Trial Outcomes & Findings for Carbon Monoxide Breath Sensor System Performance, Human Factors, and Usability Assessment Conducted in a Single Visit (NCT NCT03148613)

NCT ID: NCT03148613

Last Updated: 2022-10-18

Results Overview

CO concentrations in parts per million (ppm) in exhaled breath from subjects at a single visit will be assessed and reported in order to compare the CO measurements from COBSS to a control (predicate) device. The correlation will be reported as an R value. For each participant, the single visit includes the consent process, entry into the study, the study session, and conclusion of the study. The study session is completed on a single day requiring up to 90 minutes. COBSS breath sample will take approximately 1 minute. Breath sample with the control device will take approximately 1 minute. The breath samples will be collected within 5-10 minutes of each other.

Recruitment status

COMPLETED

Study phase

NA

Target enrollment

70 participants

Primary outcome timeframe

1 minute

Results posted on

2022-10-18

Participant Flow

Participant milestones

Participant milestones
Measure
18-49 Year Old Smokers Using COBSS
Younger smokers - Assessment of intended user performance in use scenarios, user documentation assessment, subjective feedback and rating scales. NOTE: 18-49 year Age Arm intended to evaluate human factors and usability of COBSS in younger smokers Carbon Monoxide Breath Sensor System (COBSS): Carbon Monoxide Breath Sensor System (COBSS): Device to measure carbon monoxide levels in one's breath and inform the user. Bedfont - Micro+™ Smokerlyzer®: Bedfont - Micro+™ Smokerlyzer® : Predicate Device to measure carbon monoxide levels in one's breath.
≥50 Year Old Smokers Using COBSS
Older smokers- Assessment of intended user performance in use scenarios, user documentation assessment, subjective feedback and rating scales. NOTE:≥50 year Age Arm intended to evaluate human factors and usability of COBSS in younger smokers Carbon Monoxide Breath Sensor System (COBSS): Carbon Monoxide Breath Sensor System (COBSS): Device to measure carbon monoxide levels in one's breath and inform the user. Bedfont - Micro+™ Smokerlyzer®: Bedfont - Micro+™ Smokerlyzer® : Predicate Device to measure carbon monoxide levels in one's breath.
Overall Study
STARTED
39
31
Overall Study
COMPLETED
39
31
Overall Study
NOT COMPLETED
0
0

Reasons for withdrawal

Withdrawal data not reported

Baseline Characteristics

Carbon Monoxide Breath Sensor System Performance, Human Factors, and Usability Assessment Conducted in a Single Visit

