Trial Outcomes & Findings for Evaluation of an Open-source Speech-processing Platform (NCT NCT03686046)

NCT ID: NCT03686046

Last Updated: 2026-07-02

Results Overview

5-point rating scale. 5 = definitely would use. 0 = definitely would not use

Recruitment status

COMPLETED

Target enrollment

22 participants

Primary outcome timeframe

1 hour after use of device in lab and field

Results posted on

2026-07-02

Participant Flow

By invitation from SDSU audiology clinic clientele. Based on inclusion/exclusion criteria

Participant milestones

Participant milestones
Measure
Exploratory Use
Brief (2 hour) exploratory use of prototype self-adjusted wearable Master Hearing Aid Exploratory use: Participants will be accompanied by a researcher while wearing the open-source speech processing platform as a hearing aid - adjusted to their needs using a widely accepted Fitting prescription. Using a hand-held controller, they will readjust to their liking as the acoustic conditions change. After returning to the laboratory, they will complete speech perception tests using under their self-selected amplification conditions and respond to a structured interview dealing with perceived performance of the device and its acceptability as a hearing aid during possible field studies.
Overall Study
STARTED
22
Overall Study
COMPLETED
22
Overall Study
NOT COMPLETED
0

Reasons for withdrawal

Withdrawal data not reported

Baseline Characteristics

Evaluation of an Open-source Speech-processing Platform

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
Exploratory Use
n=22 Participants
Brief (2 hour) exploratory use of prototype self-adjusted wearable Master Hearing Aid Exploratory use: Participants will be accompanied by a researcher while wearing the open-source speech processing platform as a hearing aid - adjusted to their needs using a widely accepted Fitting prescription. Using a hand-held controller, they will readjust to their liking as the acoustic conditions change. After returning to the laboratory, they will complete speech perception tests using under their self-selected amplification conditions and respond to a structured interview dealing with perceived performance of the device and its acceptability as a hearing aid during possible field studies.
Age, Categorical
<=18 years
0 Participants
n=20 Participants
Age, Categorical
Between 18 and 65 years
1 Participants
n=20 Participants
Age, Categorical
>=65 years
21 Participants
n=20 Participants
Sex: Female, Male
Female
10 Participants
n=20 Participants
Sex: Female, Male
Male
12 Participants
n=20 Participants
Ethnicity (NIH/OMB)
Hispanic or Latino
1 Participants
n=20 Participants
Ethnicity (NIH/OMB)
Not Hispanic or Latino
21 Participants
n=20 Participants
Ethnicity (NIH/OMB)
Unknown or Not Reported
0 Participants
n=20 Participants
Race (NIH/OMB)
American Indian or Alaska Native
0 Participants
n=20 Participants
Race (NIH/OMB)
Asian
1 Participants
n=20 Participants
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
0 Participants
n=20 Participants
Race (NIH/OMB)
Black or African American
0 Participants
n=20 Participants
Race (NIH/OMB)
White
21 Participants
n=20 Participants
Race (NIH/OMB)
More than one race
0 Participants
n=20 Participants
Race (NIH/OMB)
Unknown or Not Reported
0 Participants
n=20 Participants
Region of Enrollment
United States
22 participants
n=20 Participants
Hearing aid use
22 Participants
n=20 Participants

PRIMARY outcome

Timeframe: 1 hour after use of device in lab and field

Population: Adult hearing-aid users with mild-to-moderate hearing loss.

5-point rating scale. 5 = definitely would use. 0 = definitely would not use

Outcome measures

Outcome measures
Measure
Exploratory Use
n=20 Participants
Brief (2 hour) exploratory use of prototype self-adjusted wearable Master Hearing Aid Exploratory use: Participants will be accompanied by a researcher while wearing the open-source speech processing platform as a hearing aid - adjusted to their needs using a widely accepted Fitting prescription. Using a hand-held controller, they will readjust to their liking as the acoustic conditions change. After returning to the laboratory, they will complete speech perception tests using under their self-selected amplification conditions and respond to a structured interview dealing with perceived performance of the device and its acceptability as a hearing aid during possible field studies.
Willingness to Use During Possible Unsupervised Field Research Study?
Probably, would wear
4 Participants
Willingness to Use During Possible Unsupervised Field Research Study?
Not sure
1 Participants
Willingness to Use During Possible Unsupervised Field Research Study?
Probably would not wear
3 Participants
Willingness to Use During Possible Unsupervised Field Research Study?
Definitely would not wear
11 Participants
Willingness to Use During Possible Unsupervised Field Research Study?
YES,would wear
1 Participants

SECONDARY outcome

Timeframe: 1 hour after self-fitting

Population: Adult hearing-aid users with mild-to-moderate hearing loss

Real-ear device output level in dB SPL was measured, as a function of frequency after self-fitting under 2 conditions: 1. After self-adjustment while listening to conversational-level, speech, prerecorded by a female talker. 2. After a second self-adjustment following a brief, accompanied, field experience including conversation in a variety of acoustic experiences. These results were compared with prescriptions based on each participants audiogram. Note: Although dB outputs were measured at several frequencies, only the average of values at 2, 3, and 4 kHz was used, this being the frequency region at which the greatest effects effects of self-adjustment were observed in earlier studies.

