Trial Outcomes & Findings for Effect of the Built Operating Room Environment on Patient Outcome (NCT NCT03037905)
NCT ID: NCT03037905
Last Updated: 2020-07-16
Results Overview
Change (difference) in anxiety level as measured by a 100 millimeter line visual analog scale from baseline (pre-operative) to arrival in post anesthesia care unit (post-operative).The anxiety visual analog scale quantifies current state of anxiety. It is measured on a 100 millimeter line with anchors "No anxiety" to "The most anxiety I can imagine." Participants make a mark across the line to represent their current state of anxiety. A millimeter ruler is used to measure from left to right with the number of millimeters recorded where the mark crosses the line, minimum score is 0 and maximum score is 100. Higher number means more anxiety.
COMPLETED
NA
100 participants
Pre-operatively to post-operatively
2020-07-16
Participant Flow
Participant milestones
| Measure |
SignatureSuite OR
Patients who's surgeon has been randomized into the SignatureSuite operating room. The intervention is the unique operating room environment created by the product.
SignatureSuite™ OR Integration System by STERIS Corporation: SignatureSuite™ OR Integration System by STERIS Corporation is a product that includes software to control the audio and visual components of the operating room in order to provide a more relaxed and calming environment for patients prior to their surgery. It automatically dims lights, plays pleasant music or sounds of nature over the in-room speakers, and provides scenes of nature on video screens stationed in viewing range of the patient. As anesthesia is administered to the patient, the system automatically adjusts the light and sound systems to come up to surgeon-preferred preset stations. The process happens in reverse as the patient emerges from general anesthesia so that they awaken in the same environment as when they arrived in the OR.
|
Standard OR
Patients who's surgeon has been randomized into the standard operating room.
|
|---|---|---|
|
Overall Study
STARTED
|
50
|
50
|
|
Overall Study
COMPLETED
|
50
|
50
|
|
Overall Study
NOT COMPLETED
|
0
|
0
|
Reasons for withdrawal
Withdrawal data not reported
Baseline Characteristics
Effect of the Built Operating Room Environment on Patient Outcome
Baseline characteristics by cohort
| Measure |
SignatureSuite OR
n=50 Participants
Patients who's surgeon has been randomized into the SignatureSuite operating room. The intervention is the unique operating room environment created by the product.
SignatureSuite™ OR Integration System by STERIS Corporation: SignatureSuite™ OR Integration System by STERIS Corporation is a product that includes software to control the audio and visual components of the operating room in order to provide a more relaxed and calming environment for patients prior to their surgery. It automatically dims lights, plays pleasant music or sounds of nature over the in-room speakers, and provides scenes of nature on video screens stationed in viewing range of the patient. As anesthesia is administered to the patient, the system automatically adjusts the light and sound systems to come up to surgeon-preferred preset stations. The process happens in reverse as the patient emerges from general anesthesia so that they awaken in the same environment as when they arrived in the OR.
|
Standard OR
n=50 Participants
Patients who's surgeon has been randomized into the standard operating room.
|
Total
n=100 Participants
Total of all reporting groups
|
|---|---|---|---|
|
Age, Continuous
|
51.7 years
STANDARD_DEVIATION 14.3 • n=99 Participants
|
53.3 years
STANDARD_DEVIATION 14.3 • n=107 Participants
|
52.5 years
STANDARD_DEVIATION 14.2 • n=206 Participants
|
|
Sex: Female, Male
Female
|
41 Participants
n=99 Participants
|
43 Participants
n=107 Participants
|
84 Participants
n=206 Participants
|
|
Sex: Female, Male
Male
|
9 Participants
n=99 Participants
|
7 Participants
n=107 Participants
|
16 Participants
n=206 Participants
|
|
Ethnicity (NIH/OMB)
Hispanic or Latino
|
0 Participants
n=99 Participants
|
1 Participants
n=107 Participants
|
1 Participants
n=206 Participants
|
|
Ethnicity (NIH/OMB)
Not Hispanic or Latino
|
49 Participants
n=99 Participants
|
