Trial Outcomes & Findings for Clinical Evaluation Of MP26 Features in Adults (NCT NCT02927327)

NCT ID: NCT02927327

Last Updated: 2018-12-06

Results Overview

scored on a 5-pt Likert scale: 1. Unacceptable 2. Poor 3. Acceptable 4. Good 5. Excellent

Recruitment status

COMPLETED

Study phase

NA

Target enrollment

61 participants

Primary outcome timeframe

through study completion, an average of 2 months

Results posted on

2018-12-06

Participant Flow

Participant milestones

Participant milestones
Measure
PET/MR ZTE MRAC
PET/MR zero echo time (ZTE) scan for head attenuation and Q.Static . PET/MR ZTE MRAC images to be acquired for post-processing and reader assessment. Zero Echo Time (ZTE) scan for head attenuation Zero Echo Time (ZTE) scan for head attenuation: The GE SIGNA PET/MR MP26 software platform includes the zero echo time (ZTE) scan for head attenuation and Q.Static. The ZTE MR software feature has the potential to enable better visualization of bones, including those in the head, by employing optimal head attenuation correction in PET/MR. PET/MR ZTE MRAC
PET/MR Q Static (Q. MRAC)
Q.Static (Q. MRAC) for respiratory motion correction. PET/MR Q Static (Q. MRAC) images to be acquired for post-processing and reader assessment. PET/MR Q Static (Q. MRAC) Q.Static (Q. MRAC) for respiratory motion correction: An improved Q.Static feature with Q. MRAC, where phase matching the MRAC with the quiescent phase is employed to get more accurate attenuation correction for our Q.Static PET images. PET/MR Q Static (Q. MRAC)
Overall Study
STARTED
30
31
Overall Study
COMPLETED
30
31
Overall Study
NOT COMPLETED
0
0

Reasons for withdrawal

Withdrawal data not reported

Baseline Characteristics

Race and Ethnicity were not collected from any participant.

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
PET/MR ZTE MRAC
n=30 Participants
PET/MR zero echo time (ZTE) scan for head attenuation and Q.Static . PET/MR ZTE MRAC images to be acquired for post-processing and reader assessment. Zero Echo Time (ZTE) scan for head attenuation Zero Echo Time (ZTE) scan for head attenuation: The GE SIGNA PET/MR MP26 software platform includes the zero echo time (ZTE) scan for head attenuation and Q.Static. The ZTE MR software feature has the potential to enable better visualization of bones, including those in the head, by employing optimal head attenuation correction in PET/MR. PET/MR ZTE MRAC
PET/MR Q Static (Q. MRAC)
n=31 Participants
Q.Static (Q. MRAC) for respiratory motion correction. PET/MR Q Static (Q. MRAC) images to be acquired for post-processing and reader assessment. PET/MR Q Static (Q. MRAC) Q.Static (Q. MRAC) for respiratory motion correction: An improved Q.Static feature with Q. MRAC, where phase matching the MRAC with the quiescent phase is employed to get more accurate attenuation correction for our Q.Static PET images. PET/MR Q Static (Q. MRAC)
Total
n=61 Participants
Total of all reporting groups
Age, Categorical
<=18 years
0 Participants
n=30 Participants
0 Participants
n=31 Participants
0 Participants
n=61 Participants
Age, Categorical
Between 18 and 65 years
17 Participants
n=30 Participants
11 Participants
n=31 Participants
28 Participants
n=61 Participants
Age, Categorical
>=65 years
13 Participants
n=30 Participants
20 Participants
n=31 Participants
33 Participants
n=61 Participants
Age, Continuous
60.6 years
STANDARD_DEVIATION 12.86 • n=30 Participants
67.3 years
STANDARD_DEVIATION 10.76 • n=31 Participants
64.0 years
STANDARD_DEVIATION 12.21 • n=61 Participants
Sex: Female, Male
Female
12 Participants
n=30 Participants
14 Participants
n=31 Participants
26 Participants
n=61 Participants
Sex: Female, Male
Male
18 Participants
n=30 Participants
17 Participants
n=31 Participants
35 Participants
n=61 Participants
Race and Ethnicity Not Collected
0 Participants
Race and Ethnicity were not collected from any participant.
Region of Enrollment
United States
30 Participants
n=30 Participants
31 Participants
n=31 Participants
61 Participants
n=61 Participants

