Trial Outcomes & Findings for Omnipod® 5 With Libre 2 vs. MDI for Type 1 Diabetes in Children and Adults (NCT NCT05923827)
NCT ID: NCT05923827
Last Updated: 2026-04-14
Results Overview
The change in HbA1c at 13 weeks from baseline between the Intervention and Control groups.
COMPLETED
NA
189 participants
Comparing intervention group with control group during the 13-week study phase
2026-04-14
Participant Flow
A total of 189 participants were enrolled; however, one participant was withdrawn prior to randomization, resulting in 188 participants started and assigned to study arms - 125 for intervention and 63 for control.
Participant milestones
| Measure |
Intervention Group
Omnipod 5 System with FreeStyle Libre 2 continuous glucose monitor
Omnipod 5 System: The Omnipod 5 system is a tubeless patch pump that receives glucose values and trend data from the glucose sensor every 5 minutes, automatically calculates insulin dose, and sends delivery commands to the Pod for the delivery of insulin.
|
Control Group
Multiple daily injections of insulin with FreeStyle Libre 2 continuous glucose monitor
|
|---|---|---|
|
Overall Study
STARTED
|
125
|
63
|
|
Overall Study
COMPLETED
|
124
|
63
|
|
Overall Study
NOT COMPLETED
|
1
|
0
|
Reasons for withdrawal
| Measure |
Intervention Group
Omnipod 5 System with FreeStyle Libre 2 continuous glucose monitor
Omnipod 5 System: The Omnipod 5 system is a tubeless patch pump that receives glucose values and trend data from the glucose sensor every 5 minutes, automatically calculates insulin dose, and sends delivery commands to the Pod for the delivery of insulin.
|
Control Group
Multiple daily injections of insulin with FreeStyle Libre 2 continuous glucose monitor
|
|---|---|---|
|
Overall Study
Withdrawal by Subject
|
1
|
0
|
Baseline Characteristics
Race and Ethnicity were not collected from any participant.
Baseline characteristics by cohort
| Measure |
Intervention Group
n=125 Participants
Omnipod 5 System with FreeStyle Libre 2 continuous glucose monitor
Omnipod 5 System: The Omnipod 5 system is a tubeless patch pump that receives glucose values and trend data from the glucose sensor every 5 minutes, automatically calculates insulin dose, and sends delivery commands to the Pod for the delivery of insulin.
|
Control Group
n=63 Participants
Multiple daily injections of insulin with FreeStyle Libre 2 continuous glucose monitor
|
Total
n=188 Participants
Total of all reporting groups
|
|---|---|---|---|
|
Age, Continuous
|
24 years
STANDARD_DEVIATION 16 • n=125 Participants
|
24 years
STANDARD_DEVIATION 18 • n=63 Participants
|
24 years
STANDARD_DEVIATION 17 • n=188 Participants
|
|
Sex: Female, Male
Female
|
54 Participants
n=125 Participants
|
32 Participants
n=63 Participants
|
86 Participants
n=188 Participants
|
|
Sex: Female, Male
Male
|
71 Participants
n=125 Participants
|
31 Participants
n=63 Participants
|
102 Participants
n=188 Participants
|
|
Race and Ethnicity Not Collected
|
—
|
—
|
0 Participants
Race and Ethnicity were not collected from any participant.
