Trial Outcomes & Findings for PARTNER II Trial: Placement of AoRTic TraNscathetER Valves II - Nested Registry 3/Valve-in-Valve (NCT NCT03225001)

NCT ID: NCT03225001

Last Updated: 2021-03-09

Results Overview

The primary endpoint of all-cause mortality, all stroke, moderate or severe obstruction, or moderate or severe paravalvular leak

Recruitment status

COMPLETED

Study phase

NA

Target enrollment

197 participants

Primary outcome timeframe

30-day

Results posted on

2021-03-09

Participant Flow

Participant milestones

Participant milestones
Measure
Edwards SAPIEN XT THV in Patients With Failing SAVR
Patients with a failing surgical bioprosthetic valve (SAVR) in the aortic position demonstrating stenosis and/or insufficiency will be treated with Edwards SAPIEN XT transcatheter valve. Edwards SAPIEN XT transcatheter valve, Model 9300TFX: Edwards SAPIEN XT THV system Model 9300TFX with the associated delivery systems.
Overall Study
STARTED
197
Overall Study
Valve-Implanted
195
Overall Study
COMPLETED
188
Overall Study
NOT COMPLETED
9

Reasons for withdrawal

Withdrawal data not reported

Baseline Characteristics

PARTNER II Trial: Placement of AoRTic TraNscathetER Valves II - Nested Registry 3/Valve-in-Valve

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
Edwards SAPIEN XT THV in Patients With Failing SAVR
n=197 Participants
Patients with a failing surgical bioprosthetic valve (SAVR) in the aortic position demonstrating stenosis and/or insufficiency will be treated with Edwards SAPIEN XT transcatheter valve. Edwards SAPIEN XT transcatheter valve, Model 9300TFX: Edwards SAPIEN XT THV system Model 9300TFX with the associated delivery systems.
Age, Continuous
78.5 Years
STANDARD_DEVIATION 11.0 • n=99 Participants
Sex: Female, Male
Female
78 Participants
n=99 Participants
Sex: Female, Male
Male
119 Participants
n=99 Participants
STS Score
9.7 units on a scale
STANDARD_DEVIATION 5.1 • n=99 Participants
NYHA III/IV
188 Participants
n=99 Participants

PRIMARY outcome

Timeframe: 30-day

Population: Twenty-nine patients had missing echocardiography data.

The primary endpoint of all-cause mortality, all stroke, moderate or severe obstruction, or moderate or severe paravalvular leak

Outcome measures

Outcome measures
Measure
Edwards SAPIEN XT THV in Patients With Failing SAVR
n=166 Participants
Patients with a failing surgical bioprosthetic valve (SAVR) in the aortic position demonstrating stenosis and/or insufficiency will be treated with Edwards SAPIEN XT transcatheter valve. Edwards SAPIEN XT transcatheter valve, Model 9300TFX: Edwards SAPIEN XT THV system Model 9300TFX with the associated delivery systems.
Number of Participants With All-Cause Mortality, All Stroke, Moderate or Severe Obstruction, or Moderate or Severe Paravalvular Leak (Composite)
28 Participants

SECONDARY outcome

Timeframe: 30 Days

Cardiovascular cause is also included

Outcome measures

Outcome measures
Measure
Edwards SAPIEN XT THV in Patients With Failing SAVR
n=195 Participants
Patients with a failing surgical bioprosthetic valve (SAVR) in the aortic position demonstrating stenosis and/or insufficiency will be treated with Edwards SAPIEN XT transcatheter valve. Edwards SAPIEN XT transcatheter valve, Model 9300TFX: Edwards SAPIEN XT THV system Model 9300TFX with the associated delivery systems.
Number of Participants With Mortality From Any Cause
8 Participants

Adverse Events

Edwards SAPIEN XT THV in Patients With Failing SAVR

Serious events: 109 serious events
Other events: 142 other events
Deaths: 8 deaths

