Trial Outcomes & Findings for The Set-Point Study: Evaluating Effects of Changing Glucose Target on Bionic Pancreas Performance (NCT NCT02509065)

NCT ID: NCT02509065

Last Updated: 2018-11-20

Results Overview

Glucose values were collected from the Dexcom G4 CGM device every 5 minutes and an average (calculated mean with associated standard deviation) CGM glucose level (mg/dl) was calculated from days 2 and 3 of the study arm.

Recruitment status

COMPLETED

Study phase

NA

Target enrollment

24 participants

Primary outcome timeframe

Days 2 and 3

Results posted on

2018-11-20

Participant Flow

27 subjects were recruited and found eligible. Of the 27, only 24 subjects participated in the trial.

There were 24 subjects and a total of 9 study arms.The order of the arms was assigned in a psuedo random cross-over design. The 110 and 120 mg/dl arms were added later and were not included in the original randomization. 11 of the same subjects completed all 9 arms. Each arm had 20 subjects total, but these 20 were not the same cohort in each arm.

Participant milestones

Participant milestones
Measure
110B, 110IO
Arms were completed in the following order: 110 mg/dl bihormonal bionic pancreas, 110 mg/dl insulin only bionic pancreas
110B, 110IO, 120IO
Arms were completed in the following order: 110 mg/dl bihormonal bionic pancreas 110 mg/dl insulin only bionic pancreas 120 mg/dl insulin only bionic pancreas
110IO, 110B, 120IO
Arms were completed in the following order: 110 mg/dl insulin only bionic pancreas, 110 mg/dl bihormonal bionic pancreas, 120 mg/dl insulin only bionic pancreas
UC, 115B, 100B, 145IO, 130B, 130IO
Arms were completed in the following order: usual care, 115 mg/dl bihormonal bionic pancreas, 100 mg/dl bihormonal bionic pancreas, 145 mg/dl insulin only bionic pancreas, 130 mg/dl bihormonal bionic pancreas, 130 mg/dl insulin only bionic pancreas
145IO, 100B, 130B, 130IO, UC, 115
Arms were completed in the following order: 145 mg/dl insulin only bionic pancreas, 100 mg/dl bihormonal bionic pancreas, 130 mg/dl bihormonal bionic pancreas, 130 mg/dl insulin only bionic pancreas, usual care, 115 mg/dl bihormonal bionic pancreas
UC, 115B, 145IO, 100B, 130IO, 130B
Arms were completed in the following order: usual care, 115 mg/dl bihormonal bionic pancreas, 145 mg/dl insulin only bionic pancreas, 100 mg/dl bihormonal bionic pancreas, 130 mg/dl insulin only bionic pancreas, 130 mg/dl bihormonal bionic pancreas
130B, 130IO, 115B, UC, 100B, 145IO, 120IO
Arms were completed in the following order: 130 mg/dl bihormonal bionic pancreas, 130 mg/dl insulin only bionic pancreas, 115 mg/dl bihormonal bionic pancreas, usual care, 100 mg/dl bihormonal bionic pancreas, 145 mg/dl insulin only bionic pancreas, 120 mg/dl insulin only bionic pancreas
145IO, 100B, UC, 115B, 130IO, 130B, 110B, 110IO
Arms were completed in the following order: 145 mg/dl insulin only bionic pancreas, 100 mg/dl bihormonal bionic pancreas, usual care, 115 mg/dl bihormonal bionic pancreas, 130 mg/dl insulin only bionic pancreas, 130 mg/dl bihormonal bionic pancreas, 110 mg/dl bihormonal bionic pancreas, 110 mg/dl insulin only bionic pancreas
UC, 115B, 100B, 145IO, 130IO, 130B, 110IO, 110B
Arms were completed int he following order: usual care, 115 mg/dl bihormonal bionic pancreas, 100 mg/dl bihormonal bionic pancreas, 145 mg/dl insulin only bionic pancreas, 130 mg/dl insulin only bionic pancreas, 130 mg/dl bihormonal bionic pancreas, 110 mg/dl insulin only bionic pancreas, 110 mg/dl bihormonal bionic pancreas
130IO, 130B, UC, 145IO, 115B, 100B, 110B, 110IO
Arms were completed in the following order: 130 mg/dl insulin only bionic pancreas, 130 mg/dl bihormonal bionic pancreas, usual care, 145 mg/dl insulin only bionic pancreas, 115 mg/dl bihormonal bionic pancreas, 100 mg/dl bihormonal bionic pancreas, 110 mg/dl bihormonal bionic pancreas, 110 mg/dl insulin only bionic pancreas
130B, 130IO, UC, 145IO, 115B, 100B, 110B, 110IO
Arms were completed in the following order: 130 mg/dl bihormonal bionic pancreas,130 mg/dl insulin only bionic pancreas, usual care, 145 mg/dl insulin only bionic pancreas, 115 mg/dl bihormonal bionic pancreas, 100 mg/dl bihormonal bionic pancreas, 110 mg/dl insulin only bionic pancreas, 110 mg/dl bihormonal bionic pancreas
100B, 145IO, UC, 115B, 130IO, 130B, 110IO, 110B
Arms were completed in the following order: 100 mg/dl bihormonal bionic pancreas, 145 mg/dl insulin only bionic pancreas, usual care, 115 mg/dl bihormonal bionic pancreas, 130 mg/dl insulin only bionic pancreas, 130 mg/dl bihormonal bionic pancreas, 110 mg/dl insulin only bionic pancreas, 110 mg/dl bihormonal bionic pancreas
130IO, 130B, 145IO, 100B, 115B, UC, 110B, 110IO, 120IO
