Trial Outcomes & Findings for User Performance Evaluation of an Investigational Blood Glucose Monitoring System (NCT NCT01410773)

NCT ID: NCT01410773

Last Updated: 2016-02-29

Results Overview

Untrained subjects with diabetes self-test fingerstick blood using an investigational Blood Glucose Meter (BGM). BGM results are compared with capillary plasma BG results obtained with a reference lab glucose method - Yellow Springs Instrument (YSI) Analyzer. BG meter results are used to calculate the number of BG results within +/- 15mg/dL (for reference BG results \<75mg/dL) or within +/- 20%(for reference BG results \>=75mg/dL) of the reference method results (YSI capillary plasma).

Recruitment status

COMPLETED

Study phase

NA

Target enrollment

110 participants

Primary outcome timeframe

1 hour

Results posted on

2016-02-29

Participant Flow

Participant milestones

Participant milestones
Measure
Intended Users of the System
Untrained subjects with diabetes (at least 70% of subjects will be insulin users) use an investigational blood glucose monitoring system (Ninja 2) to self-test capillary blood obtained from fingerstick and palm. Ninja 2 Investigational Blood Glucose Monitoring System : The Ninja 2 meter is a Bayer investigational meter that uses an investigational sensor.
Overall Study
STARTED
110
Overall Study
COMPLETED
110
Overall Study
NOT COMPLETED
0

Reasons for withdrawal

Withdrawal data not reported

Baseline Characteristics

User Performance Evaluation of an Investigational Blood Glucose Monitoring System

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
Intended Users of the System
n=110 Participants
Untrained subjects with diabetes (at least 70% of subjects will be insulin users) use an investigational blood glucose monitoring system (Ninja 2) to self-test capillary blood obtained from fingerstick and palm. Ninja 2 Investigational Blood Glucose Monitoring System : The Ninja 2 meter is a Bayer investigational meter that uses an investigational sensor.
Age, Customized
Mean (Full range of ages)
53.1 years
n=99 Participants
Sex: Female, Male
Female
58 Participants
n=99 Participants
Sex: Female, Male
Male
52 Participants
n=99 Participants
Region of Enrollment
United States
110 participants
n=99 Participants

PRIMARY outcome

Timeframe: 1 hour

Population: 110 subjects tested one of 3 test strip lots on the BGM system. 110 (1x110) test results are available.

Untrained subjects with diabetes self-test fingerstick blood using an investigational Blood Glucose Meter (BGM). BGM results are compared with capillary plasma BG results obtained with a reference lab glucose method - Yellow Springs Instrument (YSI) Analyzer. BG meter results are used to calculate the number of BG results within +/- 15mg/dL (for reference BG results \<75mg/dL) or within +/- 20%(for reference BG results \>=75mg/dL) of the reference method results (YSI capillary plasma).

Outcome measures

Outcome measures
Measure
Intended Users of the System
n=110 Participants
Untrained subjects with diabetes (at least 70% of subjects will be insulin users) use an investigational blood glucose monitoring system (Ninja 2) to self-test capillary blood obtained from fingerstick and palm. Ninja 2 Investigational Blood Glucose Monitoring System : The Ninja 2 meter is a Bayer investigational meter that uses an investigational sensor.
Number of Self-Test Fingerstick Blood Glucose (BG) Results Within +/-15mg/dL(<75 mg/dL) or Within +/- 20% (>=75 mg/dL) of Laboratory Glucose Method
# within +/-15mg/dL(<75mg/dL)or 20% (>=75mg/dL)
110 participants
Number of Self-Test Fingerstick Blood Glucose (BG) Results Within +/-15mg/dL(<75 mg/dL) or Within +/- 20% (>=75 mg/dL) of Laboratory Glucose Method
# within +/-15mg/dL(<75mg/dL)or 15% (>=75mg/dL)
109 participants
Number of Self-Test Fingerstick Blood Glucose (BG) Results Within +/-15mg/dL(<75 mg/dL) or Within +/- 20% (>=75 mg/dL) of Laboratory Glucose Method
# within +/-10mg/dL(<75mg/dL)or 10% (>=75mg/dL)
109 participants
Number of Self-Test Fingerstick Blood Glucose (BG) Results Within +/-15mg/dL(<75 mg/dL) or Within +/- 20% (>=75 mg/dL) of Laboratory Glucose Method
# within +/-5mg/dL(<75mg/dL) or 5% (>=75mg/dL)
90 participants

SECONDARY outcome

Timeframe: 1 hour

Population: One subject had low blood sugar. The protocol (and User Guide) do not allow alternative site testing when blood sugar is low. The remaining 109 subjects tested one strip lot on the BGM system. 109 test results were available.

Untrained subjects with diabetes self-test subject Alternative Site (AST) Palm blood using an investigational Blood Glucose Monitoring System (BGMS). BGM meter results are compared with capillary plasma BG results obtained with a reference lab glucose method - Yellow Springs Instrument (YSI) Analyzer. BG meter results are used to calculate the number of BG results within +/- 15mg/dL (for reference BG results \<75mg/dL) or +/- 20% (for reference BG results \>=75mg/dL) of the YSI capillary plasma reference method results.

Outcome measures

Outcome measures
Measure
Intended Users of the System
n=109 Participants
Untrained subjects with diabetes (at least 70% of subjects will be insulin users) use an investigational blood glucose monitoring system (Ninja 2) to self-test capillary blood obtained from fingerstick and palm. Ninja 2 Investigational Blood Glucose Monitoring System : The Ninja 2 meter is a Bayer investigational meter that uses an investigational sensor.
Number of Alternative Site (AST) Palm Blood Glucose (BG) Results Within +/- 15mg/dL (<75mg/dL) or Within +/- 20% (>=75mg/dL) of Laboratory Glucose Method
# within +/- 15mg/dL(<75mg/dL) or 20% (>=75mg/dL)
106 participants
Number of Alternative Site (AST) Palm Blood Glucose (BG) Results Within +/- 15mg/dL (<75mg/dL) or Within +/- 20% (>=75mg/dL) of Laboratory Glucose Method
# within +/- 15mg/dL(<75mg/dL) or 15% (>=75mg/dL)
105 participants
Number of Alternative Site (AST) Palm Blood Glucose (BG) Results Within +/- 15mg/dL (<75mg/dL) or Within +/- 20% (>=75mg/dL) of Laboratory Glucose Method
# within +/- 10mg/dL(<75mg/dL) or 10% (>=75mg/dL)
97 participants
Number of Alternative Site (AST) Palm Blood Glucose (BG) Results Within +/- 15mg/dL (<75mg/dL) or Within +/- 20% (>=75mg/dL) of Laboratory Glucose Method
# within +/- 5mg/dL(<75mg/dL) or 5% (>=75mg/dL)
67 participants

Adverse Events

Intended Users of the System

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Serious adverse events

Adverse event data not reported

Other adverse events

Adverse event data not reported