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
18-49 Year Old Smokers Using COBSS
n=39 Participants
Younger smokers - Assessment of intended user performance in use scenarios, user documentation assessment, subjective feedback and rating scales. NOTE: 18-49 year Age Arm intended to evaluate human factors and usability of COBSS in younger smokers Carbon Monoxide Breath Sensor System (COBSS): Carbon Monoxide Breath Sensor System (COBSS): Device to measure carbon monoxide levels in one's breath and inform the user. Bedfont - Micro+™ Smokerlyzer®: Bedfont - Micro+™ Smokerlyzer® : Predicate Device to measure carbon monoxide levels in one's breath.
≥50 Year Old Smokers Using COBSS
n=31 Participants
Older smokers- Assessment of intended user performance in use scenarios, user documentation assessment, subjective feedback and rating scales. NOTE:≥50 year Age Arm intended to evaluate human factors and usability of COBSS in younger smokers Carbon Monoxide Breath Sensor System (COBSS): Carbon Monoxide Breath Sensor System (COBSS): Device to measure carbon monoxide levels in one's breath and inform the user. Bedfont - Micro+™ Smokerlyzer®: Bedfont - Micro+™ Smokerlyzer® : Predicate Device to measure carbon monoxide levels in one's breath.
Total
n=70 Participants
Total of all reporting groups
Age, Categorical
<=18 years
0 Participants
n=99 Participants
0 Participants
n=107 Participants
0 Participants
n=206 Participants
Age, Categorical
Between 18 and 65 years
39 Participants
n=99 Participants
30 Participants
n=107 Participants
69 Participants
n=206 Participants
Age, Categorical
>=65 years
0 Participants
n=99 Participants
1 Participants
n=107 Participants
1 Participants
n=206 Participants
Sex: Female, Male
Female
18 Participants
n=99 Participants
14 Participants
n=107 Participants
32 Participants
n=206 Participants
Sex: Female, Male
Male
21 Participants
n=99 Participants
17 Participants
n=107 Participants
38 Participants
n=206 Participants
Race (NIH/OMB)
American Indian or Alaska Native
0 Participants
n=99 Participants
0 Participants
n=107 Participants
0 Participants
n=206 Participants
Race (NIH/OMB)
Asian
3 Participants
n=99 Participants
2 Participants
n=107 Participants
5 Participants
n=206 Participants
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
1 Participants
n=99 Participants
1 Participants
n=107 Participants
2 Participants
n=206 Participants
Race (NIH/OMB)
Black or African American
8 Participants
n=99 Participants
9 Participants
n=107 Participants
17 Participants
n=206 Participants
Race (NIH/OMB)
White
16 Participants
n=99 Participants
16 Participants
n=107 Participants
32 Participants
n=206 Participants
Race (NIH/OMB)
More than one race
0 Participants
n=99 Participants
0 Participants
n=107 Participants
0 Participants
n=206 Participants
Race (NIH/OMB)
Unknown or Not Reported
11 Participants
n=99 Participants
3 Participants
n=107 Participants
14 Participants
n=206 Participants
Region of Enrollment
United States
39 participants
n=99 Participants
31 participants
n=107 Participants
70 participants
n=206 Participants
smoke frequency
1 per day
2 Participants
n=99 Participants
0 Participants
n=107 Participants
2 Participants
n=206 Participants
smoke frequency
2 - 7 per day
21 Participants
n=99 Participants
10 Participants
n=107 Participants
31 Participants
n=206 Participants
smoke frequency
8 or more per day
16 Participants
n=99 Participants
21 Participants
n=107 Participants
37 Participants
n=206 Participants

PRIMARY outcome

Timeframe: 1 minute

Population: The performance outcome was designed to evaluate all the participants together, thus the cohorts (arms/groups) were not evaluated separately for this outcome. The cohorts were used for the human factors and usability portion of the study, but are not applicable for this outcome. The R value reported is the correlation coefficient between 70 paired CO measurement values taken with the COBSS and predicate devices, by the 70 participants.

CO concentrations in parts per million (ppm) in exhaled breath from subjects at a single visit will be assessed and reported in order to compare the CO measurements from COBSS to a control (predicate) device. The correlation will be reported as an R value. For each participant, the single visit includes the consent process, entry into the study, the study session, and conclusion of the study. The study session is completed on a single day requiring up to 90 minutes. COBSS breath sample will take approximately 1 minute. Breath sample with the control device will take approximately 1 minute. The breath samples will be collected within 5-10 minutes of each other.

Outcome measures

Outcome measures
Measure
All Participants
n=70 Participants
Paired breath samples from each participant Carbon Monoxide Breath Sensor System (COBSS): Carbon Monoxide Breath Sensor System (COBSS): Device to measure carbon monoxide levels in one's breath and compared to predicate device Each participant measured carbon monoxide with both the COBSS and the predicate device. These CO measures were paired and used in a regression analysis.
≥50 Year Old Smokers Using COBSS
Older smokers- Assessment of intended user performance in use scenarios, user documentation assessment, subjective feedback and rating scales. NOTE:≥50 year Age Arm intended to evaluate human factors and usability of COBSS in younger smokers Carbon Monoxide Breath Sensor System (COBSS): Carbon Monoxide Breath Sensor System (COBSS): Device to measure carbon monoxide levels in one's breath and inform the user. Bedfont - Micro+™ Smokerlyzer®: Bedfont - Micro+™ Smokerlyzer® : Predicate Device to measure carbon monoxide levels in one's breath.
All Participants
Carbon Monoxide Breath Sensor System (COBSS): Carbon Monoxide Breath Sensor System (COBSS): Device to measure carbon monoxide levels in one's breath and compared to predicate device
Performance Variable - Correlation of Carbon Monoxide (CO) Measurements Between COBSS and Predicate Devices
0.9878 Correlation coefficient (R-value)

PRIMARY outcome

Timeframe: 1 minute for each question; up to 90 minutes for study session

Population: Outcome measure data table is tally of participants who were 'Successful' in each use scenario.

