Trial Outcomes & Findings for A Prospective Multi-center Study on Total Hip Arthroplasty With E1 (NCT NCT01883492)

NCT ID: NCT01883492

Last Updated: 2020-02-17

Results Overview

Vector wear, which is defined as movement (difference of femoral prosthetic head positions) between immediate postoperative and 2 years postoperative periods. There is no lower and upper limit, actual measured results. The outcome can be negative value.

Recruitment status

TERMINATED

Study phase

NA

Target enrollment

153 participants

Primary outcome timeframe

Immediate postoperative and 2 year postoperatively

Results posted on

2020-02-17

Participant Flow

Participant milestones

Participant milestones
Measure
BIOLOX Delta Head
Uncemented Femoral Stem is inserted into femoral medullary canal. Uncemented Acetabular Cup is inserted into acetabular cavity with or without fixation screw(s). Acetabular Liner, which is made of E1 highly crosslinked polyethylene, is fitted into Acetabular cup. Biolox delta head is inserted onto the taper of the Femoral stem, and is articulating against Acetabular Liner made of E1. Femoral Stem: JMDN classification/Class III device Acetabular Cup: JMDN classification: Class III device Acetabular Liner: JMDN classification: Class III device
CoCr Head
Uncemented Femoral Stem is inserted into femoral medullary canal. Uncemented Acetabular Cup is inserted into acetabular cavity with or without fixation screw(s). Acetabular Liner, which is made of E1 highly crosslinked polyethylene, is fitted into Acetabular cup. CoCr head is inserted onto the taper of the Femoral stem, and is articulating against Acetabular Liner made of E1. Femoral Stem: JMDN classification/Class III device Acetabular Cup: JMDN classification: Class III device Acetabular Liner: JMDN classification: Class III device
Overall Study
STARTED
75
78
Overall Study
COMPLETED
62
67
Overall Study
NOT COMPLETED
13
11

Reasons for withdrawal

Reasons for withdrawal
Measure
BIOLOX Delta Head
Uncemented Femoral Stem is inserted into femoral medullary canal. Uncemented Acetabular Cup is inserted into acetabular cavity with or without fixation screw(s). Acetabular Liner, which is made of E1 highly crosslinked polyethylene, is fitted into Acetabular cup. Biolox delta head is inserted onto the taper of the Femoral stem, and is articulating against Acetabular Liner made of E1. Femoral Stem: JMDN classification/Class III device Acetabular Cup: JMDN classification: Class III device Acetabular Liner: JMDN classification: Class III device
CoCr Head
Uncemented Femoral Stem is inserted into femoral medullary canal. Uncemented Acetabular Cup is inserted into acetabular cavity with or without fixation screw(s). Acetabular Liner, which is made of E1 highly crosslinked polyethylene, is fitted into Acetabular cup. CoCr head is inserted onto the taper of the Femoral stem, and is articulating against Acetabular Liner made of E1. Femoral Stem: JMDN classification/Class III device Acetabular Cup: JMDN classification: Class III device Acetabular Liner: JMDN classification: Class III device
Overall Study
Lost to Follow-up
3
4
Overall Study
Surgery on contra-lateral hip
0
1
Overall Study
Poor X-ray quality
7
4
Overall Study
Withdrawal by Subject
1
0
Overall Study
inappropriate implant used
1
1
Overall Study
bone anatomy did not meet study criteria
1
1