48 Participants
n=107 Participants
|
97 Participants
n=206 Participants
|
|
Ethnicity (NIH/OMB)
Unknown or Not Reported
|
1 Participants
n=99 Participants
|
1 Participants
n=107 Participants
|
2 Participants
n=206 Participants
|
|
Race (NIH/OMB)
American Indian or Alaska Native
|
0 Participants
n=99 Participants
|
0 Participants
n=107 Participants
|
0 Participants
n=206 Participants
|
|
Race (NIH/OMB)
Asian
|
2 Participants
n=99 Participants
|
0 Participants
n=107 Participants
|
2 Participants
n=206 Participants
|
|
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
|
0 Participants
n=99 Participants
|
0 Participants
n=107 Participants
|
0 Participants
n=206 Participants
|
|
Race (NIH/OMB)
Black or African American
|
3 Participants
n=99 Participants
|
12 Participants
n=107 Participants
|
15 Participants
n=206 Participants
|
|
Race (NIH/OMB)
White
|
41 Participants
n=99 Participants
|
36 Participants
n=107 Participants
|
77 Participants
n=206 Participants
|
|
Race (NIH/OMB)
More than one race
|
1 Participants
n=99 Participants
|
2 Participants
n=107 Participants
|
3 Participants
n=206 Participants
|
|
Race (NIH/OMB)
Unknown or Not Reported
|
3 Participants
n=99 Participants
|
0 Participants
n=107 Participants
|
3 Participants
n=206 Participants
|
|
Surgical site
Parathyroid
|
16 Participants
n=99 Participants
|
22 Participants
n=107 Participants
|
38 Participants
n=206 Participants
|
|
Surgical site
Thyroid
|
34 Participants
n=99 Participants
|
28 Participants
n=107 Participants
|
62 Participants
n=206 Participants
|
|
Kindler Preoperative Anxiety Scale Total Score
|
29.8 units on a scale
n=99 Participants
|
26.6 units on a scale
n=107 Participants
|
28.9 units on a scale
n=206 Participants
|
|
Past medical history diagnosis groups
cardiovascular/respiratory
|
13 participants
n=99 Participants
|
19 participants
n=107 Participants
|
32 participants
n=206 Participants
|
|
Past medical history diagnosis groups
Other
|
15 participants
n=99 Participants
|
19 participants
n=107 Participants
|
34 participants
n=206 Participants
|
|
Past medical history diagnosis groups
endocrine
|
49 participants
n=99 Participants
|
48 participants
n=107 Participants
|
97 participants
n=206 Participants
|
|
Past medical history diagnosis groups
gastrointestinal/genitourinary
|
20 participants
n=99 Participants
|
18 participants
n=107 Participants
|
38 participants
n=206 Participants
|
|
Past medical history diagnosis groups
neuromuscular
|
16 participants
n=99 Participants
|
25 participants
n=107 Participants
|
41 participants
n=206 Participants
|
|
Preoperative anxiety visual analog scale
|
36.3 units on a scale
STANDARD_DEVIATION 27.9 • n=99 Participants
|
32.0 units on a scale
STANDARD_DEVIATION 28.9 • n=107 Participants
|
34.13 units on a scale
STANDARD_DEVIATION 28.4 • n=206 Participants
|
|
Preoperative systolic blood pressure
|
137.50 mm Hg
STANDARD_DEVIATION 17.4 • n=99 Participants
|
142.6 mm Hg
STANDARD_DEVIATION 19.8 • n=107 Participants
|
140.0 mm Hg
STANDARD_DEVIATION 18.7 • n=206 Participants
|
|
Preoperative heart rate
|
77.9 beats per minute
STANDARD_DEVIATION 12.7 • n=99 Participants
|
78.8 beats per minute
STANDARD_DEVIATION 15.3 • n=107 Participants
|
78.4 beats per minute
STANDARD_DEVIATION 14.0 • n=206 Participants
|
PRIMARY outcome
Timeframe: Pre-operatively to post-operativelyPopulation: thyroid and/or parathyroid surgery patients
Change (difference) in anxiety level as measured by a 100 millimeter line visual analog scale from baseline (pre-operative) to arrival in post anesthesia care unit (post-operative).The anxiety visual analog scale quantifies current state of anxiety. It is measured on a 100 millimeter line with anchors "No anxiety" to "The most anxiety I can imagine." Participants make a mark across the line to represent their current state of anxiety. A millimeter ruler is used to measure from left to right with the number of millimeters recorded where the mark crosses the line, minimum score is 0 and maximum score is 100. Higher number means more anxiety.
Outcome measures
| Measure |
SignatureSuite OR
n=50 Participants
Patients who's surgeon has been randomized into the SignatureSuite operating room. The intervention is the unique operating room environment created by the product.