User performance of participants in intended use scenarios will be assessed at a single visit. Participants will be observed if they can set up the device, use the device and interpret results. Observer ratings include 'Successful' and 'Unsuccessful'. Specific attention will be given to assess if any harms can occur from observed use errors. Outcome measure data table is tally of participants who were 'Successful' in each use scenario. For each participant, the single visit includes the consent process, entry into the study, the study session, and conclusion of the study. The study session is completed on a single day requiring up to 90 minutes. Observations are made throughout the study session. Assessments for individual tasks will take approximately 1 minute and assessments will be made over the study session which can last up to 90 minutes.

Outcome measures

Outcome measures
Measure
All Participants
n=39 Participants
Paired breath samples from each participant Carbon Monoxide Breath Sensor System (COBSS): Carbon Monoxide Breath Sensor System (COBSS): Device to measure carbon monoxide levels in one's breath and compared to predicate device Each participant measured carbon monoxide with both the COBSS and the predicate device. These CO measures were paired and used in a regression analysis.
≥50 Year Old Smokers Using COBSS
n=31 Participants
Older smokers- Assessment of intended user performance in use scenarios, user documentation assessment, subjective feedback and rating scales. NOTE:≥50 year Age Arm intended to evaluate human factors and usability of COBSS in younger smokers Carbon Monoxide Breath Sensor System (COBSS): Carbon Monoxide Breath Sensor System (COBSS): Device to measure carbon monoxide levels in one's breath and inform the user. Bedfont - Micro+™ Smokerlyzer®: Bedfont - Micro+™ Smokerlyzer® : Predicate Device to measure carbon monoxide levels in one's breath.
All Participants
n=70 Participants
Carbon Monoxide Breath Sensor System (COBSS): Carbon Monoxide Breath Sensor System (COBSS): Device to measure carbon monoxide levels in one's breath and compared to predicate device
Human Factors - User Performance in Use Scenarios Assessment With Observer Ratings
set up the device- Pair Breath Sensor
34 Participants
25 Participants
59 Participants
Human Factors - User Performance in Use Scenarios Assessment With Observer Ratings
interpret results
38 Participants
31 Participants
69 Participants
Human Factors - User Performance in Use Scenarios Assessment With Observer Ratings
use the device - submit a breath sample
37 Participants
29 Participants
66 Participants

SECONDARY outcome

Timeframe: Up to 90 minutes

Population: How clear or unclear were the User Manual Reference Card, and Quick Start Guide?

Participant understanding of user documentation will be assessed at single visit. User documentation includes packaging information, quick start guide, reference card, and user manual. Observer ratings include 'Successful' and 'Unsuccessful'. For each participant, the single visit includes the consent process, entry into the study, the study session, and conclusion of the study. The study session is completed on a single day requiring up to 90 minutes. Observations are made throughout the study session. Assessments for individual tasks will take approximately 1 minute each. The User documentation assessment portion of the study session will last up to 30 minutes. The 5-Point Likert Scale for Clarity has the following answer options: 5 - Very Clear; 4 - Clear; 3 - Neither Clear nor Unclear; 2 - Unclear; and 1 - Very Unclear. Higher values of this scale represent a better outcome, which is greater clarity/understanding.

