Trial Outcomes & Findings for Comparing Conventional Dental Implants and Trabecular Metal™ Dental Implants After Sinus Floor Elevation (NCT NCT02404649)

NCT ID: NCT02404649

Last Updated: 2020-07-22

Results Overview

Implant stability quotient (ISQ) is the value on a scale that indicates the level of stability and osseointegration in dental implants. The scale ranges from 1 to 100 and is measured by implant stability meters instruments using resonance frequency analysis (RFA) technique. The acceptable stability range lies between 55-85 ISQ. Lower initial stability will normally increase with time due to the lower mechanical stability being enforced by the bone remodeling process (osseointegration). The overall average ISQ value of all implants over time is approximately 70. A significant decrease in ISQ indicates a potential problem and should be considered an early warning. For each time period three measurements were taken from three different positions on the dental implant and the measurements were averaged for each time period.

Recruitment status

COMPLETED

Study phase

NA

Target enrollment

33 participants

Primary outcome timeframe

8 months

Results posted on

2020-07-22

Participant Flow

Recruitment of patients were fulfilled in a timely manner. Patients were were down from the existing patient pool at the Harvard School of Dental medicine.

Three participants were excluded from the trial due to early dental implant failures. These three events were caused by local infections and thus subsequent readings were not taken.

Participant milestones

Participant milestones
Measure
Bone Augmention
Placement of two dental implants in Sinus augmented with Puros Cortico-Cancellous Particulate Allograft (70% Cortico and 30% Cancelleous) (Zimmer Dental- Carlsbad, CA, USA)- one implant will be TMDI (Zimmer Dental- Carlsbad, CA, USA) and the second will be Tapered Screw-Vent TSV-MTX (Zimmer Dental- Carlsbad, CA,USA). Implant stability will be determined by the initial insertion torque and RFV values at time of implant placement and after 1 month of healing on a monthly basis up until 12 months post implant placement. Thickness of bone surrounding implant body positioned in the sinus will be measured. Comparing Conventional Dental Implants and Trabecular Metal™ Dental Implants (Zimmer) after Staged Sinus Bone Augmentation Procedures: Comparing conventional implant versus trabecular metal implant biological behavior in a sinus bone augmentation environment. sinus bone augmentation
Sinus Elevation Only
Placement of two dental implants as mentioned above in sinus augmented by sinus elevation procedure, blood clot and CopiOs Pericardium Membrane (Zimmer Dental- Carlsbad, CA, USA) only. One implant will be TMDI (Zimmer Dental- Carlsbad, CA, USA) and the second will be Tapered Screw-Vent TSV-MTX (Zimmer Dental- Carlsbad, CA,USA). Implant stability will be determined by the insertion torque and RFV values at time of implant placement and after 1 month of healing on a monthly basis up until 12 months post implant placement. Thickness of bone surrounding implant body positioned in the sinus will be measured. Comparing Conventional Dental Implants and Trabecular Metal™ Dental Implants (Zimmer) after Staged Sinus Floor Elevation Procedures: Comparing conventional implant versus trabecular metal implant biological behavior in a sinus elevation only environment. sinus elevation only
Overall Study
STARTED
17
16
Overall Study
COMPLETED
16
14
Overall Study
NOT COMPLETED
1
2

Reasons for withdrawal

Reasons for withdrawal
Measure
Bone Augmention
Placement of two dental implants in Sinus augmented with Puros Cortico-Cancellous Particulate Allograft (70% Cortico and 30% Cancelleous) (Zimmer Dental- Carlsbad, CA, USA)- one implant will be TMDI (Zimmer Dental- Carlsbad, CA, USA) and the second will be Tapered Screw-Vent TSV-MTX (Zimmer Dental- Carlsbad, CA,USA). Implant stability will be determined by the initial insertion torque and RFV values at time of implant placement and after 1 month of healing on a monthly basis up until 12 months post implant placement. Thickness of bone surrounding implant body positioned in the sinus will be measured. Comparing Conventional Dental Implants and Trabecular Metal™ Dental Implants (Zimmer) after Staged Sinus Bone Augmentation Procedures: Comparing conventional implant versus trabecular metal implant biological behavior in a sinus bone augmentation environment. sinus bone augmentation
Sinus Elevation Only
Placement of two dental implants as mentioned above in sinus augmented by sinus elevation procedure, blood clot and CopiOs Pericardium Membrane (Zimmer Dental- Carlsbad, CA, USA) only. One implant will be TMDI (Zimmer Dental- Carlsbad, CA, USA) and the second will be Tapered Screw-Vent TSV-MTX (Zimmer Dental- Carlsbad, CA,USA). Implant stability will be determined by the insertion torque and RFV values at time of implant placement and after 1 month of healing on a monthly basis up until 12 months post implant placement. Thickness of bone surrounding implant body positioned in the sinus will be measured. Comparing Conventional Dental Implants and Trabecular Metal™ Dental Implants (Zimmer) after Staged Sinus Floor Elevation Procedures: Comparing conventional implant versus trabecular metal implant biological behavior in a sinus elevation only environment. sinus elevation only
Overall Study
Adverse Event
1
2