|
|
Region of Enrollment
Belgium
|
8 participants
n=125 Participants
|
3 participants
n=63 Participants
|
11 participants
n=188 Participants
|
|
Region of Enrollment
United Kingdom
|
24 participants
n=125 Participants
|
13 participants
n=63 Participants
|
37 participants
n=188 Participants
|
|
Region of Enrollment
France
|
93 participants
n=125 Participants
|
47 participants
n=63 Participants
|
140 participants
n=188 Participants
|
|
Height (cm)
|
163 cm
STANDARD_DEVIATION 16 • n=125 Participants
|
159 cm
STANDARD_DEVIATION 16 • n=63 Participants
|
162 cm
STANDARD_DEVIATION 16 • n=188 Participants
|
|
Weight (kg)
|
63 kg
STANDARD_DEVIATION 21 • n=125 Participants
|
59 kg
STANDARD_DEVIATION 21 • n=63 Participants
|
62 kg
STANDARD_DEVIATION 21 • n=188 Participants
|
|
BMI (kg/m^2)
|
23 kg/m^2
STANDARD_DEVIATION 5 • n=125 Participants
|
22 kg/m^2
STANDARD_DEVIATION 6 • n=63 Participants
|
23 kg/m^2
STANDARD_DEVIATION 5 • n=188 Participants
|
|
POC A1C (%)
|
8.3 percentage of A1C
STANDARD_DEVIATION 0.7 • n=125 Participants
|
8.2 percentage of A1C
STANDARD_DEVIATION 0.5 • n=63 Participants
|
8.3 percentage of A1C
STANDARD_DEVIATION 0.7 • n=188 Participants
|
|
Duration of Type I Diagnosis
|
11 years
STANDARD_DEVIATION 11 • n=125 Participants
|
11 years
STANDARD_DEVIATION 10 • n=63 Participants
|
11 years
STANDARD_DEVIATION 10 • n=188 Participants
|
|
Has the subject used an insulin pump in the 6 months before study? (n/N (%))
Yes
|
0 Participants
n=125 Participants
|
0 Participants
n=63 Participants
|
0 Participants
n=188 Participants
|
|
Has the subject used an insulin pump in the 6 months before study? (n/N (%))
No
|
125 Participants
n=125 Participants
|
63 Participants
n=63 Participants
|
188 Participants
n=188 Participants
|
|
Brands of Insulin currently being used by the subject - Rapid Acting Insulin
Aspart (NovoLog/NovoRapid)
|
59 Participants
n=125 Participants
|
29 Participants
n=63 Participants
|
88 Participants
n=188 Participants
|
|
Brands of Insulin currently being used by the subject - Rapid Acting Insulin
Aspart (Fiasp)
|
26 Participants
n=125 Participants
|
19 Participants
n=63 Participants
|
45 Participants
n=188 Participants
|
|
Brands of Insulin currently being used by the subject - Rapid Acting Insulin
Glulisine (Apidra)
|
1 Participants
n=125 Participants
|
0 Participants
n=63 Participants
|
1 Participants
n=188 Participants
|
|
Brands of Insulin currently being used by the subject - Rapid Acting Insulin
Lispro (Humalog)
|
31 Participants
n=125 Participants
|
11 Participants
n=63 Participants
|
42 Participants
n=188 Participants
|
|
Brands of Insulin currently being used by the subject - Rapid Acting Insulin
Lispro-aabc (Lyumjev)
|
6 Participants
n=125 Participants
|
4 Participants
n=63 Participants
|
10 Participants
n=188 Participants
|
|
Brands of Insulin currently being used by the subject - Rapid Acting Insulin
Aspart (NovoLog/NovoRapid) and Lispro-aabc (Lyumjev)
|
1 Participants
n=125 Participants
|
0 Participants
n=63 Participants
|
1 Participants
n=188 Participants
|
|
Brands of Insulin currently being used by the subject - Rapid Acting Insulin
Aspart (NovoLog/NovoRapid) and Aspart (Fiasp)
|
1 Participants
n=125 Participants
|
0 Participants
n=63 Participants
|
1 Participants
n=188 Participants
|
|
Brands of Insulin currently being used by the subject - Rapid Acting Insulin
None
|
0 Participants
n=125 Participants
|
0 Participants
n=63 Participants
|
0 Participants
n=188 Participants
|
|
Brands of Insulin currently being used by the subject - Long Acting Insulin
Degludec (Tresiba)
|
84 Participants
n=125 Participants
|
41 Participants
n=63 Participants
|
125 Participants
n=188 Participants
|
|
Brands of Insulin currently being used by the subject - Long Acting Insulin
Detemir (Levemir)
|
4 Participants
n=125 Participants
|
5 Participants
n=63 Participants
|
9 Participants
n=188 Participants
|
|
Brands of Insulin currently being used by the subject - Long Acting Insulin
Glargine (Abasaglar/Lantus/Toujeo)
|
37 Participants
n=125 Participants
|
17 Participants
n=63 Participants
|
54 Participants
n=188 Participants
|
|
Brands of Insulin currently being used by the subject - Long Acting Insulin
None
|
0 Participants
n=125 Participants
|
0 Participants
n=63 Participants
|
0 Participants
n=188 Participants
|
|
Systolic Blood Pressure (mmHg)
|
121 mmHg
STANDARD_DEVIATION 16 • n=125 Participants
|
120 mmHg
STANDARD_DEVIATION 13 • n=63 Participants
|
120 mmHg
STANDARD_DEVIATION 15 • n=188 Participants
|
|
Heart Rate (bpm)
|
80 bpm
STANDARD_DEVIATION 14 • n=125 Participants
|
79 bpm
STANDARD_DEVIATION 14 • n=63 Participants
|
80 bpm
STANDARD_DEVIATION 14 • n=188 Participants
|
|
Diastolic Blood Pressure (mmHg)
|
73 mmHg
STANDARD_DEVIATION 11 • n=125 Participants
|
75 mmHg
STANDARD_DEVIATION 10 • n=63 Participants
|
73 mmHg
STANDARD_DEVIATION 10 • n=188 Participants
|
|
Respiratory Rate (bpm)
|
18 bpm
STANDARD_DEVIATION 4 • n=125 Participants
|
18 bpm
STANDARD_DEVIATION 4 • n=63 Participants
|
18 bpm
STANDARD_DEVIATION 4 • n=188 Participants
|
|
Temperature (C°)
|
36.5 C°
STANDARD_DEVIATION 0.4 • n=125 Participants
|
36.6 C°
STANDARD_DEVIATION 0.6 • n=63 Participants
|
36.5 C°
STANDARD_DEVIATION 0.5 • n=188 Participants
|
PRIMARY outcome
Timeframe: Comparing intervention group with control group during the 13-week study phasePopulation: Subjects are included in the Primary Endpoint Analysis even if data for one period is missing.
