Trial Outcomes & Findings for Accuracy of Methemoglobin Measurement of Various Rainbow Pulse Oximeter Sensors (NCT NCT03124836)
NCT ID: NCT03124836
Last Updated: 2017-07-27
Results Overview
Accuracy willl be determined by comparing the noninvasive blood methemoglobin measurement (expressed as a percentage of total hemoglobin) of the pulse oximeter to that obtained from a blood sample and calculating the arithmetic root mean square(Arms) error value.In order to obtain the Arms value, the blood sample methemoglobin value is subtracted from the pulse oximeter methemoglobin value for a number of samples, the average of this difference is computed as the bias. The standard deviation of the differences is computed as the precision. The square root of the sum of the squares of bias and precision is computed as the Arms Error value.
COMPLETED
NA
28 participants
1-5 hours
2017-07-27
Participant Flow
Participant milestones
| Measure |
R2-25 Sensor
All subjects will be enrolled in the test group and will receive R2-25 Pulse Oximeter Sensor
Pulse Oximeter Sensor
|
|---|---|
|
Overall Study
STARTED
|
28
|
|
Overall Study
COMPLETED
|
28
|
|
Overall Study
NOT COMPLETED
|
0
|
Reasons for withdrawal
Withdrawal data not reported
Baseline Characteristics
Accuracy of Methemoglobin Measurement of Various Rainbow Pulse Oximeter Sensors
Baseline characteristics by cohort
| Measure |
R2-25 Sensor
n=28 Participants
All subjects will be enrolled in the test group and will receive R2-25 Pulse Oximeter Sensor
Pulse Oximeter Sensor
|
|---|---|
|
Age, Categorical
<=18 years
|
0 Participants
n=99 Participants
|
|
Age, Categorical
Between 18 and 65 years
|
28 Participants
n=99 Participants
|
|
Age, Categorical
>=65 years
|
0 Participants
n=99 Participants
|
|
Sex: Female, Male
Female
|
10 Participants
n=99 Participants
|
|
Sex: Female, Male
Male
|
18 Participants
n=99 Participants
|
|
Region of Enrollment
United States
|
28 participants
n=99 Participants
|
PRIMARY outcome
Timeframe: 1-5 hoursAccuracy willl be determined by comparing the noninvasive blood methemoglobin measurement (expressed as a percentage of total hemoglobin) of the pulse oximeter to that obtained from a blood sample and calculating the arithmetic root mean square(Arms) error value.In order to obtain the Arms value, the blood sample methemoglobin value is subtracted from the pulse oximeter methemoglobin value for a number of samples, the average of this difference is computed as the bias. The standard deviation of the differences is computed as the precision. The square root of the sum of the squares of bias and precision is computed as the Arms Error value.
Outcome measures
| Measure |
R2-25 Sensor
n=28 Participants
All subjects will be enrolled in the test group and will receive R2-25 Pulse Oximeter Sensor
Pulse Oximeter Sensor
|
|---|---|
|
Accuracy of Methemoglobin Measurement by Arms Calculation
|
0.48 percentage of total hemoglobin
|
Adverse Events
R2-25 Sensor
Serious adverse events
Adverse event data not reported
Other adverse events
Adverse event data not reported