Trial Outcomes & Findings for Accuracy and Precision Comparison of Temptraq (NCT NCT03264833)
NCT ID: NCT03264833
Last Updated: 2026-02-12
Results Overview
Core temperature monitoring (degree Celsius) with temp-Traq thermometer compared to Pulmonary Artery (PA) catheter
COMPLETED
60 participants
Multiple time points up to 6 hours post thermometer placement
2026-02-12
Participant Flow
Patients were recruited in a medical ICU over a 6 month period
Participant milestones
| Measure |
Longitudinal Single Arm Study
Patients were assessed at baseline, hour 2, hour 4 and hour 6 Temperature (Celsius) data were compared between TempTraq reading and pulmonary artery (central, internal; used as the gold standard) readings at 4 time points
|
|---|---|
|
Overall Study
STARTED
|
60
|
|
Overall Study
COMPLETED
|
60
|
|
Overall Study
NOT COMPLETED
|
0
|
Reasons for withdrawal
Withdrawal data not reported
Baseline Characteristics
Accuracy and Precision Comparison of Temptraq
Baseline characteristics by cohort
| Measure |
Longitudinal Single Arm Study
n=60 Participants
Medical ICU patients with a pulmonary artery catheter; obtained core temp (PA catheter) and compared to TempTraq temperatures in Celsius
|
|---|---|
|
Age, Continuous
|
65.8 years
STANDARD_DEVIATION 11.13 • n=41 Participants
|
|
Sex: Female, Male
Female
|
20 Participants
n=41 Participants
|
|
Sex: Female, Male
Male
|
40 Participants
n=41 Participants
|
|
Race/Ethnicity, Customized
White
|
40 participants
n=41 Participants
|
|
Race/Ethnicity, Customized
Black
|
20 participants
n=41 Participants
|
|
pulmonary artery catheter core temperature sensor
PA-Celsius
|
36.84 Degree Celsius
n=41 Participants
|
|
pulmonary artery catheter core temperature sensor
TempTraq- Celsius
|
36.3 Degree Celsius
n=41 Participants
|
PRIMARY outcome
Timeframe: Multiple time points up to 6 hours post thermometer placementPopulation: Adults in a medical intensive care unit with pulmonary artery catheter monitoring that included core temperature
Core temperature monitoring (degree Celsius) with temp-Traq thermometer compared to Pulmonary Artery (PA) catheter
Outcome measures
| Measure |
Longitudinal Single Arm Study
n=60 Participants
Mean differences in Core temperature (degree Celsius) monitoring with temp-Traq thermometer compared to Pulmonary Artery (PA) catheter
|
|---|---|
|
Temperature Comparison
Initial placement
|
-0.60 Degree Celsius
Interval -0.78 to -0.42
|
|
Temperature Comparison
2 hours post placement
|
-0.24 Degree Celsius
Interval -0.39 to -0.097
|
|
Temperature Comparison
4 hours post placement
|
-0.16 Degree Celsius
Interval -0.3 to -0.03
|
|
Temperature Comparison
6 hours post placement
|
-0.19 Degree Celsius
Interval -0.37 to -0.02
|
|
Temperature Comparison
Cumulative change
|
-0.30 Degree Celsius
Interval -0.39 to -0.22
|
Adverse Events
Longitudinal Single Arm Study
Serious adverse events
Adverse event data not reported
Other adverse events
Adverse event data not reported
Additional Information
Results disclosure agreements
- Principal investigator is a sponsor employee
- Publication restrictions are in place