Arms were completed in the following order: 130 mg/dl insulin only bionic pancreas 130 mg/dl bihormonal bionic pancreas 145 mg/dl insulin only bionic pancreas 100 mg/dl bihormonal bionic pancreas 115 mg/dl bihormonal bionic pancreas Usual Care 110 mg/dl bihormonal bionic pancreas 110 mg/dl insulin only bionic pancreas 120 mg/dl insulin only bionic pancreas
115B, UC, 130B, 130IO, 145IO, 100B, 110IO, 110B, 120IO
Arms were completed in the following order: 115 mg/dl bihormonal bionic pancreas Usual Care 130 mg/dl bihormonal bionic pancreas 130 mg/dl insulin only bionic pancreas 145 mg/dl insulin only bionic pancreas 100 mg/dl bihormonal bionic pancreas 110 mg/dl insulin only bionic pancreas 110 mg/dl bihormonal bionic pancreas 120 mg/dl insulin only bionic pancreas
UC, 145IO, 130B, 130IO, 100B, 115B, 110IO, 110B, 120IO
Arms were completed in the following order: Usual Care 145 mg/dl insulin only bionic pancreas 130 mg/dl bihormonal bionic pancreas 130 mg/dl insulin only bionic pancreas 100 mg/dl bihormonal bionic pancreas 115 mg/dl bihormonal bionic pancreas 110 mg/dl insulin only bionic pancreas 110 mg/dl bihormonal bionic pancreas 120 mg/dl insulin only bionic pancreas
UC, 145IO, 130B, 130IO, 100B, 115B, 110B, 110IO, 120IO
Arms were completed in the following order: Usual Care 145 mg/dl insulin only bionic pancreas 130 mg/dl bihormonal bionic pancreas 130 mg/dl insulin only bionic pancreas 100 mg/dl bihormonal bionic pancreas 115 mg/dl bihormonal bionic pancreas 110 mg/dl bihormonal bionic pancreas 110 mg/dl insulin only bionic pancreas 120 mg/dl insulin only bionic pancreas
145IO, 100B, UC, 115B, 130B, 130IO, 110IO, 110B, 120IO
Arms were completed in the following order: 145 mg/dl insulin only bionic pancreas 100 mg/dl bihormonal bionic pancreas Usual Care 115 mg/dl bihormonal bionic pancreas 130 mg/dl bihormonal bionic pancreas 130 mg/dl insulin only bionic pancreas 110 mg/dl insulin only bionic pancreas 110 mg/dl bihormonal bionic pancreas 120 mg/dl insulin only bionic pancreas
115B, UC, 130B, 130IO, 145IO, 100B, 110B, 110IO, 120IO
Arms were completed in the following order: 115 mg/dl bihormonal bionic pancreas Usual Care 130 mg/dl bihormonal bionic pancreas 130 mg/dl insulin only bionic pancreas 145 mg/dl insulin only bionic pancreas 100 mg/dl bihormonal bionic pancreas 110 mg/dl bihormonal bionic pancreas 110 mg/dl insulin only bionic pancreas 120 mg/dl insulin only bionic pancreas
130IO, 130B, 145IO, 100B, 115B, UC, 110IO, 110B, 120IO
Arms were completed in the following order: 130 mg/dl insulin only bionic pancreas 130 mg/dl bihormonal bionic pancreas 145 mg/dl insulin only bionic pancreas 100 mg/dl bihormonal bionic pancreas 115 mg/dl bihormonal bionic pancreas Usual Care 110 mg/dl insulin only bionic pancreas 110 mg/dl bihormonal bionic pancreas 120 mg/dl insulin only bionic pancreas
100B, 145IO, UC, 115B, 130IO, 130B, 110IO, 110B, 120IO
Arms were completed in the following order: 100 mg/dl bihormonal bionic pancreas 145 mg/dl insulin only bionic pancreas Usual Care 115 mg/dl bihormonal bionic pancreas 130 mg/dl insulin only bionic pancreas 130 mg/dl bihormonal bionic pancreas 110 mg/dl insulin only bionic pancreas 110 mg/dl bihormonal bionic pancreas 120 mg/dl insulin only bionic pancreas
UC, 115B, 145IO, 100B, 130B, 130IO, 110B, 110IO, 120
Arms were completed in the following order: Usual Care 115 mg/dl bihormonal bionic pancreas 145 mg/dl insulin only bionic pancreas 100 mg/dl bihormonal bionic pancreas 130 mg/dl bihormonal bionic pancreas 130 mg/dl insulin only bionic pancreas 110 mg/dl bihormonal bionic pancreas 110 mg/dl insulin only bionic pancreas 120 mg/dl insulin only bionic pancreas
145IO, 100B, 130B, 130IO, UC, 115B, 110IO, 110B, 120IO
Arms were completed in the following order: 145 mg/dl insulin only bionic pancreas 100 mg/dl bihormonal bionic pancreas 130 mg/dl bihormonal bionic pancreas 130 mg/dl insulin only bionic pancreas Usual Care 115 mg/dl bihormonal bionic pancreas 110 mg/dl insulin only bionic pancreas 110 mg/dl bihormonal bionic pancreas 120 mg/dl insulin only bionic pancreas
130IO, 130B, 115B, UC, 100B, 145IO, 110B, 110IO, 120IO
Arms were completed in the following order: 130 mg/dl insulin only bionic pancreas 130 mg/dl bihormonal bionic pancreas 115 mg/dl bihormonal bionic pancreas Usual Care 100 mg/dl bihormonal bionic pancreas 145 mg/dl insulin only bionic pancreas 110 mg/dl bihormonal bionic pancreas 110 mg/dl insulin only bionic pancreas 120 mg/dl insulin only bionic pancreas
Overall Study
STARTED
1
1
2
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
Overall Study
COMPLETED
1
1
2
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
Overall Study
NOT COMPLETED
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0

