Trial Outcomes & Findings for Carbon Monoxide Breath Sensor System Performance, Human Factors, and Usability Assessment Conducted in a Single Visit (NCT NCT03148613)
NCT ID: NCT03148613
Last Updated: 2022-10-18
Results Overview
CO concentrations in parts per million (ppm) in exhaled breath from subjects at a single visit will be assessed and reported in order to compare the CO measurements from COBSS to a control (predicate) device. The correlation will be reported as an R value. For each participant, the single visit includes the consent process, entry into the study, the study session, and conclusion of the study. The study session is completed on a single day requiring up to 90 minutes. COBSS breath sample will take approximately 1 minute. Breath sample with the control device will take approximately 1 minute. The breath samples will be collected within 5-10 minutes of each other.
COMPLETED
NA
70 participants
1 minute
2022-10-18
Participant Flow
Participant milestones
| Measure |
18-49 Year Old Smokers Using COBSS
Younger smokers - Assessment of intended user performance in use scenarios, user documentation assessment, subjective feedback and rating scales.
NOTE: 18-49 year Age Arm intended to evaluate human factors and usability of COBSS in younger smokers
Carbon Monoxide Breath Sensor System (COBSS): Carbon Monoxide Breath Sensor System (COBSS): Device to measure carbon monoxide levels in one's breath and inform the user.
Bedfont - Micro+™ Smokerlyzer®: Bedfont - Micro+™ Smokerlyzer® : Predicate Device to measure carbon monoxide levels in one's breath.
|
≥50 Year Old Smokers Using COBSS
Older smokers- Assessment of intended user performance in use scenarios, user documentation assessment, subjective feedback and rating scales.
NOTE:≥50 year Age Arm intended to evaluate human factors and usability of COBSS in younger smokers
Carbon Monoxide Breath Sensor System (COBSS): Carbon Monoxide Breath Sensor System (COBSS): Device to measure carbon monoxide levels in one's breath and inform the user.
Bedfont - Micro+™ Smokerlyzer®: Bedfont - Micro+™ Smokerlyzer® : Predicate Device to measure carbon monoxide levels in one's breath.
|
|---|---|---|
|
Overall Study
STARTED
|
39
|
31
|
|
Overall Study
COMPLETED
|
39
|
31
|
|
Overall Study
NOT COMPLETED
|
0
|
0
|
Reasons for withdrawal
Withdrawal data not reported
Baseline Characteristics
Carbon Monoxide Breath Sensor System Performance, Human Factors, and Usability Assessment Conducted in a Single Visit
Baseline characteristics by cohort
| Measure |
18-49 Year Old Smokers Using COBSS
n=39 Participants
Younger smokers - Assessment of intended user performance in use scenarios, user documentation assessment, subjective feedback and rating scales.
NOTE: 18-49 year Age Arm intended to evaluate human factors and usability of COBSS in younger smokers
Carbon Monoxide Breath Sensor System (COBSS): Carbon Monoxide Breath Sensor System (COBSS): Device to measure carbon monoxide levels in one's breath and inform the user.
Bedfont - Micro+™ Smokerlyzer®: Bedfont - Micro+™ Smokerlyzer® : Predicate Device to measure carbon monoxide levels in one's breath.
|
≥50 Year Old Smokers Using COBSS
n=31 Participants
Older smokers- Assessment of intended user performance in use scenarios, user documentation assessment, subjective feedback and rating scales.
NOTE:≥50 year Age Arm intended to evaluate human factors and usability of COBSS in younger smokers
Carbon Monoxide Breath Sensor System (COBSS): Carbon Monoxide Breath Sensor System (COBSS): Device to measure carbon monoxide levels in one's breath and inform the user.