Baseline Characteristics

Comparing Conventional Dental Implants and Trabecular Metal™ Dental Implants After Sinus Floor Elevation

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
Bone Augmention
n=16 Participants
Placement of two dental implants in Sinus augmented with Puros Cortico-Cancellous Particulate Allograft (70% Cortico and 30% Cancelleous) (Zimmer Dental- Carlsbad, CA, USA)- one implant will be TMDI (Zimmer Dental- Carlsbad, CA, USA) and the second will be Tapered Screw-Vent TSV-MTX (Zimmer Dental- Carlsbad, CA,USA). Implant stability will be determined by the initial insertion torque and RFV values at time of implant placement and after 1 month of healing on a monthly basis up until 12 months post implant placement. Thickness of bone surrounding implant body positioned in the sinus will be measured. Comparing Conventional Dental Implants and Trabecular Metal™ Dental Implants (Zimmer) after Staged Sinus Bone Augmentation Procedures: Comparing conventional implant versus trabecular metal implant biological behavior in a sinus bone augmentation environment. sinus bone augmentation
Sinus Elevation Only
n=14 Participants
Placement of two dental implants as mentioned above in sinus augmented by sinus elevation procedure, blood clot and CopiOs Pericardium Membrane (Zimmer Dental- Carlsbad, CA, USA) only. One implant will be TMDI (Zimmer Dental- Carlsbad, CA, USA) and the second will be Tapered Screw-Vent TSV-MTX (Zimmer Dental- Carlsbad, CA,USA). Implant stability will be determined by the insertion torque and RFV values at time of implant placement and after 1 month of healing on a monthly basis up until 12 months post implant placement. Thickness of bone surrounding implant body positioned in the sinus will be measured. Comparing Conventional Dental Implants and Trabecular Metal™ Dental Implants (Zimmer) after Staged Sinus Floor Elevation Procedures: Comparing conventional implant versus trabecular metal implant biological behavior in a sinus elevation only environment. sinus elevation only
Total
n=30 Participants
Total of all reporting groups
Age, Categorical
<=18 years
0 Participants
n=99 Participants
0 Participants
n=107 Participants
0 Participants
n=206 Participants
Age, Categorical
Between 18 and 65 years
16 Participants
n=99 Participants
14 Participants
n=107 Participants
30 Participants
n=206 Participants
Age, Categorical
>=65 years
0 Participants
n=99 Participants
0 Participants
n=107 Participants
0 Participants
n=206 Participants
Sex: Female, Male
Female
9 Participants
n=99 Participants
6 Participants
n=107 Participants
15 Participants
n=206 Participants
Sex: Female, Male
Male
7 Participants
n=99 Participants
8 Participants
n=107 Participants
15 Participants
n=206 Participants
Region of Enrollment
United States
16 participants
n=99 Participants
14 participants
n=107 Participants
30 participants
n=206 Participants

PRIMARY outcome

Timeframe: 8 months

Implant stability quotient (ISQ) is the value on a scale that indicates the level of stability and osseointegration in dental implants. The scale ranges from 1 to 100 and is measured by implant stability meters instruments using resonance frequency analysis (RFA) technique. The acceptable stability range lies between 55-85 ISQ. Lower initial stability will normally increase with time due to the lower mechanical stability being enforced by the bone remodeling process (osseointegration). The overall average ISQ value of all implants over time is approximately 70. A significant decrease in ISQ indicates a potential problem and should be considered an early warning. For each time period three measurements were taken from three different positions on the dental implant and the measurements were averaged for each time period.