Reasons for withdrawal

Withdrawal data not reported

Baseline Characteristics

The Set-Point Study: Evaluating Effects of Changing Glucose Target on Bionic Pancreas Performance

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
All Enrolled Subjects
n=24 Participants
All subjects enrolled into the study
Age, Continuous
38.6 years
STANDARD_DEVIATION 15.5 • n=99 Participants
Sex: Female, Male
Female
13 Participants
n=99 Participants
Sex: Female, Male
Male
11 Participants
n=99 Participants
Race (NIH/OMB)
American Indian or Alaska Native
0 Participants
n=99 Participants
Race (NIH/OMB)
Asian
0 Participants
n=99 Participants
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
0 Participants
n=99 Participants
Race (NIH/OMB)
Black or African American
0 Participants
n=99 Participants
Race (NIH/OMB)
White
23 Participants
n=99 Participants
Race (NIH/OMB)
More than one race
0 Participants
n=99 Participants
Race (NIH/OMB)
Unknown or Not Reported
1 Participants
n=99 Participants
Hemoglobin A1c
7.3 percent
STANDARD_DEVIATION 0.9 • n=99 Participants

PRIMARY outcome

Timeframe: Days 2 and 3

Glucose values were collected from the Dexcom G4 CGM device every 5 minutes and an average (calculated mean with associated standard deviation) CGM glucose level (mg/dl) was calculated from days 2 and 3 of the study arm.

Outcome measures

Outcome measures
Measure
Usual Care
n=20 Participants
Comparator week to all closed-loop control, utilizing usual diabetes care and the subject's own insulin pump.
145 mg/dl Set Point - Insulin Only
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 145 mg/dl and using only an insulin pump.
130 mg/dl Set Point - Insulin Only
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 130 mg/dl and using only an insulin pump. This arm is for subjects with type 1 diabetes only. This arm is double blinded; neither the subject nor the study team know if it is insulin only or bihormonal.
130 mg/dl Set Point - Bihormonal
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 130 mg/dl using both an insulin and a glucagon pump. This arm is double blinded; neither the subject nor the study team know if it is insulin only or bihormonal.
115 mg/dl Set Point - Bihormonal
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 115 mg/dl using both an insulin and a glucagon pump.
100 mg/dl Set Point - Bihormonal
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 100 mg/dl using both an insulin and a glucagon pump.
110 mg/dl Set Point - Insulin Only
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 110 mg/dl and using only an insulin pump. This arm is double blinded; neither the subject nor the study team know if it is insulin only or bihormonal.
110 mg/dl Set Point - Bihormonal
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 110 mg/dl using both an insulin and a glucagon pump. This arm is double blinded; neither the subject nor the study team know if it is insulin only or bihormonal.
120 mg/dl Set Point - Insulin Only
n=14 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 120 mg/dl and using only an insulin pump.
Mean Continuous Glucose Monitor (CGM) Glucose Values
158 mg/dl
Standard Deviation 31
174 mg/dl
Standard Deviation 23
161 mg/dl
Standard Deviation 17
156 mg/dl
Standard Deviation 12
146 mg/dl
Standard Deviation 15
136 mg/dl
Standard Deviation 14
153 mg/dl
Standard Deviation 15
148 mg/dl
Standard Deviation 17
156 mg/dl
Standard Deviation 15

PRIMARY outcome

Timeframe: Days 2 and 3

For this calculation we analyzed data from every 5 minutes of CGM data for days 2 and 3 of each study arm. We looked specifically at the percentage time of those roughly 48 hours that had any glucose values less than 60 mg/dl. The final value presented is the percent of time per 2 days of the study arm spent in a specific hypoglycemia range of less than 60 mg/dl with an associated standard deviation.

Outcome measures

Outcome measures
Measure
Usual Care
n=20 Participants
Comparator week to all closed-loop control, utilizing usual diabetes care and the subject's own insulin pump.
145 mg/dl Set Point - Insulin Only
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 145 mg/dl and using only an insulin pump.
130 mg/dl Set Point - Insulin Only
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 130 mg/dl and using only an insulin pump. This arm is for subjects with type 1 diabetes only. This arm is double blinded; neither the subject nor the study team know if it is insulin only or bihormonal.
130 mg/dl Set Point - Bihormonal
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 130 mg/dl using both an insulin and a glucagon pump. This arm is double blinded; neither the subject nor the study team know if it is insulin only or bihormonal.
115 mg/dl Set Point - Bihormonal
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 115 mg/dl using both an insulin and a glucagon pump.
100 mg/dl Set Point - Bihormonal
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 100 mg/dl using both an insulin and a glucagon pump.
110 mg/dl Set Point - Insulin Only
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 110 mg/dl and using only an insulin pump. This arm is double blinded; neither the subject nor the study team know if it is insulin only or bihormonal.
110 mg/dl Set Point - Bihormonal
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 110 mg/dl using both an insulin and a glucagon pump. This arm is double blinded; neither the subject nor the study team know if it is insulin only or bihormonal.
120 mg/dl Set Point - Insulin Only
n=14 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 120 mg/dl and using only an insulin pump.
Percentage of Time With CGM < 60 mg/dl
1.38 percentage of time
Standard Deviation 2.6
0.96 percentage of time
Standard Deviation 1.5
0.79 percentage of time
Standard Deviation 1.4
0.55 percentage of time
Standard Deviation 1.0
0.88 percentage of time
Standard Deviation 1.2
0.76 percentage of time
Standard Deviation 1.1
1.34 percentage of time
Standard Deviation 1.3
0.4 percentage of time
Standard Deviation 0.6
1.05 percentage of time
Standard Deviation 1.7