Bedfont - Micro+™ Smokerlyzer®: Bedfont - Micro+™ Smokerlyzer® : Predicate Device to measure carbon monoxide levels in one's breath.
|
Total
n=70 Participants
Total of all reporting groups
|
|---|---|---|---|
|
Age, Categorical
<=18 years
|
0 Participants
n=99 Participants
|
0 Participants
n=107 Participants
|
0 Participants
n=206 Participants
|
|
Age, Categorical
Between 18 and 65 years
|
39 Participants
n=99 Participants
|
30 Participants
n=107 Participants
|
69 Participants
n=206 Participants
|
|
Age, Categorical
>=65 years
|
0 Participants
n=99 Participants
|
1 Participants
n=107 Participants
|
1 Participants
n=206 Participants
|
|
Sex: Female, Male
Female
|
18 Participants
n=99 Participants
|
14 Participants
n=107 Participants
|
32 Participants
n=206 Participants
|
|
Sex: Female, Male
Male
|
21 Participants
n=99 Participants
|
17 Participants
n=107 Participants
|
38 Participants
n=206 Participants
|
|
Race (NIH/OMB)
American Indian or Alaska Native
|
0 Participants
n=99 Participants
|
0 Participants
n=107 Participants
|
0 Participants
n=206 Participants
|
|
Race (NIH/OMB)
Asian
|
3 Participants
n=99 Participants
|
2 Participants
n=107 Participants
|
5 Participants
n=206 Participants
|
|
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
|
1 Participants
n=99 Participants
|
1 Participants
n=107 Participants
|
2 Participants
n=206 Participants
|
|
Race (NIH/OMB)
Black or African American
|
8 Participants
n=99 Participants
|
9 Participants
n=107 Participants
|
17 Participants
n=206 Participants
|
|
Race (NIH/OMB)
White
|
16 Participants
n=99 Participants
|
16 Participants
n=107 Participants
|
32 Participants
n=206 Participants
|
|
Race (NIH/OMB)
More than one race
|
0 Participants
n=99 Participants
|
0 Participants
n=107 Participants
|
0 Participants
n=206 Participants
|
|
Race (NIH/OMB)
Unknown or Not Reported
|
11 Participants
n=99 Participants
|
3 Participants
n=107 Participants
|
14 Participants
n=206 Participants
|
|
Region of Enrollment
United States
|
39 participants
n=99 Participants
|
31 participants
n=107 Participants
|
70 participants
n=206 Participants
|
|
smoke frequency
1 per day
|
2 Participants
n=99 Participants
|
0 Participants
n=107 Participants
|
2 Participants
n=206 Participants
|
|
smoke frequency
2 - 7 per day
|
21 Participants
n=99 Participants
|
10 Participants
n=107 Participants
|
31 Participants
n=206 Participants
|
|
smoke frequency
8 or more per day
|
16 Participants
n=99 Participants
|
21 Participants
n=107 Participants
|
37 Participants
n=206 Participants
|
PRIMARY outcome
Timeframe: 1 minutePopulation: The performance outcome was designed to evaluate all the participants together, thus the cohorts (arms/groups) were not evaluated separately for this outcome. The cohorts were used for the human factors and usability portion of the study, but are not applicable for this outcome. The R value reported is the correlation coefficient between 70 paired CO measurement values taken with the COBSS and predicate devices, by the 70 participants.
CO concentrations in parts per million (ppm) in exhaled breath from subjects at a single visit will be assessed and reported in order to compare the CO measurements from COBSS to a control (predicate) device. The correlation will be reported as an R value. For each participant, the single visit includes the consent process, entry into the study, the study session, and conclusion of the study. The study session is completed on a single day requiring up to 90 minutes. COBSS breath sample will take approximately 1 minute. Breath sample with the control device will take approximately 1 minute. The breath samples will be collected within 5-10 minutes of each other.
Outcome measures
| Measure |
All Participants
n=70 Participants
Paired breath samples from each participant
Carbon Monoxide Breath Sensor System (COBSS): Carbon Monoxide Breath Sensor System (COBSS): Device to measure carbon monoxide levels in one's breath and compared to predicate device
Each participant measured carbon monoxide with both the COBSS and the predicate device. These CO measures were paired and used in a regression analysis.
|
≥50 Year Old Smokers Using COBSS
Older smokers- Assessment of intended user performance in use scenarios, user documentation assessment, subjective feedback and rating scales.
NOTE:≥50 year Age Arm intended to evaluate human factors and usability of COBSS in younger smokers
Carbon Monoxide Breath Sensor System (COBSS): Carbon Monoxide Breath Sensor System (COBSS): Device to measure carbon monoxide levels in one's breath and inform the user.
