Trial Outcomes & Findings for Clinical Trial Testing the BioSerenity Pulse Oximeter (EOXY) Efficacy, Performance and Safety in Healthy Volunteer (NCT NCT03614416)
NCT ID: NCT03614416
Last Updated: 2021-01-12
Results Overview
For each desaturation plateau, SpO2 ARMS are calculated by the difference between the investigational pulse oximeter SpO2 measurement and the co-oximeter's SaO2 measurements
COMPLETED
NA
13 participants
1 day
2021-01-12
Participant Flow
Participant milestones
| Measure |
EOXY Device and Gold Standard Oximeter and SaO2 Measures
Heart rate and SPO2 measures provided from EOXY device Heart rate measures provided from gold standard oximeter. SaO2 measures provided from blood sampling There is only one arm: all subjects have simultaneously three interventions (as required in the European standard ISO 80601-2-61), to qualify a pulse oximeter
EOXY device: Sp02 and HR measures from EOXY device are simulateously compared to gold standard measures : Sa02 sampling measures and pulse oximeter gold standard
Gold standard oximeter: Sp02 and HR measures from EOXY device are simulateously compared to gold standard measures : Sa02 sampling measures and pulse oximeter gold standard
SaO2 sampling: Sp02 and HR measures from EOXY device are simulateously compared to gold standard measures : Sa02 sampling measures and pulse oximeter gold standard
|
|---|---|
|
Overall Study
STARTED
|
13
|
|
Overall Study
COMPLETED
|
13
|
|
Overall Study
NOT COMPLETED
|
0
|
Reasons for withdrawal
Withdrawal data not reported
Baseline Characteristics
Clinical Trial Testing the BioSerenity Pulse Oximeter (EOXY) Efficacy, Performance and Safety in Healthy Volunteer
Baseline characteristics by cohort
| Measure |
EOXY Device and Gold Standard Oximeter and SaO2 Measures
n=10 Participants
Heart rate and SPO2 measures provided from EOXY device Heart rate measures provided from gold standard oximeter. SaO2 measures provided from blood sampling There is only one arm: all subjects have simultaneously three interventions (as required in the European standard ISO 80601-2-61), to qualify a pulse oximeter
EOXY device: Sp02 and HR measures from EOXY device are simulateously compared to gold standard measures : Sa02 sampling measures and pulse oximeter gold standard
Gold standard oximeter: Sp02 and HR measures from EOXY device are simulateously compared to gold standard measures : Sa02 sampling measures and pulse oximeter gold standard
SaO2 sampling: Sp02 and HR measures from EOXY device are simulateously compared to gold standard measures : Sa02 sampling measures and pulse oximeter gold standard
|
|---|---|
|
Age, Categorical
<=18 years
|
0 Participants
n=99 Participants
|
|
Age, Categorical
Between 18 and 65 years
|
10 Participants
n=99 Participants
|
|
Age, Categorical
>=65 years
|
0 Participants
n=99 Participants
|
|
Age, Continuous
|
23.5 years
n=99 Participants
|
|
Sex: Female, Male
Female
|
3 Participants
n=99 Participants
|
|
Sex: Female, Male
Male
|
7 Participants
n=99 Participants
|
|
Race (NIH/OMB)
American Indian or Alaska Native
|
0 Participants
n=99 Participants
|
|
Race (NIH/OMB)
Asian
|
0 Participants
n=99 Participants
|
|
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
|
0 Participants
n=99 Participants
|
|
Race (NIH/OMB)
Black or African American
|
2 Participants
n=99 Participants
|
|
Race (NIH/OMB)
White
|
6 Participants
n=99 Participants
|
|
Race (NIH/OMB)
More than one race
|
2 Participants
n=99 Participants
|
|
Race (NIH/OMB)
Unknown or Not Reported
|
0 Participants
n=99 Participants
|
|
Region of Enrollment
Belgium
|
10 participants
n=99 Participants
|
PRIMARY outcome
Timeframe: 1 dayPopulation: Subjects respecting the study conditions (no movements, good irrigation) 3 subjects excluded from analysis(no respect of study conditions) so 10 subjects analyzed
For each desaturation plateau, SpO2 ARMS are calculated by the difference between the investigational pulse oximeter SpO2 measurement and the co-oximeter's SaO2 measurements