PRIMARY outcome

Timeframe: Day 4 in-clinic Exercise Visit (110 and 130 mg/dl arms only)

During both 110 mg/dl and 130 mg/dl arms, subjects will report for a fasted in-clinic exercise visit, with plasma blood glucose measurements obtained at least every 10 minutes

Outcome measures

Outcome measures
Measure
Usual Care
n=20 Participants
Comparator week to all closed-loop control, utilizing usual diabetes care and the subject's own insulin pump.
145 mg/dl Set Point - Insulin Only
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 145 mg/dl and using only an insulin pump.
130 mg/dl Set Point - Insulin Only
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 130 mg/dl and using only an insulin pump. This arm is for subjects with type 1 diabetes only. This arm is double blinded; neither the subject nor the study team know if it is insulin only or bihormonal.
130 mg/dl Set Point - Bihormonal
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 130 mg/dl using both an insulin and a glucagon pump. This arm is double blinded; neither the subject nor the study team know if it is insulin only or bihormonal.
115 mg/dl Set Point - Bihormonal
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 115 mg/dl using both an insulin and a glucagon pump.
100 mg/dl Set Point - Bihormonal
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 100 mg/dl using both an insulin and a glucagon pump.
110 mg/dl Set Point - Insulin Only
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 110 mg/dl and using only an insulin pump. This arm is double blinded; neither the subject nor the study team know if it is insulin only or bihormonal.
110 mg/dl Set Point - Bihormonal
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 110 mg/dl using both an insulin and a glucagon pump. This arm is double blinded; neither the subject nor the study team know if it is insulin only or bihormonal.
120 mg/dl Set Point - Insulin Only
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 120 mg/dl and using only an insulin pump.
Number of Subjects Discordant for Reaching a BG < 60 mg/dl for > 2 Consecutive Plasma Glucose Measurements During Inpatient Exercise Visit
0 Participants
0 Participants
3 Participants
1 Participants

SECONDARY outcome

Timeframe: Days 1 through 3

Average glucose according to continuous glucose monitor readings including the 1 day washout

Outcome measures

Outcome measures
Measure
Usual Care
n=20 Participants
Comparator week to all closed-loop control, utilizing usual diabetes care and the subject's own insulin pump.
145 mg/dl Set Point - Insulin Only
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 145 mg/dl and using only an insulin pump.
130 mg/dl Set Point - Insulin Only
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 130 mg/dl and using only an insulin pump. This arm is for subjects with type 1 diabetes only. This arm is double blinded; neither the subject nor the study team know if it is insulin only or bihormonal.
130 mg/dl Set Point - Bihormonal
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 130 mg/dl using both an insulin and a glucagon pump. This arm is double blinded; neither the subject nor the study team know if it is insulin only or bihormonal.
115 mg/dl Set Point - Bihormonal
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 115 mg/dl using both an insulin and a glucagon pump.
100 mg/dl Set Point - Bihormonal
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 100 mg/dl using both an insulin and a glucagon pump.
110 mg/dl Set Point - Insulin Only
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 110 mg/dl and using only an insulin pump. This arm is double blinded; neither the subject nor the study team know if it is insulin only or bihormonal.
110 mg/dl Set Point - Bihormonal
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 110 mg/dl using both an insulin and a glucagon pump. This arm is double blinded; neither the subject nor the study team know if it is insulin only or bihormonal.
120 mg/dl Set Point - Insulin Only
n=14 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 120 mg/dl and using only an insulin pump.
Mean CGM Glucose
160.8 mg/dl
Standard Deviation 30.4
177.8 mg/dl
Standard Deviation 19.5
164.1 mg/dl
Standard Deviation 15.3
159.4 mg/dl
Standard Deviation 13.4
151.6 mg/dl
Standard Deviation 16.1
140.7 mg/dl
Standard Deviation 15.4
158.8 mg/dl
Standard Deviation 18.4
155.0 mg/dl
Standard Deviation 19.4
160.2 mg/dl
Standard Deviation 18.9

SECONDARY outcome

Timeframe: Days 1-3

The average severity of nausea, as recorded on a 10 centimeter visual analog scale by the daily e-mail survey they receive. Scores were determined by how many centimeters the subject marked as a surrogate for their nausea level on a daily basis. Any value more than 0 cm indicates some degree of nausea with a score of 10 cm being the worst nausea imaginable. A score less than 3 cm indicates a mild level of nausea.