Baseline Characteristics

A Prospective Multi-center Study on Total Hip Arthroplasty With E1

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
BIOLOX Delta Head
n=62 Participants
Uncemented Femoral Stem is inserted into femoral medullary canal. Uncemented Acetabular Cup is inserted into acetabular cavity with or without fixation screw(s). Acetabular Liner, which is made of E1 highly crosslinked polyethylene, is fitted into Acetabular cup. Biolox delta head is inserted onto the taper of the Femoral stem, and is articulating against Acetabular Liner made of E1. Femoral Stem: JMDN classification/Class III device Acetabular Cup: JMDN classification: Class III device Acetabular Liner: JMDN classification: Class III device
CoCr Head
n=67 Participants
Uncemented Femoral Stem is inserted into femoral medullary canal. Uncemented Acetabular Cup is inserted into acetabular cavity with or without fixation screw(s). Acetabular Liner, which is made of E1 highly crosslinked polyethylene, is fitted into Acetabular cup. CoCr head is inserted onto the taper of the Femoral stem, and is articulating against Acetabular Liner made of E1. Femoral Stem: JMDN classification/Class III device Acetabular Cup: JMDN classification: Class III device Acetabular Liner: JMDN classification: Class III device
Total
n=129 Participants
Total of all reporting groups
Age, Continuous
61.90 years
STANDARD_DEVIATION 7.16 • n=99 Participants
61.94 years
STANDARD_DEVIATION 7.26 • n=107 Participants
61.92 years
STANDARD_DEVIATION 7.18 • n=206 Participants
Sex: Female, Male
Female
62 Participants
n=99 Participants
67 Participants
n=107 Participants
129 Participants
n=206 Participants
Sex: Female, Male
Male
0 Participants
n=99 Participants
0 Participants
n=107 Participants
0 Participants
n=206 Participants
Race (NIH/OMB)
American Indian or Alaska Native
0 Participants
n=99 Participants
0 Participants
n=107 Participants
0 Participants
n=206 Participants
Race (NIH/OMB)
Asian
62 Participants
n=99 Participants
67 Participants
n=107 Participants
129 Participants
n=206 Participants
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
0 Participants
n=99 Participants
0 Participants
n=107 Participants
0 Participants
n=206 Participants
Race (NIH/OMB)
Black or African American
0 Participants
n=99 Participants
0 Participants
n=107 Participants
0 Participants
n=206 Participants
Race (NIH/OMB)
White
0 Participants
n=99 Participants
0 Participants
n=107 Participants
0 Participants
n=206 Participants
Race (NIH/OMB)
More than one race
0 Participants
n=99 Participants
0 Participants
n=107 Participants
0 Participants
n=206 Participants
Race (NIH/OMB)
Unknown or Not Reported
0 Participants
n=99 Participants
0 Participants
n=107 Participants
0 Participants
n=206 Participants
Region of Enrollment
Japan
62 participants
n=99 Participants
67 participants
n=107 Participants
129 participants
n=206 Participants
Harris Hip Score
55.11 score on scale
STANDARD_DEVIATION 11.65 • n=99 Participants
52.63 score on scale
STANDARD_DEVIATION 11.67 • n=107 Participants
53.83 score on scale
STANDARD_DEVIATION 11.68 • n=206 Participants
WOMAC osteoarthritis index
40.75 score on scale
STANDARD_DEVIATION 17.09 • n=99 Participants
41.24 score on scale
STANDARD_DEVIATION 15.29 • n=107 Participants
41.00 score on scale
STANDARD_DEVIATION 16.13 • n=206 Participants
UCLA Activity Score
2.95 score on scale
STANDARD_DEVIATION 0.95 • n=99 Participants
3.28 score on scale
STANDARD_DEVIATION 1.10 • n=107 Participants
3.12 score on scale
STANDARD_DEVIATION 1.04 • n=206 Participants

PRIMARY outcome

Timeframe: Immediate postoperative and 2 year postoperatively

Population: Participants, whose x-ray taken at 2 year follow-up visit being available.

Vector wear, which is defined as movement (difference of femoral prosthetic head positions) between immediate postoperative and 2 years postoperative periods. There is no lower and upper limit, actual measured results. The outcome can be negative value.

Outcome measures

Outcome measures
Measure
BIOLOX Delta Head
n=57 Participants
Uncemented Femoral Stem is inserted into femoral medullary canal. Uncemented Acetabular Cup is inserted into acetabular cavity with or without fixation screw(s). Acetabular Liner, which is made of E1 highly crosslinked polyethylene, is fitted into Acetabular cup. Biolox delta head is inserted onto the taper of the Femoral stem, and is articulating against Acetabular Liner made of E1. Femoral Stem: JMDN classification/Class III device Acetabular Cup: JMDN classification: Class III device Acetabular Liner: JMDN classification: Class III device
CoCr Head
n=63 Participants
Uncemented Femoral Stem is inserted into femoral medullary canal. Uncemented Acetabular Cup is inserted into acetabular cavity with or without fixation screw(s). Acetabular Liner, which is made of E1 highly crosslinked polyethylene, is fitted into Acetabular cup. CoCr head is inserted onto the taper of the Femoral stem, and is articulating against Acetabular Liner made of E1. Femoral Stem: JMDN classification/Class III device Acetabular Cup: JMDN classification: Class III device Acetabular Liner: JMDN classification: Class III device
Polyethylene Wear Between Immediate Postoperative and 2 Year Postoperative Period
0.42 millimeter
Standard Deviation 1.08
0.15 millimeter
Standard Deviation 0.70