The change in HbA1c at 13 weeks from baseline between the Intervention and Control groups.
Outcome measures
| Measure |
Intervention Group
n=125 Participants
Omnipod 5 System with FreeStyle Libre 2 continuous glucose monitor
Omnipod 5 System: The Omnipod 5 system is a tubeless patch pump that receives glucose values and trend data from the glucose sensor every 5 minutes, automatically calculates insulin dose, and sends delivery commands to the Pod for the delivery of insulin.
|
Control Group
n=63 Participants
Multiple daily injections of insulin with FreeStyle Libre 2 continuous glucose monitor
|
|---|---|---|
|
Change in HbA1c (%)
Baseline
|
8.13 Percentage of HbA1C
Standard Deviation 0.70
|
8.10 Percentage of HbA1C
Standard Deviation 0.59
|
|
Change in HbA1c (%)
13 Weeks
|
7.20 Percentage of HbA1C
Standard Deviation 0.60
|
8.00 Percentage of HbA1C
Standard Deviation 0.73
|
SECONDARY outcome
Timeframe: Comparing intervention group with control group during the 13-week study phasePopulation: Subjects are included in the Secondary Endpoint Analysis even if data for one period is missing.
Glucose metric from study provided continuous glucose monitor (CGM)
Outcome measures
| Measure |
Intervention Group
n=125 Participants
Omnipod 5 System with FreeStyle Libre 2 continuous glucose monitor
Omnipod 5 System: The Omnipod 5 system is a tubeless patch pump that receives glucose values and trend data from the glucose sensor every 5 minutes, automatically calculates insulin dose, and sends delivery commands to the Pod for the delivery of insulin.
|
Control Group
n=63 Participants
Multiple daily injections of insulin with FreeStyle Libre 2 continuous glucose monitor
|
|---|---|---|
|
Percentage of Time in Range 70-180 mg/dL
Baseline
|
39.42 Percentage of time
Standard Deviation 11.75
|
40.85 Percentage of time
Standard Deviation 13.16
|
|
Percentage of Time in Range 70-180 mg/dL
13 Weeks
|
65.26 Percentage of time
Standard Deviation 7.56
|
43.01 Percentage of time
Standard Deviation 10.99
|
SECONDARY outcome
Timeframe: Comparing intervention group with control group at the end of the 13-week study phasePopulation: Subjects are included in the Secondary Endpoint Analysis even if data for one period is missing.
Glucose metric from CGM
Outcome measures
| Measure |
Intervention Group
n=125 Participants
Omnipod 5 System with FreeStyle Libre 2 continuous glucose monitor
Omnipod 5 System: The Omnipod 5 system is a tubeless patch pump that receives glucose values and trend data from the glucose sensor every 5 minutes, automatically calculates insulin dose, and sends delivery commands to the Pod for the delivery of insulin.
|
Control Group
n=63 Participants
Multiple daily injections of insulin with FreeStyle Libre 2 continuous glucose monitor
|
|---|---|---|
|
Percentage of Time <54 mg/dL (Non-inferior)
Baseline
|
0.61 Percentage of time
Standard Deviation 1.49
|
0.95 Percentage of time
Standard Deviation 2.23
|
|
Percentage of Time <54 mg/dL (Non-inferior)
13 Weeks
|
0.49 Percentage of time
Standard Deviation 0.52
|
0.63 Percentage of time
Standard Deviation 1.03
|
SECONDARY outcome
Timeframe: Comparing intervention group with control group at the end of the 13-week study phasePopulation: Subjects are included in the Secondary Endpoint Analysis even if data for one period is missing.
