Trial Outcomes & Findings for A Multi-Center Prospective Open Label Study of a Web-based Application for Pulse Rate in Adult Patients (NCT NCT05853380)
NCT ID: NCT05853380
Last Updated: 2025-11-03
Results Overview
The primary endpoint is a comparison to an FDA-cleared reference device to establish that PR measured with the IVC App is accurate.
Recruitment status
COMPLETED
Target enrollment
86 participants
Primary outcome timeframe
120 seconds
Results posted on
2025-11-03
Participant Flow
The study recruited subjects from clinics, hospitals, and the general public. The study was conducted in the United States.
Participant milestones
| Measure |
Single Arm
Open Label, Single Group Assignment
|
|---|---|
|
Overall Study
STARTED
|
86
|
|
Overall Study
COMPLETED
|
82
|
|
Overall Study
NOT COMPLETED
|
4
|
Reasons for withdrawal
Withdrawal data not reported
Baseline Characteristics
ECG HR is reported for the 69 subjects with comparator data.
Baseline characteristics by cohort
| Measure |
Single Arm
n=82 Participants
Open Label, Single Group Assignment
|
|---|---|
|
Age, Continuous
|
48.7 Years
n=82 Participants
|
|
Sex: Female, Male
Female
|
42 Participants
n=82 Participants
|
|
Sex: Female, Male
Male
|
40 Participants
n=82 Participants
|
|
Ethnicity (NIH/OMB)
Hispanic or Latino
|
7 Participants
n=82 Participants
|
|
Ethnicity (NIH/OMB)
Not Hispanic or Latino
|
75 Participants
n=82 Participants
|
|
Ethnicity (NIH/OMB)
Unknown or Not Reported
|
0 Participants
n=82 Participants
|
|
Race (NIH/OMB)
American Indian or Alaska Native
|
2 Participants
n=82 Participants
|
|
Race (NIH/OMB)
Asian
|
5 Participants
n=82 Participants
|
|
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
|
1 Participants
n=82 Participants
|
|
Race (NIH/OMB)
Black or African American
|
32 Participants
n=82 Participants
|
|
Race (NIH/OMB)
White
|
41 Participants
n=82 Participants
|
|
Race (NIH/OMB)
More than one race
|
1 Participants
n=82 Participants
|
|
Race (NIH/OMB)
Unknown or Not Reported
|
0 Participants
n=82 Participants
|
|
Region of Enrollment
United States
|
82 participants
n=82 Participants
|
|
IVC App Pulse Rate
|
69.0 Beats per Minute
n=69 Participants • ECG HR is reported for the 69 subjects with comparator data.
|
|
IVC App Report Rate
|
69 Participants
n=82 Participants • IVC App successfully reported the Pulse Rate for 69 of 82 subjects
|
PRIMARY outcome
Timeframe: 120 secondsPopulation: Intent to Capture
The primary endpoint is a comparison to an FDA-cleared reference device to establish that PR measured with the IVC App is accurate.
Outcome measures
| Measure |
Single Arm
n=82 Participants
Open Label, Single Group Assignment
|
|---|---|
|
Pulse Rate Accuracy
|
67 Participants
|
Adverse Events
Single Arm
Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths
Serious adverse events
Adverse event data not reported
Other adverse events
Adverse event data not reported
Additional Information
Results disclosure agreements
- Principal investigator is a sponsor employee
- Publication restrictions are in place