Trial Outcomes & Findings for Evaluation of an Ascensia Diabetes Care ONYX PLUS Investigational Blood Glucose Monitoring System (NCT NCT02371876)
NCT ID: NCT02371876
Last Updated: 2019-10-30
Results Overview
Untrained subjects with diabetes self-tested fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 15mg/dL (\<100mg/dL YSI capillary plasma) and +/- 15% (\>= 100mg/dL YSI capillary plasma).
COMPLETED
NA
134 participants
1 hour
2019-10-30
Participant Flow
Participant milestones
| Measure |
Users of the Monitoring System
Untrained subjects with diabetes used the ONYX PLUS Investigational Blood Glucose Monitoring System.
ONYX PLUS Investigational Blood Glucose Monitoring System: Untrained subjects with diabetes performed self Blood Glucose (BG) tests with capillary fingerstick and palm blood (and study staff tested subject fingerstick blood) using the ONYX PLUS Investigational Blood Glucose Monitoring System. All BG results were compared to reference method results obtained from subject capillary blood. Study staff tested subject venous blood and BG results were compared to reference method results obtained from subject venous plasma.
|
|---|---|
|
Overall Study
STARTED
|
134
|
|
Overall Study
COMPLETED
|
134
|
|
Overall Study
NOT COMPLETED
|
0
|
Reasons for withdrawal
Withdrawal data not reported
Baseline Characteristics
Evaluation of an Ascensia Diabetes Care ONYX PLUS Investigational Blood Glucose Monitoring System
Baseline characteristics by cohort
| Measure |
Users of the Monitoring System
n=134 Participants
Untrained subjects with diabetes used the ONYX PLUS Investigational Blood Glucose Monitoring System.
ONYX PLUS Investigational Blood Glucose Monitoring System: Untrained subjects with diabetes performed self Blood Glucose (BG) tests with capillary fingerstick and palm blood (and study staff tested subject fingerstick blood) using the ONYX PLUS Investigational Blood Glucose Monitoring System. All BG results were compared to reference method results obtained from subject capillary blood. Study staff tested subject venous blood and BG results were compared to reference method results obtained from subject venous plasma.
|
|---|---|
|
Age, Continuous
|
54.0 years
n=99 Participants
|
|
Sex: Female, Male
Female
|
70 Participants
n=99 Participants
|
|
Sex: Female, Male
Male
|
64 Participants
n=99 Participants
|
|
Ethnicity (NIH/OMB)
Hispanic or Latino
|
14 Participants
n=99 Participants
|
|
Ethnicity (NIH/OMB)
Not Hispanic or Latino
|
119 Participants
n=99 Participants
|
|
Ethnicity (NIH/OMB)
Unknown or Not Reported
|
1 Participants
n=99 Participants
|
|
Race/Ethnicity, Customized
American Indian or Alaska Native
|
3 participants
n=99 Participants
|
|
Race/Ethnicity, Customized
Asian
|
4 participants
n=99 Participants
|
|
Race/Ethnicity, Customized
Native Hawaiian or Other Pacific Islander
|
2 participants
n=99 Participants
|
|
Race/Ethnicity, Customized
Black or African American
|
14 participants
n=99 Participants
|
|
Race/Ethnicity, Customized
White
|
112 participants
n=99 Participants
|
|
Race/Ethnicity, Customized
More than one race
|
3 participants
n=99 Participants
|
|
Race/Ethnicity, Customized
Unknown or Not Reported
|
2 participants
n=99 Participants
|
|
Region of Enrollment
United States
|
134 participants
n=99 Participants
|
PRIMARY outcome
Timeframe: 1 hourUntrained subjects with diabetes self-tested fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 15mg/dL (\<100mg/dL YSI capillary plasma) and +/- 15% (\>= 100mg/dL YSI capillary plasma).
Outcome measures
| Measure |
Users of the Monitoring System
n=134 Participants
Untrained subjects with diabetes used the ONYX PLUS Investigational Blood Glucose Monitoring System.
ONYX PLUS Investigational Blood Glucose Monitoring System: Untrained subjects with diabetes performed self Blood Glucose (BG) tests with capillary fingerstick blood using the ONYX PLUS Investigational Blood Glucose Monitoring System. All BG results were compared to reference method results obtained from subject capillary blood.
|
|---|---|
|
Percent of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 15mg/dL (<100mg/dL) and Within +/- 15% (>=100mg/dL) of Laboratory Glucose Method
|
99.2 Percentage of BG Results
|
SECONDARY outcome
Timeframe: 1 hourPopulation: 126 (134-8) Blood glucose results were analyzed. Four subjects with low blood sugar did not attempt palm testing per protocol. Four subjects had low blood sugar; their palm results were not evaluable per protocol.
Untrained subjects with diabetes self-tested Alternate Site (AST) palm blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 15mg/dL (\<100mg/dL YSI capillary plasma) and +/- 15% (\>= 100mg/dL YSI capillary plasma).
