Trial Outcomes & Findings for Comparing NGENUITY vs Conventional Microscope in a New Hydrus User (NCT NCT05529966)

NCT ID: NCT05529966

Last Updated: 2025-04-17

Results Overview

Recruitment status

COMPLETED

Study phase

NA

Target enrollment

32 participants

Primary outcome timeframe

up to 1 hour

Results posted on

2025-04-17

Participant Flow

32 participants signed informed consent, 28 were randomized to either study arm.

Unit of analysis: Eyes

Participant milestones

Participant milestones
Measure
Conventional Microscope
Participants randomized to the use of a conventional microscope for the first eye procedure. Participants who had a procedure for their second eye were randomized again to either the conventional or NGENUITY microscope.
NGENUITY
Participants randomized to the use of a NGENUITY microscope for the first eye procedure. Participants who had a procedure for their second eye were randomized again to either the conventional or NGENUITY microscope.
Conventional Microscope (Re-randomized)
Participants who had a procedure for their second eye were re-randomized to use of a conventional microscope for the procedure.
NGENUITY (Re-randomized)
Participants who had a procedure for their second eye were re-randomized to use of the NGENUITY microscope for the procedure.
First Eye
STARTED
14 14
14 14
0 0
0 0
First Eye
Stent Placement Could Not be Performed
1 1
1 1
0 0
0 0
First Eye
Had Missing Data
1 1
0 0
0 0
0 0
First Eye
Evaluable for Primary and Secondary Outcome Measure Analysis
12 12
13 13
0 0
0 0
First Eye
COMPLETED
13 13
13 13
0 0
0 0
First Eye
NOT COMPLETED
1 1
1 1
0 0
0 0
Second Eye
STARTED
0 0
0 0
1 1
3 3
Second Eye
Randomized to Use of Opposite Microscope for Second Eye Procedure
0 0
0 0
1 1
2 2
Second Eye
Evaluable for Primary and Secondary Outcome Measure Analysis
0 0
0 0
1 1
3 3
Second Eye
COMPLETED
0 0
0 0
1 1
3 3
Second Eye
NOT COMPLETED
0 0
0 0
0 0
0 0

Reasons for withdrawal

Withdrawal data not reported

Baseline Characteristics

Comparing NGENUITY vs Conventional Microscope in a New Hydrus User

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
Conventional Microscope
n=12 Participants
Participants randomized to the use of a conventional microscope for the first eye procedure. Participants who had a procedure for their second eye were randomized again to either the conventional or NGENUITY microscope.
NGENUITY
n=13 Participants
Participants randomized to the use of a NGENUITY microscope for the first eye procedure. Participants who had a procedure for their second eye were randomized again to either the conventional or NGENUITY microscope.
Total
n=25 Participants
Total of all reporting groups
Age, Continuous
75.6 years
STANDARD_DEVIATION 8.1 • n=99 Participants
76.1 years
STANDARD_DEVIATION 7.3 • n=107 Participants
75.9 years
STANDARD_DEVIATION 7.5 • n=206 Participants
Sex: Female, Male
Female
7 Participants
n=99 Participants
6 Participants
n=107 Participants
13 Participants
n=206 Participants
Sex: Female, Male
Male
5 Participants
n=99 Participants
7 Participants
n=107 Participants
12 Participants
n=206 Participants
Ethnicity (NIH/OMB)
Hispanic or Latino
0 Participants
n=99 Participants
1 Participants
n=107 Participants
1 Participants
n=206 Participants
Ethnicity (NIH/OMB)
Not Hispanic or Latino
12 Participants
n=99 Participants
12 Participants
n=107 Participants
24 Participants
n=206 Participants
Ethnicity (NIH/OMB)
Unknown or Not Reported
0 Participants
n=99 Participants
0 Participants
n=107 Participants
0 Participants
n=206 Participants
Race/Ethnicity, Customized
American Indian
0 Participants
n=99 Participants
1 Participants
n=107 Participants
1 Participants
n=206 Participants
Race/Ethnicity, Customized
Asian
6 Participants
n=99 Participants
2 Participants
n=107 Participants
8 Participants
n=206 Participants
Race/Ethnicity, Customized
Black
0 Participants
n=99 Participants
1 Participants
n=107 Participants
1 Participants
n=206 Participants
Race/Ethnicity, Customized
Other
2 Participants
n=99 Participants
0 Participants
n=107 Participants
2 Participants
n=206 Participants
Race/Ethnicity, Customized
White
4 Participants
n=99 Participants
9 Participants
n=107 Participants
13 Participants
n=206 Participants
Region of Enrollment
United States
12 Participants
n=99 Participants
13 Participants
n=107 Participants
25 Participants
n=206 Participants

