Trial Outcomes & Findings for Alternate Methodology of Pulse Oximeter Validation (NCT NCT03628560)

NCT ID: NCT03628560

Last Updated: 2026-08-11

Results Overview

The unit of measure of accuracy is called Accuracy Root Mean Square Error (A\[RMS\]) and is a composite value of bias and precision. A\[RMS\] is reported as a single percentage value representing the overall accuracy of a medical device with lower values indicating better performance.

Recruitment status

TERMINATED

Study phase

NA

Target enrollment

2 participants

Primary outcome timeframe

1 study day

Results posted on

2026-08-11

Participant Flow

Participant milestones

Participant milestones
Measure
Desaturation
The study will consist of five controlled desaturation runs with SaO2 reduced from 100 to 70% followed by a return to 100%: Run #1: RespirAct step \[normocapnia\] Run #2: RespirAct slope \[normocapnia\] Run #3: RespirAct step \[hypocapnia\] Run #4: RespirAct slope \[hypocapnia\] Run #5: ROBD step
Run #1: RespirAct Step [Normocapnia]
STARTED
2
Run #1: RespirAct Step [Normocapnia]
COMPLETED
2
Run #1: RespirAct Step [Normocapnia]
NOT COMPLETED
0
Run #2: RespirAct Slope [Normocapnia]
STARTED
2
Run #2: RespirAct Slope [Normocapnia]
COMPLETED
2
Run #2: RespirAct Slope [Normocapnia]
NOT COMPLETED
0
Run #3: RespirAct Step [Hypocapnia]
STARTED
2
Run #3: RespirAct Step [Hypocapnia]
COMPLETED
2
Run #3: RespirAct Step [Hypocapnia]
NOT COMPLETED
0
Run #4: RespirAct Slope [Hypocapnia]
STARTED
2
Run #4: RespirAct Slope [Hypocapnia]
COMPLETED
2
Run #4: RespirAct Slope [Hypocapnia]
NOT COMPLETED
0
Run #5: ROBD Step
STARTED
2
Run #5: ROBD Step
COMPLETED
2
Run #5: ROBD Step
NOT COMPLETED
0

Reasons for withdrawal

Withdrawal data not reported

Baseline Characteristics

Alternate Methodology of Pulse Oximeter Validation

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
Desaturation
n=2 Participants
The study will consist of five controlled desaturation runs with SaO2 reduced from 100 to 70% followed by a return to 100%: Run #1: RespirAct step \[normocapnia\] Run #2: RespirAct slope \[normocapnia\] Run #3: RespirAct step \[hypocapnia\] Run #4: RespirAct slope \[hypocapnia\] Run #5: ROBD step
Age, Categorical
<=18 years
0 Participants
n=54 Participants
Age, Categorical
Between 18 and 65 years
2 Participants
n=54 Participants
Age, Categorical
>=65 years
0 Participants
n=54 Participants
Sex: Female, Male
Female
1 Participants
n=54 Participants
Sex: Female, Male
Male
1 Participants
n=54 Participants
Race (NIH/OMB)
American Indian or Alaska Native
0 Participants
n=54 Participants
Race (NIH/OMB)
Asian
0 Participants
n=54 Participants
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
0 Participants
n=54 Participants
Race (NIH/OMB)
Black or African American
0 Participants
n=54 Participants
Race (NIH/OMB)
White
1 Participants
n=54 Participants
Race (NIH/OMB)
More than one race
1 Participants
n=54 Participants
Race (NIH/OMB)
Unknown or Not Reported
0 Participants
n=54 Participants
Ethnicity (NIH/OMB)
Hispanic or Latino
1 Participants
n=54 Participants
Ethnicity (NIH/OMB)
Not Hispanic or Latino
1 Participants
n=54 Participants
Ethnicity (NIH/OMB)
Unknown or Not Reported
0 Participants
n=54 Participants
Region of Enrollment
United States
2 Participants
n=54 Participants

PRIMARY outcome

Timeframe: 1 study day

Population: This analysis only applies to Runs #1-4 as the ROBD device is distinct from the RespirAct and cannot provide the measured parameter.