Glucose metric from CGM
Outcome measures
| Measure |
Intervention Group
n=125 Participants
Omnipod 5 System with FreeStyle Libre 2 continuous glucose monitor
Omnipod 5 System: The Omnipod 5 system is a tubeless patch pump that receives glucose values and trend data from the glucose sensor every 5 minutes, automatically calculates insulin dose, and sends delivery commands to the Pod for the delivery of insulin.
|
Control Group
n=63 Participants
Multiple daily injections of insulin with FreeStyle Libre 2 continuous glucose monitor
|
|---|---|---|
|
Percentage of Time >180 mg/dL
Baseline
|
57.14 Percentage of time
Standard Deviation 13.20
|
55.55 Percentage of time
Standard Deviation 14.81
|
|
Percentage of Time >180 mg/dL
13 Weeks
|
31.90 Percentage of time
Standard Deviation 7.85
|
53.49 Percentage of time
Standard Deviation 11.86
|
SECONDARY outcome
Timeframe: Comparing intervention group with control group at the end of the 13-week study phasePopulation: Subjects are included in the Secondary Endpoint Analysis even if data for one period is missing.
Glucose metric from CGM
Outcome measures
| Measure |
Intervention Group
n=125 Participants
Omnipod 5 System with FreeStyle Libre 2 continuous glucose monitor
Omnipod 5 System: The Omnipod 5 system is a tubeless patch pump that receives glucose values and trend data from the glucose sensor every 5 minutes, automatically calculates insulin dose, and sends delivery commands to the Pod for the delivery of insulin.
|
Control Group
n=63 Participants
Multiple daily injections of insulin with FreeStyle Libre 2 continuous glucose monitor
|
|---|---|---|
|
Percentage of Time >300 mg/dL
Baseline
|
15.21 Percentage of time
Standard Deviation 9.93
|
13.97 Percentage of time
Standard Deviation 9.46
|
|
Percentage of Time >300 mg/dL
13 Weeks
|
3.50 Percentage of time
Standard Deviation 2.60
|
13.29 Percentage of time
Standard Deviation 9.05
|
SECONDARY outcome
Timeframe: Comparing intervention group with control group at the end of the 13-week study phasePopulation: Subjects are included in the Secondary Endpoint Analysis even if data for one period is missing.
Glucose metric from CGM
Outcome measures
| Measure |
Intervention Group
n=125 Participants
Omnipod 5 System with FreeStyle Libre 2 continuous glucose monitor
Omnipod 5 System: The Omnipod 5 system is a tubeless patch pump that receives glucose values and trend data from the glucose sensor every 5 minutes, automatically calculates insulin dose, and sends delivery commands to the Pod for the delivery of insulin.
|
Control Group
n=63 Participants
Multiple daily injections of insulin with FreeStyle Libre 2 continuous glucose monitor
|
|---|---|---|
|
Percentage of Time <70 mg/dL (Non-inferior)
Baseline
|
3.45 Percentage of time
Standard Deviation 3.71
|
3.59 Percentage of time
Standard Deviation 5.38
|
|
Percentage of Time <70 mg/dL (Non-inferior)
13 Weeks
|
2.84 Percentage of time
Standard Deviation 1.73
|
3.50 Percentage of time
Standard Deviation 3.53
|
SECONDARY outcome
Timeframe: Comparing intervention group with control group at the end of the 13-week study phasePopulation: Subjects are included in the Secondary Endpoint Analysis even if data for one period is missing.
Glucose metric from study CGM
Outcome measures
| Measure |
Intervention Group
n=52 Participants
Omnipod 5 System with FreeStyle Libre 2 continuous glucose monitor
Omnipod 5 System: The Omnipod 5 system is a tubeless patch pump that receives glucose values and trend data from the glucose sensor every 5 minutes, automatically calculates insulin dose, and sends delivery commands to the Pod for the delivery of insulin.
|
Control Group
n=27 Participants
Multiple daily injections of insulin with FreeStyle Libre 2 continuous glucose monitor
|
|---|---|---|
|
Change From Baseline in HbA1c for Participants ≥18 Years of Age at Baseline
Baseline
|
8.27 Percentage of HbA1C
Standard Error 0.76
|
8.17 Percentage of HbA1C
Standard Error 0.62
|
|
Change From Baseline in HbA1c for Participants ≥18 Years of Age at Baseline
13 Weeks
|
7.32 Percentage of HbA1C
Standard Error 0.57
|
8.09 Percentage of HbA1C
Standard Error 0.62
|
SECONDARY outcome
Timeframe: Comparing intervention group with control group at the end of the 13-week study phasePopulation: Subjects are included in the Secondary Endpoint Analysis even if data for one period is missing.