PRIMARY outcome

Timeframe: through study completion, an average of 2 months

Population: Subjects that had a clinical indication for PET/CT or PET/MR with radiotracer injection prescribed for their medical care outside of the study.

scored on a 5-pt Likert scale: 1. Unacceptable 2. Poor 3. Acceptable 4. Good 5. Excellent

Outcome measures

Outcome measures
Measure
PET/MR ZTE MRAC
n=30 Participants
PET/MR zero echo time (ZTE) scan for head attenuation and Q.Static . PET/MR ZTE MRAC images to be acquired for post-processing and reader assessment. Zero Echo Time (ZTE) scan for head attenuation Zero Echo Time (ZTE) scan for head attenuation: The GE SIGNA PET/MR MP26 software platform includes the zero echo time (ZTE) scan for head attenuation and Q.Static. The ZTE MR software feature has the potential to enable better visualization of bones, including those in the head, by employing optimal head attenuation correction in PET/MR. PET/MR ZTE MRAC
PET/MR Q Static (Q. MRAC)
n=31 Participants
Q.Static (Q. MRAC) for respiratory motion correction. PET/MR Q Static (Q. MRAC) images to be acquired for post-processing and reader assessment. PET/MR Q Static (Q. MRAC) Q.Static (Q. MRAC) for respiratory motion correction: An improved Q.Static feature with Q. MRAC, where phase matching the MRAC with the quiescent phase is employed to get more accurate attenuation correction for our Q.Static PET images. PET/MR Q Static (Q. MRAC)
Diagnostic Image Quality
4.4 scores on a scale
Standard Deviation 0.54
3.9 scores on a scale
Standard Deviation 0.68

SECONDARY outcome

Timeframe: through study completion, an average of 2 months

Population: Diagnostic acceptability in the form of binary responses (Y/N) were collected from 61 subjects who completed the study.

Site level determination of diagnostic acceptability (Y/N) based on consensus between radiologist(s) and nuclear medicine physician(s) at the site. Diagnostic acceptability in the form of binary responses (Y/N) were collected from 61 subjects who completed the study.

Outcome measures

Outcome measures
Measure
PET/MR ZTE MRAC
n=30 Participants
PET/MR zero echo time (ZTE) scan for head attenuation and Q.Static . PET/MR ZTE MRAC images to be acquired for post-processing and reader assessment. Zero Echo Time (ZTE) scan for head attenuation Zero Echo Time (ZTE) scan for head attenuation: The GE SIGNA PET/MR MP26 software platform includes the zero echo time (ZTE) scan for head attenuation and Q.Static. The ZTE MR software feature has the potential to enable better visualization of bones, including those in the head, by employing optimal head attenuation correction in PET/MR. PET/MR ZTE MRAC
PET/MR Q Static (Q. MRAC)
n=31 Participants
Q.Static (Q. MRAC) for respiratory motion correction. PET/MR Q Static (Q. MRAC) images to be acquired for post-processing and reader assessment. PET/MR Q Static (Q. MRAC) Q.Static (Q. MRAC) for respiratory motion correction: An improved Q.Static feature with Q. MRAC, where phase matching the MRAC with the quiescent phase is employed to get more accurate attenuation correction for our Q.Static PET images. PET/MR Q Static (Q. MRAC)
Determination of Diagnostic Acceptability
30 Participants
31 Participants

SECONDARY outcome

Timeframe: through study completion, an average of 2 months

Ease of use ratings, based on a 5-point Likert scale (e.g. 1=unacceptable, 2=poor, 3=acceptable, 4=good, 5=excellent), were collected from the population of subjects who completed the study.