PRIMARY outcome

Timeframe: up to 1 hour

Population: Participants/eyes evaluable for primary and secondary outcome measure analysis

Outcome measures

Outcome measures
Measure
Conventional Microscope
n=13 Eyes
Conventional microscope used during micro-stent placement.
NGENUITY
n=16 Eyes
NGENUITY microscope used during micro-stent placement.
Time of First Cannula in Until Satisfactory Stent Placement and Cannula Out
2.55 minutes
Standard Deviation 1.36
3.06 minutes
Standard Deviation 2.17

SECONDARY outcome

Timeframe: up to 1 hour

Population: Participants/eyes evaluable for primary and secondary outcome measure analysis

Outcome measures

Outcome measures
Measure
Conventional Microscope
n=13 Eyes
Conventional microscope used during micro-stent placement.
NGENUITY
n=16 Eyes
NGENUITY microscope used during micro-stent placement.
Number of Trabecular Meshwork Piercing Attempts Needed to Place Stent
1.50 piercing
Standard Deviation 0.94
1.53 piercing
Standard Deviation 0.92

SECONDARY outcome

Timeframe: up to 1 hour

Population: Participants/eyes evaluable for primary and secondary outcome measure analysis

Outcome measures

Outcome measures
Measure
Conventional Microscope
n=13 Eyes
Conventional microscope used during micro-stent placement.
NGENUITY
n=16 Eyes
NGENUITY microscope used during micro-stent placement.
Number of Anterior Chamber Entry and Reentry Attempts During Stent Placement
1.14 Number of Anterior Chamber Entries
Standard Deviation 0.36
1.13 Number of Anterior Chamber Entries
Standard Deviation 0.35

SECONDARY outcome

Timeframe: up to 1 hour

Population: Participants/eyes evaluable for primary and secondary outcome measure analysis

Outcome measures

Outcome measures
Measure
Conventional Microscope
n=15 Eyes
Conventional microscope used during micro-stent placement.
NGENUITY
n=16 Eyes
NGENUITY microscope used during micro-stent placement.
Number of Microscope Focus or Position Readjustments Needed During Stent Placement
1.43 Number of Focus/Microscope adjustments
Standard Deviation 1.28
1.93 Number of Focus/Microscope adjustments
Standard Deviation 1.79

SECONDARY outcome

Timeframe: Up to 5 minutes

Population: Participants/eyes evaluable for primary and secondary outcome measure analysis

Providers rate their comfort, visibility, image quality, focus, depth perception, staff engagement, teaching and level of confidence in using the system and system performance. Ratings are on a Likert scale of 1 to 5. Higher scores correspond to excellent ratings, higher confidence in using the system, and better system performance; for Scope Preference only, lower numbers (either group) correspond to a preference for NGenuity visualization.