Outcome measures

Outcome measures
Measure
Usual Care
n=20 Participants
Comparator week to all closed-loop control, utilizing usual diabetes care and the subject's own insulin pump.
145 mg/dl Set Point - Insulin Only
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 145 mg/dl and using only an insulin pump.
130 mg/dl Set Point - Insulin Only
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 130 mg/dl and using only an insulin pump. This arm is for subjects with type 1 diabetes only. This arm is double blinded; neither the subject nor the study team know if it is insulin only or bihormonal.
130 mg/dl Set Point - Bihormonal
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 130 mg/dl using both an insulin and a glucagon pump. This arm is double blinded; neither the subject nor the study team know if it is insulin only or bihormonal.
115 mg/dl Set Point - Bihormonal
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 115 mg/dl using both an insulin and a glucagon pump.
100 mg/dl Set Point - Bihormonal
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 100 mg/dl using both an insulin and a glucagon pump.
110 mg/dl Set Point - Insulin Only
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 110 mg/dl and using only an insulin pump. This arm is double blinded; neither the subject nor the study team know if it is insulin only or bihormonal.
110 mg/dl Set Point - Bihormonal
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 110 mg/dl using both an insulin and a glucagon pump. This arm is double blinded; neither the subject nor the study team know if it is insulin only or bihormonal.
120 mg/dl Set Point - Insulin Only
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 120 mg/dl and using only an insulin pump.
Nausea Severity
0.3 scores on a scale 0-10 cm
Standard Deviation 0.5
0.5 scores on a scale 0-10 cm
Standard Deviation 1.6
0.4 scores on a scale 0-10 cm
Standard Deviation 0.6
0.4 scores on a scale 0-10 cm
Standard Deviation 0.6
0.8 scores on a scale 0-10 cm
Standard Deviation 1.6
1.2 scores on a scale 0-10 cm
Standard Deviation 1.8
0.64 scores on a scale 0-10 cm
Standard Deviation 1.25
0.80 scores on a scale 0-10 cm
Standard Deviation 1.38
0.26 scores on a scale 0-10 cm
Standard Deviation 0.26

SECONDARY outcome

Timeframe: Days 2-3

The fraction of time spent in each of these ranges according to continuous glucose monitor readings

Outcome measures

Outcome measures
Measure
Usual Care
n=20 Participants
Comparator week to all closed-loop control, utilizing usual diabetes care and the subject's own insulin pump.
145 mg/dl Set Point - Insulin Only
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 145 mg/dl and using only an insulin pump.
130 mg/dl Set Point - Insulin Only
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 130 mg/dl and using only an insulin pump. This arm is for subjects with type 1 diabetes only. This arm is double blinded; neither the subject nor the study team know if it is insulin only or bihormonal.
130 mg/dl Set Point - Bihormonal
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 130 mg/dl using both an insulin and a glucagon pump. This arm is double blinded; neither the subject nor the study team know if it is insulin only or bihormonal.
115 mg/dl Set Point - Bihormonal
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 115 mg/dl using both an insulin and a glucagon pump.
100 mg/dl Set Point - Bihormonal
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 100 mg/dl using both an insulin and a glucagon pump.
110 mg/dl Set Point - Insulin Only
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 110 mg/dl and using only an insulin pump. This arm is double blinded; neither the subject nor the study team know if it is insulin only or bihormonal.
110 mg/dl Set Point - Bihormonal
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 110 mg/dl using both an insulin and a glucagon pump. This arm is double blinded; neither the subject nor the study team know if it is insulin only or bihormonal.
120 mg/dl Set Point - Insulin Only
n=14 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 120 mg/dl and using only an insulin pump.
Percentage of Time Spent in: < 50 mg/dl, < 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, >250 mg/dl
< 50 mg/dl
0.3 percentage of time
Standard Deviation 0.6
0.5 percentage of time
Standard Deviation 1.0
0.2 percentage of time
Standard Deviation 0.8
0.1 percentage of time
Standard Deviation 0.3
0.2 percentage of time
Standard Deviation 0.4
0.2 percentage of time
Standard Deviation 0.7
0.5 percentage of time
Standard Deviation 0.7
0.1 percentage of time
Standard Deviation 0.2
0.3 percentage of time
Standard Deviation 0.6
Percentage of Time Spent in: < 50 mg/dl, < 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, >250 mg/dl
< 70 mg/dl
3.8 percentage of time
Standard Deviation 6.2
2.4 percentage of time
Standard Deviation 2.2
2.3 percentage of time
Standard Deviation 2.1
1.3 percentage of time
Standard Deviation 2.0
2.3 percentage of time
Standard Deviation 2.3
2.4 percentage of time
Standard Deviation 2.0
3.6 percentage of time
Standard Deviation 2.2
1.2 percentage of time
Standard Deviation 1.3
2.4 percentage of time
Standard Deviation 3.3
Percentage of Time Spent in: < 50 mg/dl, < 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, >250 mg/dl
70-120 mg/dl
29.3 percentage of time
Standard Deviation 18.8
16.9 percentage of time
Standard Deviation 5.8
21.3 percentage of time
Standard Deviation 7.4
25.9 percentage of time
Standard Deviation 10.2
34.3 percentage of time
Standard Deviation 8.2
45.3 percentage of time
Standard Deviation 11.1
30.2 percentage of time
Standard Deviation 7.1
37.5 percentage of time
Standard Deviation 10.5
25.1 percentage of time
Standard Deviation 7.6
Percentage of Time Spent in: < 50 mg/dl, < 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, >250 mg/dl
70-180 mg/dl
62.5 percentage of time
Standard Deviation 19.0
57.0 percentage of time
Standard Deviation 11.2
67.2 percentage of time
Standard Deviation 13.6
70.8 percentage of time
Standard Deviation 10.2
76.1 percentage of time
Standard Deviation 11.0
80.7 percentage of time
Standard Deviation 8.2
70.4 percentage of time
Standard Deviation 10.3
76.5 percentage of time
Standard Deviation 11.2
69.5 percentage of time
Standard Deviation 14.0
Percentage of Time Spent in: < 50 mg/dl, < 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, >250 mg/dl
> 180 mg/dl
33.7 percentage of time
Standard Deviation 21.5
40.6 percentage of time
Standard Deviation 40.9
30.5 percentage of time
Standard Deviation 13.1
27.9 percentage of time
Standard Deviation 9.8
21.6 percentage of time
Standard Deviation 10.3
16.9 percentage of time
Standard Deviation 8.5
26.0 percentage of time
Standard Deviation 10.2
22.2 percentage of time
Standard Deviation 11.6
28.2 percentage of time
Standard Deviation 12.9
Percentage of Time Spent in: < 50 mg/dl, < 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, >250 mg/dl
> 250 mg/dl
8.2 percentage of time
Standard Deviation 8.4
11.3 percentage of time
Standard Deviation 12.0
6.6 percentage of time
Standard Deviation 7.3
5.1 percentage of time
Standard Deviation 3.8
5.7 percentage of time
Standard Deviation 6.7
3.4 percentage of time
Standard Deviation 4.5
7.7 percentage of time
Standard Deviation 6.2
6.5 percentage of time
Standard Deviation 7.0
5.6 percentage of time
Standard Deviation 6.0