The unit of measure of accuracy is called Accuracy Root Mean Square Error (A\[RMS\]) and is a composite value of bias and precision. A\[RMS\] is reported as a single percentage value representing the overall accuracy of a medical device with lower values indicating better performance.

Outcome measures

Outcome measures
Measure
Desaturation
n=2 Participants
The study will consist of five controlled desaturation runs with SaO2 reduced from 100 to 70% followed by a return to 100%: Run #1: RespirAct step \[normocapnia\] Run #2: RespirAct slope \[normocapnia\] Run #3: RespirAct step \[hypocapnia\] Run #4: RespirAct slope \[hypocapnia\] Run #5: ROBD step
Accuracy of FDA-approved Pulse Oximeter When Calculated With RespirAct-derived End-tidal PAO2 Values
Run #4: RespirAct Slope Hypocapnia desaturation sequence
6.7 percentage
Accuracy of FDA-approved Pulse Oximeter When Calculated With RespirAct-derived End-tidal PAO2 Values
Run #1: RespirAct Step-Wise Normocapnia desaturation sequence
3.1 percentage
Accuracy of FDA-approved Pulse Oximeter When Calculated With RespirAct-derived End-tidal PAO2 Values
Run #2: RespirAct Slope Normocapnia desaturation sequence
5.3 percentage
Accuracy of FDA-approved Pulse Oximeter When Calculated With RespirAct-derived End-tidal PAO2 Values
Run #3: RespirAct Step-Wise Hypocapnia desaturation sequence
5.4 percentage

PRIMARY outcome

Timeframe: 1 study day

The unit of measure of accuracy is called Accuracy Root Mean Square Error (A\[RMS\]) and is a composite value of bias and precision. A\[RMS\] is reported as a single percentage value representing the overall accuracy of a medical device with lower values indicating better performance.

Outcome measures

Outcome measures
Measure
Desaturation
n=2 Participants
The study will consist of five controlled desaturation runs with SaO2 reduced from 100 to 70% followed by a return to 100%: Run #1: RespirAct step \[normocapnia\] Run #2: RespirAct slope \[normocapnia\] Run #3: RespirAct step \[hypocapnia\] Run #4: RespirAct slope \[hypocapnia\] Run #5: ROBD step
Accuracy of FDA-approved Pulse Oximeter When Calculated With Co-oximeter Measured Arterial PaO2 Values
Run #1: RespirAct Step-Wise Normocapnia desaturation sequence
2.9 percentage
Accuracy of FDA-approved Pulse Oximeter When Calculated With Co-oximeter Measured Arterial PaO2 Values
Run #2: RespirAct Slope Normocapnia desaturation sequence
2.2 percentage
Accuracy of FDA-approved Pulse Oximeter When Calculated With Co-oximeter Measured Arterial PaO2 Values
Run #3: RespirAct Step-Wise Hypocapnia desaturation sequence
2.2 percentage
Accuracy of FDA-approved Pulse Oximeter When Calculated With Co-oximeter Measured Arterial PaO2 Values
Run #4: RespirAct Slope Hypocapnia desaturation sequence
1.3 percentage
Accuracy of FDA-approved Pulse Oximeter When Calculated With Co-oximeter Measured Arterial PaO2 Values
Run #5: ROBD Step-Wise desaturation sequence
2.2 percentage

PRIMARY outcome

Timeframe: 1 study day

Population: This analysis only applies to Runs #1-4 as the ROBD device is distinct from the RespirAct and cannot provide the measured parameter.

A\[RMS\] of RespirAct-derived end-tidal PAO2 values with co-oximeter measured arterial PaO2 values. The unit of measure of accuracy is called Accuracy Root Mean Square Error (A\[RMS\]) and is a composite value of bias and precision. A\[RMS\] is reported as a single percentage value representing the overall accuracy of a medical device with lower values indicating better performance.