SECONDARY outcome

Timeframe: 6 month postoperative

Population: 1 patient in BIOLOX delta head group and 13 patients in CoCr head group missed collecting Harris Hip Score at 6 month visit.

Harris Hip Score is Physician rating score to evaluate hip disabilities. Score domains are "pain", "function", "absence of deformity" and "range of hip motion". Minimum possible score is 0 and maximum possible score is 100. Higher score means better outcome.

Outcome measures

Outcome measures
Measure
BIOLOX Delta Head
n=61 Participants
Uncemented Femoral Stem is inserted into femoral medullary canal. Uncemented Acetabular Cup is inserted into acetabular cavity with or without fixation screw(s). Acetabular Liner, which is made of E1 highly crosslinked polyethylene, is fitted into Acetabular cup. Biolox delta head is inserted onto the taper of the Femoral stem, and is articulating against Acetabular Liner made of E1. Femoral Stem: JMDN classification/Class III device Acetabular Cup: JMDN classification: Class III device Acetabular Liner: JMDN classification: Class III device
CoCr Head
n=54 Participants
Uncemented Femoral Stem is inserted into femoral medullary canal. Uncemented Acetabular Cup is inserted into acetabular cavity with or without fixation screw(s). Acetabular Liner, which is made of E1 highly crosslinked polyethylene, is fitted into Acetabular cup. CoCr head is inserted onto the taper of the Femoral stem, and is articulating against Acetabular Liner made of E1. Femoral Stem: JMDN classification/Class III device Acetabular Cup: JMDN classification: Class III device Acetabular Liner: JMDN classification: Class III device
Harris Hip Score at 6 Month Follow-up Visit
92.67 score on a scale
Standard Deviation 8.05
92.61 score on a scale
Standard Deviation 7.92

SECONDARY outcome

Timeframe: 1 year postoperative

Population: 1 patient in BIOLOX delta head group and 6 patients in CoCr head group missed collecting Harris Hip Score at 1 year visit.

Harris Hip Score is Physician rating score to evaluate hip disabilities. Score domains are "pain", "function", "absence of deformity" and "range of hip motion". Minimum possible score is 0 and maximum possible score is 100. Higher score means better outcome.

Outcome measures

Outcome measures
Measure
BIOLOX Delta Head
n=61 Participants
Uncemented Femoral Stem is inserted into femoral medullary canal. Uncemented Acetabular Cup is inserted into acetabular cavity with or without fixation screw(s). Acetabular Liner, which is made of E1 highly crosslinked polyethylene, is fitted into Acetabular cup. Biolox delta head is inserted onto the taper of the Femoral stem, and is articulating against Acetabular Liner made of E1. Femoral Stem: JMDN classification/Class III device Acetabular Cup: JMDN classification: Class III device Acetabular Liner: JMDN classification: Class III device
CoCr Head
n=61 Participants
Uncemented Femoral Stem is inserted into femoral medullary canal. Uncemented Acetabular Cup is inserted into acetabular cavity with or without fixation screw(s). Acetabular Liner, which is made of E1 highly crosslinked polyethylene, is fitted into Acetabular cup. CoCr head is inserted onto the taper of the Femoral stem, and is articulating against Acetabular Liner made of E1. Femoral Stem: JMDN classification/Class III device Acetabular Cup: JMDN classification: Class III device Acetabular Liner: JMDN classification: Class III device
Harris Hip Score at 1 Year Follow-up Visit
95.23 score on a scale
Standard Deviation 6.10
94.52 score on a scale
Standard Deviation 6.76

SECONDARY outcome

Timeframe: 2 year postoperative

Population: 3 patients in BIOLOX delta head group and 9 patients in CoCr head group missed collecting Harris Hip Score at 2 year visit.

