Trial Outcomes & Findings for User Performance Evaluation of an Investigational Blood Glucose Monitoring System (NCT NCT01410773)
NCT ID: NCT01410773
Last Updated: 2016-02-29
Results Overview
Untrained subjects with diabetes self-test fingerstick blood using an investigational Blood Glucose Meter (BGM). BGM results are compared with capillary plasma BG results obtained with a reference lab glucose method - Yellow Springs Instrument (YSI) Analyzer. BG meter results are used to calculate the number of BG results within +/- 15mg/dL (for reference BG results \<75mg/dL) or within +/- 20%(for reference BG results \>=75mg/dL) of the reference method results (YSI capillary plasma).
COMPLETED
NA
110 participants
1 hour
2016-02-29
Participant Flow
Participant milestones
| Measure |
Intended Users of the System
Untrained subjects with diabetes (at least 70% of subjects will be insulin users) use an investigational blood glucose monitoring system (Ninja 2) to self-test capillary blood obtained from fingerstick and palm.
Ninja 2 Investigational Blood Glucose Monitoring System : The Ninja 2 meter is a Bayer investigational meter that uses an investigational sensor.
|
|---|---|
|
Overall Study
STARTED
|
110
|
|
Overall Study
COMPLETED
|
110
|
|
Overall Study
NOT COMPLETED
|
0
|
Reasons for withdrawal
Withdrawal data not reported
Baseline Characteristics
User Performance Evaluation of an Investigational Blood Glucose Monitoring System
Baseline characteristics by cohort
| Measure |
Intended Users of the System
n=110 Participants
Untrained subjects with diabetes (at least 70% of subjects will be insulin users) use an investigational blood glucose monitoring system (Ninja 2) to self-test capillary blood obtained from fingerstick and palm.
Ninja 2 Investigational Blood Glucose Monitoring System : The Ninja 2 meter is a Bayer investigational meter that uses an investigational sensor.
|
|---|---|
|
Age, Customized
Mean (Full range of ages)
|
53.1 years
n=99 Participants
|
|
Sex: Female, Male
Female
|
58 Participants
n=99 Participants
|
|
Sex: Female, Male
Male
|
52 Participants
n=99 Participants
|
|
Region of Enrollment
United States
|
110 participants
n=99 Participants
|
PRIMARY outcome
Timeframe: 1 hourPopulation: 110 subjects tested one of 3 test strip lots on the BGM system. 110 (1x110) test results are available.
Untrained subjects with diabetes self-test fingerstick blood using an investigational Blood Glucose Meter (BGM). BGM results are compared with capillary plasma BG results obtained with a reference lab glucose method - Yellow Springs Instrument (YSI) Analyzer. BG meter results are used to calculate the number of BG results within +/- 15mg/dL (for reference BG results \<75mg/dL) or within +/- 20%(for reference BG results \>=75mg/dL) of the reference method results (YSI capillary plasma).
Outcome measures
| Measure |
Intended Users of the System
n=110 Participants
Untrained subjects with diabetes (at least 70% of subjects will be insulin users) use an investigational blood glucose monitoring system (Ninja 2) to self-test capillary blood obtained from fingerstick and palm.
Ninja 2 Investigational Blood Glucose Monitoring System : The Ninja 2 meter is a Bayer investigational meter that uses an investigational sensor.
|
|---|---|
|
Number of Self-Test Fingerstick Blood Glucose (BG) Results Within +/-15mg/dL(<75 mg/dL) or Within +/- 20% (>=75 mg/dL) of Laboratory Glucose Method
# within +/-15mg/dL(<75mg/dL)or 20% (>=75mg/dL)
|
110 participants
|
|
Number of Self-Test Fingerstick Blood Glucose (BG) Results Within +/-15mg/dL(<75 mg/dL) or Within +/- 20% (>=75 mg/dL) of Laboratory Glucose Method
# within +/-15mg/dL(<75mg/dL)or 15% (>=75mg/dL)
|
109 participants
|
|
Number of Self-Test Fingerstick Blood Glucose (BG) Results Within +/-15mg/dL(<75 mg/dL) or Within +/- 20% (>=75 mg/dL) of Laboratory Glucose Method
# within +/-10mg/dL(<75mg/dL)or 10% (>=75mg/dL)
|
109 participants
|
|
Number of Self-Test Fingerstick Blood Glucose (BG) Results Within +/-15mg/dL(<75 mg/dL) or Within +/- 20% (>=75 mg/dL) of Laboratory Glucose Method
# within +/-5mg/dL(<75mg/dL) or 5% (>=75mg/dL)
|
90 participants
|
SECONDARY outcome
Timeframe: 1 hourPopulation: One subject had low blood sugar. The protocol (and User Guide) do not allow alternative site testing when blood sugar is low. The remaining 109 subjects tested one strip lot on the BGM system. 109 test results were available.
Untrained subjects with diabetes self-test subject Alternative Site (AST) Palm blood using an investigational Blood Glucose Monitoring System (BGMS). BGM meter results are compared with capillary plasma BG results obtained with a reference lab glucose method - Yellow Springs Instrument (YSI) Analyzer. BG meter results are used to calculate the number of BG results within +/- 15mg/dL (for reference BG results \<75mg/dL) or +/- 20% (for reference BG results \>=75mg/dL) of the YSI capillary plasma reference method results.
Outcome measures
| Measure |
Intended Users of the System
n=109 Participants
Untrained subjects with diabetes (at least 70% of subjects will be insulin users) use an investigational blood glucose monitoring system (Ninja 2) to self-test capillary blood obtained from fingerstick and palm.
Ninja 2 Investigational Blood Glucose Monitoring System : The Ninja 2 meter is a Bayer investigational meter that uses an investigational sensor.
|
|---|---|
|
Number of Alternative Site (AST) Palm Blood Glucose (BG) Results Within +/- 15mg/dL (<75mg/dL) or Within +/- 20% (>=75mg/dL) of Laboratory Glucose Method
# within +/- 15mg/dL(<75mg/dL) or 20% (>=75mg/dL)
|
106 participants
|
|
Number of Alternative Site (AST) Palm Blood Glucose (BG) Results Within +/- 15mg/dL (<75mg/dL) or Within +/- 20% (>=75mg/dL) of Laboratory Glucose Method
# within +/- 15mg/dL(<75mg/dL) or 15% (>=75mg/dL)
|
105 participants
|
|
Number of Alternative Site (AST) Palm Blood Glucose (BG) Results Within +/- 15mg/dL (<75mg/dL) or Within +/- 20% (>=75mg/dL) of Laboratory Glucose Method
# within +/- 10mg/dL(<75mg/dL) or 10% (>=75mg/dL)
|
97 participants
|
|
Number of Alternative Site (AST) Palm Blood Glucose (BG) Results Within +/- 15mg/dL (<75mg/dL) or Within +/- 20% (>=75mg/dL) of Laboratory Glucose Method
# within +/- 5mg/dL(<75mg/dL) or 5% (>=75mg/dL)
|
67 participants
|