Outcome measures

Outcome measures
Measure
Desaturation
n=2 Participants
The study will consist of five controlled desaturation runs with SaO2 reduced from 100 to 70% followed by a return to 100%: Run #1: RespirAct step \[normocapnia\] Run #2: RespirAct slope \[normocapnia\] Run #3: RespirAct step \[hypocapnia\] Run #4: RespirAct slope \[hypocapnia\] Run #5: ROBD step
PAO2 Values
Run #1: RespirAct Step-Wise Normocapnia desaturation sequence
6.5 percentage
PAO2 Values
Run #2: RespirAct Slope Normocapnia desaturation sequence
14.2 percentage
PAO2 Values
Run #3: RespirAct Step-Wise Hypocapnia desaturation sequence
1.8 percentage
PAO2 Values
Run #4: RespirAct Slope Hypocapnia desaturation sequence
3.4 percentage

SECONDARY outcome

Timeframe: 1 study day

Population: This analysis only applies to Runs #1-4 as the ROBD device is distinct from the RespirAct and cannot provide the measured parameter. It is necessary to determine which of the four runs were STEP and which were SLOPE for this outcome. Runs #1 \& 3 were the two in which STEP-WISE desaturation was performed; similarly, Runs #2 \& 4 were SLOPE. Note that the amalgamation of the data from the two Runs is based upon STEP/SLOPE and is independent of carbon dioxide value (i.e. Hypocapnia \& Normocapnia).

The unit of measure of accuracy is called Accuracy Root Mean Square Error (A\[RMS\]) and is a composite value of bias and precision. A\[RMS\] is reported as a single percentage value representing the overall accuracy of a medical device with lower values indicating better performance.

Outcome measures

Outcome measures
Measure
Desaturation
n=2 Participants
The study will consist of five controlled desaturation runs with SaO2 reduced from 100 to 70% followed by a return to 100%: Run #1: RespirAct step \[normocapnia\] Run #2: RespirAct slope \[normocapnia\] Run #3: RespirAct step \[hypocapnia\] Run #4: RespirAct slope \[hypocapnia\] Run #5: ROBD step
Accuracy, A[RMS], When a Step-wise Desaturation Sequence is Used Against a Slope Desaturation Sequence.
Step-wise Desaturation (Runs #1 and #3)
2.4 percentage
Accuracy, A[RMS], When a Step-wise Desaturation Sequence is Used Against a Slope Desaturation Sequence.
Slope Desaturation (Runs #2 and #4)
2.2 percentage

SECONDARY outcome

Timeframe: 1 study day

Population: This analysis only applies to Runs #1-4 as the ROBD device is distinct from the RespirAct and cannot provide the measured parameter. It is necessary to determine which of the four runs were NORMOCAPNIA and which were HYPOCAPNIA for this outcome. Runs #1 \& 2 were the two in which NORMOCAPNIA condition was performed; similarly, Runs #3 \& 4 were HYPOCAPNIA. Runs #1 \& 2 are amalgamated to represent the condition of NORMOCAPNIA and Runs #3 \& 4 represent HYPOCAPNIA.

A\[RMS\] when hypocapnia desaturation is compared to normocapnia desaturation. A\[RMS\] is reported as a single percentage value representing the overall accuracy of a medical device with lower values indicating better performance.

Outcome measures

Outcome measures
Measure
Desaturation
n=2 Participants
The study will consist of five controlled desaturation runs with SaO2 reduced from 100 to 70% followed by a return to 100%: Run #1: RespirAct step \[normocapnia\] Run #2: RespirAct slope \[normocapnia\] Run #3: RespirAct step \[hypocapnia\] Run #4: RespirAct slope \[hypocapnia\] Run #5: ROBD step
Level of PaCO2 (Hypocapnia Versus Normocapnia)
Hypocapnia Desaturation (Runs #1 and #2)
2.5 percentage
Level of PaCO2 (Hypocapnia Versus Normocapnia)
Normocapnia Desaturation (Runs #3 and #4)
1.7 percentage

Adverse Events

Desaturation

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Serious adverse events

Adverse event data not reported

Other adverse events

Adverse event data not reported