Serious adverse events

Serious adverse events
Measure
Edwards SAPIEN XT THV in Patients With Failing SAVR
n=195 participants at risk
Patients with a failing surgical bioprosthetic valve (SAVR) in the aortic position demonstrating stenosis and/or insufficiency will be treated with Edwards SAPIEN XT transcatheter valve. Edwards SAPIEN XT transcatheter valve, Model 9300TFX: Edwards SAPIEN XT THV system Model 9300TFX with the associated delivery systems.
Investigations
Abnormal Lab Value
1.5%
3/195 • Number of events 3 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Vascular disorders
Anemia/Hemolytic Anemia
1.5%
3/195 • Number of events 4 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Cardiac disorders
Angina/Cardiac Chest Pain
2.1%
4/195 • Number of events 5 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Cardiac disorders
Arrhythmia/Conduction System Injury (Defect)
5.6%
11/195 • Number of events 12 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Infections and infestations
Bacteremia
3.1%
6/195 • Number of events 8 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Neoplasms benign, malignant and unspecified (incl cysts and polyps)
Cancer
2.1%
4/195 • Number of events 4 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Cardiac disorders
Cardiac Arrest
2.6%
5/195 • Number of events 5 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Cardiac disorders
Cardiogenic Shock
0.51%
1/195 • Number of events 1 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Nervous system disorders
Change in level of consciousness
0.51%
1/195 • Number of events 1 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Injury, poisoning and procedural complications
Compartment syndrome
0.51%
1/195 • Number of events 1 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Injury, poisoning and procedural complications
Coronary Flow Obstruction/Transvalvular flow disturbance
1.0%
2/195 • Number of events 2 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Respiratory, thoracic and mediastinal disorders
Dyspnea
2.6%
5/195 • Number of events 7 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Injury, poisoning and procedural complications
Embolization including air/calcific valve material or Thrombus
0.51%
1/195 • Number of events 1 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Nervous system disorders
Encephalopathy
1.0%
2/195 • Number of events 2 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Infections and infestations
Endocarditis
0.51%
1/195 • Number of events 1 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Injury, poisoning and procedural complications
Failure of percutaneous access site closure resulting in intervention
0.51%
1/195 • Number of events 1 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Injury, poisoning and procedural complications
Fracture
3.1%
6/195 • Number of events 6 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Gastrointestinal disorders
Gastro Intestinal Event
2.6%
5/195 • Number of events 6 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Gastrointestinal disorders
Gastro Intestinal Event Nonhemorrhagic
3.1%
6/195 • Number of events 7 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Renal and urinary disorders
Genito Urinary Event
0.51%
1/195 • Number of events 1 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Renal and urinary disorders
Genito Urinary Event Nonhemorrhagic
0.51%
1/195 • Number of events 1 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Cardiac disorders
Heart Failure / CHF / Low output Failure
13.3%
26/195 • Number of events 28 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Vascular disorders
Hemorrhage
11.3%
22/195 • Number of events 22 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Nervous system disorders
Hemorrhagic Stroke
1.0%
2/195 • Number of events 2 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Vascular disorders
Hypertension
0.51%
1/195 • Number of events 1 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Vascular disorders
Hypotension
1.5%
3/195 • Number of events 3 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Vascular disorders
Ischemic Stroke
2.1%
4/195 • Number of events 5 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Injury, poisoning and procedural complications
Mitral Valve Injury or Insufficiency
0.51%
1/195 • Number of events 1 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Investigations
Multi Organ Failure
0.51%
1/195 • Number of events 1 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Cardiac disorders
Myocardial Infarction
1.5%
3/195 • Number of events 3 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Injury, poisoning and procedural complications
Non Structural Dysfunction
1.5%
3/195 • Number of events 3 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Nervous system disorders
Pain
3.6%
7/195 • Number of events 7 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Injury, poisoning and procedural complications
Perforation
1.0%
2/195 • Number of events 2 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Respiratory, thoracic and mediastinal disorders
Pneumothorax
0.51%
1/195 • Number of events 1 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Vascular disorders
Pseudoaneurysm
0.51%
1/195 • Number of events 1 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Respiratory, thoracic and mediastinal disorders
Pulmonary Edema
2.1%
4/195 • Number of events 4 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Respiratory, thoracic and mediastinal disorders
Pulmonary Effusion
4.6%
9/195 • Number of events 10 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Renal and urinary disorders
Renal Failure
4.6%
9/195 • Number of events 10 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Respiratory, thoracic and mediastinal disorders
Respiratory Failure
9.2%
18/195 • Number of events 19 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Respiratory, thoracic and mediastinal disorders
Respiratory Infection
3.1%
6/195 • Number of events 6 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Vascular disorders
Restenosis
1.5%
3/195 • Number of events 3 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Injury, poisoning and procedural complications
Rupture
0.51%
1/195 • Number of events 1 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Infections and infestations
Sepsis
4.1%
8/195 • Number of events 8 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Vascular disorders
Syncope
1.5%
3/195 • Number of events 3 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Blood and lymphatic system disorders
Thrombocytopenia
1.0%
2/195 • Number of events 2 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Investigations
Unknown cause of Death
2.1%
4/195 • Number of events 4 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Investigations
Other
7.7%
15/195 • Number of events 18 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Infections and infestations
Urinary Tract Infection
0.51%
1/195 • Number of events 1 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Investigations
Undetermined Stroke
0.51%
1/195 • Number of events 1 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Vascular disorders
TIA
0.51%
1/195 • Number of events 1 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Vascular disorders
Hematoma
0.51%
1/195 • Number of events 1 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Injury, poisoning and procedural complications
Fall
1.5%
3/195 • Number of events 4 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Psychiatric disorders
Confusion
1.0%
2/195 • Number of events 2 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Infections and infestations
Access site infection
0.51%
1/195 • Number of events 1 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Infections and infestations
Cough
0.51%
1/195 • Number of events 1 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Psychiatric disorders
Dementia
0.51%
1/195 • Number of events 2 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Injury, poisoning and procedural complications
Device Migration/malposition requiring intervention
1.0%
2/195 • Number of events 2 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Investigations
Drug Reaction
0.51%
1/195 • Number of events 1 • 30-day
Site reported events Only Valve-Implant patients are analyzed.