Glucose metric from CGM
Outcome measures
| Measure |
Intervention Group
n=73 Participants
Omnipod 5 System with FreeStyle Libre 2 continuous glucose monitor
Omnipod 5 System: The Omnipod 5 system is a tubeless patch pump that receives glucose values and trend data from the glucose sensor every 5 minutes, automatically calculates insulin dose, and sends delivery commands to the Pod for the delivery of insulin.
|
Control Group
n=35 Participants
Multiple daily injections of insulin with FreeStyle Libre 2 continuous glucose monitor
|
|---|---|---|
|
Change From Baseline in HbA1c for Participants <18 Years of Age at Baseline
Baseline
|
8.02 Percentage of HbA1C
Standard Error 0.65
|
8.06 Percentage of HbA1C
Standard Error 0.57
|
|
Change From Baseline in HbA1c for Participants <18 Years of Age at Baseline
13 Weeks
|
7.11 Percentage of HbA1C
Standard Error 0.62
|
7.92 Percentage of HbA1C
Standard Error 0.82
|
SECONDARY outcome
Timeframe: Comparing intervention group with control group at the end of the 13-week study phasePopulation: Subjects are included in the Secondary Endpoint Analysis even if data for one period is missing.
Glucose metric from CGM
Outcome measures
| Measure |
Intervention Group
n=52 Participants
Omnipod 5 System with FreeStyle Libre 2 continuous glucose monitor
Omnipod 5 System: The Omnipod 5 system is a tubeless patch pump that receives glucose values and trend data from the glucose sensor every 5 minutes, automatically calculates insulin dose, and sends delivery commands to the Pod for the delivery of insulin.
|
Control Group
n=27 Participants
Multiple daily injections of insulin with FreeStyle Libre 2 continuous glucose monitor
|
|---|---|---|
|
Percentage of Time in Range 70-180 mg/dL for Participants ≥18 Years of Age at Baseline
Baseline
|
40.67 Percentage of time
Standard Error 12.48
|
43.11 Percentage of time
Standard Error 14.23
|
|
Percentage of Time in Range 70-180 mg/dL for Participants ≥18 Years of Age at Baseline
13 Weeks
|
64.65 Percentage of time
Standard Error 8.05
|
41.66 Percentage of time
Standard Error 13.76
|
SECONDARY outcome
Timeframe: Comparing intervention group with control group at the end of the 13-week study phaseGlucose metric from CGM
Outcome measures
| Measure |
Intervention Group
n=73 Participants
Omnipod 5 System with FreeStyle Libre 2 continuous glucose monitor
Omnipod 5 System: The Omnipod 5 system is a tubeless patch pump that receives glucose values and trend data from the glucose sensor every 5 minutes, automatically calculates insulin dose, and sends delivery commands to the Pod for the delivery of insulin.
|
Control Group
n=35 Participants
Multiple daily injections of insulin with FreeStyle Libre 2 continuous glucose monitor
|
|---|---|---|
|
Percentage of Time in Range 70-180 mg/dL for Participants <18 Years of Age at Baseline
Baseline
|
38.55 Percentage of time
Standard Error 11.22
|
39.11 Percentage of time
Standard Error 12.20
|
|
Percentage of Time in Range 70-180 mg/dL for Participants <18 Years of Age at Baseline
13 Weeks
|
65.69 Percentage of time
Standard Error 7.22
|
44.07 Percentage of time
Standard Error 8.23
|
SECONDARY outcome
Timeframe: Comparing intervention group with control group at the end of the 13-week study phasePopulation: Subjects are included in the Secondary Endpoint Analysis even if data for one period is missing.
Measured at baseline and 13 weeks in both intervention and control groups
Outcome measures
| Measure |
Intervention Group
n=64 Participants
Omnipod 5 System with FreeStyle Libre 2 continuous glucose monitor
Omnipod 5 System: The Omnipod 5 system is a tubeless patch pump that receives glucose values and trend data from the glucose sensor every 5 minutes, automatically calculates insulin dose, and sends delivery commands to the Pod for the delivery of insulin.
|
Control Group
n=31 Participants
Multiple daily injections of insulin with FreeStyle Libre 2 continuous glucose monitor
|
|---|---|---|
|
Change From Baseline in HbA1c in Participants With Baseline HbA1c ≥8%
Baseline
|
8.65 Percent of change
Standard Deviation 0.57
|
8.60 Percent of change
Standard Deviation 0.40
|
|
Change From Baseline in HbA1c in Participants With Baseline HbA1c ≥8%
13 Weeks
|
7.44 Percent of change
Standard Deviation 0.62
|
8.37 Percent of change
Standard Deviation 0.79
|
SECONDARY outcome
Timeframe: Comparing intervention group with control group at the end of the 13-week study phasePopulation: Subjects are included in the Secondary Endpoint Analysis even if data for one period is missing.