Outcome measures
| Measure |
Users of the Monitoring System
n=126 Participants
Untrained subjects with diabetes used the ONYX PLUS Investigational Blood Glucose Monitoring System.
ONYX PLUS Investigational Blood Glucose Monitoring System: Untrained subjects with diabetes performed self Blood Glucose (BG) tests with capillary fingerstick blood using the ONYX PLUS Investigational Blood Glucose Monitoring System. All BG results were compared to reference method results obtained from subject capillary blood.
|
|---|---|
|
Percent of Alternate Site Palm Blood Glucose (BG) Results Within +/- 15mg/dL (<100mg/dL) and Within +/- 15% (>=100mg/dL) of Laboratory Glucose Method
|
99.2 Percentage of BG Results
|
SECONDARY outcome
Timeframe: 1 hourStudy staff tested subject fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 15mg/dL (\<100mg/dL YSI capillary plasma) and +/- 15% (\>= 100mg/dL YSI capillary plasma).
Outcome measures
| Measure |
Users of the Monitoring System
n=134 Participants
Untrained subjects with diabetes used the ONYX PLUS Investigational Blood Glucose Monitoring System.
ONYX PLUS Investigational Blood Glucose Monitoring System: Untrained subjects with diabetes performed self Blood Glucose (BG) tests with capillary fingerstick blood using the ONYX PLUS Investigational Blood Glucose Monitoring System. All BG results were compared to reference method results obtained from subject capillary blood.
|
|---|---|
|
Percent of Subject Fingerstick Blood Glucose (BG) Results Within +/- 15mg/dL (<100mg/dL) and Within +/- 15% (>=100mg/dL) of Laboratory Glucose Method When Tested by Study Staff
|
99.2 Percentage of BG Results
|
SECONDARY outcome
Timeframe: 1 hourPopulation: 132 (134-2) Blood glucose results were analyzed. Venipunctures were unsuccessful for two subjects.
Study staff tested subject venous blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject venous plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 15mg/dL (\<100mg/dL YSI venous plasma) and +/- 15% (\>= 100mg/dL YSI venous plasma).
Outcome measures
| Measure |
Users of the Monitoring System
n=132 Participants
Untrained subjects with diabetes used the ONYX PLUS Investigational Blood Glucose Monitoring System.
ONYX PLUS Investigational Blood Glucose Monitoring System: Untrained subjects with diabetes performed self Blood Glucose (BG) tests with capillary fingerstick blood using the ONYX PLUS Investigational Blood Glucose Monitoring System. All BG results were compared to reference method results obtained from subject capillary blood.
|
|---|---|
|
Percent of Venous Blood Glucose (BG) Results Within +/- 15mg/dL (<100mg/dL) and Within +/- 15% (>=100mg/dL) of Laboratory Glucose Method
|
100 Percentage of BG Results
|
SECONDARY outcome
Timeframe: 1 hourUntrained subjects with diabetes self-tested fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 12.5mg/dL (\<100mg/dL YSI capillary plasma) and +/- 12.5% (\>= 100mg/dL YSI capillary plasma).
Outcome measures
| Measure |
Users of the Monitoring System
n=134 Participants
Untrained subjects with diabetes used the ONYX PLUS Investigational Blood Glucose Monitoring System.
ONYX PLUS Investigational Blood Glucose Monitoring System: Untrained subjects with diabetes performed self Blood Glucose (BG) tests with capillary fingerstick blood using the ONYX PLUS Investigational Blood Glucose Monitoring System. All BG results were compared to reference method results obtained from subject capillary blood.
|
|---|---|
|
Percent of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 12.5mg/dL (<100mg/dL) and Within +/- 12.5% (>=100mg/dL) of Laboratory Glucose Method
|
98.5 Percentage of BG Results
|
SECONDARY outcome
Timeframe: 1 hourStaff obtained responses from persons with diabetes using short questionnaires to provide feedback on instructions for use and the basic operation of the BGMS. Subjects could respond 'Strongly Agree' or 'Agree' or 'Neutral' or 'Disagree' or 'Strongly Disagree'. The percent of subjects who provided responses that were 'Strongly Agree','Agree','Neutral' about each statement was calculated.
Outcome measures
| Measure |
Users of the Monitoring System
n=134 Participants
Untrained subjects with diabetes used the ONYX PLUS Investigational Blood Glucose Monitoring System.