Outcome measures

Outcome measures
Measure
Conventional Microscope
n=13 Eyes
Conventional microscope used during micro-stent placement.
NGENUITY
n=16 Eyes
NGENUITY microscope used during micro-stent placement.
Change in Attending Questionnaire Score
Comfort
3.14 score on a scale
Standard Deviation 0.53
3.93 score on a scale
Standard Deviation 0.80
Change in Attending Questionnaire Score
Visibility
3.57 score on a scale
Standard Deviation 0.85
4.53 score on a scale
Standard Deviation 0.64
Change in Attending Questionnaire Score
Image Quality
3.14 score on a scale
Standard Deviation 0.77
4.6 score on a scale
Standard Deviation 0.74
Change in Attending Questionnaire Score
Focus
3.14 score on a scale
Standard Deviation 0.66
4.0 score on a scale
Standard Deviation 1.25
Change in Attending Questionnaire Score
Depth Perception
2.93 score on a scale
Standard Deviation 0.73
4.67 score on a scale
Standard Deviation 0.62
Change in Attending Questionnaire Score
Staff Engagement
3.36 score on a scale
Standard Deviation 0.50
4.27 score on a scale
Standard Deviation 0.80
Change in Attending Questionnaire Score
Teaching
3.29 score on a scale
Standard Deviation 0.73
4.67 score on a scale
Standard Deviation 0.62
Change in Attending Questionnaire Score
En Face Confidence
3.43 score on a scale
Standard Deviation 0.51
4.07 score on a scale
Standard Deviation 0.96
Change in Attending Questionnaire Score
Hydrus Placement Confidence
3.71 score on a scale
Standard Deviation 0.47
4.20 score on a scale
Standard Deviation 1.08
Change in Attending Questionnaire Score
Scope Preference (1 = 3D.)
2.71 score on a scale
Standard Deviation 0.61
1.80 score on a scale
Standard Deviation 0.86

SECONDARY outcome

Timeframe: Up to 5 minutes

Population: Participants/eyes evaluable for primary and secondary outcome measure analysis

Providers rate their comfort, visibility, image quality, focus, depth perception, staff engagement, teaching and level of confidence in using the system and system performance. Ratings are on a Likert scale of 1 to 5. Higher scores correspond to excellent ratings, higher confidence in using the system, and better system performance; for Scope Preference only, lower numbers (either group) correspond to a preference for NGenuity visualization.

Outcome measures

Outcome measures
Measure
Conventional Microscope
n=13 Eyes
Conventional microscope used during micro-stent placement.
NGENUITY
n=16 Eyes
NGENUITY microscope used during micro-stent placement.
Change in Fellow Questionnaire Score
Comfort
4.21 score on a scale
Standard Deviation 0.89
4.20 score on a scale
Standard Deviation 0.56
Change in Fellow Questionnaire Score
Image Quality
4.21 score on a scale
Standard Deviation 0.80
4.60 score on a scale
Standard Deviation 0.83
Change in Fellow Questionnaire Score
Focus
3.93 score on a scale
Standard Deviation 0.83
4.40 score on a scale
Standard Deviation 0.91
Change in Fellow Questionnaire Score
Depth Perception
4.29 score on a scale
Standard Deviation 0.61
4.73 score on a scale
Standard Deviation 0.59
Change in Fellow Questionnaire Score
Staff Engagement
3.57 score on a scale
Standard Deviation 0.94
4.73 score on a scale
Standard Deviation 0.59
Change in Fellow Questionnaire Score
Teaching
4.00 score on a scale
Standard Deviation 0.78
4.80 score on a scale
Standard Deviation 0.56
Change in Fellow Questionnaire Score
En Face Confidence
4.00 score on a scale
Standard Deviation 0.88
4.33 score on a scale
Standard Deviation 0.82
Change in Fellow Questionnaire Score
Hydrus Placement Confidence
4.29 score on a scale
Standard Deviation 0.61
4.53 score on a scale
Standard Deviation 0.83
Change in Fellow Questionnaire Score
Scope Preference (1 = 3D..)
3.00 score on a scale
Standard Deviation 1.11
1.93 score on a scale
Standard Deviation 1.10
Change in Fellow Questionnaire Score
Visibility
4.21 score on a scale
Standard Deviation 0.80
4.60 score on a scale
Standard Deviation 0.83

Adverse Events

NGENUITY

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Conventional Microscope (Re-randomized)

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

NGENUITY (Re-randomized)

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Conventional Microscope

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Serious adverse events

Adverse event data not reported

Other adverse events

Adverse event data not reported

Additional Information

Robert T. Chang, MD

Stanford University

Phone: 650-725-5743

Results disclosure agreements

  • Principal investigator is a sponsor employee
  • Publication restrictions are in place