Outcome measures

Outcome measures
Measure
Exploratory Use
n=19 Participants
Brief (2 hour) exploratory use of prototype self-adjusted wearable Master Hearing Aid Exploratory use: Participants will be accompanied by a researcher while wearing the open-source speech processing platform as a hearing aid - adjusted to their needs using a widely accepted Fitting prescription. Using a hand-held controller, they will readjust to their liking as the acoustic conditions change. After returning to the laboratory, they will complete speech perception tests using under their self-selected amplification conditions and respond to a structured interview dealing with perceived performance of the device and its acceptability as a hearing aid during possible field studies.
Self-fitted Hearing-aid Output Averaged at 2, 3, and 4 kHz.
72.5 dB SPL
Standard Error 7.5

SECONDARY outcome

Timeframe: 1 hour after use of Master Hearing Aid

Population: Adults with mild-to-moderate hearing loss

% Recognition of speech sounds in lists of monosyllabic words after self-fitting of the Master Hearing Aid

Outcome measures

Outcome measures
Measure
Exploratory Use
n=20 Participants
Brief (2 hour) exploratory use of prototype self-adjusted wearable Master Hearing Aid Exploratory use: Participants will be accompanied by a researcher while wearing the open-source speech processing platform as a hearing aid - adjusted to their needs using a widely accepted Fitting prescription. Using a hand-held controller, they will readjust to their liking as the acoustic conditions change. After returning to the laboratory, they will complete speech perception tests using under their self-selected amplification conditions and respond to a structured interview dealing with perceived performance of the device and its acceptability as a hearing aid during possible field studies.
Aided Speech Perception
Percent recognition below 80%
3 Participants
Aided Speech Perception
Percent recognition of 80% or more
17 Participants

SECONDARY outcome

Timeframe: 1 hour after final self-fitting in the lab

Population: Adult hearing-aid users with mild-to-moderate hearing loss

Adjustment time

Outcome measures

Outcome measures
Measure
Exploratory Use
n=19 Participants
Brief (2 hour) exploratory use of prototype self-adjusted wearable Master Hearing Aid Exploratory use: Participants will be accompanied by a researcher while wearing the open-source speech processing platform as a hearing aid - adjusted to their needs using a widely accepted Fitting prescription. Using a hand-held controller, they will readjust to their liking as the acoustic conditions change. After returning to the laboratory, they will complete speech perception tests using under their self-selected amplification conditions and respond to a structured interview dealing with perceived performance of the device and its acceptability as a hearing aid during possible field studies.
Self-fitting Behavior
77.6 sec
Standard Deviation 28.2

SECONDARY outcome

Timeframe: 1 hour after final self-fitting in the lab

Population: Adult hearing-aid users with mild-to-moderate hearing loss

Number of adjustment steps the participant took during self adjustment in the laboratory. The available changes were in overall output level (a single button labelled "Loudness"), in low-frequency output alone (a single button labeled "Fullness"), and high-frequency output alone (a single button labeled "Crispness"). Each increase or decrease was in 3 dB steps.

Outcome measures

Outcome measures
Measure
Exploratory Use
n=19 Participants
Brief (2 hour) exploratory use of prototype self-adjusted wearable Master Hearing Aid Exploratory use: Participants will be accompanied by a researcher while wearing the open-source speech processing platform as a hearing aid - adjusted to their needs using a widely accepted Fitting prescription. Using a hand-held controller, they will readjust to their liking as the acoustic conditions change. After returning to the laboratory, they will complete speech perception tests using under their self-selected amplification conditions and respond to a structured interview dealing with perceived performance of the device and its acceptability as a hearing aid during possible field studies.
Self-fitting Behavior
14.6 Number of adjustment steps
Standard Deviation 7.9

SECONDARY outcome

Timeframe: 1 hour after final self-fitting in the lab

Population: Adult hearing-aid users with mild-to-moderate hearing loss

Ranges of output

Outcome measures

Outcome measures
Measure
Exploratory Use
n=19 Participants
Brief (2 hour) exploratory use of prototype self-adjusted wearable Master Hearing Aid Exploratory use: Participants will be accompanied by a researcher while wearing the open-source speech processing platform as a hearing aid - adjusted to their needs using a widely accepted Fitting prescription. Using a hand-held controller, they will readjust to their liking as the acoustic conditions change. After returning to the laboratory, they will complete speech perception tests using under their self-selected amplification conditions and respond to a structured interview dealing with perceived performance of the device and its acceptability as a hearing aid during possible field studies.
Self-fitting Behavior
2.5 dB
Standard Deviation 2.2

SECONDARY outcome

Timeframe: 1 hour after final self-fitting in the lab

Population: Adult hearing-aid users with mild-to-moderate hearing loss

Spectral tilt explored: The difference between self-adjusted outputs at 750 and 1500 Hz, divided by 2, was defined as "spectral tilt" and recorded in dB per octave. Individual values were averaged and a standard deviation clculated.

Outcome measures

Outcome measures
Measure
Exploratory Use
n=19 Participants
Brief (2 hour) exploratory use of prototype self-adjusted wearable Master Hearing Aid Exploratory use: Participants will be accompanied by a researcher while wearing the open-source speech processing platform as a hearing aid - adjusted to their needs using a widely accepted Fitting prescription. Using a hand-held controller, they will readjust to their liking as the acoustic conditions change. After returning to the laboratory, they will complete speech perception tests using under their self-selected amplification conditions and respond to a structured interview dealing with perceived performance of the device and its acceptability as a hearing aid during possible field studies.
Self-fitting Behavior
4.5 dB/octave
Standard Deviation 5.4

Adverse Events

Exploratory Use

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Serious adverse events

Adverse event data not reported

Other adverse events

Adverse event data not reported

Additional Information

Dr. Arthur Boothroyd

San Diego State University

Phone: 6195508951

Results disclosure agreements

  • Principal investigator is a sponsor employee
  • Publication restrictions are in place