SECONDARY outcome

Timeframe: Days 2-3

The percentage of subjects who's mean CGM glucose level is \< 154 mg/dl, which is an estimated hemoglobin a1c \< 7%, which is the ADA goal for therapy

Outcome measures

Outcome measures
Measure
Usual Care
n=20 Participants
Comparator week to all closed-loop control, utilizing usual diabetes care and the subject's own insulin pump.
145 mg/dl Set Point - Insulin Only
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 145 mg/dl and using only an insulin pump.
130 mg/dl Set Point - Insulin Only
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 130 mg/dl and using only an insulin pump. This arm is for subjects with type 1 diabetes only. This arm is double blinded; neither the subject nor the study team know if it is insulin only or bihormonal.
130 mg/dl Set Point - Bihormonal
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 130 mg/dl using both an insulin and a glucagon pump. This arm is double blinded; neither the subject nor the study team know if it is insulin only or bihormonal.
115 mg/dl Set Point - Bihormonal
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 115 mg/dl using both an insulin and a glucagon pump.
100 mg/dl Set Point - Bihormonal
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 100 mg/dl using both an insulin and a glucagon pump.
110 mg/dl Set Point - Insulin Only
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 110 mg/dl and using only an insulin pump. This arm is double blinded; neither the subject nor the study team know if it is insulin only or bihormonal.
110 mg/dl Set Point - Bihormonal
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 110 mg/dl using both an insulin and a glucagon pump. This arm is double blinded; neither the subject nor the study team know if it is insulin only or bihormonal.
120 mg/dl Set Point - Insulin Only
n=14 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 120 mg/dl and using only an insulin pump.
Percentage of Subjects With Mean CGM < 154 mg/dl
7 Participants
1 Participants
7 Participants
6 Participants
16 Participants
18 Participants
12 Participants
13 Participants
5 Participants

SECONDARY outcome

Timeframe: Day 4 in-clinic Exercise Visit (110 and 130 mg/dl arms only)

Plasma glucose values to create this curve included time points from time 0 to a maximum of 360 minutes, with a sampling frequency of at least every 10 minutes throughout.

Outcome measures

Outcome measures
Measure
Usual Care
n=20 Participants
Comparator week to all closed-loop control, utilizing usual diabetes care and the subject's own insulin pump.
145 mg/dl Set Point - Insulin Only
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 145 mg/dl and using only an insulin pump.
130 mg/dl Set Point - Insulin Only
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 130 mg/dl and using only an insulin pump. This arm is for subjects with type 1 diabetes only. This arm is double blinded; neither the subject nor the study team know if it is insulin only or bihormonal.
130 mg/dl Set Point - Bihormonal
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 130 mg/dl using both an insulin and a glucagon pump. This arm is double blinded; neither the subject nor the study team know if it is insulin only or bihormonal.
115 mg/dl Set Point - Bihormonal
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 115 mg/dl using both an insulin and a glucagon pump.
100 mg/dl Set Point - Bihormonal
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 100 mg/dl using both an insulin and a glucagon pump.
110 mg/dl Set Point - Insulin Only
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 110 mg/dl and using only an insulin pump. This arm is double blinded; neither the subject nor the study team know if it is insulin only or bihormonal.
110 mg/dl Set Point - Bihormonal
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 110 mg/dl using both an insulin and a glucagon pump. This arm is double blinded; neither the subject nor the study team know if it is insulin only or bihormonal.
120 mg/dl Set Point - Insulin Only
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 120 mg/dl and using only an insulin pump.
Area Between the Glucose Curve and 60 mg/dl Calculated From BG Measurements
0 min*mg/dl
Standard Deviation 0
0 min*mg/dl
Standard Deviation 0
26.3 min*mg/dl
Standard Deviation 78.5
5.15 min*mg/dl
Standard Deviation 23.03