Harris Hip Score is Physician rating score to evaluate hip disabilities. Score domains are "pain", "function", "absence of deformity" and "range of hip motion". Minimum possible score is 0 and maximum possible score is 100. Higher score means better outcome.

Outcome measures

Outcome measures
Measure
BIOLOX Delta Head
n=59 Participants
Uncemented Femoral Stem is inserted into femoral medullary canal. Uncemented Acetabular Cup is inserted into acetabular cavity with or without fixation screw(s). Acetabular Liner, which is made of E1 highly crosslinked polyethylene, is fitted into Acetabular cup. Biolox delta head is inserted onto the taper of the Femoral stem, and is articulating against Acetabular Liner made of E1. Femoral Stem: JMDN classification/Class III device Acetabular Cup: JMDN classification: Class III device Acetabular Liner: JMDN classification: Class III device
CoCr Head
n=58 Participants
Uncemented Femoral Stem is inserted into femoral medullary canal. Uncemented Acetabular Cup is inserted into acetabular cavity with or without fixation screw(s). Acetabular Liner, which is made of E1 highly crosslinked polyethylene, is fitted into Acetabular cup. CoCr head is inserted onto the taper of the Femoral stem, and is articulating against Acetabular Liner made of E1. Femoral Stem: JMDN classification/Class III device Acetabular Cup: JMDN classification: Class III device Acetabular Liner: JMDN classification: Class III device
Harris Hip Score at 2 Year Follow-up Visit
96.15 score on a scale
Standard Deviation 5.44
96.28 score on a scale
Standard Deviation 5.69

SECONDARY outcome

Timeframe: 3 year postoperative

Population: 12 patients in BIOLOX delta head group and 12 patients in CoCr head group missed collecting Harris Hip Score at 3 year visit.

Harris Hip Score is Physician rating score to evaluate hip disabilities. Score domains are "pain", "function", "absence of deformity" and "range of hip motion". Minimum possible score is 0 and maximum possible score is 100. Higher score means better outcome.

Outcome measures

Outcome measures
Measure
BIOLOX Delta Head
n=50 Participants
Uncemented Femoral Stem is inserted into femoral medullary canal. Uncemented Acetabular Cup is inserted into acetabular cavity with or without fixation screw(s). Acetabular Liner, which is made of E1 highly crosslinked polyethylene, is fitted into Acetabular cup. Biolox delta head is inserted onto the taper of the Femoral stem, and is articulating against Acetabular Liner made of E1. Femoral Stem: JMDN classification/Class III device Acetabular Cup: JMDN classification: Class III device Acetabular Liner: JMDN classification: Class III device
CoCr Head
n=55 Participants
Uncemented Femoral Stem is inserted into femoral medullary canal. Uncemented Acetabular Cup is inserted into acetabular cavity with or without fixation screw(s). Acetabular Liner, which is made of E1 highly crosslinked polyethylene, is fitted into Acetabular cup. CoCr head is inserted onto the taper of the Femoral stem, and is articulating against Acetabular Liner made of E1. Femoral Stem: JMDN classification/Class III device Acetabular Cup: JMDN classification: Class III device Acetabular Liner: JMDN classification: Class III device
Harris Hip Score at 3 Year Follow-up Visit
96.30 score on a scale
Standard Deviation 4.56
96.82 score on a scale
Standard Deviation 6.03

SECONDARY outcome

Timeframe: 6 month postoperative

Population: 6 patients in BIOLOX delta head group and 11 patients in CoCr head group missed colleting WOMAC questionnaire at 6 month visit.

The Western Ontario and McMaster Universities Arthritis Index (WOMAC) is used in evaluation of Hip and Knee Osteoarthritis. It is a self-administered questionnaire consisting of 24 items. The scores for each subscale are summed up, with a possible score range of 0-20 for "Pain", 0-8 for Stiffness, and 0-68 for Physical Function. Usually a sum of the scores for all three subscales gives a The total WOMAC score ranged from 0 to 96. Higher scores on the WOMAC indicate worse pain, stiffness, and functional limitations. Higher scores on the WOMAC indicate worse pain, stiffness, and functional limitations.

