Trial Outcomes & Findings for Duchenne Muscular Dystrophy Clinical Trial (NCT NCT01874275)

NCT ID: NCT01874275

Last Updated: 2015-05-04

Results Overview

Range of Motion (ROM) testing (Wireless Tracker System) - All muscle and joint testing was performed using the JTech computerized testing system, including Goniometry, Grip Testing, Inclinometry, Muscle Testing and Joint Range of Motion. Testing occurred at Baseline (Day 0), 30, 60, 90 and 180 days to determine changes in joint Range of Motion. JTech computerized testing system determined the joint range of motion by radio signals from a goniometer measuring movement of the participants' joints. 44 separate tests were performed for Range of Motion for each participant at each time interval. The results from the range of motion tests were averaged the percent change from baseline to 180 days. Efficacy is defined as an increase in range of motion.

Recruitment status

COMPLETED

Study phase

NA

Target enrollment

6 participants

Primary outcome timeframe

Baseline to 180 days

Results posted on

2015-05-04

Participant Flow

Participant milestones

Participant milestones
Measure
VECTTOR
muscle stimulator treatment twice daily for duration of study - 365 days VECTTOR: The VT-200, or VECTTOR system, delivers electrical stimulation via electrodes on the acupuncture points of a patient's feet/legs and hands/arms to provide symptomatic relief of chronic intractable pain and/or management of post-surgical pain. Once these electrodes are placed, the machine determines the appropriate choice of stimulation by automatically measuring body temperatures through the use of thermistors placed on the fingers during a testing sequence.
Device - Sham
placebo treatment - no electrical stimulation treatment twice daily for 180 days followed by active treatment for the duration of study, 365 days VECTTOR: The VT-200, or VECTTOR system, delivers electrical stimulation via electrodes on the acupuncture points of a patient's feet/legs and hands/arms to provide symptomatic relief of chronic intractable pain and/or management of post-surgical pain. Once these electrodes are placed, the machine determines the appropriate choice of stimulation by automatically measuring body temperatures through the use of thermistors placed on the fingers during a testing sequence.
Overall Study
STARTED
4
2
Overall Study
COMPLETED
3
2
Overall Study
NOT COMPLETED
1
0

Reasons for withdrawal

Reasons for withdrawal
Measure
VECTTOR
muscle stimulator treatment twice daily for duration of study - 365 days VECTTOR: The VT-200, or VECTTOR system, delivers electrical stimulation via electrodes on the acupuncture points of a patient's feet/legs and hands/arms to provide symptomatic relief of chronic intractable pain and/or management of post-surgical pain. Once these electrodes are placed, the machine determines the appropriate choice of stimulation by automatically measuring body temperatures through the use of thermistors placed on the fingers during a testing sequence.
Device - Sham
placebo treatment - no electrical stimulation treatment twice daily for 180 days followed by active treatment for the duration of study, 365 days VECTTOR: The VT-200, or VECTTOR system, delivers electrical stimulation via electrodes on the acupuncture points of a patient's feet/legs and hands/arms to provide symptomatic relief of chronic intractable pain and/or management of post-surgical pain. Once these electrodes are placed, the machine determines the appropriate choice of stimulation by automatically measuring body temperatures through the use of thermistors placed on the fingers during a testing sequence.
Overall Study
Lost to Follow-up
1
0

Baseline Characteristics

Duchenne Muscular Dystrophy Clinical Trial

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
VECTTOR - Active
n=4 Participants
nerve stimulator treatment twice daily for duration of study - 365 days ...
Device Sham
n=2 Participants
placebo treatment - no nerve stimulator treatment twice daily for 180 days of study
Total
n=6 Participants
Total of all reporting groups
Age, Categorical
<=18 years
4 Participants
n=99 Participants
2 Participants
n=107 Participants
6 Participants
n=206 Participants
Age, Categorical
Between 18 and 65 years
0 Participants
n=99 Participants
0 Participants
n=107 Participants
0 Participants
n=206 Participants
Age, Categorical
>=65 years
0 Participants
n=99 Participants
0 Participants
n=107 Participants
0 Participants
n=206 Participants
Sex: Female, Male
Female
0 Participants
n=99 Participants
0 Participants
n=107 Participants
0 Participants
n=206 Participants
Sex: Female, Male
Male
4 Participants
n=99 Participants
2 Participants
n=107 Participants
6 Participants
n=206 Participants
Region of Enrollment
United States
4 participants
n=99 Participants
2 participants
n=107 Participants
6 participants
n=206 Participants