SECONDARY outcome

Timeframe: Day 4 in-clinic Exercise Visit (type 1 diabetes only, 110 and 130 mg/dl arms only)

CGM glucose values to create this curve included time points from time 0 to a maximum of 360 minutes, with a sampling frequency of every 5 minutes

Outcome measures

Outcome measures
Measure
Usual Care
n=20 Participants
Comparator week to all closed-loop control, utilizing usual diabetes care and the subject's own insulin pump.
145 mg/dl Set Point - Insulin Only
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 145 mg/dl and using only an insulin pump.
130 mg/dl Set Point - Insulin Only
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 130 mg/dl and using only an insulin pump. This arm is for subjects with type 1 diabetes only. This arm is double blinded; neither the subject nor the study team know if it is insulin only or bihormonal.
130 mg/dl Set Point - Bihormonal
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 130 mg/dl using both an insulin and a glucagon pump. This arm is double blinded; neither the subject nor the study team know if it is insulin only or bihormonal.
115 mg/dl Set Point - Bihormonal
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 115 mg/dl using both an insulin and a glucagon pump.
100 mg/dl Set Point - Bihormonal
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 100 mg/dl using both an insulin and a glucagon pump.
110 mg/dl Set Point - Insulin Only
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 110 mg/dl and using only an insulin pump. This arm is double blinded; neither the subject nor the study team know if it is insulin only or bihormonal.
110 mg/dl Set Point - Bihormonal
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 110 mg/dl using both an insulin and a glucagon pump. This arm is double blinded; neither the subject nor the study team know if it is insulin only or bihormonal.
120 mg/dl Set Point - Insulin Only
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 120 mg/dl and using only an insulin pump.
Area Between the Glucose Curve and 60 mg/dl Calculated From CGM Measurements
14.8 min*mg/dl
Standard Deviation 66.2
0 min*mg/dl
Standard Deviation 0
29.6 min*mg/dl
Standard Deviation 91.5
0 min*mg/dl
Standard Deviation 0

SECONDARY outcome

Timeframe: Day 4 in-clinic Exercise Visit (type 1 diabetes only, 110 and 130 mg/dl arms only)

During both 110 mg/dl and 130 mg/dl arms, subjects will report for a fasted in-clinic exercise visit, with plasma blood glucose measurements obtained at least every 10 minutes

Outcome measures

Outcome measures
Measure
Usual Care
n=20 Participants
Comparator week to all closed-loop control, utilizing usual diabetes care and the subject's own insulin pump.
145 mg/dl Set Point - Insulin Only
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 145 mg/dl and using only an insulin pump.
130 mg/dl Set Point - Insulin Only
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 130 mg/dl and using only an insulin pump. This arm is for subjects with type 1 diabetes only. This arm is double blinded; neither the subject nor the study team know if it is insulin only or bihormonal.
130 mg/dl Set Point - Bihormonal
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 130 mg/dl using both an insulin and a glucagon pump. This arm is double blinded; neither the subject nor the study team know if it is insulin only or bihormonal.
115 mg/dl Set Point - Bihormonal
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 115 mg/dl using both an insulin and a glucagon pump.
100 mg/dl Set Point - Bihormonal
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 100 mg/dl using both an insulin and a glucagon pump.
110 mg/dl Set Point - Insulin Only
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 110 mg/dl and using only an insulin pump. This arm is double blinded; neither the subject nor the study team know if it is insulin only or bihormonal.
110 mg/dl Set Point - Bihormonal
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 110 mg/dl using both an insulin and a glucagon pump. This arm is double blinded; neither the subject nor the study team know if it is insulin only or bihormonal.
120 mg/dl Set Point - Insulin Only
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 120 mg/dl and using only an insulin pump.
Time From Start of Exercise to First BG Measurement < 60 mg/dl
NA minutes
Standard Deviation NA
No BG values \< 60 were reached during 130 arms
NA minutes
Standard Deviation NA
No BG values \< 60 were reached during 130 arms
46.7 minutes
Standard Deviation 59.2
30 minutes
Standard Deviation 0

SECONDARY outcome

Timeframe: Day 4 in-clinic Exercise Visit (110 and 130 mg/dl arms only)

During both 110 mg/dl and 130 mg/dl arms, subjects will report for a fasted in-clinic exercise visit, with plasma blood glucose measurements obtained at least every 10 minutes