Outcome measures

Outcome measures
Measure
BIOLOX Delta Head
n=56 Participants
Uncemented Femoral Stem is inserted into femoral medullary canal. Uncemented Acetabular Cup is inserted into acetabular cavity with or without fixation screw(s). Acetabular Liner, which is made of E1 highly crosslinked polyethylene, is fitted into Acetabular cup. Biolox delta head is inserted onto the taper of the Femoral stem, and is articulating against Acetabular Liner made of E1. Femoral Stem: JMDN classification/Class III device Acetabular Cup: JMDN classification: Class III device Acetabular Liner: JMDN classification: Class III device
CoCr Head
n=56 Participants
Uncemented Femoral Stem is inserted into femoral medullary canal. Uncemented Acetabular Cup is inserted into acetabular cavity with or without fixation screw(s). Acetabular Liner, which is made of E1 highly crosslinked polyethylene, is fitted into Acetabular cup. CoCr head is inserted onto the taper of the Femoral stem, and is articulating against Acetabular Liner made of E1. Femoral Stem: JMDN classification/Class III device Acetabular Cup: JMDN classification: Class III device Acetabular Liner: JMDN classification: Class III device
WOMAC Osteoarthritis Index at 6 Month Follow-up Visit
9.27 score on a scale
Standard Deviation 9.75
10.43 score on a scale
Standard Deviation 9.62

SECONDARY outcome

Timeframe: 1 year postoperative

Population: 6 patients in BIOLOX delta head group and 13 patients in CoCr head group missed colleting WOMAC questionnaire at 1 year visit.

The Western Ontario and McMaster Universities Arthritis Index (WOMAC) is used in evaluation of Hip and Knee Osteoarthritis. It is a self-administered questionnaire consisting of 24 items. The scores for each subscale are summed up, with a possible score range of 0-20 for "Pain", 0-8 for Stiffness, and 0-68 for Physical Function. Usually a sum of the scores for all three subscales gives a The total WOMAC score ranged from 0 to 96. Higher scores on the WOMAC indicate worse pain, stiffness, and functional limitations.

Outcome measures

Outcome measures
Measure
BIOLOX Delta Head
n=56 Participants
Uncemented Femoral Stem is inserted into femoral medullary canal. Uncemented Acetabular Cup is inserted into acetabular cavity with or without fixation screw(s). Acetabular Liner, which is made of E1 highly crosslinked polyethylene, is fitted into Acetabular cup. Biolox delta head is inserted onto the taper of the Femoral stem, and is articulating against Acetabular Liner made of E1. Femoral Stem: JMDN classification/Class III device Acetabular Cup: JMDN classification: Class III device Acetabular Liner: JMDN classification: Class III device
CoCr Head
n=54 Participants
Uncemented Femoral Stem is inserted into femoral medullary canal. Uncemented Acetabular Cup is inserted into acetabular cavity with or without fixation screw(s). Acetabular Liner, which is made of E1 highly crosslinked polyethylene, is fitted into Acetabular cup. CoCr head is inserted onto the taper of the Femoral stem, and is articulating against Acetabular Liner made of E1. Femoral Stem: JMDN classification/Class III device Acetabular Cup: JMDN classification: Class III device Acetabular Liner: JMDN classification: Class III device
WOMAC Osteoarthritis Index at 1 Year Follow-up Visit
9.86 score on a scale
Standard Deviation 13.52
9.57 score on a scale
Standard Deviation 9.23

SECONDARY outcome

Timeframe: 2 year postoperative

Population: 9 patients in BIOLOX delta head group and 16 patients in CoCr head group missed colleting WOMAC questionnaire at 2 year visit.

The Western Ontario and McMaster Universities Arthritis Index (WOMAC) is used in evaluation of Hip and Knee Osteoarthritis. It is a self-administered questionnaire consisting of 24 items. The scores for each subscale are summed up, with a possible score range of 0-20 for "Pain", 0-8 for Stiffness, and 0-68 for Physical Function. Usually a sum of the scores for all three subscales gives a The total WOMAC score ranged from 0 to 96. Higher scores on the WOMAC indicate worse pain, stiffness, and functional limitations.

