Trial Outcomes & Findings for Accuracy of the New Barrett TAL Formula With the Argos Measurements: A Multicenter Prospective Study (NCT NCT06085131)

NCT ID: NCT06085131

Last Updated: 2026-05-27

Results Overview

Recruitment status

COMPLETED

Target enrollment

523 participants

Primary outcome timeframe

2 months postoperative

Results posted on

2026-05-27

Participant Flow

Participant milestones

Participant milestones
Measure
Argos Biometer
Preoperative biometry with the Argos device Argos: Argos Biometer
Overall Study
STARTED
523
Overall Study
COMPLETED
523
Overall Study
NOT COMPLETED
0

Reasons for withdrawal

Withdrawal data not reported

Baseline Characteristics

Race and Ethnicity were not collected from any participant.

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
Argos Biometer
n=523 Participants
Preoperative biometry with the Argos device Argos: Argos Biometer
Age, Continuous
71.6 years
STANDARD_DEVIATION 7.23 • n=523 Participants
Sex: Female, Male
Female
316 Participants
n=523 Participants
Sex: Female, Male
Male
207 Participants
n=523 Participants

PRIMARY outcome

Timeframe: 2 months postoperative

Outcome measures

Outcome measures
Measure
Argos Biometer
n=523 Eyes
Preoperative biometry with the Argos device Argos: Argos Biometer
BUII
Preoperative planning with the Argos biometer and BUII formula
The Percentage of Eyes That Achieve a Postoperative Manifest Spherical Equivalent Refraction ≤ ±0.50 D.
489 Eyes

SECONDARY outcome

Timeframe: 2 months postoperative

Outcome measures

Outcome measures
Measure
Argos Biometer
n=523 Eyes
Preoperative biometry with the Argos device Argos: Argos Biometer
BUII
Preoperative planning with the Argos biometer and BUII formula
The Percentage of Eyes That Achieve a Postoperative Manifest Spherical Equivalent Refraction ≤ ±0.250D, ≤ ±0.75D and ≤ ±1.00D.
≤ ±0.250D
352 Eyes
The Percentage of Eyes That Achieve a Postoperative Manifest Spherical Equivalent Refraction ≤ ±0.250D, ≤ ±0.75D and ≤ ±1.00D.
≤ ±0.750D
518 Eyes
The Percentage of Eyes That Achieve a Postoperative Manifest Spherical Equivalent Refraction ≤ ±0.250D, ≤ ±0.75D and ≤ ±1.00D.
≤ ±1.00D
523 Eyes

SECONDARY outcome

Timeframe: 2 months postoperative

Population: Overall absolute prediction error of two formulas were compared, the BTAL and BUII. The same eyes were used for both formulas. BTAL = Barrett true axial length; BUII = Barrett universal II

Prediction error is the difference in diopters (D) between the prediction made by the BTAL or BUII preoperative planning formulas, and the measured postoperative refraction. A value closer to 0 is better.

Outcome measures

Outcome measures
Measure
Argos Biometer
n=523 Eyes
Preoperative biometry with the Argos device Argos: Argos Biometer
BUII
n=523 Eyes
Preoperative planning with the Argos biometer and BUII formula
Mean Absolute Prediction Error (in Diopters)
0.21 D
Standard Deviation 0.16
0.22 D
Standard Deviation 0.17

SECONDARY outcome

Timeframe: 2 months postoperative

Population: Absolute prediction error of two formulas were compared in the BTAL and BUII, in different length eyes. The same eyes were used in both formulas. BTAL = Barrett true axial length; BUII = Barrett universal II

The Mean Predictive Errors in Short AL\<22.5mm, Long AL \>24.5mm and Medium AL 22.6mm to 24.4mm eyes. Prediction error is the difference in diopters (D) between the prediction made by the BTAL or BUII preoperative planning formulas, and the measured postoperative refraction. A value closer to 0 is better.

Outcome measures

Outcome measures
Measure
Argos Biometer
n=523 Eyes
Preoperative biometry with the Argos device Argos: Argos Biometer
BUII
n=523 Eyes
Preoperative planning with the Argos biometer and BUII formula
Mean Absolute Prediction Error (in Diopters) in Eyes With Short AL<22.5mm, Long AL >24.5mm and Medium AL 22.6mm to 24.4mm
Long Eyes
0.19 D
Standard Deviation 0.15
0.20 D
Standard Deviation 0.15
Mean Absolute Prediction Error (in Diopters) in Eyes With Short AL<22.5mm, Long AL >24.5mm and Medium AL 22.6mm to 24.4mm
Medium Eyes
0.21 D
Standard Deviation 0.16
0.16 D
Standard Deviation 0.17
Mean Absolute Prediction Error (in Diopters) in Eyes With Short AL<22.5mm, Long AL >24.5mm and Medium AL 22.6mm to 24.4mm
Short Eyes
0.19 D
Standard Deviation 0.23
0.29 D
Standard Deviation 0.22

Adverse Events

Argos Biometer

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Serious adverse events

Adverse event data not reported

Other adverse events

Adverse event data not reported