Outcome measures

Outcome measures
Measure
Usual Care
n=20 Participants
Comparator week to all closed-loop control, utilizing usual diabetes care and the subject's own insulin pump.
145 mg/dl Set Point - Insulin Only
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 145 mg/dl and using only an insulin pump.
130 mg/dl Set Point - Insulin Only
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 130 mg/dl and using only an insulin pump. This arm is for subjects with type 1 diabetes only. This arm is double blinded; neither the subject nor the study team know if it is insulin only or bihormonal.
130 mg/dl Set Point - Bihormonal
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 130 mg/dl using both an insulin and a glucagon pump. This arm is double blinded; neither the subject nor the study team know if it is insulin only or bihormonal.
115 mg/dl Set Point - Bihormonal
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 115 mg/dl using both an insulin and a glucagon pump.
100 mg/dl Set Point - Bihormonal
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 100 mg/dl using both an insulin and a glucagon pump.
110 mg/dl Set Point - Insulin Only
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 110 mg/dl and using only an insulin pump. This arm is double blinded; neither the subject nor the study team know if it is insulin only or bihormonal.
110 mg/dl Set Point - Bihormonal
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 110 mg/dl using both an insulin and a glucagon pump. This arm is double blinded; neither the subject nor the study team know if it is insulin only or bihormonal.
120 mg/dl Set Point - Insulin Only
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 120 mg/dl and using only an insulin pump.
Time From Start of Exercise to First CGM Measurement < 60 mg/dl
25.0 minutes
Standard Deviation 0
NA minutes
Standard Deviation NA
There were no CGM values \< 60 mg/dl in this arm
53.3 minutes
Standard Deviation 31.7
NA minutes
Standard Deviation NA
There were no CGM values \< 60 mg/dl in this arm

SECONDARY outcome

Timeframe: Day 4 in-clinic Exercise Visit (110 and 130 mg/dl arms only)

During both 110 mg/dl and 130 mg/dl arms, subjects will report for a fasted in-clinic exercise visit, with plasma blood glucose measurements obtained at least every 10 minutes

Outcome measures

Outcome measures
Measure
Usual Care
n=20 Participants
Comparator week to all closed-loop control, utilizing usual diabetes care and the subject's own insulin pump.
145 mg/dl Set Point - Insulin Only
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 145 mg/dl and using only an insulin pump.
130 mg/dl Set Point - Insulin Only
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 130 mg/dl and using only an insulin pump. This arm is for subjects with type 1 diabetes only. This arm is double blinded; neither the subject nor the study team know if it is insulin only or bihormonal.
130 mg/dl Set Point - Bihormonal
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 130 mg/dl using both an insulin and a glucagon pump. This arm is double blinded; neither the subject nor the study team know if it is insulin only or bihormonal.
115 mg/dl Set Point - Bihormonal
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 115 mg/dl using both an insulin and a glucagon pump.
100 mg/dl Set Point - Bihormonal
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 100 mg/dl using both an insulin and a glucagon pump.
110 mg/dl Set Point - Insulin Only
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 110 mg/dl and using only an insulin pump. This arm is double blinded; neither the subject nor the study team know if it is insulin only or bihormonal.
110 mg/dl Set Point - Bihormonal
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 110 mg/dl using both an insulin and a glucagon pump. This arm is double blinded; neither the subject nor the study team know if it is insulin only or bihormonal.
120 mg/dl Set Point - Insulin Only
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 120 mg/dl and using only an insulin pump.
Grams of Carbohydrates Given to the Subject to Treat Hypoglycemia
0 grams of carbohydrates
Standard Deviation 0
0 grams of carbohydrates
Standard Deviation 0
1.77 grams of carbohydrates
Standard Deviation 5.51
0.86 grams of carbohydrates
Standard Deviation 3.83

SECONDARY outcome

Timeframe: Day 4 in-clinic Exercise Visit (110 and 130 mg/dl arms only)

During both 110 mg/dl and 130 mg/dl arms, subjects will report for a fasted in-clinic exercise visit, with plasma blood glucose measurements obtained at least every 10 minutes

Outcome measures

Outcome measures
Measure
Usual Care
n=20 Participants
Comparator week to all closed-loop control, utilizing usual diabetes care and the subject's own insulin pump.
145 mg/dl Set Point - Insulin Only
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 145 mg/dl and using only an insulin pump.
130 mg/dl Set Point - Insulin Only
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 130 mg/dl and using only an insulin pump. This arm is for subjects with type 1 diabetes only. This arm is double blinded; neither the subject nor the study team know if it is insulin only or bihormonal.
130 mg/dl Set Point - Bihormonal
n=20 Participants
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 130 mg/dl using both an insulin and a glucagon pump. This arm is double blinded; neither the subject nor the study team know if it is insulin only or bihormonal.
115 mg/dl Set Point - Bihormonal
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 115 mg/dl using both an insulin and a glucagon pump.
100 mg/dl Set Point - Bihormonal
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 100 mg/dl using both an insulin and a glucagon pump.
110 mg/dl Set Point - Insulin Only
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 110 mg/dl and using only an insulin pump. This arm is double blinded; neither the subject nor the study team know if it is insulin only or bihormonal.
110 mg/dl Set Point - Bihormonal
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 110 mg/dl using both an insulin and a glucagon pump. This arm is double blinded; neither the subject nor the study team know if it is insulin only or bihormonal.
120 mg/dl Set Point - Insulin Only
Automated blood glucose control via a closed-loop bionic pancreas device targeting a blood glucose level of 120 mg/dl and using only an insulin pump.
Total Glucagon Dosing by the Bihormonal Bionic Pancreas From the Start of Exercise Until the End of the Visit
0.0 mcg/kg
Standard Deviation 0.0
1.36 mcg/kg
Standard Deviation 1.06
0.0 mcg/kg
Standard Deviation 0.0
0.70 mcg/kg
Standard Deviation 0.53

Adverse Events

All Enrolled Subjects

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Serious adverse events

Adverse event data not reported

Other adverse events

Adverse event data not reported

Additional Information

Courtney A Balliro

Massachusetts General Hospital

Phone: 617-726-1242

Results disclosure agreements

  • Principal investigator is a sponsor employee
  • Publication restrictions are in place