The Pittsburgh Sleep Quality Index (PSQI) is a self-report scale assessing sleep quality over the past month. It produces a 0-21 total score, with higher scores indicating worse sleep quality. Results are reported as scores on a scale.
Outcome measures
| Measure |
Intervention Group
n=125 Participants
Omnipod 5 System with FreeStyle Libre 2 continuous glucose monitor
Omnipod 5 System: The Omnipod 5 system is a tubeless patch pump that receives glucose values and trend data from the glucose sensor every 5 minutes, automatically calculates insulin dose, and sends delivery commands to the Pod for the delivery of insulin.
|
Control Group
n=63 Participants
Multiple daily injections of insulin with FreeStyle Libre 2 continuous glucose monitor
|
|---|---|---|
|
Change From Baseline in PSQI Total Score
13 Weeks
|
5.03 Scores on a scale
Standard Deviation 2.82
|
5.91 Scores on a scale
Standard Deviation 3.61
|
|
Change From Baseline in PSQI Total Score
Baseline
|
5.05 Scores on a scale
Standard Deviation 2.65
|
5.71 Scores on a scale
Standard Deviation 3.49
|
SECONDARY outcome
Timeframe: Comparing intervention group with control group at the end of the 13-week study phasePopulation: Subjects are included in the Secondary Endpoint Analysis even if data for one period is missing.
A questionnaire that measures seven critical dimensions of distress (28-item scale with 6 choices that range from 1 (Not a Problem) to 6 (A Very Serious Problem)). The total score can range from 1 to 6, with a lower score indicating a better outcome.
Outcome measures
| Measure |
Intervention Group
n=52 Participants
Omnipod 5 System with FreeStyle Libre 2 continuous glucose monitor
Omnipod 5 System: The Omnipod 5 system is a tubeless patch pump that receives glucose values and trend data from the glucose sensor every 5 minutes, automatically calculates insulin dose, and sends delivery commands to the Pod for the delivery of insulin.
|
Control Group
n=27 Participants
Multiple daily injections of insulin with FreeStyle Libre 2 continuous glucose monitor
|
|---|---|---|
|
Change From Baseline in T1-DDS Total Score for Participants ≥18 Years of Age at Baseline
Baseline
|
1.87 Scores on a scale
Standard Deviation 0.56
|
1.86 Scores on a scale
Standard Deviation 0.61
|
|
Change From Baseline in T1-DDS Total Score for Participants ≥18 Years of Age at Baseline
13 Weeks
|
1.59 Scores on a scale
Standard Deviation 0.49
|
1.72 Scores on a scale
Standard Deviation 0.51
|
SECONDARY outcome
Timeframe: Comparing intervention group with control group at the end of the 13-week study phasePopulation: Subjects are included in the Secondary Endpoint Analysis even if data for one period is missing.
A questionnaire that examines the degree to which people with diabetes feel able, secure, and comfortable regarding their ability to stay safe from hypoglycemic-related problems (9-item scale with 4 choices that range from 1 (Not Confident At All) to 4 (Very Confident)). The total score can range from 1 to 4, with a higher score indicating a better outcome.
Outcome measures
| Measure |
Intervention Group
n=52 Participants
Omnipod 5 System with FreeStyle Libre 2 continuous glucose monitor
Omnipod 5 System: The Omnipod 5 system is a tubeless patch pump that receives glucose values and trend data from the glucose sensor every 5 minutes, automatically calculates insulin dose, and sends delivery commands to the Pod for the delivery of insulin.
|
Control Group
n=27 Participants
Multiple daily injections of insulin with FreeStyle Libre 2 continuous glucose monitor
|
|---|---|---|
|
Change From Baseline in HCS Total Score for Participants ≥18 Years of Age at Baseline
Baseline
|
3.46 Score on a scale
Standard Deviation 0.47
|
3.30 Score on a scale
Standard Deviation 0.54
|
|
Change From Baseline in HCS Total Score for Participants ≥18 Years of Age at Baseline
13 Weeks
|
3.48 Score on a scale
Standard Deviation 0.46
|
3.35 Score on a scale
Standard Deviation 0.57
|
SECONDARY outcome
Timeframe: Comparing intervention group with control group at the end of the 13-week study phasePopulation: Subjects are included in the Secondary Endpoint Analysis even if data for one period is missing.