Bedfont - Micro+™ Smokerlyzer®: Bedfont - Micro+™ Smokerlyzer® : Predicate Device to measure carbon monoxide levels in one's breath.
|
All Participants
Carbon Monoxide Breath Sensor System (COBSS): Carbon Monoxide Breath Sensor System (COBSS): Device to measure carbon monoxide levels in one's breath and compared to predicate device
|
|---|---|---|---|
|
Performance Variable - Correlation of Carbon Monoxide (CO) Measurements Between COBSS and Predicate Devices
|
0.9878 Correlation coefficient (R-value)
|
—
|
—
|
PRIMARY outcome
Timeframe: 1 minute for each question; up to 90 minutes for study sessionPopulation: Outcome measure data table is tally of participants who were 'Successful' in each use scenario.
User performance of participants in intended use scenarios will be assessed at a single visit. Participants will be observed if they can set up the device, use the device and interpret results. Observer ratings include 'Successful' and 'Unsuccessful'. Specific attention will be given to assess if any harms can occur from observed use errors. Outcome measure data table is tally of participants who were 'Successful' in each use scenario. For each participant, the single visit includes the consent process, entry into the study, the study session, and conclusion of the study. The study session is completed on a single day requiring up to 90 minutes. Observations are made throughout the study session. Assessments for individual tasks will take approximately 1 minute and assessments will be made over the study session which can last up to 90 minutes.
Outcome measures
| Measure |
All Participants
n=39 Participants
Paired breath samples from each participant
Carbon Monoxide Breath Sensor System (COBSS): Carbon Monoxide Breath Sensor System (COBSS): Device to measure carbon monoxide levels in one's breath and compared to predicate device
Each participant measured carbon monoxide with both the COBSS and the predicate device. These CO measures were paired and used in a regression analysis.
|
≥50 Year Old Smokers Using COBSS
n=31 Participants
Older smokers- Assessment of intended user performance in use scenarios, user documentation assessment, subjective feedback and rating scales.
NOTE:≥50 year Age Arm intended to evaluate human factors and usability of COBSS in younger smokers
Carbon Monoxide Breath Sensor System (COBSS): Carbon Monoxide Breath Sensor System (COBSS): Device to measure carbon monoxide levels in one's breath and inform the user.
Bedfont - Micro+™ Smokerlyzer®: Bedfont - Micro+™ Smokerlyzer® : Predicate Device to measure carbon monoxide levels in one's breath.
|
All Participants
n=70 Participants
Carbon Monoxide Breath Sensor System (COBSS): Carbon Monoxide Breath Sensor System (COBSS): Device to measure carbon monoxide levels in one's breath and compared to predicate device
|
|---|---|---|---|
|
Human Factors - User Performance in Use Scenarios Assessment With Observer Ratings
set up the device- Pair Breath Sensor
|
34 Participants
|
25 Participants
|
59 Participants
|
|
Human Factors - User Performance in Use Scenarios Assessment With Observer Ratings
interpret results
|
38 Participants
|
31 Participants
|
69 Participants
|
|
Human Factors - User Performance in Use Scenarios Assessment With Observer Ratings
use the device - submit a breath sample
|
37 Participants
|
29 Participants
|
66 Participants
|
SECONDARY outcome
Timeframe: Up to 90 minutesPopulation: How clear or unclear were the User Manual Reference Card, and Quick Start Guide?
Participant understanding of user documentation will be assessed at single visit. User documentation includes packaging information, quick start guide, reference card, and user manual. Observer ratings include 'Successful' and 'Unsuccessful'. For each participant, the single visit includes the consent process, entry into the study, the study session, and conclusion of the study. The study session is completed on a single day requiring up to 90 minutes. Observations are made throughout the study session. Assessments for individual tasks will take approximately 1 minute each. The User documentation assessment portion of the study session will last up to 30 minutes. The 5-Point Likert Scale for Clarity has the following answer options: 5 - Very Clear; 4 - Clear; 3 - Neither Clear nor Unclear; 2 - Unclear; and 1 - Very Unclear. Higher values of this scale represent a better outcome, which is greater clarity/understanding.
Outcome measures
| Measure |
All Participants
n=38 Participants
Paired breath samples from each participant
Carbon Monoxide Breath Sensor System (COBSS): Carbon Monoxide Breath Sensor System (COBSS): Device to measure carbon monoxide levels in one's breath and compared to predicate device
Each participant measured carbon monoxide with both the COBSS and the predicate device. These CO measures were paired and used in a regression analysis.
|
≥50 Year Old Smokers Using COBSS
n=31 Participants
Older smokers- Assessment of intended user performance in use scenarios, user documentation assessment, subjective feedback and rating scales.
