Trial Outcomes & Findings for TopQ Cut-off Score Validation Study for 3D OCT-1 Maestro (NCT NCT02447133)

NCT ID: NCT02447133

Last Updated: 2022-06-29

Results Overview

To validate the values of the predetermined TopQ score by showing the variability above and below the TopQ score of 25 for 12x9 Wide, 28 for 6x6 Macula, and 30 for 6x6 Disc scans.

Recruitment status

COMPLETED

Target enrollment

12 participants

Primary outcome timeframe

1 hour

Results posted on

2022-06-29

Participant Flow

Participant milestones

Participant milestones
Measure
Subjects Presenting With Normal Eyes
Subjects with no known ocular diseases will be scanned on the Maestro device 3D OCT-1 Maestro: OCT Machine used for diagnostic purposes
Overall Study
STARTED
12
Overall Study
COMPLETED
12
Overall Study
NOT COMPLETED
0

Reasons for withdrawal

Withdrawal data not reported

Baseline Characteristics

TopQ Cut-off Score Validation Study for 3D OCT-1 Maestro

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
Subjects Presenting With Normal Eyes
n=12 Participants
Subjects with no known ocular diseases will be scanned on the Maestro device 3D OCT-1 Maestro: OCT Machine used for diagnostic purposes
Age, Categorical
<=18 years
0 Participants
n=99 Participants
Age, Categorical
Between 18 and 65 years
11 Participants
n=99 Participants
Age, Categorical
>=65 years
1 Participants
n=99 Participants
Age, Continuous
43.17 years
STANDARD_DEVIATION 13.48 • n=99 Participants
Sex: Female, Male
Female
7 Participants
n=99 Participants
Sex: Female, Male
Male
5 Participants
n=99 Participants
Region of Enrollment
United States
12 participants
n=99 Participants

PRIMARY outcome

Timeframe: 1 hour

Population: use of 12 subjects for 3 different scan patterns

To validate the values of the predetermined TopQ score by showing the variability above and below the TopQ score of 25 for 12x9 Wide, 28 for 6x6 Macula, and 30 for 6x6 Disc scans.

Outcome measures

Outcome measures
Measure
12x9 Wide Scan
n=12 Participants
TopQ of 12x9 Wide Scan
6x6 Macula Scan
n=12 Participants
TopQ of 6x6 Macular Scan
6x6 Disc Scan
n=12 Participants
TopQ of 6x6 Disc Scan
TopQ Cut Off
Standard Deviation Below Threshold
17.54 microns
13.67 microns
13.67 microns
TopQ Cut Off
Standard Deviation Above Threshold
2.94 microns
1.5 microns
2.14 microns

Adverse Events

Subjects Presenting With Normal Eyes

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Serious adverse events

Adverse event data not reported

Other adverse events

Adverse event data not reported

Additional Information

Charles Reisman

Topcon

Phone: 201-599-5283

Results disclosure agreements

  • Principal investigator is a sponsor employee
  • Publication restrictions are in place