Glucose metric from CGM
Outcome measures
| Measure |
Intervention Group
n=125 Participants
Omnipod 5 System with FreeStyle Libre 2 continuous glucose monitor
Omnipod 5 System: The Omnipod 5 system is a tubeless patch pump that receives glucose values and trend data from the glucose sensor every 5 minutes, automatically calculates insulin dose, and sends delivery commands to the Pod for the delivery of insulin.
|
Control Group
n=63 Participants
Multiple daily injections of insulin with FreeStyle Libre 2 continuous glucose monitor
|
|---|---|---|
|
Percentage of Time <70 mg/dL
13 Weeks
|
2.84 Percent of time
Standard Deviation 1.73
|
3.50 Percent of time
Standard Deviation 3.53
|
|
Percentage of Time <70 mg/dL
Baseline
|
3.45 Percent of time
Standard Deviation 3.71
|
3.59 Percent of time
Standard Deviation 5.38
|
SECONDARY outcome
Timeframe: Comparing intervention group with control group at the end of the 13-week study phasePopulation: Subjects are included in the Secondary Endpoint Analysis even if data for one period is missing.
Participants achieving HbA1C \< 7% \[N\]
Outcome measures
| Measure |
Intervention Group
n=122 Participants
Omnipod 5 System with FreeStyle Libre 2 continuous glucose monitor
Omnipod 5 System: The Omnipod 5 system is a tubeless patch pump that receives glucose values and trend data from the glucose sensor every 5 minutes, automatically calculates insulin dose, and sends delivery commands to the Pod for the delivery of insulin.
|
Control Group
n=60 Participants
Multiple daily injections of insulin with FreeStyle Libre 2 continuous glucose monitor
|
|---|---|---|
|
Percentage of Participants Achieving HbA1c <7%
Participants achieving HbA1c <7%
|
34 Participants
|
4 Participants
|
|
Percentage of Participants Achieving HbA1c <7%
Participants not achieving HbA1c <7%
|
88 Participants
|
56 Participants
|
SECONDARY outcome
Timeframe: Comparing intervention group with control group at the end of the 13-week study phasePopulation: Subjects are included in the Secondary Endpoint Analysis even if data for one period is missing.
Change from baseline in the EQ-5D-3L index score, a standardized measure of health-related quality of life. Scores range from -0.594 to 1 (full health), with higher scores indicating better health status.
Outcome measures
| Measure |
Intervention Group
n=125 Participants
Omnipod 5 System with FreeStyle Libre 2 continuous glucose monitor
Omnipod 5 System: The Omnipod 5 system is a tubeless patch pump that receives glucose values and trend data from the glucose sensor every 5 minutes, automatically calculates insulin dose, and sends delivery commands to the Pod for the delivery of insulin.
|
Control Group
n=63 Participants
Multiple daily injections of insulin with FreeStyle Libre 2 continuous glucose monitor
|
|---|---|---|
|
Change From Baseline in EQ-5D-3L Index Score
Baseline
|
0.94 Score on a scale
Standard Deviation 0.12
|
0.93 Score on a scale
Standard Deviation 0.16
|
|
Change From Baseline in EQ-5D-3L Index Score
13 Weeks
|
0.93 Score on a scale
Standard Deviation 0.13
|
0.92 Score on a scale
Standard Deviation 0.14
|
SECONDARY outcome
Timeframe: Comparing intervention group with control group at the end of the 13-week study phasePopulation: Subjects are included in the Secondary Endpoint Analysis even if data for one period is missing.
Measure of insulin requirements measured at 13 weeks in both intervention and control groups
Outcome measures
| Measure |
Intervention Group
n=52 Participants
Omnipod 5 System with FreeStyle Libre 2 continuous glucose monitor
Omnipod 5 System: The Omnipod 5 system is a tubeless patch pump that receives glucose values and trend data from the glucose sensor every 5 minutes, automatically calculates insulin dose, and sends delivery commands to the Pod for the delivery of insulin.
|
Control Group
n=27 Participants
Multiple daily injections of insulin with FreeStyle Libre 2 continuous glucose monitor
|
|---|---|---|
|
Change From Baseline in Total Daily Insulin (TDI) (Units/kg) for Participants ≥18 Years of Age
Baseline
|
0.72 Change from Baseline (units/kg)
Standard Deviation 0.24
|
0.62 Change from Baseline (units/kg)
Standard Deviation 0.22
|
|
Change From Baseline in Total Daily Insulin (TDI) (Units/kg) for Participants ≥18 Years of Age
13 Weeks
|
0.67 Change from Baseline (units/kg)
Standard Deviation 0.23
|
0.63 Change from Baseline (units/kg)
Standard Deviation 0.24
|
SECONDARY outcome
Timeframe: Comparing intervention group with control group at the end of the 13-week study phasePopulation: Subjects are included in the Secondary Endpoint Analysis even if data for one period is missing.
