Refractix DSP and TS Meter-DSP Measurement Comparison
NCT05717036 · Status: UNKNOWN · Type: OBSERVATIONAL · Enrollment: 36
Last updated 2023-02-08
Summary
The goal of this clinical trial is to compare the serum plasma measurement results between the FDA cleared TS Meter-DSP and investigational Refractix DSP. Study participants will be individuals being screened for plasma donation in a plasmaphereris center. The main question the study aims to answer is whether the measurement taken on the investigational Refractix DSP device is equivalent, defined as within +/- 0.3g/100ml, of the measurement taken on the TS Meter-DSP. Participants will be asked to provide a second capillary tube of blood (\<0.085ml) from the fingerstick that is part of their routine pre-screening for plasma donation.
Conditions
- Healthy
Interventions
- DEVICE
-
Serum Total Protein Measurement with TS Meter-DSP and Refractix DSP
The two blood samples will be centrifuged and hematocrit measured per the plasma center's standard operating procedures. One participant sample will be dispensed onto the cleared TS Meter-DSP for a serum total protein measurement. This measurement will be used, along with other pre-screening information, to determine if the participant can donate plasma. The second participant sample will be dispensed onto the investigational Refractix DSP for a serum total protein measurement. It will be compared to the sample measurement taken on the cleared TS Meter-DSP.
Sponsors & Collaborators
-
Reichert, Inc.
lead INDUSTRY
Eligibility
- Min Age
- 18 Years
- Max Age
- 90 Years
- Sex
- ALL
- Healthy Volunteers
- Yes
Timeline & Regulatory
- Start
- 2023-02-13
- Primary Completion
- 2023-02-15
- Completion
- 2023-02-15
- FDA Device
- Yes
Countries
- United States
Study Locations
More Related Trials
-
Clinical Evaluation of i-STAT Analyzer Glucose, Hematocrit and Sodium for Capillary Samples
NCT02897154 ·Status: COMPLETED
-
To Validate an Analytical Method to Measure Concentration of Tacrolimus in Blood Taken From Finger Pricks
NCT02377609 ·Status: TERMINATED
-
Validation of Diagnostics to Identify Glucose-6-phosphate Dehydrogenase Activity in the US
NCT04010695 ·Status: COMPLETED ·Phase: NA
-
Initial Case Series With Exalt Single-Use Duodenoscope - Expanded User Experience
NCT04223830 ·Status: COMPLETED ·Phase: NA
-
Neck-Worn Monitoring Sensor for - A Study for Monitoring Subjects With Fluid-Management Issues During Dialysis
NCT02140905 ·Status: COMPLETED
-
Healthy.io Method Comparison & User Performance Study
NCT02785445 ·Status: COMPLETED ·Phase: NA
-
Tolerability and Background Fluorescence of the MediBeacon Transdermal GFR Measurement System
NCT03810833 ·Status: COMPLETED ·Phase: NA
-
Bias Source of Signal in SCOUT DS
NCT01550328 ·Status: TERMINATED
-
TRIP-Patch vs the Rigiscan
NCT05219071 ·Status: COMPLETED
-
Low Cost QTc Meter for Long QT Syndrome Screening in Primary Care
NCT04441892 ·Status: TERMINATED ·Phase: NA
-
Evaluation of the LumiraDx Point of Care D-Dimer and CRP Tests
NCT04375982 ·Status: ACTIVE_NOT_RECRUITING ·Phase: NA
-
Wetness Sensing System
NCT05454813 ·Status: COMPLETED ·Phase: NA
-
QIAstat-Dx® BCID Plus AMR Panels Clinical Performance Study
NCT05305534 ·Status: COMPLETED
-
AngioSafe Peripheral CTO Crossing System Study (RESTOR-1 Study)
NCT04663867 ·Status: COMPLETED ·Phase: NA
-
Transdermal Optical Imaging (TOI) as a Non-invasive Measure to Detect Changes in Hydration Status in Healthy Adults and Athletes
NCT06212518 ·Status: RECRUITING ·Phase: NA
-
Research & Evaluation of LumiraDx SARS-CoV-2 (COVID-19) Point of Care Ab Test
NCT04597047 ·Status: UNKNOWN ·Phase: NA
-
Evaluation of the Spectra Optia® Apheresis System Mononuclear Cell (MNC) Collection Procedure in Healthy Blood Donors
NCT01277549 ·Status: COMPLETED
-
A Human Clinical Study to Collect Calibration and Performance Data for the RBA-1 and KBS-1
NCT04305938 ·Status: UNKNOWN
-
Pilot Evaluation of the StarGuide Next Generation SPECT/CT System
NCT07272161 ·Status: ENROLLING_BY_INVITATION ·Phase: NA
-
To Investigate Agreement Between the EM-3000 and the Predicate Device Noncon Robo Pachy (F&A) (CellChek XL)
NCT02293122 ·Status: COMPLETED
-
Reference Range Study for the Quantra Analyzer With the Quantra Surgical Cartridge
NCT03139097 ·Status: COMPLETED
-
A Normal Donor Sample Collection Study in Healthy Adults
NCT01318863 ·Status: COMPLETED
-
Performance Evaluation of the Nanomix eLab® CRP, PCT, and LAC Assays With the Nanomix eLab System
NCT03876535 ·Status: COMPLETED
-
REduction of Contrast Via DyeVert Used in CTO Procedures
NCT03118544 ·Status: TERMINATED
-
Measuring Alterations of DNA in Human Blood
NCT00178308 ·Status: COMPLETED