NOTE:≥50 year Age Arm intended to evaluate human factors and usability of COBSS in younger smokers
Carbon Monoxide Breath Sensor System (COBSS): Carbon Monoxide Breath Sensor System (COBSS): Device to measure carbon monoxide levels in one's breath and inform the user.
Bedfont - Micro+™ Smokerlyzer®: Bedfont - Micro+™ Smokerlyzer® : Predicate Device to measure carbon monoxide levels in one's breath.
|
All Participants
n=69 Participants
Carbon Monoxide Breath Sensor System (COBSS): Carbon Monoxide Breath Sensor System (COBSS): Device to measure carbon monoxide levels in one's breath and compared to predicate device
|
|---|---|---|---|
|
User Documentation Assessment With Questionnaire With Observer Ratings
5 - Very Clear
|
23 Participants
|
17 Participants
|
40 Participants
|
|
User Documentation Assessment With Questionnaire With Observer Ratings
4 - Clear
|
13 Participants
|
10 Participants
|
23 Participants
|
|
User Documentation Assessment With Questionnaire With Observer Ratings
3 - Neither Clear nor Unclear
|
2 Participants
|
1 Participants
|
3 Participants
|
|
User Documentation Assessment With Questionnaire With Observer Ratings
2 - Unclear
|
0 Participants
|
2 Participants
|
2 Participants
|
|
User Documentation Assessment With Questionnaire With Observer Ratings
1 - Very Unclear
|
0 Participants
|
1 Participants
|
1 Participants
|
SECONDARY outcome
Timeframe: Up to 90 minutesPopulation: Outcome measure data table is tally of participants answering No.
Subjective feedback focusing on any concerns (i.e. 'yes', 'no') on using device will be assessed at single visit. For each participant, the single visit includes the consent process, entry into the study, the study session, and conclusion of the study. The study session is completed on a single day requiring up to 90 minutes. \*Row title truncated due to character limit. Actual question asked: Were there any points during this portion of the study that you experienced difficulty or had concerns about the product? Subjective feedback focusing on any concerns prior to the conclusion of the study session. Subjective feedback questions will take approximately 1-3 minutes each. The Subjective feedback portion of the study session will last up to 5 to 15 minutes.
Outcome measures
| Measure |
All Participants
n=38 Participants
Paired breath samples from each participant
Carbon Monoxide Breath Sensor System (COBSS): Carbon Monoxide Breath Sensor System (COBSS): Device to measure carbon monoxide levels in one's breath and compared to predicate device
Each participant measured carbon monoxide with both the COBSS and the predicate device. These CO measures were paired and used in a regression analysis.
|
≥50 Year Old Smokers Using COBSS
n=31 Participants
Older smokers- Assessment of intended user performance in use scenarios, user documentation assessment, subjective feedback and rating scales.
NOTE:≥50 year Age Arm intended to evaluate human factors and usability of COBSS in younger smokers
Carbon Monoxide Breath Sensor System (COBSS): Carbon Monoxide Breath Sensor System (COBSS): Device to measure carbon monoxide levels in one's breath and inform the user.
Bedfont - Micro+™ Smokerlyzer®: Bedfont - Micro+™ Smokerlyzer® : Predicate Device to measure carbon monoxide levels in one's breath.
|
All Participants
n=69 Participants
Carbon Monoxide Breath Sensor System (COBSS): Carbon Monoxide Breath Sensor System (COBSS): Device to measure carbon monoxide levels in one's breath and compared to predicate device
|
|---|---|---|---|
|
Number of Participants Answering "No" for the Subjective Feedback With Questionnaire
Were there any points during ... that you experienced difficulty or had concerns about the product?*
|
30 Participants
|
23 Participants
|
53 Participants
|
|
Number of Participants Answering "No" for the Subjective Feedback With Questionnaire
Did you experience any moments of hesitation while using the product?
|
33 Participants
|
27 Participants
|
60 Participants
|
|
Number of Participants Answering "No" for the Subjective Feedback With Questionnaire
Did you have any safety concerns regarding the product or labeling?