ONYX PLUS Investigational Blood Glucose Monitoring System: Untrained subjects with diabetes performed self Blood Glucose (BG) tests with capillary fingerstick blood using the ONYX PLUS Investigational Blood Glucose Monitoring System. All BG results were compared to reference method results obtained from subject capillary blood.
|
|---|---|
|
Percent of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMS
The meter display was easy to see and read.
|
98.5 Percentage of Participants Who Responded
|
|
Percent of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMS
It was easy to understand my test results.
|
100 Percentage of Participants Who Responded
|
|
Percent of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMS
I like the overall meter design.
|
97.8 Percentage of Participants Who Responded
|
|
Percent of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMS
I find the meter easy to use.
|
100 Percentage of Participants Who Responded
|
|
Percent of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMS
The instructions were easy to understand.
|
96.3 Percentage of Participants Who Responded
|
|
Percent of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMS
Instructions clearly explain how to run a test.
|
97.8 Percentage of Participants Who Responded
|
|
Percent of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMS
Instructions clearly explain meter error messages.
|
99.2 Percentage of Participants Who Responded
|
|
Percent of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMS
I find it easy to do a blood test with this meter.
|
99.2 Percentage of Participants Who Responded
|
SECONDARY outcome
Timeframe: 1 hourPopulation: Percent (of those who responded) who 'Strongly Agree', 'Agree', 'Neutral' was calculated for each statement. Abbreviations used: 'HCP' is HealthCare Professional and 'Acc' is Accuracy
Staff obtained responses from persons with diabetes using short questionnaires to provide feedback on views and behaviors related to managing their diabetes. Subjects could respond 'Strongly Agree' or 'Agree' or 'Neutral' or 'Disagree' or 'Strongly Disagree' or 'No Answer or Not Applicable'. The percents (of subjects who responded) that were 'Strongly Agree','Agree','Neutral' about views/behaviors related to self-monitoring blood glucose were calculated.
Outcome measures
| Measure |
Users of the Monitoring System
n=134 Participants
Untrained subjects with diabetes used the ONYX PLUS Investigational Blood Glucose Monitoring System.
ONYX PLUS Investigational Blood Glucose Monitoring System: Untrained subjects with diabetes performed self Blood Glucose (BG) tests with capillary fingerstick blood using the ONYX PLUS Investigational Blood Glucose Monitoring System. All BG results were compared to reference method results obtained from subject capillary blood.
|
|---|---|
|
Percent of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding Views/Behaviors Related to Self-Monitoring Blood Glucose
Acc:important to help prevent low blood sugar
|
100 Percentage of Participants Who Responded
|
|
Percent of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding Views/Behaviors Related to Self-Monitoring Blood Glucose
Acc:important to understand food/exercise effects.
|
100 Percentage of Participants Who Responded
|
|
Percent of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding Views/Behaviors Related to Self-Monitoring Blood Glucose
Acc:help w using results to better control diabete
|
100 Percentage of Participants Who Responded
|
|
Percent of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding Views/Behaviors Related to Self-Monitoring Blood Glucose
Acc:important to achieve greater peace of mind.
|
100 Percentage of Participants Who Responded
|
|
Percent of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding Views/Behaviors Related to Self-Monitoring Blood Glucose
I prefer the meter I just used to my regular meter
|
86.5 Percentage of Participants Who Responded
|
|
Percent of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding Views/Behaviors Related to Self-Monitoring Blood Glucose
I use my meter - my care provider gave it to me.
|
84.1 Percentage of Participants Who Responded
|
|
Percent of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding Views/Behaviors Related to Self-Monitoring Blood Glucose
I use my current meter -my insurance covers strips
|
80 Percentage of Participants Who Responded
|
|
Percent of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding Views/Behaviors Related to Self-Monitoring Blood Glucose
I use my current meter-I think it is most accurate
|
86.9 Percentage of Participants Who Responded
|
|
Percent of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding Views/Behaviors Related to Self-Monitoring Blood Glucose
Acc:important to help me talk w my HCP.
|
99.25 Percentage of Participants Who Responded
|
|
Percent of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding Views/Behaviors Related to Self-Monitoring Blood Glucose
Acc:important to help w self-monitor satisfaction.
|
99.25 Percentage of Participants Who Responded
|
|
Percent of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding Views/Behaviors Related to Self-Monitoring Blood Glucose
Acc:important to help w managing my diabetes.
|
100 Percentage of Participants Who Responded
|
Adverse Events
Users of the Monitoring System
Serious adverse events
Adverse event data not reported
Other adverse events
| Measure |
Users of the Monitoring System
n=134 participants at risk
Untrained subjects with diabetes used the ONYX PLUS Investigational Blood Glucose Monitoring System.
ONYX PLUS Investigational Blood Glucose Monitoring System: Untrained subjects with diabetes performed self Blood Glucose (BG) tests with capillary fingerstick and palm blood (and study staff tested subject fingerstick blood) using the ONYX PLUS Investigational Blood Glucose Monitoring System. All BG results were compared to reference method results obtained from subject capillary blood. Study staff tested subject venous blood and BG results were compared to reference method results obtained from subject venous plasma.
|
|---|---|
|
Endocrine disorders
Hypoglycemia
|
3.0%
4/134 • Number of events 4 • Subject visits were approximately 1 hour.
|
Additional Information
Jane Wallace, Deputy Director for Global Clinical Affairs
Ascensia Diabetes Care
Results disclosure agreements
- Principal investigator is a sponsor employee
- Publication restrictions are in place
Restriction type: GT60