Outcome measures

Outcome measures
Measure
Bone Augmention
n=16 Participants
Placement of two dental implants in Sinus augmented with Puros Cortico-Cancellous Particulate Allograft (70% Cortico and 30% Cancelleous) (Zimmer Dental- Carlsbad, CA, USA)- one implant will be TMDI (Zimmer Dental- Carlsbad, CA, USA) and the second will be Tapered Screw-Vent TSV-MTX (Zimmer Dental- Carlsbad, CA,USA). Implant stability will be determined by the initial insertion torque and RFV values at time of implant placement and after 1 month of healing on a monthly basis up until 12 months post implant placement. Thickness of bone surrounding implant body positioned in the sinus will be measured. Comparing Conventional Dental Implants and Trabecular Metal™ Dental Implants (Zimmer) after Staged Sinus Bone Augmentation Procedures: Comparing conventional implant versus trabecular metal implant biological behavior in a sinus bone augmentation environment. sinus bone augmentation
Sinus Elevation Only
n=14 Participants
Placement of two dental implants as mentioned above in sinus augmented by sinus elevation procedure, blood clot and CopiOs Pericardium Membrane (Zimmer Dental- Carlsbad, CA, USA) only. One implant will be TMDI (Zimmer Dental- Carlsbad, CA, USA) and the second will be Tapered Screw-Vent TSV-MTX (Zimmer Dental- Carlsbad, CA,USA). Implant stability will be determined by the insertion torque and RFV values at time of implant placement and after 1 month of healing on a monthly basis up until 12 months post implant placement. Thickness of bone surrounding implant body positioned in the sinus will be measured. Comparing Conventional Dental Implants and Trabecular Metal™ Dental Implants (Zimmer) after Staged Sinus Floor Elevation Procedures: Comparing conventional implant versus trabecular metal implant biological behavior in a sinus elevation only environment. sinus elevation only
Resonance Frequency Values of Dental Implants
83.04 units on a scale
Standard Deviation 3.94
80.59 units on a scale
Standard Deviation 7.69

SECONDARY outcome

Timeframe: 2 years

Evaluating crestal bone levels were performed by measuring (millimeters) vertical bone height gain in the maxillary 2 years postoperatively from CBCT images between control and test group.

Outcome measures

Outcome measures
Measure
Bone Augmention
n=13 Participants
Placement of two dental implants in Sinus augmented with Puros Cortico-Cancellous Particulate Allograft (70% Cortico and 30% Cancelleous) (Zimmer Dental- Carlsbad, CA, USA)- one implant will be TMDI (Zimmer Dental- Carlsbad, CA, USA) and the second will be Tapered Screw-Vent TSV-MTX (Zimmer Dental- Carlsbad, CA,USA). Implant stability will be determined by the initial insertion torque and RFV values at time of implant placement and after 1 month of healing on a monthly basis up until 12 months post implant placement. Thickness of bone surrounding implant body positioned in the sinus will be measured. Comparing Conventional Dental Implants and Trabecular Metal™ Dental Implants (Zimmer) after Staged Sinus Bone Augmentation Procedures: Comparing conventional implant versus trabecular metal implant biological behavior in a sinus bone augmentation environment. sinus bone augmentation
Sinus Elevation Only
n=11 Participants
Placement of two dental implants as mentioned above in sinus augmented by sinus elevation procedure, blood clot and CopiOs Pericardium Membrane (Zimmer Dental- Carlsbad, CA, USA) only. One implant will be TMDI (Zimmer Dental- Carlsbad, CA, USA) and the second will be Tapered Screw-Vent TSV-MTX (Zimmer Dental- Carlsbad, CA,USA). Implant stability will be determined by the insertion torque and RFV values at time of implant placement and after 1 month of healing on a monthly basis up until 12 months post implant placement. Thickness of bone surrounding implant body positioned in the sinus will be measured. Comparing Conventional Dental Implants and Trabecular Metal™ Dental Implants (Zimmer) after Staged Sinus Floor Elevation Procedures: Comparing conventional implant versus trabecular metal implant biological behavior in a sinus elevation only environment. sinus elevation only
Crestal Bone Levels
13.05 millimeters
Standard Deviation 1.73
11.68 millimeters
Standard Deviation 1.4

Adverse Events

Bone Augmention

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Sinus Elevation Only

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Serious adverse events

Adverse event data not reported

Other adverse events

Adverse event data not reported

Additional Information