|
35 Participants
|
26 Participants
|
61 Participants
|
SECONDARY outcome
Timeframe: Up to 90 minutesParticipant will be asked to provide ratings on the following: * How would you rate the ease or difficulty using the device and phone app? 5 - Very Clear; 4 - Clear; 3 - Neither Clear nor Unclear; 2 - Unclear; 1 - Very Unclear * How easy or difficult was it to understand and interpret the test results? For each participant, the single visit includes the consent process, entry into the study, the study session, and conclusion of the study. The study session is completed on a single day requiring up to 90 minutes. Participant will provide the ratings prior to the conclusion of the study session. Participant assessment with questionnaire rating scales will take approximately 1-2 minutes each. The Subjective feedback portion of the study session will last up to 5 minutes.
Outcome measures
| Measure |
All Participants
n=38 Participants
Paired breath samples from each participant
Carbon Monoxide Breath Sensor System (COBSS): Carbon Monoxide Breath Sensor System (COBSS): Device to measure carbon monoxide levels in one's breath and compared to predicate device
Each participant measured carbon monoxide with both the COBSS and the predicate device. These CO measures were paired and used in a regression analysis.
|
≥50 Year Old Smokers Using COBSS
n=31 Participants
Older smokers- Assessment of intended user performance in use scenarios, user documentation assessment, subjective feedback and rating scales.
NOTE:≥50 year Age Arm intended to evaluate human factors and usability of COBSS in younger smokers
Carbon Monoxide Breath Sensor System (COBSS): Carbon Monoxide Breath Sensor System (COBSS): Device to measure carbon monoxide levels in one's breath and inform the user.
Bedfont - Micro+™ Smokerlyzer®: Bedfont - Micro+™ Smokerlyzer® : Predicate Device to measure carbon monoxide levels in one's breath.
|
All Participants
n=69 Participants
Carbon Monoxide Breath Sensor System (COBSS): Carbon Monoxide Breath Sensor System (COBSS): Device to measure carbon monoxide levels in one's breath and compared to predicate device
|
|---|---|---|---|
|
Participant Assessment With Questionnaire Rating Scales
How would you rate the ease or difficulty using the device and phone app? · 1 - Very difficult
|
0 Participants
|
1 Participants
|
1 Participants
|
|
Participant Assessment With Questionnaire Rating Scales
How easy or difficult was it to understand and interpret the test results? · 5 - Very Easy
|
29 Participants
|
17 Participants
|
46 Participants
|
|
Participant Assessment With Questionnaire Rating Scales
How easy or difficult was it to understand and interpret the test results? · 4 - Easy
|
8 Participants
|
10 Participants
|
18 Participants
|
|
Participant Assessment With Questionnaire Rating Scales
How easy or difficult was it to understand and interpret the test results? · 3 - Neither easy nor difficult
|
1 Participants
|
2 Participants
|
3 Participants
|
|
Participant Assessment With Questionnaire Rating Scales
How easy or difficult was it to understand and interpret the test results? · 2 - Difficult
|
0 Participants
|
2 Participants
|
2 Participants
|
|
Participant Assessment With Questionnaire Rating Scales
How easy or difficult was it to understand and interpret the test results? · 1 - Very difficult
|
0 Participants
|
0 Participants
|
0 Participants
|
|
Participant Assessment With Questionnaire Rating Scales
How would you rate the ease or difficulty using the device and phone app? · 4 - Easy
|
9 Participants
|
18 Participants
|
27 Participants
|
|
Participant Assessment With Questionnaire Rating Scales
How would you rate the ease or difficulty using the device and phone app? · 5 - Very Easy
|
26 Participants
|
8 Participants
|
34 Participants
|
|
Participant Assessment With Questionnaire Rating Scales
How would you rate the ease or difficulty using the device and phone app? · 3 - Neither easy nor difficult
|
3 Participants
|
2 Participants
|
5 Participants
|
|
Participant Assessment With Questionnaire Rating Scales
How would you rate the ease or difficulty using the device and phone app? · 2 - Difficult
|
0 Participants
|
2 Participants
|
2 Participants
|
Adverse Events
18-49 Year Old Smokers Using COBSS
≥50 Year Old Smokers Using COBSS
All Participants
Serious adverse events
Adverse event data not reported
Other adverse events
Adverse event data not reported
Additional Information
Results disclosure agreements
- Principal investigator is a sponsor employee
- Publication restrictions are in place