Outcome measures
| Measure |
EOXY Device and Gold Standard Oximeter and SaO2 Measures
n=10 Participants
Heart rate and SPO2 measures provided from EOXY device Heart rate measures provided from gold standard oximeter. SaO2 measures provided from blood sampling There is only one arm: all subjects have simultaneously three interventions (as required in the European standard ISO 80601-2-61), to qualify a pulse oximeter
EOXY device: Sp02 and HR measures from EOXY device are simulateously compared to gold standard measures : Sa02 sampling measures and pulse oximeter gold standard
Gold standard oximeter: Sp02 and HR measures from EOXY device are simulateously compared to gold standard measures : Sa02 sampling measures and pulse oximeter gold standard
SaO2 sampling: Sp02 and HR measures from EOXY device are simulateously compared to gold standard measures : Sa02 sampling measures and pulse oximeter gold standard
|
|---|---|
|
Percentage SPO2 Accuracy Root-Mean Square (ARMS) at Forehead Position
Overall range 70-100% of O2
|
3.91 percentage of ARMS (accuracy)
|
|
Percentage SPO2 Accuracy Root-Mean Square (ARMS) at Forehead Position
97-100% of O2
|
3.08 percentage of ARMS (accuracy)
|
|
Percentage SPO2 Accuracy Root-Mean Square (ARMS) at Forehead Position
97-92% of O2
|
3.36 percentage of ARMS (accuracy)
|
|
Percentage SPO2 Accuracy Root-Mean Square (ARMS) at Forehead Position
92-85% of O2
|
2.85 percentage of ARMS (accuracy)
|
|
Percentage SPO2 Accuracy Root-Mean Square (ARMS) at Forehead Position
84-78% of O2
|
3.48 percentage of ARMS (accuracy)
|
|
Percentage SPO2 Accuracy Root-Mean Square (ARMS) at Forehead Position
77-70% of O2
|
6.21 percentage of ARMS (accuracy)
|
PRIMARY outcome
Timeframe: 1 dayPopulation: Subjects respecting the study conditions (no movements, good irrigation) 3 subjects excluded from analysis(no respect of study conditions) so 10 subjects analyzed
For each desaturation plateau, SpO2 ARMS (Accuracy Root-Mean Square) are calculated by the difference between the investigational pulse oximeter SpO2 measurement and the co-oximeter's SaO2 measurements
Outcome measures
| Measure |
EOXY Device and Gold Standard Oximeter and SaO2 Measures
n=10 Participants
Heart rate and SPO2 measures provided from EOXY device Heart rate measures provided from gold standard oximeter. SaO2 measures provided from blood sampling There is only one arm: all subjects have simultaneously three interventions (as required in the European standard ISO 80601-2-61), to qualify a pulse oximeter
EOXY device: Sp02 and HR measures from EOXY device are simulateously compared to gold standard measures : Sa02 sampling measures and pulse oximeter gold standard
Gold standard oximeter: Sp02 and HR measures from EOXY device are simulateously compared to gold standard measures : Sa02 sampling measures and pulse oximeter gold standard
SaO2 sampling: Sp02 and HR measures from EOXY device are simulateously compared to gold standard measures : Sa02 sampling measures and pulse oximeter gold standard
|
|---|---|
|
Percentage SPO2 Accuracy Root-Mean Square (ARMS) at Torso Position
Overall range 70-100% of O2
|
11.14 percentage of ARMS (accuracy)
|
|
Percentage SPO2 Accuracy Root-Mean Square (ARMS) at Torso Position
97-100% of O2
|
12.99 percentage of ARMS (accuracy)
|
|
Percentage SPO2 Accuracy Root-Mean Square (ARMS) at Torso Position
97-92% of O2
|
12.51 percentage of ARMS (accuracy)
|
|
Percentage SPO2 Accuracy Root-Mean Square (ARMS) at Torso Position
92-85% of O2
|
9.91 percentage of ARMS (accuracy)
|
|
Percentage SPO2 Accuracy Root-Mean Square (ARMS) at Torso Position
84-78% of O2
|
7.31 percentage of ARMS (accuracy)
|
|
Percentage SPO2 Accuracy Root-Mean Square (ARMS) at Torso Position
77-70% of O2
|
10.76 percentage of ARMS (accuracy)
|
PRIMARY outcome
Timeframe: 1 dayFor each desaturation plateau, SpO2 ARMS are calculated by the difference between the investigational pulse oximeter SpO2 measurement and the co-oximeter's SaO2 measurements