PRIMARY outcome

Timeframe: Baseline to 180 days

Range of Motion (ROM) testing (Wireless Tracker System) - All muscle and joint testing was performed using the JTech computerized testing system, including Goniometry, Grip Testing, Inclinometry, Muscle Testing and Joint Range of Motion. Testing occurred at Baseline (Day 0), 30, 60, 90 and 180 days to determine changes in joint Range of Motion. JTech computerized testing system determined the joint range of motion by radio signals from a goniometer measuring movement of the participants' joints. 44 separate tests were performed for Range of Motion for each participant at each time interval. The results from the range of motion tests were averaged the percent change from baseline to 180 days. Efficacy is defined as an increase in range of motion.

Outcome measures

Outcome measures
Measure
VECTTOR
n=4 Participants
Muscle strength testing (Wireless Tracker system) - All muscle strength testing was performed using the JTech computerized testing system, including Goniometry, Grip Testing, Inclinometry, Muscle Testing and Joint Range of Motion. Testing occurred at Baseline (Day 0), 30, 60, 90, 180 and 365 days to determine changes in muscle strength. JTech computerized testing system determined the muscle strength by radio signals from a strain gauge measuring strength of the participant's muscles. 44 separate tests were performed for muscle strength for each participant at each time interval. Efficacy is defined as an increase in the strength measurement (pounds) from Baseline to 365 days.
Device - Sham
n=2 Participants
placebo treatment - no electrical stimulation treatment twice daily for 180 days followed by cross-over with active treatment for the next 180 days
Percent Change in Range of Motion From Baseline to 180 Days
67 percentage of change
Standard Deviation 44.3
33 percentage of change
Standard Deviation 49.2

SECONDARY outcome

Timeframe: Baseline to 180 days

Muscle strength testing (Wireless Tracker system) - All muscle strength testing was performed using the JTech computerized testing system, including Goniometry, Grip Testing, Inclinometry, Muscle Testing and Joint Range of Motion. Testing occurred at Baseline (Day 0), 30, 60, 90, and 180 days to determine changes in muscle strength. JTech computerized testing system determined the muscle strength by radio signals from a strain gauge measuring strength of the participant's muscles. 44 separate tests were performed for muscle strength for each participant at each time interval. Efficacy is defined as an increase in the strength measurement (pounds) from Baseline to 180 days.

Outcome measures

Outcome measures
Measure
VECTTOR
n=4 Participants
Muscle strength testing (Wireless Tracker system) - All muscle strength testing was performed using the JTech computerized testing system, including Goniometry, Grip Testing, Inclinometry, Muscle Testing and Joint Range of Motion. Testing occurred at Baseline (Day 0), 30, 60, 90, 180 and 365 days to determine changes in muscle strength. JTech computerized testing system determined the muscle strength by radio signals from a strain gauge measuring strength of the participant's muscles. 44 separate tests were performed for muscle strength for each participant at each time interval. Efficacy is defined as an increase in the strength measurement (pounds) from Baseline to 365 days.
Device - Sham
n=2 Participants
placebo treatment - no electrical stimulation treatment twice daily for 180 days followed by cross-over with active treatment for the next 180 days
Percent Change in Muscle Strength
95 percentage of change
Standard Deviation 2.16
48 percentage of change
Standard Deviation 1.08

SECONDARY outcome

Timeframe: Baseline to 365 days

Range of Motion (ROM) testing (Wireless Tracker System) - All muscle and joint testing was performed using the JTech computerized testing system, including Goniometry, Grip Testing, Inclinometry, Muscle Testing and Joint Range of Motion. Testing occurred at Baseline (Day 0), 30, 60, 90, 180 and 365 days to determine changes in joint Range of Motion. JTech computerized testing system determined the joint range of motion by radio signals from a goniometer measuring movement of the participants' joints. 44 separate tests were performed for Range of Motion for each participant at each time interval. The results from the range of motion tests were averaged the percent change from baseline to 365 days. Efficacy is defined as an increase in range of motion.

