples (plasma) will eventually be used to develop other coagulation related parameters.
* von Willebrand Factor antigen levels
* von Willebrand Factor ristocetin activity levels
* Prothrombin Fragment 1+2 levels
* ADAMTS13 activity
* FVIII antigen levels
* blood group
Conditions
Interventions
- DIAGNOSTIC_TEST
-
Several assays
Standard blood draw by venipuncture is performed in all study participants to collect blood samples in four 2.7 mL citrated blood tubes. Tube one is a dummy tube and is discarded. Tube two is used for the FVIII activity and thrombin generation assay on the EnzySystem. Tube three is used to perform conventional FVIII activity and thrombin generation assays. Tube four is used for duplicate measurements of the conventional assays and measurement of prothrombin F1+2 antigen levels, ADAMTS13 activity, FVIII antigen, VWF (von Willebrand Factor) antigen and von Willebrand ristocetin cofactor activity levels.
Sponsors & Collaborators
-
Enzyre B.V.
collaborator INDUSTRY -
Radboud University Medical Center
lead OTHER
Principal Investigators
-
Saskia Schols, PhD · Radboud University Medical Center
Eligibility
- Min Age
- 20 Years
- Max Age
- 70 Years
- Sex
- ALL
- Healthy Volunteers
- Yes
Timeline & Regulatory
- Start
- 2024-05-29
- Primary Completion
- 2026-03-01
- Completion
- 2026-05-01
Countries
- Netherlands
Study Locations
More Related Trials
-
Study of Emicizumab Prophylaxis in Participants With Hemophilia A With or Without Inhibitors Undergoing Minor Surgical Procedures
NCT03361137 ·Status: TERMINATED ·Phase: PHASE4
-
Improved Factor VIII Inhibitor Evaluation
NCT02726139 ·Status: WITHDRAWN
-
Hemophilia Inhibitor Previously Untreated Patient Study
NCT01652027 ·Status: COMPLETED
-
TGA (Thrombin Generation Assay) and Prophylaxis in Haemophilia
NCT02781766 ·Status: UNKNOWN ·Phase: NA
-
PK Driven Prophylaxis for Hemophilia A
NCT02634424 ·Status: TERMINATED ·Phase: NA
-
TFPI Levels in Haemophilia A and B Patients
NCT04570696 ·Status: UNKNOWN
-
Hemophilia Inhibitor Genetics Study (HIGS)
NCT00166387 ·Status: COMPLETED
-
Research Study to Look at How Well the Drug Concizumab Works in Your Body if You Have Haemophilia With Inhibitors
NCT04083781 ·Status: ACTIVE_NOT_RECRUITING ·Phase: PHASE3
-
Prevention of Bleeding in Patients With Moderate and Severe Hemophilia A Playing Sports: A Comparison Between Factor VIII and Emicizumab Prophylaxis
NCT05022459 ·Status: RECRUITING
-
Individualizing Hemophilia Prophylaxis Using Thromboelastography
NCT02582060 ·Status: COMPLETED ·Phase: NA
-
The Dynamic Interplay Between Bleeding Phenotype and Baseline Factor Level in Moderate and Mild Hemophilia A and B
NCT03623295 ·Status: COMPLETED
-
A Study Comparing Factor Level and Inhibitor Titer Testing Results Drawn From Central Venous Lines and Venipuncture
NCT02402829 ·Status: TERMINATED
-
Low Dose Emicizumab vs Low Dose Factor VIII in Prophylaxis in Hemophilia A Patients
NCT06938659 ·Status: NOT_YET_RECRUITING ·Phase: PHASE2/PHASE3
-
Inhibitor Development in Patients With Hemophilia A Undergoing Surgery
NCT01571934 ·Status: COMPLETED
-
Global Hemostatic Methods in Hemophilia and Von Willebrand's Disease
NCT02061033 ·Status: UNKNOWN
-
Does the Thrombin Generation Test Performed During the Pharmacokinetic Profile of the Substitutive Factor VIII Bring Benefits to the Personalized Treatment of Pediatric Patients and Adult Hemophilia A Patients Under Prophylaxis ?
NCT02803502 ·Status: UNKNOWN ·Phase: NA
-
Rare Bleeding Disorders in the Netherlands
NCT03347591 ·Status: UNKNOWN
-
Joint Status in Subjects With Severe Hemophilia A in Relation to Different Treatment Regimens
NCT00927667 ·Status: COMPLETED ·Phase: PHASE4
-
Study Evaluating Inhibitor Specificity in Hemophilia A
NCT00151385 ·Status: WITHDRAWN ·Phase: PHASE4
-
A Study to Evaluate the Efficacy, Safety, and Pharmacokinetics of Prophylactic Emicizumab Versus no Prophylaxis in Hemophilia A Participants With Inhibitors
NCT02622321 ·Status: COMPLETED ·Phase: PHASE3
-
In Vitro Correction of Thrombin Generation by Concizumab (Anti-TFPI) for Severe Hemophilia Patients
NCT05617209 ·Status: COMPLETED
-
Research Study to Look at How Well the Drug Concizumab Works in Your Body if You Have Haemophilia Without Inhibitors
NCT04082429 ·Status: ACTIVE_NOT_RECRUITING ·Phase: PHASE3
-
Monitoring of Anti-TFPI in Hemophilia
NCT07101926 ·Status: RECRUITING
-
Assessment of the Risk of Inhibitor Formation in Previously Treated Patients With Severe Hemophilia A
NCT00621673 ·Status: TERMINATED ·Phase: PHASE4
-
Trial of Rituximab Versus Oral Cyclophosphamide to Eradicate or Suppress Autoimmune Anti-Factor VIII Antibodies in Acquired Hemophilia A
NCT00306670 ·Status: TERMINATED ·Phase: PHASE2/PHASE3