Outcome measures

Outcome measures
Measure
BIOLOX Delta Head
n=53 Participants
Uncemented Femoral Stem is inserted into femoral medullary canal. Uncemented Acetabular Cup is inserted into acetabular cavity with or without fixation screw(s). Acetabular Liner, which is made of E1 highly crosslinked polyethylene, is fitted into Acetabular cup. Biolox delta head is inserted onto the taper of the Femoral stem, and is articulating against Acetabular Liner made of E1. Femoral Stem: JMDN classification/Class III device Acetabular Cup: JMDN classification: Class III device Acetabular Liner: JMDN classification: Class III device
CoCr Head
n=51 Participants
Uncemented Femoral Stem is inserted into femoral medullary canal. Uncemented Acetabular Cup is inserted into acetabular cavity with or without fixation screw(s). Acetabular Liner, which is made of E1 highly crosslinked polyethylene, is fitted into Acetabular cup. CoCr head is inserted onto the taper of the Femoral stem, and is articulating against Acetabular Liner made of E1. Femoral Stem: JMDN classification/Class III device Acetabular Cup: JMDN classification: Class III device Acetabular Liner: JMDN classification: Class III device
WOMAC Osteoarthritis Index at 2 Year Follow-up Visit
7.85 score on a scale
Standard Deviation 8.88
7.80 score on a scale
Standard Deviation 10.32

SECONDARY outcome

Timeframe: 3 year postoperative

Population: 14 patients in BIOLOX delta head group and 17 patients in CoCr head group missed colleting WOMAC questionnaire at 3 year visit.

The Western Ontario and McMaster Universities Arthritis Index (WOMAC) is used in evaluation of Hip and Knee Osteoarthritis. It is a self-administered questionnaire consisting of 24 items. The scores for each subscale are summed up, with a possible score range of 0-20 for "Pain", 0-8 for Stiffness, and 0-68 for Physical Function. Usually a sum of the scores for all three subscales gives a The total WOMAC score ranged from 0 to 96. Higher scores on the WOMAC indicate worse pain, stiffness, and functional limitations.

Outcome measures

Outcome measures
Measure
BIOLOX Delta Head
n=48 Participants
Uncemented Femoral Stem is inserted into femoral medullary canal. Uncemented Acetabular Cup is inserted into acetabular cavity with or without fixation screw(s). Acetabular Liner, which is made of E1 highly crosslinked polyethylene, is fitted into Acetabular cup. Biolox delta head is inserted onto the taper of the Femoral stem, and is articulating against Acetabular Liner made of E1. Femoral Stem: JMDN classification/Class III device Acetabular Cup: JMDN classification: Class III device Acetabular Liner: JMDN classification: Class III device
CoCr Head
n=50 Participants
Uncemented Femoral Stem is inserted into femoral medullary canal. Uncemented Acetabular Cup is inserted into acetabular cavity with or without fixation screw(s). Acetabular Liner, which is made of E1 highly crosslinked polyethylene, is fitted into Acetabular cup. CoCr head is inserted onto the taper of the Femoral stem, and is articulating against Acetabular Liner made of E1. Femoral Stem: JMDN classification/Class III device Acetabular Cup: JMDN classification: Class III device Acetabular Liner: JMDN classification: Class III device
WOMAC Osteoarthritis Index at 3 Year Follow-up Visit
8.29 score on a scale
Standard Deviation 9.31
6.84 score on a scale
Standard Deviation 8.00

SECONDARY outcome

Timeframe: 6 month postoperative

Population: 4 patients in BIOLOX delta head group and 12 patients in CoCr head group missed colleting UCLA Activity score at 6 month visit.

University of California, Los Angeles (UCLA) activity score is questionnaire to assess intensity of patients' activity. The score ranges from 1 to 10 (1 point increment), higher score means higher activity level.