Outcome measures
| Measure |
EOXY Device and Gold Standard Oximeter and SaO2 Measures
n=10 Participants
Heart rate and SPO2 measures provided from EOXY device Heart rate measures provided from gold standard oximeter. SaO2 measures provided from blood sampling There is only one arm: all subjects have simultaneously three interventions (as required in the European standard ISO 80601-2-61), to qualify a pulse oximeter
EOXY device: Sp02 and HR measures from EOXY device are simulateously compared to gold standard measures : Sa02 sampling measures and pulse oximeter gold standard
Gold standard oximeter: Sp02 and HR measures from EOXY device are simulateously compared to gold standard measures : Sa02 sampling measures and pulse oximeter gold standard
SaO2 sampling: Sp02 and HR measures from EOXY device are simulateously compared to gold standard measures : Sa02 sampling measures and pulse oximeter gold standard
|
|---|---|
|
Percentage SPO2 Accuracy Root-Mean Square (ARMS) at BACK Position
Overall range 70-100% of O2
|
6.02 percentage of ARMS (accuracy)
|
|
Percentage SPO2 Accuracy Root-Mean Square (ARMS) at BACK Position
97-100% of O2
|
7.09 percentage of ARMS (accuracy)
|
|
Percentage SPO2 Accuracy Root-Mean Square (ARMS) at BACK Position
97-92% of O2
|
6.63 percentage of ARMS (accuracy)
|
|
Percentage SPO2 Accuracy Root-Mean Square (ARMS) at BACK Position
92-85% of O2
|
6.57 percentage of ARMS (accuracy)
|
|
Percentage SPO2 Accuracy Root-Mean Square (ARMS) at BACK Position
84-78% of O2
|
5.06 percentage of ARMS (accuracy)
|
|
Percentage SPO2 Accuracy Root-Mean Square (ARMS) at BACK Position
77-70% of O2
|
3.96 percentage of ARMS (accuracy)
|
PRIMARY outcome
Timeframe: 1 dayPopulation: People respecting study conditions (good irrigation and no movements)
For each desaturation plateau, HR ARMS are calculated by the difference between the investigational pulse oximeter HR measurement and the HR gold standard measurements
Outcome measures
| Measure |
EOXY Device and Gold Standard Oximeter and SaO2 Measures
n=10 Participants
Heart rate and SPO2 measures provided from EOXY device Heart rate measures provided from gold standard oximeter. SaO2 measures provided from blood sampling There is only one arm: all subjects have simultaneously three interventions (as required in the European standard ISO 80601-2-61), to qualify a pulse oximeter
EOXY device: Sp02 and HR measures from EOXY device are simulateously compared to gold standard measures : Sa02 sampling measures and pulse oximeter gold standard
Gold standard oximeter: Sp02 and HR measures from EOXY device are simulateously compared to gold standard measures : Sa02 sampling measures and pulse oximeter gold standard
SaO2 sampling: Sp02 and HR measures from EOXY device are simulateously compared to gold standard measures : Sa02 sampling measures and pulse oximeter gold standard
|
|---|---|
|
HR Accuracy Root-Mean Square (ARMS) at Forehead Position
92-85% of O2
|
11.5 percentage of ARMS (accuracy)
|
|
HR Accuracy Root-Mean Square (ARMS) at Forehead Position
Overall range 70-100% of O2
|
12.2 percentage of ARMS (accuracy)
|
|
HR Accuracy Root-Mean Square (ARMS) at Forehead Position
97-100% of O2
|
10.0 percentage of ARMS (accuracy)
|
|
HR Accuracy Root-Mean Square (ARMS) at Forehead Position
97-92% of O2
|
11.7 percentage of ARMS (accuracy)
|
|
HR Accuracy Root-Mean Square (ARMS) at Forehead Position
84-78% of O2
|
14.3 percentage of ARMS (accuracy)
|
|
HR Accuracy Root-Mean Square (ARMS) at Forehead Position
77-70% of O2
|
13.0 percentage of ARMS (accuracy)
|
PRIMARY outcome
Timeframe: 1 dayPopulation: People respecting study conditions (good irrigation, no movements)
For each desaturation plateau, HR ARMS are calculated by the difference between the investigational pulse oximeter HR measurement and the HR gold standard measurements
Outcome measures