Glucose metric from CGM
Outcome measures
| Measure |
Intervention Group
n=125 Participants
Omnipod 5 System with FreeStyle Libre 2 continuous glucose monitor
Omnipod 5 System: The Omnipod 5 system is a tubeless patch pump that receives glucose values and trend data from the glucose sensor every 5 minutes, automatically calculates insulin dose, and sends delivery commands to the Pod for the delivery of insulin.
|
Control Group
n=63 Participants
Multiple daily injections of insulin with FreeStyle Libre 2 continuous glucose monitor
|
|---|---|---|
|
Percentage of Time <54 mg/dL
Baseline
|
0.61 Percentage of time
Standard Deviation 1.49
|
0.95 Percentage of time
Standard Deviation 2.23
|
|
Percentage of Time <54 mg/dL
13 Weeks
|
0.49 Percentage of time
Standard Deviation 0.52
|
0.63 Percentage of time
Standard Deviation 1.03
|
Adverse Events
Intervention Group
Control Group
Serious adverse events
| Measure |
Intervention Group
n=125 participants at risk
Omnipod 5 System with FreeStyle Libre 2 continuous glucose monitor
Omnipod 5 System: The Omnipod 5 system is a tubeless patch pump that receives glucose values and trend data from the glucose sensor every 5 minutes, automatically calculates insulin dose, and sends delivery commands to the Pod for the delivery of insulin.
|
Control Group
n=63 participants at risk
Multiple daily injections of insulin with FreeStyle Libre 2 continuous glucose monitor
|
|---|---|---|
|
Vascular disorders
Other
|
0.80%
1/125 • Number of events 1 • Start time of the standard therapy (14-days prior to Study Day 1) and continued until the participant's participation ended, up to 13 weeks.
|
0.00%
0/63 • Start time of the standard therapy (14-days prior to Study Day 1) and continued until the participant's participation ended, up to 13 weeks.
|
|
Cardiac disorders
Other
|
0.80%
1/125 • Number of events 1 • Start time of the standard therapy (14-days prior to Study Day 1) and continued until the participant's participation ended, up to 13 weeks.
|
0.00%
0/63 • Start time of the standard therapy (14-days prior to Study Day 1) and continued until the participant's participation ended, up to 13 weeks.
|
Other adverse events
| Measure |
Intervention Group
n=125 participants at risk
Omnipod 5 System with FreeStyle Libre 2 continuous glucose monitor
Omnipod 5 System: The Omnipod 5 system is a tubeless patch pump that receives glucose values and trend data from the glucose sensor every 5 minutes, automatically calculates insulin dose, and sends delivery commands to the Pod for the delivery of insulin.
|
Control Group
n=63 participants at risk
Multiple daily injections of insulin with FreeStyle Libre 2 continuous glucose monitor
|
|---|---|---|
|
Investigations
Other
|
13.6%
17/125 • Number of events 21 • Start time of the standard therapy (14-days prior to Study Day 1) and continued until the participant's participation ended, up to 13 weeks.
|
4.8%
3/63 • Number of events 3 • Start time of the standard therapy (14-days prior to Study Day 1) and continued until the participant's participation ended, up to 13 weeks.
|
|
Endocrine disorders
Hyperglycemia
|
4.0%
5/125 • Number of events 8 • Start time of the standard therapy (14-days prior to Study Day 1) and continued until the participant's participation ended, up to 13 weeks.
|
0.00%
0/63 • Start time of the standard therapy (14-days prior to Study Day 1) and continued until the participant's participation ended, up to 13 weeks.
|
|
Endocrine disorders
Prolonged Hyperglycemia
|
6.4%
8/125 • Number of events 8 • Start time of the standard therapy (14-days prior to Study Day 1) and continued until the participant's participation ended, up to 13 weeks.
|
0.00%
0/63 • Start time of the standard therapy (14-days prior to Study Day 1) and continued until the participant's participation ended, up to 13 weeks.
|
Additional Information
Results disclosure agreements
- Principal investigator is a sponsor employee
- Publication restrictions are in place