Outcome measures

Outcome measures
Measure
BIOLOX Delta Head
n=58 Participants
Uncemented Femoral Stem is inserted into femoral medullary canal. Uncemented Acetabular Cup is inserted into acetabular cavity with or without fixation screw(s). Acetabular Liner, which is made of E1 highly crosslinked polyethylene, is fitted into Acetabular cup. Biolox delta head is inserted onto the taper of the Femoral stem, and is articulating against Acetabular Liner made of E1. Femoral Stem: JMDN classification/Class III device Acetabular Cup: JMDN classification: Class III device Acetabular Liner: JMDN classification: Class III device
CoCr Head
n=55 Participants
Uncemented Femoral Stem is inserted into femoral medullary canal. Uncemented Acetabular Cup is inserted into acetabular cavity with or without fixation screw(s). Acetabular Liner, which is made of E1 highly crosslinked polyethylene, is fitted into Acetabular cup. CoCr head is inserted onto the taper of the Femoral stem, and is articulating against Acetabular Liner made of E1. Femoral Stem: JMDN classification/Class III device Acetabular Cup: JMDN classification: Class III device Acetabular Liner: JMDN classification: Class III device
UCLA Activity Score at 6 Month Follow-up Visit
5.19 score on a scale
Standard Deviation 1.42
4.75 score on a scale
Standard Deviation 1.28

SECONDARY outcome

Timeframe: 1 year postoperative

Population: 2 patients in BIOLOX delta head group and 8 patients in CoCr head group missed colleting UCLA Activity score at 1 year visit.

University of California, Los Angeles (UCLA) activity score is questionnaire to assess intensity of patients' activity. The score ranges from 1 to 10 (1 point increment), higher score means higher activity level.

Outcome measures

Outcome measures
Measure
BIOLOX Delta Head
n=60 Participants
Uncemented Femoral Stem is inserted into femoral medullary canal. Uncemented Acetabular Cup is inserted into acetabular cavity with or without fixation screw(s). Acetabular Liner, which is made of E1 highly crosslinked polyethylene, is fitted into Acetabular cup. Biolox delta head is inserted onto the taper of the Femoral stem, and is articulating against Acetabular Liner made of E1. Femoral Stem: JMDN classification/Class III device Acetabular Cup: JMDN classification: Class III device Acetabular Liner: JMDN classification: Class III device
CoCr Head
n=59 Participants
Uncemented Femoral Stem is inserted into femoral medullary canal. Uncemented Acetabular Cup is inserted into acetabular cavity with or without fixation screw(s). Acetabular Liner, which is made of E1 highly crosslinked polyethylene, is fitted into Acetabular cup. CoCr head is inserted onto the taper of the Femoral stem, and is articulating against Acetabular Liner made of E1. Femoral Stem: JMDN classification/Class III device Acetabular Cup: JMDN classification: Class III device Acetabular Liner: JMDN classification: Class III device
UCLA Activity Score at 1 Year Follow-up Visit
5.45 score on a scale
Standard Deviation 1.52
5.03 score on a scale
Standard Deviation 1.27

SECONDARY outcome

Timeframe: 2 year postoperative

Population: 3 patients in BIOLOX delta head group and 9 patients in CoCr head group missed colleting UCLA Activity score at 2 year visit.

University of California, Los Angeles (UCLA) activity score is questionnaire to assess intensity of patients' activity. The score ranges from 1 to 10 (1 point increment), higher score means higher activity level.

Outcome measures

Outcome measures
Measure
BIOLOX Delta Head
n=59 Participants
Uncemented Femoral Stem is inserted into femoral medullary canal. Uncemented Acetabular Cup is inserted into acetabular cavity with or without fixation screw(s). Acetabular Liner, which is made of E1 highly crosslinked polyethylene, is fitted into Acetabular cup. Biolox delta head is inserted onto the taper of the Femoral stem, and is articulating against Acetabular Liner made of E1. Femoral Stem: JMDN classification/Class III device Acetabular Cup: JMDN classification: Class III device Acetabular Liner: JMDN classification: Class III device
CoCr Head
n=58 Participants
Uncemented Femoral Stem is inserted into femoral medullary canal. Uncemented Acetabular Cup is inserted into acetabular cavity with or without fixation screw(s). Acetabular Liner, which is made of E1 highly crosslinked polyethylene, is fitted into Acetabular cup. CoCr head is inserted onto the taper of the Femoral stem, and is articulating against Acetabular Liner made of E1. Femoral Stem: JMDN classification/Class III device Acetabular Cup: JMDN classification: Class III device Acetabular Liner: JMDN classification: Class III device
UCLA Activity Score at 2 Year Follow-up Visit
5.53 score on a scale
Standard Deviation 1.38
5.31 score on a scale
Standard Deviation 1.30

SECONDARY outcome

Timeframe: 3 year postoperative

Population: 12 patients in BIOLOX delta head group and 11 patients in CoCr head group missed colleting UCLA Activity score at 3 year visit.