Outcome measures

Outcome measures
Measure
VECTTOR
n=4 Participants
Muscle strength testing (Wireless Tracker system) - All muscle strength testing was performed using the JTech computerized testing system, including Goniometry, Grip Testing, Inclinometry, Muscle Testing and Joint Range of Motion. Testing occurred at Baseline (Day 0), 30, 60, 90, 180 and 365 days to determine changes in muscle strength. JTech computerized testing system determined the muscle strength by radio signals from a strain gauge measuring strength of the participant's muscles. 44 separate tests were performed for muscle strength for each participant at each time interval. Efficacy is defined as an increase in the strength measurement (pounds) from Baseline to 365 days.
Device - Sham
placebo treatment - no electrical stimulation treatment twice daily for 180 days followed by cross-over with active treatment for the next 180 days
Percent Change in Percent Range of Motion From Baseline to 365 Days
59 percentage of change
Standard Deviation 45.1

SECONDARY outcome

Timeframe: Baseline to 365 days

Muscle strength testing (Wireless Tracker system) - All muscle strength testing was performed using the JTech computerized testing system, including Goniometry, Grip Testing, Inclinometry, Muscle Testing and Joint Range of Motion. Testing occurred at Baseline (Day 0), 30, 60, 90, 180 and 365 days to determine changes in muscle strength. JTech computerized testing system determined the muscle strength by radio signals from a strain gauge measuring strength of the participant's muscles. 44 separate tests were performed for muscle strength for each participant at each time interval. Efficacy is defined as an increase in the strength measurement (pounds) from Baseline to 365 days.

Outcome measures

Outcome measures
Measure
VECTTOR
n=4 Participants
Muscle strength testing (Wireless Tracker system) - All muscle strength testing was performed using the JTech computerized testing system, including Goniometry, Grip Testing, Inclinometry, Muscle Testing and Joint Range of Motion. Testing occurred at Baseline (Day 0), 30, 60, 90, 180 and 365 days to determine changes in muscle strength. JTech computerized testing system determined the muscle strength by radio signals from a strain gauge measuring strength of the participant's muscles. 44 separate tests were performed for muscle strength for each participant at each time interval. Efficacy is defined as an increase in the strength measurement (pounds) from Baseline to 365 days.
Device - Sham
placebo treatment - no electrical stimulation treatment twice daily for 180 days followed by cross-over with active treatment for the next 180 days
Percent Change in Muscle Strength
110 percentage of change
Standard Deviation 1.93

OTHER_PRE_SPECIFIED outcome

Timeframe: Baseline to 180 days

Sleep / Arousal Statistics Index (ASI) description: Sleep Studies were obtained twice before beginning the study. The second sleep study data was used as baseline. A follow up sleep study was obtained after 6 months (180 days) of Active or Placebo VECTTOR treatment. Sleep, breathing, arousal(s), and limb movements were scored manually according to guidelines of the American Academy of Sleep Medicine. The efficacy of the outcome measure of ASI is demonstrated by a percent improvement between Baseline and 180 days. The ASI is measured by the number of times sleep is interrupted per hour.

Outcome measures

Outcome measures
Measure
VECTTOR
n=4 Participants
Muscle strength testing (Wireless Tracker system) - All muscle strength testing was performed using the JTech computerized testing system, including Goniometry, Grip Testing, Inclinometry, Muscle Testing and Joint Range of Motion. Testing occurred at Baseline (Day 0), 30, 60, 90, 180 and 365 days to determine changes in muscle strength. JTech computerized testing system determined the muscle strength by radio signals from a strain gauge measuring strength of the participant's muscles. 44 separate tests were performed for muscle strength for each participant at each time interval. Efficacy is defined as an increase in the strength measurement (pounds) from Baseline to 365 days.
Device - Sham
n=2 Participants
placebo treatment - no electrical stimulation treatment twice daily for 180 days followed by cross-over with active treatment for the next 180 days
Percent Change in Sleep Quality Arousal Statistics Index (ASI) From Baseline to 180 Days
38 percentage of change
Standard Deviation 1.31
-53 percentage of change
Standard Deviation 1.77

OTHER_PRE_SPECIFIED outcome

Timeframe: Baseline to 365 days

Sleep / Arousal Statistics Index (ASI) description: Sleep Studies were obtained twice before beginning the study. The second sleep study data was used as baseline. A follow up sleep study was obtained after 6 months (180 days) of Active or Placebo VECTTOR treatment. Sleep, breathing, arousal(s), and limb movements were scored manually according to guidelines of the American Academy of Sleep Medicine. The efficacy of the outcome measure of ASI is demonstrated by a decrease in the value between Baseline and 180 days. The ASI is measured by the number of times sleep is interrupted per hour. Lower ASI values/numbers indicate improved sleep quality.

Outcome measures

Outcome data not reported

Adverse Events

VECTTOR

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Device - Sham

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Serious adverse events

Adverse event data not reported

Other adverse events

Adverse event data not reported