Outcome measures

Outcome measures
Measure
All Participants
n=38 Participants
Paired breath samples from each participant Carbon Monoxide Breath Sensor System (COBSS): Carbon Monoxide Breath Sensor System (COBSS): Device to measure carbon monoxide levels in one's breath and compared to predicate device Each participant measured carbon monoxide with both the COBSS and the predicate device. These CO measures were paired and used in a regression analysis.
≥50 Year Old Smokers Using COBSS
n=31 Participants
Older smokers- Assessment of intended user performance in use scenarios, user documentation assessment, subjective feedback and rating scales. NOTE:≥50 year Age Arm intended to evaluate human factors and usability of COBSS in younger smokers Carbon Monoxide Breath Sensor System (COBSS): Carbon Monoxide Breath Sensor System (COBSS): Device to measure carbon monoxide levels in one's breath and inform the user. Bedfont - Micro+™ Smokerlyzer®: Bedfont - Micro+™ Smokerlyzer® : Predicate Device to measure carbon monoxide levels in one's breath.
All Participants
n=69 Participants
Carbon Monoxide Breath Sensor System (COBSS): Carbon Monoxide Breath Sensor System (COBSS): Device to measure carbon monoxide levels in one's breath and compared to predicate device
User Documentation Assessment With Questionnaire With Observer Ratings
5 - Very Clear
23 Participants
17 Participants
40 Participants
User Documentation Assessment With Questionnaire With Observer Ratings
4 - Clear
13 Participants
10 Participants
23 Participants
User Documentation Assessment With Questionnaire With Observer Ratings
3 - Neither Clear nor Unclear
2 Participants
1 Participants
3 Participants
User Documentation Assessment With Questionnaire With Observer Ratings
2 - Unclear
0 Participants
2 Participants
2 Participants
User Documentation Assessment With Questionnaire With Observer Ratings
1 - Very Unclear
0 Participants
1 Participants
1 Participants

SECONDARY outcome

Timeframe: Up to 90 minutes

Population: Outcome measure data table is tally of participants answering No.

Subjective feedback focusing on any concerns (i.e. 'yes', 'no') on using device will be assessed at single visit. For each participant, the single visit includes the consent process, entry into the study, the study session, and conclusion of the study. The study session is completed on a single day requiring up to 90 minutes. \*Row title truncated due to character limit. Actual question asked: Were there any points during this portion of the study that you experienced difficulty or had concerns about the product? Subjective feedback focusing on any concerns prior to the conclusion of the study session. Subjective feedback questions will take approximately 1-3 minutes each. The Subjective feedback portion of the study session will last up to 5 to 15 minutes.

Outcome measures

Outcome measures
Measure
All Participants
n=38 Participants
Paired breath samples from each participant Carbon Monoxide Breath Sensor System (COBSS): Carbon Monoxide Breath Sensor System (COBSS): Device to measure carbon monoxide levels in one's breath and compared to predicate device Each participant measured carbon monoxide with both the COBSS and the predicate device. These CO measures were paired and used in a regression analysis.
≥50 Year Old Smokers Using COBSS
n=31 Participants
Older smokers- Assessment of intended user performance in use scenarios, user documentation assessment, subjective feedback and rating scales. NOTE:≥50 year Age Arm intended to evaluate human factors and usability of COBSS in younger smokers Carbon Monoxide Breath Sensor System (COBSS): Carbon Monoxide Breath Sensor System (COBSS): Device to measure carbon monoxide levels in one's breath and inform the user. Bedfont - Micro+™ Smokerlyzer®: Bedfont - Micro+™ Smokerlyzer® : Predicate Device to measure carbon monoxide levels in one's breath.
All Participants
n=69 Participants
Carbon Monoxide Breath Sensor System (COBSS): Carbon Monoxide Breath Sensor System (COBSS): Device to measure carbon monoxide levels in one's breath and compared to predicate device
Number of Participants Answering "No" for the Subjective Feedback With Questionnaire
Were there any points during ... that you experienced difficulty or had concerns about the product?*
30 Participants
23 Participants
53 Participants
Number of Participants Answering "No" for the Subjective Feedback With Questionnaire
Did you experience any moments of hesitation while using the product?
33 Participants
27 Participants
60 Participants
Number of Participants Answering "No" for the Subjective Feedback With Questionnaire
Did you have any safety concerns regarding the product or labeling?
35 Participants
26 Participants
61 Participants