University of California, Los Angeles (UCLA) activity score is questionnaire to assess intensity of patients' activity. The score ranges from 1 to 10 (1 point increment), higher score means higher activity level.

Outcome measures

Outcome measures
Measure
BIOLOX Delta Head
n=50 Participants
Uncemented Femoral Stem is inserted into femoral medullary canal. Uncemented Acetabular Cup is inserted into acetabular cavity with or without fixation screw(s). Acetabular Liner, which is made of E1 highly crosslinked polyethylene, is fitted into Acetabular cup. Biolox delta head is inserted onto the taper of the Femoral stem, and is articulating against Acetabular Liner made of E1. Femoral Stem: JMDN classification/Class III device Acetabular Cup: JMDN classification: Class III device Acetabular Liner: JMDN classification: Class III device
CoCr Head
n=56 Participants
Uncemented Femoral Stem is inserted into femoral medullary canal. Uncemented Acetabular Cup is inserted into acetabular cavity with or without fixation screw(s). Acetabular Liner, which is made of E1 highly crosslinked polyethylene, is fitted into Acetabular cup. CoCr head is inserted onto the taper of the Femoral stem, and is articulating against Acetabular Liner made of E1. Femoral Stem: JMDN classification/Class III device Acetabular Cup: JMDN classification: Class III device Acetabular Liner: JMDN classification: Class III device
UCLA Activity Score at 3 Year Follow-up Visit
5.42 score on a scale
Standard Deviation 1.49
5.43 score on a scale
Standard Deviation 1.36

Adverse Events

BIOLOX Delta Head

Serious events: 0 serious events
Other events: 2 other events
Deaths: 0 deaths

CoCr Head

Serious events: 0 serious events
Other events: 1 other events
Deaths: 0 deaths

Serious adverse events

Adverse event data not reported

Other adverse events

Other adverse events
Measure
BIOLOX Delta Head
n=62 participants at risk
Uncemented Femoral Stem is inserted into femoral medullary canal. Uncemented Acetabular Cup is inserted into acetabular cavity with or without fixation screw(s). Acetabular Liner, which is made of E1 highly crosslinked polyethylene, is fitted into Acetabular cup. Biolox delta head is inserted onto the taper of the Femoral stem, and is articulating against Acetabular Liner made of E1. Femoral Stem: JMDN classification/Class III device Acetabular Cup: JMDN classification: Class III device Acetabular Liner: JMDN classification: Class III device
CoCr Head
n=67 participants at risk
Uncemented Femoral Stem is inserted into femoral medullary canal. Uncemented Acetabular Cup is inserted into acetabular cavity with or without fixation screw(s). Acetabular Liner, which is made of E1 highly crosslinked polyethylene, is fitted into Acetabular cup. CoCr head is inserted onto the taper of the Femoral stem, and is articulating against Acetabular Liner made of E1. Femoral Stem: JMDN classification/Class III device Acetabular Cup: JMDN classification: Class III device Acetabular Liner: JMDN classification: Class III device
Musculoskeletal and connective tissue disorders
Intraoperative Trochanteric Fracture
1.6%
1/62 • Number of events 1 • 3 year after surgery
0.00%
0/67 • 3 year after surgery
Nervous system disorders
Femoral Nerve Palsy
1.6%
1/62 • Number of events 1 • 3 year after surgery
0.00%
0/67 • 3 year after surgery
Musculoskeletal and connective tissue disorders
Inappropriate femoral stem position
0.00%
0/62 • 3 year after surgery
1.5%
1/67 • Number of events 1 • 3 year after surgery

Additional Information

Senior Manager of Clinical Affairs

Zimmer Biomet G.K.

Phone: +81-3-6402-6610

Results disclosure agreements

  • Principal investigator is a sponsor employee
  • Publication restrictions are in place