Other adverse events

Other adverse events
Measure
Edwards SAPIEN XT THV in Patients With Failing SAVR
n=195 participants at risk
Patients with a failing surgical bioprosthetic valve (SAVR) in the aortic position demonstrating stenosis and/or insufficiency will be treated with Edwards SAPIEN XT transcatheter valve. Edwards SAPIEN XT transcatheter valve, Model 9300TFX: Edwards SAPIEN XT THV system Model 9300TFX with the associated delivery systems.
Investigations
Abnormal Lab Value
11.8%
23/195 • Number of events 39 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Injury, poisoning and procedural complications
Anemia/Hemolytic Anemia
11.8%
23/195 • Number of events 24 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Cardiac disorders
Arrhythmia/Conduction System Injury (Defect)
19.5%
38/195 • Number of events 42 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Injury, poisoning and procedural complications
Fall
6.2%
12/195 • Number of events 13 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Cardiac disorders
Heart Failure / CHF / Low output Failure
6.2%
12/195 • Number of events 12 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Vascular disorders
Hemorrhage
20.5%
40/195 • Number of events 45 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Vascular disorders
Hypotension
5.1%
10/195 • Number of events 10 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Nervous system disorders
Pain
6.2%
12/195 • Number of events 19 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Respiratory, thoracic and mediastinal disorders
Pulmonary Effusion
5.6%
11/195 • Number of events 12 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Renal and urinary disorders
Renal Insufficiency or Renal Failure
8.2%
16/195 • Number of events 20 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Respiratory, thoracic and mediastinal disorders
Respiratory Infection
8.2%
16/195 • Number of events 16 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Blood and lymphatic system disorders
Thrombocytopenia
11.8%
23/195 • Number of events 23 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Infections and infestations
Urinary Tract Infection
9.2%
18/195 • Number of events 18 • 30-day
Site reported events Only Valve-Implant patients are analyzed.
Investigations
Other
25.6%
50/195 • Number of events 63 • 30-day
Site reported events Only Valve-Implant patients are analyzed.

Additional Information

Edwards THV Clinical Affairs

Edwards Lifesciences

Phone: (949) 250-2500

Results disclosure agreements

  • Principal investigator is a sponsor employee