SECONDARY outcome

Timeframe: Up to 90 minutes

Participant will be asked to provide ratings on the following: * How would you rate the ease or difficulty using the device and phone app? 5 - Very Clear; 4 - Clear; 3 - Neither Clear nor Unclear; 2 - Unclear; 1 - Very Unclear * How easy or difficult was it to understand and interpret the test results? For each participant, the single visit includes the consent process, entry into the study, the study session, and conclusion of the study. The study session is completed on a single day requiring up to 90 minutes. Participant will provide the ratings prior to the conclusion of the study session. Participant assessment with questionnaire rating scales will take approximately 1-2 minutes each. The Subjective feedback portion of the study session will last up to 5 minutes.

Outcome measures

Outcome measures
Measure
All Participants
n=38 Participants
Paired breath samples from each participant Carbon Monoxide Breath Sensor System (COBSS): Carbon Monoxide Breath Sensor System (COBSS): Device to measure carbon monoxide levels in one's breath and compared to predicate device Each participant measured carbon monoxide with both the COBSS and the predicate device. These CO measures were paired and used in a regression analysis.
≥50 Year Old Smokers Using COBSS
n=31 Participants
Older smokers- Assessment of intended user performance in use scenarios, user documentation assessment, subjective feedback and rating scales. NOTE:≥50 year Age Arm intended to evaluate human factors and usability of COBSS in younger smokers Carbon Monoxide Breath Sensor System (COBSS): Carbon Monoxide Breath Sensor System (COBSS): Device to measure carbon monoxide levels in one's breath and inform the user. Bedfont - Micro+™ Smokerlyzer®: Bedfont - Micro+™ Smokerlyzer® : Predicate Device to measure carbon monoxide levels in one's breath.
All Participants
n=69 Participants
Carbon Monoxide Breath Sensor System (COBSS): Carbon Monoxide Breath Sensor System (COBSS): Device to measure carbon monoxide levels in one's breath and compared to predicate device
Participant Assessment With Questionnaire Rating Scales
How would you rate the ease or difficulty using the device and phone app? · 1 - Very difficult
0 Participants
1 Participants
1 Participants
Participant Assessment With Questionnaire Rating Scales
How easy or difficult was it to understand and interpret the test results? · 5 - Very Easy
29 Participants
17 Participants
46 Participants
Participant Assessment With Questionnaire Rating Scales
How easy or difficult was it to understand and interpret the test results? · 4 - Easy
8 Participants
10 Participants
18 Participants
Participant Assessment With Questionnaire Rating Scales
How easy or difficult was it to understand and interpret the test results? · 3 - Neither easy nor difficult
1 Participants
2 Participants
3 Participants
Participant Assessment With Questionnaire Rating Scales
How easy or difficult was it to understand and interpret the test results? · 2 - Difficult
0 Participants
2 Participants
2 Participants
Participant Assessment With Questionnaire Rating Scales
How easy or difficult was it to understand and interpret the test results? · 1 - Very difficult
0 Participants
0 Participants
0 Participants
Participant Assessment With Questionnaire Rating Scales
How would you rate the ease or difficulty using the device and phone app? · 4 - Easy
9 Participants
18 Participants
27 Participants
Participant Assessment With Questionnaire Rating Scales
How would you rate the ease or difficulty using the device and phone app? · 5 - Very Easy
26 Participants
8 Participants
34 Participants
Participant Assessment With Questionnaire Rating Scales
How would you rate the ease or difficulty using the device and phone app? · 3 - Neither easy nor difficult
3 Participants
2 Participants
5 Participants
Participant Assessment With Questionnaire Rating Scales
How would you rate the ease or difficulty using the device and phone app? · 2 - Difficult
0 Participants
2 Participants
2 Participants

Adverse Events

18-49 Year Old Smokers Using COBSS

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

≥50 Year Old Smokers Using COBSS

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

All Participants

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Serious adverse events

Adverse event data not reported

Other adverse events

Adverse event data not reported

Additional Information

Director Clinical Affairs

Carrot Inc.

Phone: 650-906-4255

Results disclosure agreements

  • Principal investigator is a sponsor employee
  • Publication restrictions are in place