Outcome measures

Outcome measures
Measure
PET/MR ZTE MRAC
n=30 Participants
PET/MR zero echo time (ZTE) scan for head attenuation and Q.Static . PET/MR ZTE MRAC images to be acquired for post-processing and reader assessment. Zero Echo Time (ZTE) scan for head attenuation Zero Echo Time (ZTE) scan for head attenuation: The GE SIGNA PET/MR MP26 software platform includes the zero echo time (ZTE) scan for head attenuation and Q.Static. The ZTE MR software feature has the potential to enable better visualization of bones, including those in the head, by employing optimal head attenuation correction in PET/MR. PET/MR ZTE MRAC
PET/MR Q Static (Q. MRAC)
n=31 Participants
Q.Static (Q. MRAC) for respiratory motion correction. PET/MR Q Static (Q. MRAC) images to be acquired for post-processing and reader assessment. PET/MR Q Static (Q. MRAC) Q.Static (Q. MRAC) for respiratory motion correction: An improved Q.Static feature with Q. MRAC, where phase matching the MRAC with the quiescent phase is employed to get more accurate attenuation correction for our Q.Static PET images. PET/MR Q Static (Q. MRAC)
Ease of Use Per Procedure Rated on a 5-pt Likert Scale Score
Prescribing PET scan
4.6 scores on a scale
Standard Deviation 0.73
4.6 scores on a scale
Standard Deviation 0.73
Ease of Use Per Procedure Rated on a 5-pt Likert Scale Score
Prescribing MR scans simultaneous with PET
4.6 scores on a scale
Standard Deviation 0.73
4.6 scores on a scale
Standard Deviation 0.73
Ease of Use Per Procedure Rated on a 5-pt Likert Scale Score
Starting and monitoring scanning
4.6 scores on a scale
Standard Deviation 0.81
4.6 scores on a scale
Standard Deviation 0.81
Ease of Use Per Procedure Rated on a 5-pt Likert Scale Score
Reviewing PET and MR images
4.6 scores on a scale
Standard Deviation 0.73
4.6 scores on a scale
Standard Deviation 0.73
Ease of Use Per Procedure Rated on a 5-pt Likert Scale Score
Overall process of doing this exam
4.6 scores on a scale
Standard Deviation 0.78
4.6 scores on a scale
Standard Deviation 0.78
Ease of Use Per Procedure Rated on a 5-pt Likert Scale Score
Performing exam within acceptable duration to meet
4.6 scores on a scale
Standard Deviation 0.85
4.6 scores on a scale
Standard Deviation 0.85

SECONDARY outcome

Timeframe: through study completion, an average of 2 months

Population: Functionality ratings, based on a binary performance scale (Y/N), were collected from the population of subjects who completed the study, including both ZTE and Q.Static populations (N=61 total subjects).

The number of participants reporting "yes" on a binary performance scale (Y/N) were collected.

Outcome measures

Outcome measures
Measure
PET/MR ZTE MRAC
n=30 Participants
PET/MR zero echo time (ZTE) scan for head attenuation and Q.Static . PET/MR ZTE MRAC images to be acquired for post-processing and reader assessment. Zero Echo Time (ZTE) scan for head attenuation Zero Echo Time (ZTE) scan for head attenuation: The GE SIGNA PET/MR MP26 software platform includes the zero echo time (ZTE) scan for head attenuation and Q.Static. The ZTE MR software feature has the potential to enable better visualization of bones, including those in the head, by employing optimal head attenuation correction in PET/MR. PET/MR ZTE MRAC
PET/MR Q Static (Q. MRAC)
n=31 Participants
Q.Static (Q. MRAC) for respiratory motion correction. PET/MR Q Static (Q. MRAC) images to be acquired for post-processing and reader assessment. PET/MR Q Static (Q. MRAC) Q.Static (Q. MRAC) for respiratory motion correction: An improved Q.Static feature with Q. MRAC, where phase matching the MRAC with the quiescent phase is employed to get more accurate attenuation correction for our Q.Static PET images. PET/MR Q Static (Q. MRAC)
Number of Participants Reporting "Yes" On All Binary Performance Scale Questions
30 Participants
31 Participants

OTHER_PRE_SPECIFIED outcome

Timeframe: through study completion, an average of 2 months

Number of safety events (AEs and SAEs).

Outcome measures

Outcome data not reported

Adverse Events

PET/MR ZTE MRAC

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

PET/MR Q Static (Q. MRAC)

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Serious adverse events

Adverse event data not reported

Other adverse events

Adverse event data not reported

Additional Information