SignatureSuite™ OR Integration System by STERIS Corporation: SignatureSuite™ OR Integration System by STERIS Corporation is a product that includes software to control the audio and visual components of the operating room in order to provide a more relaxed and calming environment for patients prior to their surgery. It automatically dims lights, plays pleasant music or sounds of nature over the in-room speakers, and provides scenes of nature on video screens stationed in viewing range of the patient. As anesthesia is administered to the patient, the system automatically adjusts the light and sound systems to come up to surgeon-preferred preset stations. The process happens in reverse as the patient emerges from general anesthesia so that they awaken in the same environment as when they arrived in the OR.
|
Standard OR
n=50 Participants
Patients who's surgeon has been randomized into the standard operating room.
|
|---|---|---|
|
Change (Difference) in Anxiety
|
-24.4 units on a scale
Interval -31.7 to -17.0
|
-17.6 units on a scale
Interval -25.1 to -10.2
|
SECONDARY outcome
Timeframe: Pre-operative to post-operativePopulation: Patients undergoing thyroid and/or parathyroid surgery
Difference between pre-operative systolic blood pressure and a mean of 3 post-operative systolic blood pressures assessed at 15, 30 and 45 minutes after placement in post-anesthesia care unit
Outcome measures
| Measure |
SignatureSuite OR
n=50 Participants
Patients who's surgeon has been randomized into the SignatureSuite operating room. The intervention is the unique operating room environment created by the product.
SignatureSuite™ OR Integration System by STERIS Corporation: SignatureSuite™ OR Integration System by STERIS Corporation is a product that includes software to control the audio and visual components of the operating room in order to provide a more relaxed and calming environment for patients prior to their surgery. It automatically dims lights, plays pleasant music or sounds of nature over the in-room speakers, and provides scenes of nature on video screens stationed in viewing range of the patient. As anesthesia is administered to the patient, the system automatically adjusts the light and sound systems to come up to surgeon-preferred preset stations. The process happens in reverse as the patient emerges from general anesthesia so that they awaken in the same environment as when they arrived in the OR.
|
Standard OR
n=50 Participants
Patients who's surgeon has been randomized into the standard operating room.
|
|---|---|---|
|
Difference Between Pre-operative Systolic Blood Pressure and a Mean of 3 Post-operative Systolic Blood Pressures
|
-4.2 mm Hg
Interval -9.8 to 1.4
|
-6.4 mm Hg
Interval -12.3 to -0.5
|
SECONDARY outcome
Timeframe: Pre-operative to post-operativePopulation: Patients undergoing thyroid and/or parathyroid surgery
Difference between pre-operative heart rate and a mean of 3 post-operative heart rates assessed at 15, 30 and 45 minutes after placement in post-anesthesia care unit
Outcome measures
| Measure |
SignatureSuite OR
n=50 Participants
Patients who's surgeon has been randomized into the SignatureSuite operating room. The intervention is the unique operating room environment created by the product.
SignatureSuite™ OR Integration System by STERIS Corporation: SignatureSuite™ OR Integration System by STERIS Corporation is a product that includes software to control the audio and visual components of the operating room in order to provide a more relaxed and calming environment for patients prior to their surgery. It automatically dims lights, plays pleasant music or sounds of nature over the in-room speakers, and provides scenes of nature on video screens stationed in viewing range of the patient. As anesthesia is administered to the patient, the system automatically adjusts the light and sound systems to come up to surgeon-preferred preset stations. The process happens in reverse as the patient emerges from general anesthesia so that they awaken in the same environment as when they arrived in the OR.
|
Standard OR
n=50 Participants
Patients who's surgeon has been randomized into the standard operating room.
|
|---|---|---|
|
Difference Between Pre-operative Heart Rate and a Mean of 3 Post-operative Heart Rates
|
3.1 beats per minute
Interval -1.3 to 7.5
|
4.1 beats per minute
Interval -0.2 to 8.3
|
OTHER_PRE_SPECIFIED outcome
Timeframe: PostoperativelyPopulation: Patients that have undergone thyroid and/or parathyroid surgery
Number of people requiring analgesia in the post-anesthesia care unit
Outcome measures
| Measure |
SignatureSuite OR
n=50 Participants
Patients who's surgeon has been randomized into the SignatureSuite operating room. The intervention is the unique operating room environment created by the product.
SignatureSuite™ OR Integration System by STERIS Corporation: SignatureSuite™ OR Integration System by STERIS Corporation is a product that includes software to control the audio and visual components of the operating room in order to provide a more relaxed and calming environment for patients prior to their surgery. It automatically dims lights, plays pleasant music or sounds of nature over the in-room speakers, and provides scenes of nature on video screens stationed in viewing range of the patient. As anesthesia is administered to the patient, the system automatically adjusts the light and sound systems to come up to surgeon-preferred preset stations. The process happens in reverse as the patient emerges from general anesthesia so that they awaken in the same environment as when they arrived in the OR.
|
Standard OR
n=50 Participants
Patients who's surgeon has been randomized into the standard operating room.
|
|---|---|---|
|
Analgesic Use
|
9 Participants
|
10 Participants
|
Adverse Events
SignatureSuite OR
Standard OR
Serious adverse events
Adverse event data not reported
Other adverse events
Adverse event data not reported
Additional Information
Results disclosure agreements
- Principal investigator is a sponsor employee
- Publication restrictions are in place