| Measure |
EOXY Device and Gold Standard Oximeter and SaO2 Measures
n=10 Participants
Heart rate and SPO2 measures provided from EOXY device Heart rate measures provided from gold standard oximeter. SaO2 measures provided from blood sampling There is only one arm: all subjects have simultaneously three interventions (as required in the European standard ISO 80601-2-61), to qualify a pulse oximeter
EOXY device: Sp02 and HR measures from EOXY device are simulateously compared to gold standard measures : Sa02 sampling measures and pulse oximeter gold standard
Gold standard oximeter: Sp02 and HR measures from EOXY device are simulateously compared to gold standard measures : Sa02 sampling measures and pulse oximeter gold standard
SaO2 sampling: Sp02 and HR measures from EOXY device are simulateously compared to gold standard measures : Sa02 sampling measures and pulse oximeter gold standard
|
|---|---|
|
Percentage of HR Accuracy Root-Mean Square (ARMS) at TORSO Position
Overall range 70-100% of O2
|
32.1 percentage of ARMS (accuracy)
|
|
Percentage of HR Accuracy Root-Mean Square (ARMS) at TORSO Position
97-100% of O2
|
30.8 percentage of ARMS (accuracy)
|
|
Percentage of HR Accuracy Root-Mean Square (ARMS) at TORSO Position
97-92% of O2
|
31.5 percentage of ARMS (accuracy)
|
|
Percentage of HR Accuracy Root-Mean Square (ARMS) at TORSO Position
92-85% of O2
|
31.5 percentage of ARMS (accuracy)
|
|
Percentage of HR Accuracy Root-Mean Square (ARMS) at TORSO Position
84-78% of O2
|
30.4 percentage of ARMS (accuracy)
|
|
Percentage of HR Accuracy Root-Mean Square (ARMS) at TORSO Position
77-70% of O2
|
36.5 percentage of ARMS (accuracy)
|
PRIMARY outcome
Timeframe: 1 dayPopulation: People respecting study conditions (good irrigation and no movements)
For each desaturation plateau, HR ARMS are calculated by the difference between the investigational pulse oximeter HR measurement and the HR gold standard measurements
Outcome measures
| Measure |
EOXY Device and Gold Standard Oximeter and SaO2 Measures
n=10 Participants
Heart rate and SPO2 measures provided from EOXY device Heart rate measures provided from gold standard oximeter. SaO2 measures provided from blood sampling There is only one arm: all subjects have simultaneously three interventions (as required in the European standard ISO 80601-2-61), to qualify a pulse oximeter
EOXY device: Sp02 and HR measures from EOXY device are simulateously compared to gold standard measures : Sa02 sampling measures and pulse oximeter gold standard
Gold standard oximeter: Sp02 and HR measures from EOXY device are simulateously compared to gold standard measures : Sa02 sampling measures and pulse oximeter gold standard
SaO2 sampling: Sp02 and HR measures from EOXY device are simulateously compared to gold standard measures : Sa02 sampling measures and pulse oximeter gold standard
|
|---|---|
|
Percentage of HR Accuracy Root-Mean Square (ARMS) at BACK Position
Overall range 70-100% of O2
|
21.1 percentage of ARMS (accuracy)
|
|
Percentage of HR Accuracy Root-Mean Square (ARMS) at BACK Position
97-100% of O2
|
22.3 percentage of ARMS (accuracy)
|
|
Percentage of HR Accuracy Root-Mean Square (ARMS) at BACK Position
97-92% of O2
|
24.6 percentage of ARMS (accuracy)
|
|
Percentage of HR Accuracy Root-Mean Square (ARMS) at BACK Position
92-85% of O2
|
19.2 percentage of ARMS (accuracy)
|
|
Percentage of HR Accuracy Root-Mean Square (ARMS) at BACK Position
84-78% of O2
|
19.1 percentage of ARMS (accuracy)
|
|
Percentage of HR Accuracy Root-Mean Square (ARMS) at BACK Position
77-70% of O2
|
20.6 percentage of ARMS (accuracy)
|
Adverse Events
EOXY Device and Gold Standard Oximeter and SaO2 Measures
Serious adverse events
Adverse event data not reported
Other adverse events
Adverse event data not reported
Additional Information
Results disclosure agreements
- Principal investigator is a sponsor employee
- Publication restrictions are in place